MinXray PowerPlus HF120/60HPPWV User guide

E-mail: [email protected]
www.minxray.com
Tel. 1-847-564-0323
Toll Free Tel: 1-800-221-2245 (U.S. and Canada)
Fax: 1-847-564-9040
MinXray, Inc.
3611 Commercial Avenue
Northbrook, Illinois 60062-1822
U.S.A.
Federal law restricts this device to sale by or on the order of a
licensed physician.
This special purpose portable equipment is for diagnostic radiography only.
It is not designed, nor does it meet, the standards for any other use.
Caution:
HF120/60HPPWV PowerPlus
High Frequency Portable X-ray Unit
Installation and Operating Instruction
Version 1.8
TM


1
2
3
5
6
8
9
9
…
………… 10
14
15
17
17
19
21
23
24
X-ray Tube
…
………………………………………………
…
Schematic Diagram
…
………………………………………………
…
DR interface connector
…
…………………………………
…
Specifications
…
………………………………………………
…
Operating Instructions
Operation
…
…………………………………
…
Safety Precautions
…
…………………………………
…
Maintenance
…
…………………………………
…
Unpacking
…
…………………………………
…
A
ssembly Instructions
…
……………………
…
Pre-Operational Required Testing
Installation Check list
…
……………………
…
Warning statements, figures and symbols
…
……………………
…
Installation Instructions
Main Parts
…
………………………………………………
…
Recordkeeping Requirements
…
……………………
…
Components
…
………………………………………………
…
TABLE OF CONTENTS
Introduction
…
………………………………………………
…
Certifications
…
………………………………………………
…

A.
B.
C.
D.
E. Instructions and precautionary notices of this manual.
1
Requirements of the user’s in-house radiation protection program.
State and local regulations governing radiation protection and the use of diagnostic X-ray
e
q
ui
p
ment.
All documents relating to the Performance Standard fo
r
Diagnostic X-Ray Systems, 21 CFR
Subchapter J, Part 1020; obtainable from FDA Center for Devices and Radiological Health,
Department of HHS, 2098 Gaither Road, Rockville, MD 20850.
All personnel authorized to operate o
r
service this equipment should be fully acquainted with the
established maximum permissible doses, safety recommendations, and procedures derived from the
following sources:
National Council on Radiation Protection Report No.102 (Medical X-Ray and Gamma Ray
Protection for Energies up to 10 MEV - Equipment Design and Use); from NRCP Publications;
P.O. Box 30175, Washington, D.C. 20014.
CARELESS OR IMPROPER USE OF X-RAY EQUIPMENT CAN BE EXTREMELY HAZARDOUS.
It is imperative that this equipment be operated and serviced only by trained personnel familiar with the
safety precautions required to prevent excessive exposure to primary X-ray radiation, the dangers of
exposure to X-ray radiation, and the proper use of the equipment discussed in this manual.
The MinXray model HF120/60HPPWV PowerPlus TM is a mono-block type high frequency portable x-ray
unit. The HF120/60HPPWV PowerPlus TM can be used with conventional diagnostic film or with solid-state
digital radiographic sensors. It is supplied with the x-ray tubehead/collimator, detachable AC power cord,
and detachable exposure cord with 2-stage exposure switch. The MinXray HF120/60HPPWV
PowerPlus TM high frequency diagnostic x-ray unit is designed for use with the MinXray model XGS MKIV
LW or equivalent stand capable of providing secure, stable support that allows for proper positioning of the
x-ray unit for radiographic imaging.
INTRODUCTION

A.
B.
C.
D.
E.
DATE OF MANUFACTURE
2
on the rating label affixed to the x-ray unit.
CERTIFICATIONS
THIS PRODUCT CONFORMS TO
ALL APPLICABLE DHHS
RADIATION STANDARDS OF 21
CFR SUBCHAPTER J AS OF THE
Equipment which has been modified or tampered with in any way.
Equipment that has been connected to input voltage other than the voltage specified
Overexposure due to poor operating techniques or procedures.
Equipment not properly serviced or maintained in accordance with this manual.
Injury or danger to patient or other personnel from x-ray exposure.
Although this equipment incorporates protective design features for limiting both the direct (primary)
x-ray beam and the secondary radiation produced by this beam, design factors alone cannot prevent
human carelessness, negligence, or lack of knowledge. This apparatus is sold with the
understanding that the user assumes sole responsibility for radiation safety and that MinXray, Inc., its
agents and representatives and do not accept responsibility for:

1.
a.
b.
2.
a.
b.
3.
A
ssembler's Report
All
assem
bl
ers w
h
o
i
ns
t
a
ll
cer
tifi
e
d
componen
t
ss
h
ou
ld
fil
e a repor
t
o
f
suc
h
assem
bl
y.
All
assembler's reports shall be on Form FD-2579, which is prescribed by and available from
the Director, FDA/Department of Health and Human Services, Division of Compliance,
1390 Piccard Drive, Rockville, Maryland 20850. The original of Form FD-2579 shall be
sent to the Director and copies to the purchaser, State Agency responsible for radiation
p
rotection
,
and one ke
p
t b
y
the assembler for a
p
eriod of least 5
y
ears.
3
The deale
r
o
r
distributo
r
elects to hold and preserve such information, and to immediately
furnish it to MinXray, Inc. when advised by MinXray or Director, Department of Health and
Human Services, that such information is required for purposes of Section 359 of the
Radiation Control for Health and Safet
y
Act of 1968.
The deale
r
o
r
distributor, upon making the election unde
r
"a" above of this section, promptly
notifies MinXray and the Center for Devices and Radiological Health of such election. Such
notification shall be in writing and shall identify the dealer or distributor and the type of
e
q
ui
p
ment for which the information is bein
g
accumulated.
Records to be furnished to MinXray, Inc. by Dealers and Distributors
The information required in "1" above shall immediately be forwarded to MinXray unless:
2) Identification and brand name of the product.
3) Model number and serial or other identification number of the product.
4) Date of sale, award or lease.
Such information shall include:
1) The name and mailing address of distributor, dealer, or purchaser to whom the product
was transferred.
Dealer and Distributor Records
Dealers and distributors of x-ray equipment shall obtain and preserve fo
r
a period of five
years from the date of sale, award, or lease of each such product, such information as is
necessar
y
to
p
ermit tracin
g
of s
p
ecific
p
roducts to s
p
ecific
p
urchasers.
RECORD KEEPING REQUIREMENTS

Manufacturer Responsibility - The FDA believes that plant-based manufacturers must
include in their assembly instructions a specific requirement that the assembler perform a
test(s) for the applicable requirements of the FDA performance standard at the time of
installation. A thorough explanation of the equipment required and step-by-step
instructions must be provided by the component or system manufacturer. The instructions
should include a requirement to record key data to demonstrate at a later data that all tests
were performed and that the equipment was left in full compliance with the standard. The
FDA's Department of Health and Human Services will insure that these assembler test
instructions are provided through a close review of the information submitted by
manufacturers in initial, model change and annual reports. Plant-based manufacturers who
do not include a final compliance test in their assembler instructions could be subject to
disapproval of their quality control and testing program.
4
TO: ALL MANUFACTURERS AND ASSEMBLERS OF DIAGNOSTIC X-RAY EQUIPMENT
This lette
r
is intended to establish HHS policy relative to final testing of a newly-assembled
x-ray system or component before release to the user.
A
ssem
bl
e
r
R
espons
ibilit
y-
A
ssem
bl
ers o
f
di
agnos
ti
c x-ray equ
i
pmen
t
mus
t
per
f
orm a
t
es
t
or tests for the applicable requirements of the FDA performance standard at the time of
installation if specified in the assembly instructions provided by the component or system
manufacturer. Assemblers who do not perform and document such final compliance tests
will be considered by the FDA to have issued a false and misleading certification and will,
therefore
,
be sub
j
ect to re
g
ulator
y
action b
y
the A
g
enc
y
.
Should they be any questions concerning this Bureau policy please call X-ray Products
Branch at 301-594-4591.
SUBJECT : Final Testin
g
of Dia
g
nostic X-ra
y
S
y
stems and Com
p
onents Followin
g
Assembl
y

1. X-ray Unit (Model HF120/60HPPWV PowerPlus TM
)
1 set
1 pc.
1 pc.
1 pc.
Portable stand (Model XGS MKIV LW or equivalent)
System Transport Case (Model HF100HSTC)
Carrying case (Model HF120/60HCC)
5
OPTIONAL COMPONENTS
COMPONENTS
2. Power cord
3. Hand-held Exposure Switch and Cord
4. Instruction Manual

DR conecto
r
Main power switch
Exposure switch and cord Power cord
6
Connector for exposure switch
Connector for power cord
Fuse folder
Battery compartment
Collimato
r
A
ngle indicator
Low line voltage indicator
MAIN PARTS OF MinXray HF120/60HPPWV PowerPlus TM
mA test jack

Li
ne
18
29
310
411
5 12 Shift Button
13 Memory Storage Button
6 14 Memory Recall Buttons (M1 to M5)
7 15 Low line voltage indicator
※
Low line voltage indicator illuminates when line voltage is less than
90V during the X-ray exposure.
7
sec Indicator
kV Indicator sec/mAs Adjustment Button (-)
Exposure timer(sec)/mAs
mAs Indicator
Ready Indicator kV Adjustment Button (+)
Error Indicator sec/mAs Adjustment Button (+)
⑭⑬
⑮
X-ray Indicator kV Adjustment Button (-)
⑨⑩
⑫⑪
④⑥
⑧⑦
③
⑤
CONTROL PANEL
①②
Indicator

A
node direction
Main
p
ower of
f
Radiation warning symbol
Main power on
A
ttention, consult ACCOMPANYING DOCUMENTS
8
Warning Statements, Figures and Symbols
DANGER: HIGH VOLTAGE.
NO USER SERVICEABLE
COMPONENTS INSIDE. REFER
SERVICE TO QUALIFIED SERVICE
TECHNICIAN. FOR CONTINUED
PROTECTION AGAINST HAZARDOUS
CONDITION, REPLACE FUSE WITH
SAME TYPE AND RATING ONLY.
WARNING: THIS X-RAY UNIT MAY BE
DANGEROUS TO PATIENT AND
OPERATOR UNLESS SAFE EXPOSURE
FACTORS, OPERATING INSTRUCTIONS
AND MAINTENANCE SCHEDULES ARE
OBSERVED

1.
2.
9
A
ttach the HF120
/
60HPPWV PowerPlus
TM
to a stand designed o
r
suitable
f
o
r
use with this
x-ray unit, according to the instructions provided with the stand. The MinXray XGS series
of gas spring portable mobile stands is designed specifically for use with the
HF120/60HPPWV PowerPlus TM x-ray unit. Once the HF120/60HPPWV PowerPlusTM is
p
ro
p
erl
y
attached to the model XGS stand
,
it can remain attached for use and trans
p
ort.
This equipment was shipped in perfect condition. When the equipment is received, each
shipping container should be carefully examined for any evidence of mishandling during
shipment. Note its condition; if abnormal, carefully unpack all parts and examine for
damage. If any damage is noted, immediately report it to the carrier in the proper manner
by personally calling it to attention by phone where possible, and filing a written report.
Save all packing material for inspection by the carrier. If this equipment arrives with any
damage, it is your responsibility to report to the carrier and file a claim.
Connect the AC power cord to the socket on the back of the HF120/60HPPWV
PowerPlus TM. Note the rated voltage of the HF120/60HPPWV PowerPlus TM before
Connect the exposure cord to the RJ11 telephone-type socket on the back of the
HF120/60HPPWV PowerPlus TM.
A
ll printed matter supplied with the HF120/60HPPWV PowerPlus TM and the XGS series
stand should be saved for reference during installation and operation.
Assembly Instructions
If a stand is purchased with the HF120/60HPPWV PowerPlus TM, such as the MinXray XGS
series of gas spring portable mobile stands, instructions for assembly of the stand and the
attachment of the
HF120/60HPPWV PowerPlus
TM
are included with the stand.
Unpacking
The MinXray HF120/60HPPWV PowerPlus TM consists of an X-ray generator
(tubehead/control),continuously adjustable light beam collimator, mounting trunnion,
exposure cord with 2
-
stage exposure switch, and AC power cord.
MinXra
y
HF120/60HPPWV PowerPlus TM
INSTALLATION INSTRUCTIONS

1.
1.
2.
3.
4.
5.
(1) remove the 4 screws on the side of the top cove
r
of the HF120/60HPPWV PowerPlusTM , and lift
the top off of the body of the unit; (2) replace the M2300 PC board.
10
This measurement requires that an exposure be made. During the
measurement, close the collimator shutters fully to block the primary X-ray.
1-3. Rejection limit:
±10%+1 msec.
If the exposure time value exceeds the rejection limit, M2300 PC board needs replacement:
0.02, 0.04, 0.10, 0.20, 0.50, and 1.00 sec.
1-2. Instruments:
NERO Model 6000M x-ray beam analyzer
Manufactured by Victoreen, Inc. or equivalent
SID: 18 inches; Wheel range: 100 ~ 150 kV; Sensitivity: Low; Phase select: 1Ø;
Measurement mode: STAT; X-ray values: 120 kV, 0.01 sec.
Make exposure and read the measured value of exposure time on the NERO.
Measure value of exposure time at following settings with same procedure: 0.01,
1-1. Test method:
Set HF120/60HPPWV PowerPlus TM exposure timer to 0.01 sec.
Set HF120/60HPPWV PowerPlus TM output to 120 kV.
Set NERO to the following settings.
PRE-OPERATIONAL REQUIRED TESTING
EXPOSURE TIMER
The following test is the responsibility of the dealer or qualified end user technical personnel and must
be conducted without fail. Fill out the checklist for this report when installation of x-ray unit is
complete.
Fig.1

2.
2-1. Test method:
1.
2.
3.
4.
±10%
I
f
the peak tube potential needs adjustment:
(
1
)
remove the 4 screws on the side o
f
the top cove
r
o
f
the HF120/60HPPWV PowerPlusTM, and lift the top off of the body of the unit; (2) adjust the
potentiometer "VR3" on the printed circuit board M9152 located on top of the lower portion of the unit
(see Fig. 2) so that peak tube potential equals the indicated kV value, taking into consideration the
accurac
y
tolerances
;
(
3
)
re
p
lace the to
p
cover and 4 screws.
11
This measurement requires that an exposure be made. During the
measurement, close the collimator shutters fully to block the primary X-ray.
2-2. Instruments:
NERO Model 6000M x-ray beam analyzer
Manufactured by Victoreen, Inc. or equivalent
2-3. Rejection limit:
Measurement mode: SGL; Time delay 0.1 sec; X-ray values: 40 kV, 0.2 sec.
Make exposure and read the measured AVG. kV value on the NERO.
Measure also value of AVG. kV at 40 kV, 50 kV ,60 kV, 70 kV, 80 kV ,90 kV,
100kV, 110kV, 120 kV with same procedure.
PEAK TUBE POTENTIAL
Set NERO to the following settings:
SID: 18 inches; Wheel range: 30 ~ 60 kV; Sensitivity Hi; Phase select: 1Ø;
Set HF120/60HPPWV PowerPlus TM exposure timer to 0.2 sec., and set kV
to 40 kV.

3.
1.
2.
3.
4. Measure also Hi value of ch2 at 60, 70, 80, 90, 100, 120kV with same
p
rocedure.
1.
2.
3.
4. Measure also Hi value of ch2 at 60, 70, 80, 90, 100, 120kV with same
p
rocedure.
±20%
If the tube current needs adjustment: (1) adjust the potentiomete
r
"VR4" on the printed circuit board
M9152 (see Fig. 2) so that tube current equals 10 mA, taking into consideration the accuracy tolerances;
(2) replace the top cover and 4 screws.
12
3-1-3. Instruments:
Instruments: Digital Storage & Analog Oscilloscope 2212
Manufactured by Tektoronix Holland, N.V. or equivalent
3-1-4. Rejection limit:
3-1-2. mA at longer exposure times
Set oscilloscope as the same as above procedure in 3-1-1. 1..
Set HF120/60HPPWV PowerPlus TM to 50 kV, and timer to 0.2 sec.
Make exposure, and measure Hi value of ch 2. 0.5 V is equal to 10 mA.
trigger level: ch1 1.5 V
Open the top cove
r
of the HF120/60HPPWV PowerPlusTM by removing the 4 screws
and lifting the cover off of the unit. Connect ch 1 of oscilloscope to Ep pin of PC
board M9152 and connect ch 2 of oscilloscope to Ip pin of PC board M9152.
Connect GND line of probe to GND pin of PC board M9152.
Set HF120/60HPPWV PowerPlus TM to 50 kV, and timer to 0.05 sec.
Make exposure, and measure Hi value of ch 2. 0.5 V is equal to 10 mA.
3-1. Test method :
3-1-1. mA at shorter exposure times
Set oscilloscope to the following settings:
ch 1:Ep, ch 2:Ip, ch1 V/div: 1V, ch2 V/div: 500 mV, sec/div: 25 msec,
TUBE CURRENT
VR4 VR3
Fig.2

mA Jack
(
Sin
g
le round
p
hone
j
ack, 3.5 mm diameter
)
connection dia
g
ram
1.
2.
3.
4. Measure volta
g
e of mA Jack ‘s out
p
ut.
(
Sin
g
le round
p
hone
j
ack,3.5 mm/dm
)
5. Calculate the mA as follows : 20 mA=1.00 V
Note:
4.
Refer to Advantech collimator instruction manual for test method and ad
j
ustment
p
rocedure.
LIGHT FIELD TO X-RAY FIELD ALIGNMENT
13
3-2. Measure mA by mA jack output:
Close the collimator shutters fully.
Connect voltage measurement equipment (ex. Osilloscope) with mA jack (Single
round
p
hone
j
ack, 3.5mm/dm
)
to socket on the back of the HF120/60HPPWV
This measurement requires that an exposure be made. During the
measurement, close the collimator shutters full
y
to block the
p
rimar
y
X-ra
y
.
Only voltage value can be measured from mA jack output.
mA jack ‘s output can be measured only during x-ray exposure.
Make exposure.

40 kV±10% kV
50 kV±10% kV
60 kV±10% kV
70 kV±10% kV
80 kV±10% kV
90 kV±10% kV
100 kV±10% kV
110 kV±10% kV
120 kV±10% kV
60 / 42 mA±20% at 50kV mA mA
50 / 35 mA±20% at 60kV mA mA
45 / 31.5 mA±20% at 70kV mA mA
38 / 26.6 mA±20% at 80kV mA mA
33 / 23.1 mA±20% at 90kV mA mA
30 / 21 mA±20% at 100kV mA mA
20 / 14 mA±20% at 120kV mA mA
0.01 sec±10%+1 msec sec
0.02 sec±10%+1 msec sec
0.05 sec±10%+1 msec sec
0.1 sec±10%+1 msec sec
0.2 sec±10%+1 msec sec
0.5 sec±10%+1 msec sec
1sec±10%+1 msec sec
Model Accuracy
Dealer name Dealers Phone Number
Dealer Address
Assembler signature DATE
14
Instruments Manufacturer Last calibrated
Used
Light field to x-ray
field alignment See Collimator Operation Manual □ □
GO NO GO
All Mechanical Smooth Movement □ □
Movement GO NO GO
□ □
Exposure time GO NO GO
□ □
Tube Current GO NO GO
Peak Tube Potential GO NO GO
□ □
Line Voltage 4.80% % □ □
Regulation GO NO GO
Line Voltage 100~260V V □ □
GO NO GO
EQUIPMENT LOCATION
TEST
A
CCEPTANCE RESULTS CHECK
DESCRIPTION LIMIT
INSTALLATION CHECK LIST HF120/60HPPWV PowerPlus TM
Fill out this checklist
when installation is
complete.
SERIAL NO.
DATE
MANUFACTURED
INSTALLATION CHECK LIST

1.0
2.0
3.0
4.0
5.0
6.0
7.0
Use the retractable tape measure on the collimator to position the HF120/60HPPWV
PowerPlus TM at the proper source-image distance (SID) for the image to be taken.
Turn on the collimator light by pressing the button on the collimator. The light will remain on for
approximately 30 seconds, and then automatically turn off. The shutters of the collimator can be
adjusted at any time with the 2 knobs on the collimator. The size of radiation field will only be
p
ro
j
ected on the ima
g
e rece
p
tor when the li
g
ht is on.
Position the patient. Adjust the collimator position so the cross hair projected by the collimator
light is aligned with the anatomical point of interest and the center of the image receptor.
Collimate the size of the light field so that it stays within the dimensions of the image receptor.
15
Make sure that the main power switch is OFF. Connect the AC power cord to a grounded
electrical outlet. Check the label on the HF120/60HPPWV PowerPlus TM for the correct AC
input voltage and amperage required.
Turn on the main power switch on the HF120/60HPPWV PowerPlusTM. The kV and sec LED
displays will light and remain on. The HF120/60HPPWV PowerPlusTM has internal circuitry to
automatically compensate for variations in the input voltage to assure consistent kV and mA
Select the kV output by pressing the arrows below the kV display to choose 40 to 120 kV. mA
output is selected automatically based upon the kV output selected.
Set the exposure time for the view to be taken, and the imaging medium used. Press the arrows
below the sec. display to choose the exposure time between 0.01 and 5.0 second exposures.
The display can be changed between mAs and sec. selected by pressing the shift switch (#)
and pressing the down arrow (▽) below the kV display.
OPERATION INSTRUCTIONS
IT IS ASSUMED BY THE DISTRIBUTORS AND MANUFACTURERS OF THIS EQUIPMENT THAT
THE PERSON RESPONSIBLE FOR ITS OPERATION HAS A GENERAL KNOWLEDGE OF THE
USE OF X-RAYS, INCLUDING THE SAFETY PRECAUTIONS THAT MUST BE TAKEN.
Operation

8.0 The exposure switch has 2-stage button.
9.0
10.0
11.0
12.0
close o
p
en
p
ull
※ Error lamp will blink if the exposure switch is released before the set exposure time. Turn off
Main Power Switch and turn power on again to reset the equipment.
16
When you are finished using the HF120/60HPPWV PowerPlusTM,turn off the main power
switch, disconnect the AC power cord and exposure cord, and put the stand with the
HF120/60HPPWV PowerPlus TM attached into its correct configuration for transport.
Five different technique settings can be stored in memory for quick recall on the
HF120/60HPPWV PowerPlus TM. To store a technique setting, press the Memory Storage button
MR then press Memory Recall button M1. This will store the technique setting displayed in
memory position M1. Repeat this process for storing the different technique setting in M2-M5
as desired. Stored technique settings can be easily recalled by simply pressing M1-M5.
Memory Storage use two (2) size AA alkaline batteries inside Battery Compartment, located on
the back side of equipment. Open the Battery Compartment using a coin, turning the
compartment lid counterclockwise 90°. If there are no batteries in the Battery Compartment,
use some adhesive ta
p
e to hel
p
y
ou remove the com
p
artment lid.
Insert two (2) size AA alkaline batteries in the Battery Compartment, with the + end toward
the Battery Compartment lid. Reattach the Battery Compartment lid and lock it with a coin,
turning it clockwise 90°.
The first stage pre-heats the filament of the x-ray tube. When this stage alone is
depressed and held, " READY " indicator will light after 2.5 seconds, indicating the
filament was heated and the unit is read
y
to
g
enerate x-ra
y
s.
The second stage initiates the x-ray exposure for the time set in the sec display.
When the second stage is fully depressed, x-ray will be emitted, the X-RAY indicator
will light, and an audible signal will be heard. PRESS AND HOLD THE FIRST AND
SECOND STAGE BUTTONS TOGETHER UNTIL THE EXPOSURE HAS
TERMINATED. This is a " dead man" exposure switch -- x-ray exposure will
terminated immediatel
y
as a safet
y
feature when the buttons are released.
It is possible to depress the two stage buttons simultaneously. When this is done,
there is a
pp
roximatel
y
a 2.5 second dela
y
before the timed x-ra
y
ex
p
osure be
g
ins.
After each exposure, release the exposure buttons. The HF120/60HPPWV PowerPlusTM is
now ready for the next exposure.

1.0
2.0
3.0
4.0
5.0
1.0
2.0
3.0
4.0
5.0
6.0
7.0
8.0
9.0
10.0
1.0
2.0
3.0
Make sure the audible signal operates during an exposure.
If the ERROR INDICATOR is on, it means the unit has malfunctioned and x-ray cannot be
g
enerated.
If ERROR INDICATOR is lit even after the EXPOSURE SWITCH is released, it means an
unusual situation has occurred. Turn off the MAIN POWER SWITCH, wait 3 minutes, then start
procedures over again.
17
If the ERROR INDICATOR remains on, contact your dealer or MinXray for service.
Check all cords and cord connections.
Check the exposure buttons--the exposure must terminate if the button is released during an
ex
p
osure.
Indicators
Check the X-RAY indicator for proper operation when both exposure buttons are depressed and
held.
Check the kV adjustment for proper operation.
Check the timer for proper operation.
Check collimator shutters for proper operation.
Check collimator light bulb for proper operation.
Make sure the certification and identification labels are in place.
Check for any loose or missing screws.
Check for oil leaks.
Check for any physical damage.
Maintenance
Clean the HF120/60HPPWV PowerPlus TM as needed with a damp sponge or cloth and mild
The following must be checked before each use:
Tubehead
The operator must not stand in the primary x-ray beam.
The operator should wear a monitoring dosimeter while operating this unit. It should be worn
where it is not covered by a lead apron.
X-ray exposure should be kept as short as possible.
The presence of radio frequency (RF) or electromagnetic radiation may adversely affect the
operation of this device. Please keep this equipment away from likely sources of RF radiation.
Safety Precautions
During exposure the operator must stand as far as possible from the patient being x-rayed, and
should wear a lead apron or stand behind a lead shield.
Table of contents
Other MinXray Medical Equipment manuals
Popular Medical Equipment manuals by other brands

sorin
sorin PARADYM RF DR 9550 Implant manual

Smiths Medical
Smiths Medical level 1 HOTLINE HL-90 Operator's manual

Graham Field
Graham Field LUMEX TENDERCLOUD AQTC8035 Installation & operation instructions

Schiller
Schiller FRED easy user guide

Erbe
Erbe 20193-008 Notes on Use

Newtech
Newtech NT1A Operator's manual