molex Phillips Medisize SynfuGo User manual

1
SYNFUGO
™
Automated Infusion
Pump
Instructions for Use

Table
of Contents
Chapter
1
Important Safety
Information 5
1.1
Important information about
your SYNFUGO
6
1.2
Critical device use
8
Chapter
2
Getting to Know Your SYNFUGO
11
2.1
Introduction
12
2.2
What is included in the box
14
2.3
Features of your SYNFUGO
16
2.4
Cleaning and Storage
18
2.5
Disposing of your SYNFUGO
19
Chapter
3
Charging, Initial Setup, & Settings
21
3.1
Before turning on your
SYNFUGO for the first time
22
3.2
Charging
22
3.3
Initial Setup
26
3.4
Main Menu
27
3.5
General settings
33
3.6
Setting default infusion rate
35
3.7
Infusion mode
37
3.8
Syringe selector
38
2

Chapter
4
Performing an
Infusion
39
4.1
Before you begin
40
4.2
Turn on your SYNFUGO and
select infusion type
41
4.3
Scan medication vials
42
4.4
Prepare medication, syringe
,
and butterfly needle
46
4.5
Load syringe
47
4.6
Priming the system (removing air)
using your SYNFUGO
50
4.7
Confirm that system is primed (air
is removed)
52
4.8
Insert butterfly needle into vein and
check for blood
52
4.9
Check for blood
53
4.10
Start infusion
54
4.11
Adjust infusion rate
56
4.12
Second syringe to infuse?
57
4.13
Infuse with second syringe
58
4.14
Remove final syringe and needle
60
4.15
Turn off your SYNFUGO
60
Chapter
5
Using your SYNFUGO
with Mobile and
W
eb
apps 61
3

Chapter
6
FAQ,
Troubleshooting,
Alerts
and Errors, Labeling, Contact
Info
65
6.1
Frequently asked questions
66
6.2
Troubleshooting
69
6.3
Alert and error screens
74
6.4
Labeling
82
6.5
Important contact information
86
Chapter
7
Technical
Specifications
87
7.1
Environmental
88
7.2
Physical
88
7.3
Product classification
88
7.4
Electrical
88
7.5
Radio transmitter
88
7.6
Infusion flow rates
89
7.7
Scanning RFID
89
7.8
Charging Adapter
89
7.9
Pump identification
89
4

Chapter
1
Important Safety Informa
tion

1.1
Important information
about
your
SYNFUGO
•
Read these instructions entirely before you use your
SYNFUGO infusion pump for the first time, to quickly
resolve any issues using SYNFUGO. Keep these
instructions in a safe and easily accessible place
.
•
Before you use your SYNFUGO for the first time,
make sure you have been trained by your doctor or
healthcare provider
.
•
Only trained users shall use SYNFUGO. If
this is
your first self-administered infusion with your
SYNFUGO, it should be given under the
supervision of a doctor or health care provider
.
Training will guide you using
SYNFUGO safely.
•
You MUST have successfully
perfor
med
manual
intravenous infusions on yourself before using this
infusion
pump
.
Manual infusions is to be performed
in case of malfunction of SYNFUGO.
•
Use only with AFSTYLA or
IDEL
VION
medication
supplied by your doctor, health care provider, or
phar
macy
.
Using other medications can cause
unpredictable risk.
•
Use only syringes and butterfly needle supplied
along with your medication. Other syringes or
butterfly needles may interfere with normal infusion
pump operation and may cause incorrect infusion
rate or
dose
.

•
Refer to the instructions for use of your
AFSTYLA
or IDELVION medication for
directions on how to
prepare it for use with your SYNFUGO infusion
pump
.
•
Check expiration dates on medication and
accessories before using. Do not use the medication
beyond the expiration date printed on the vial labels.
•
Sight or hearing impaired patients should only
use SYNFUGO with the help of a caregiver who
has
been properly trained on
how to use it.
To
safely see and hear information’s or alerts from the
infusion pump.
•
Keep your SYNFUGO pump and all of its
accessories out of reach of children. Cables may
present a strangulation
risk and risk of
suffocation/asphyxiation.
•
Caution: US Federal Law restricts this infusion
pump to be for sale by or on order of a
physician.
•
DO NOT use the SYNFUGO infusion
pump
for
patients under 2 years of
age
.

1.2 Critical
device use
See
the prescribing instructions
accompanying your AFSTYLA
or
IDELVION for direction on
the
reconstitution of your medication.
•
DO NOT use your SYNFUGO if it
does
not operate as described in this instruction
manual. If you require further assistance
please refer to the contact information in
Chapter
6.5.
•
DO NOT use your SYNFUGO if you
experience a problem and cannot
resolve the issue using
Chapter
6.2
T
roubleshooting.
•
DO NOT use your SYNFUGO if your
prepared syringe cannot be inserted
or
removed according to the instructions
for
use. Perform a manual infusion.
•
DO NOT use your SYNFUGO
if it has been
exposed to extreme vibration, pressure,
or
shock (such as being dropped or stepped
on) as it may have
been damaged.
•
DO NOT use your SYNFUGO if you
observe any worn, damaged, or broken
parts, or if you hear any unusual
or
unexpected sounds during use. Inspect
your SYNFUGO regularly for damage and
wear
.
8

•
DO NOT use your SYNFUGO if you are
exposed to environments with excessive
noise and/or distractions that may
impact
your ability to use your
SYNFUGO
.
•
DO NOT use your SYNFUGO in areas
with high oxygen levels, such as when
supplemental oxygen is in use
.
•
DO NOT
allow your SYNFUGO to become
submerged or wet. It is an electronic device
and may be damaged if submerged or
exposed to excessively wet
conditions
.
•
DO NOT
attempt to use your SYNFUGO
while it is charging. Your
SYNFUGO cannot
be used while charging
.
•
DO NOT attempt to disassemble,
tamper
with, or repair your SYNFUGO. It contains
Lithium Ion batteries which
could explode
or lead to severe burns and/or electrical
shock.
•
DO NOT use your SYNFUGO
if it has been
exposed to extreme temperatures. Keep your
SYNFUGO away from heat or fire
.
Keep
your SYNFUGO out of direct sunlight.
•
DO NOT attempt to use any charging
adapter or charging cables other than
those included with your SYNFUGO. Other
adapters
or cables may create problems
and may interfere with normal device
operation.
9

•
DO NOT expose SYNFUGO to
any kind of
MRI, X-ray, CT,
or Ultrasound.
•
DO NOT use your SYNFUGO close
to
or stacked with
other electrical
equipment.
•
Portable RF
communications equipment,
including antennas,
can effect medical
electrical
equipment.
Always keep a
minimum distance of 12 inches (30 cm) to
RF equipment such as mobile phones, WIFI
transceivers,
etc
.
•
The battery in your SYNFUGO is not
replaceable. Your SYNFUGO contains a
non-replaceable battery. There are no user
serviceable parts
.
•
Always scan all vials needed for the infusion
prior to beginning the infusion.
1.3 Intended use
SYNFUGO, is intended for the following use:
1. For use with reconstituted drug (AFSTYLA and IDELVION) in
a user filled syringe.
2. To be used in the intravenous administration of
reconstituted drug solution through controlled infusion
rates.
3. To be used as a portable personal syringe infusion pump in
home environment.
4. Re-usable, personal device for one specific patient to be
used by the adults, adolcents and children.

1.4 Essential performance
-Reverse plunger movement of more than 1mL
SYNFUGO will prevent the plunger being moved more
than 1mL backwards to prevent unintended air in the syringe.
Always check that all air in the syringe is removed,
electromagnetic disturbance (e.g. from mobile phones) may result
in unintended reverse plunger movement.
-Correct scanning of the vial the dose IU
SYNFUGO will display the sum of the scanned IU of the vials.
Electromagnetic disturbance may result in wrong readings,
always make sure the scanned IU is correct.
Chapter
2
Getting to Know
Y
our
SYNFUGO

2.1 Introduction
•
Your SYNFUGO is a personal
automated
infusion pump that allows
hands-free
infusion
with user controlled infusion
speed.
•
Your SYNFUGO is a home use device with
dedicated charging adapter and micro USB
cable, for use by patients or caregivers
.
Only use your SYNFUGO
with
AFSTYLA
or
IDELVION medication
as
supplied by your doctor, health car
e
provider or
pharmacy.
•
Only use your SYNFUGO with
Becton
Dickinson (BD) 5 mL (catalog
# 309646)
and 10mL (catalog
#309604) plastic
syringe. Any changes
to compatible
butterfly needles made after date
of
publication may be confirmed using
CSL Behring SYNFUGO website or by
contacting customer
service
.
•
Only use your SYNFUGO with NIPRO
brand SafeTouch PSV 25G x
3/4” x
350mm
long butterfly needles. Any changes
to
compatible syringes made after date
of publication may be confirmed using
CSL Behring SYNFUGO website or by
contacting customer
service
.
•
Your SYNFUGO can record and store the
patient´s
infusion information and enables
easy patient´s
access to his/her complete
infusion and medication histor
y
.
12

•
Optionally, your infusion
record can be
transferred to mobile or web apps. If you
choose, you may also share your infusion
history with your healthcare provider
-
see
Chapter
5.
•
If connected with the SYNFUGO mobile
app, the date and time will automatically
be
set.(Chapter
5
)
•
SYNFUGO is for patients greater than
2 years of age and diagnosed with
Hemophilia A or B, who are trained in
manual infusions
.
•
For patients greater than 2 and less than
7 years of age, a caregiver must operate
SYNFUGO and perform the infusion.
•
For patients from 7 to 11 years old a
caregiver must be available to super
vise
or assist (as
needed) with the use of
SYNFUGO
.
•
Patients
12 years of age and older, who
can independently perform a manual
infusion, can self-infuse using SYNFUGO
if trained.
•
Patients under the age of 2 must not use
the SYNFUGO infusion
pump
.
•
Applied parts (part that comes in contact to
the skin) is the Butterfly Needle
.
13

2.2
What
is
included
in
thebox
Before
starting, check that all items
listed are
present.
Your SYNFUGO infusion pump kit contains
the following items:
A SYNFUGO infusion
pump
B Charging adapter
C Charging cable
1m
D Quick Start Guide
E
Instructions For Use (IFU)
F
Storage case (not shown)
Your medication, syringe(s)
and
butterfly needle(s)
ARE NO
T
INCLUDED in this
kit
and
will be
supplied separately from your
Docto
r
,
health care provider, or
pharmacy.
Note: The optional software for your
SYNFUGO (mobile and web apps) are
not included. See
Chapter 5 for more
infor
mation.
14

A
B
C
D
E
15

2.3
Features
of your
SYNFUGO
Study the features listed below before
attempting to use your SYNFUGO. If
you have any questions regarding these
features, consult with your doctor, health
care provider, or pharmacist
first.
You may
also contact one of our technical specialists
for assistance (See
Chapter 6.5 for
additional infor
mation).
A Charging Port:
Receives the Charging
Cable
.
B Display:
Displays all
device settings and
infor
mation.
C Power Button
:
Press to turn your
SYNFUGO’s power on and
of
f
.
D Back Button
:
Use
to step back to
previous selection.
E
Navigation Buttons
:
Use to
navigate
up or down through the display
menu.
F
Ok/Confirm
Button :
Use to confirm
a
desired input selection.
G Stop/Emergency Stop Button
:
Use
to
immediately stop an infusion.
H Grey Tab: Manually move plunger
to
Remove air
or check for blood.
I
Lid Release:
Use to unlock and open lid
and access syringe
.
J
Vial Scanner Target:
Use to scan and
record vial infor
mation.
16

K Syringe Cradle:
Receives and retains
your 5mL or 10mL
syringe
.
L
Syringe Selector:
Use to select a
5mL
syringe or 10mL syringe
size
.
M
Syringe Window:
Use to visually
monitor
the syringe
volume
.
J
A
C
D
B E
I F
H G
M
K
L
17

2.4
Cleaning and
storage
•
Always keep your SYNFUGO clean and
free of any dirt and debris
.
•
If medicine or other fluid is spilled or leaked
onto your SYNFUGO, be sure to clean it
off
immediately
.
•
To clean your SYNFUGO, wipe carefully
with a dry or slightly
damp cloth and mild
detergent.
•
Alcohol wipes (70% isopropyl) may be
used to clean your device
.
•
DO NOT place or submerge your
SYNFUGO in any liquid, and do not
sterilize
.
•
DO NOT
attempt to clean your SYNFUGO
in a
dishwasher
.
•
Always store your SYNFUGO with the lid
closed and in its storage case to protect it
from dust, dirt, or direct sunlight.
•
Keep SYNFUGO away from heat, fire, and
extreme temperatures
.
•
Keep your SYNFUGO out of direct sunlight.
•
Keep SYNFUGO and all of its accessories
out of reach of children. Cables may
present a strangulation
risk.
18

2.5
Disposing
of your
SYNFUGO
•
Your SYNFUGO contains Lithium Ion
batteries. Dispose of
this device according
to your local environmental laws and
guidelines
.
•
DO NOT dispose of in fire. Your
SYNFUGO
contains Lithium Ion batteries which
could
explode when exposed to fire or extreme
heat.
•
DO NOT attempt to disassemble your
SYNFUGO
.
•
Used butterfly needle and syringe must
be disposed of with due consideration
to National and Local safety and
environmental regulations
.
•
Always dispose of your empty syringe and
used butterfly needle immediately after you
completed your infusion into an appropriate
sharps container
.
•
Your used SYNFUGO device, charging
adapter and micro USB cable are to be
disposed of safely according to WEEE
regulation (requirement in Europe).
19

•
To protect natural resources and
to promote
material reuse, please separate
packaging
material from other types of waste and
recycle the material through your local
recycling system.
20
Table of contents
Popular Water Pump manuals by other brands

DAB PUMPS
DAB PUMPS ESYBOX DIVER 55/120 Instruction for installation and maintenance

Pentair
Pentair myers aplex Series Installation and service manual

Johnson Pump
Johnson Pump MCHZ instruction manual

TapFlo
TapFlo SPT Series Original instruction

hillvert
hillvert HT-ROBSON-SP1100-93 user manual

Pfeiffer Vacuum
Pfeiffer Vacuum HIPACE 80 MINI operating instructions