
MOLEAER LOTUS
NANOBUBBLE GENERATOR
Owner’s Manual
www.moleaer.com 17
Cleaning and Sanitizing
Nanobubble Generators (NBGs) can become
contaminated after they have been used for some time.
Pollutants such as colloids, biofilms, mineral scale, and
biological matter build up over time. Contaminants can
result in decreased performance. Periodic cleaning is thus
very important and essential to optimal system
performance.
Cleaning of an NBG system when one or all of the following
operating conditions occur:
•The system is unable to reach 80% of specified
nominal gas flow with gas flow valve wide open.
•The normal increase in dissolved oxygen in one
pass has decreased by 25%.
•The pressure of the gas injection required to
maintain gas flow exceeds supply gas pressure.
If cleaning is delayed too long, complete recovery of the unit
may not be possible.
Safety Precautions
Maximum Temperature: 100°C (212°F.)
pH Tolerance Range: 2 to 14.
Each cleaning situation is different; therefore, specific
cleaning recommendations are dependent on the type of
foulant. Consult the general cleaning instructions for
information that is common to all types of cleaning, such as
suggested equipment, pH and temperature limits and
recommended flow rates; then apply the specific
recommendation as needed.
1. When using any chemical, follow accepted safety
practices. Consult the chemical manufacturer for
detailed information about safety, handling and
disposal.
2. When preparing cleaning solutions, ensure that all
chemicals are dissolved and well mixed before
circulating the solutions through the system.
3. It is recommended that the system be flushed with
good-quality water (20°C minimum temperature)
after cleaning. City water quality is recommended.
Flush the bulk of the cleaning solution from the
system before resuming normal operation.
Cleaning chemicals may be present on the treated
water side following cleaning, to avoid
contamination, the treated water can be diverted to
a drain for a few minutes or until the water is clean
when starting up after cleaning.
4. Ensure the system power is disconnected during
cleaning procedures to avoid accidental startup of
the pump or gas production system.
NOTE: The maximum temperature limit during cleaning is
45°C (113°F) for all PVC systems.
NOTE: The minimum and maximum pH limits for all PVC
systems are 1 and 13, respectively.
General Cleaning Procedures
The cleaning procedure of an nanobubble generator
system consists of the following process steps:
1. Turn off the pump and drain any water in the
system. Isolate the Lotus from the tank.
2. Assemble CIP kit. See Figure 10.
3. Prepare the cleaning solution.The chemicals
used for the cleaning process must be dissolved
and mixed before the cleaning fluid is added to the
nanobubble generator. Household Bleach, Lysol,
CLR or other commonly available cleaning
chemicals will remove any biological growth from
the Lotus internal mechanism.
4. Introduce the cleaning solution. Release the
clamp below the check valve, see Figure 11.
Disassemble the tube and the check valve and put
the CIP tube into the tube from the nanobubble
generator, see Figure 12. Slowly inject about 25 ml
of cleaning solution into the nanobubble generator.
5. Soak the cleaning solution. Allow the cleaning
solution to soak the nanobubble generator for a
minimum of 15 minutes.
6. Rinse the system. Using the same syringe, inject
clean water into the nanobubble generator to flush
the CIP solution out. Repeat 3 times. Generally,
after 3 flushes, the remainder of the solution is very
small and can be left inside the system.
7. Restart the system. Reconnect all the tubes and
pipes. Follow the startup process as described in
the start-up section of this manual.