MSA S5000 User manual

Operating Manual
General Monitors S5000
Gas Monitor
Order No.: MANS5000/07
Print Spec: 10000005389 (EO)
CR: 800000053913
MSAsafety.com

© MSA 2020, 2022. All rights reserved
WARNING!
Read this manual carefully before using or maintaining the device. The device will perform as designed only if it is used and
maintained in accordance with the manufacturer's instructions. Otherwise, it could fail to perform as designed, and persons
who rely on this device could sustain serious injury or death.
The warranties made by MSA with respect to the product are voided if the product is not installed and used in accordance with
the instructions in this manual. Please protect yourself and your employees by following the instructions.
Please read and observe the WARNINGS and CAUTIONS inside. For additional information relative to use or repair, call
1-800-MSA-2222 during regular working hours.
For countries of Russian Federation, Republic of Kazakhstan and Republic of Belarus, the gas detector will be delivered with a
passport document that includes valid approval information. On the CD with manual instruction attached to the gas detector
the user will find the documents "Type Description" and "Test Method" - appendixes to Pattern Approval Certificate of
Measuring instrument, valid in the countries of use.
MSA is a registered trademark of MSA Technology, LLC in the US, Europe and other Countries. For all other trademarks visit
https://us.msasafety.com/Trademarks.
This product incorporates Bluetooth® wireless technology. The Bluetooth word mark and logos are registered trademarks
owned by Bluetooth SIG, Inc., and any use of such marks by MSA is under license. Other trademarks and trade names are
those of their respective owners.
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may
not cause harmful interference, and (2) this device must accept any interference received, including interference that may
cause undesired operation.
You are cautioned that changes or modifications not expressly approved by the party responsible for compliance could void the
user's authority to operate the equipment.
English:
This device complies with RSS-210 of the Industry Canada Rules. Operation is subject to the following two conditions: (1)
This device may not cause harmful interference, and (2) this device must accept any interference received, including
interference that may cause undesired operation.
French:
Ce dispositif est conforme à la norme CNR-210 d'Industrie Canada applicable aux appareils radio exempts de licence. Son
fonctionnement est sujet aux deux conditions suivantes: (1) le dispositif ne doit pas produire de brouillage préjudiciable, et (2)
ce dispositif doit accepter tout brouillage reçu, y compris un brouillage susceptible de provoquer un fonctionnement
indésirable.
General Monitors
16782 Von Karman Avenue, Unit 14
Irvine, CA 92606
1-949-581-4464
For your local MSA contacts, please go to our website www.MSAsafety.com

Contents
1 Safety Regulations 5
1.1 Correct Use 5
1.2 Product Warranty 6
2 Description 8
2.1 Display 8
2.2 No Tool Interface 8
2.3 Bluetooth® Wireless Technology 8
2.4 Dual Sensing 9
2.5 Retrofit Installations 9
2.6 XCell Sensors Optimized for Fixed Gas Applications 9
2.7 TruCal Sensing Technology for CO and H2S Electrochemical Sensors 10
2.8 SafeSwap 10
2.9 Housing 11
2.10 Component Overview 11
2.11 Label Overview 12
3 Installation 14
3.1 Installation Warnings - Read Before Installation 14
3.2 Reviewing Shipment and Identifying Product Model 15
3.3 Product Installation Check List 15
3.4 Mounting 15
3.5 Installing a Remote Sensor Junction Box 22
3.6 Electrical Power Connections 23
4 Operation 34
4.1 Startup 34
4.2 Settings 35
4.3 Info Menu - Viewing Device Status 58
4.4 Setting only configurable via Bluetooth®, Modbus, or HART 60
5 Calibration 63
5.1 Calibration Equipment 63
5.2 Calibration Frequency 66
5.3 Calibration Frequency for XCell Sensors with TruCal (H2S & CO only) 66
5.4 Calibration Types: Zero vs. Span 67
5.5 How to Zero Calibrate XCell and IR Sensors 68
5.6 How to Calibrate XCell Sensors 69
5.7 How to Calibrate an Oxygen XCell Sensor 70
5.8 How to Calibrate an IR Sensor 71
5.9 How to Calibrate a Passive Sensor (Catalytic Bead or MOS) 71
6 Maintenance 73
6.1 IR Sensor Cleaning Procedure 73
6.2 Replacing an XCell Sensor 74
6.3 Clearing a Blockage 75
6.4 Replacing a Passive Sensor (Cat Bead or MOS) 77
3
US

1 Safety Regulations
1.1 Correct Use
The S5000 Gas Monitor, hereafter also called device, is a gas monitor for measuring toxic and combustible gases as well as
oxygen. Using sensors, the device tests the ambient air and triggers the alarm as soon as the gas exceeds a specific
concentration level.
WARNING!
Do not use silicone-type lubricants in assembling the device and do not allow silicone vapors to be drawn into the flow system
while in operation. Silicone can desensitize the combustible gas sensor, thereby giving erroneously low readings.
Use only genuine MSA replacement parts when performing any maintenance procedures provided in this manual. Failure to do
so may seriously impair sensor and gas monitoring performance, alter flameproof/explosionproof characteristics or void
agency approvals.
Failure to follow these warnings can result in serious personal injury or death.
WARNING!
As with all gas monitors of this type, high levels of, or long exposure to, certain compounds in the tested atmosphere could
contaminate the sensors. In atmospheres where an S5000 Gas Monitor may be exposed to such materials, calibration must be
performed frequently for dependable operation and to confirm that display indications are accurate.
These effects include, but are not limited to:
• Passive MOS sensors may be adversely affected by prolonged exposure to certain substances. Loss of sensitivity or
corrosion may be gradual if such agents are present in low concentrations or it may be rapid at high concentrations.
Examples of these substances are as follows:
○ Halides: compounds containing fluorine, chlorine, bromine and iodine
○ Heavy metals, e.g. tetraethyl lead
○ Caustic and Acidic liquids and vapors
○ Glycol
• The H2S Digital Sensor may be adversely affected by the following substances:
○ Alcohols (methanol, ethanol, isopropanol)
○ Nitrogen dioxide (NO2)
○ Chlorine (Cl2)
○ Paint solvents (acetone, turpentine, toluene, mineral spirits, etc.)
• The CO Digital Sensor may be adversely affected by the following substances:
○ Alcohols (methanol, ethanol, isopropanol)
○ Paint solvents (acetone, turpentine, toluene, mineral spirits, etc.)
•The O2Digital Sensor may be adversely affected by the following substances:
○ Long term exposure to low levels of Acetylene
○ Paint solvents (acetone, turpentine, toluene, mineral spirits, etc.) in high concentrations larger than 1000 ppm or
prolonged exposure to lower concentrations
• Prolonged exposure of the H2S Digital Sensor to humidity levels of 5% RH or lower will result in gas measurement
readings of H2S that are greater than the actual gas concentration present.
5
1 Safety Regulations US

• It is not recommended to expose the O2, H2S and CO Digital Sensors to environments containing oxygen levels above
30% (v/v) or below 5% (v/v). The O2sensor will operate at concentrations below 5% (v/v) but not for prolonged periods of
time.
Failure to follow these warnings can result in serious personal injury or death.
This device complies with Part 15 of the FCC Rules. Operation is subject to the -following two conditions:
• this device may not cause harmful interference, and
• this device must accept any interference received, including interference that may cause undesired
operation.
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part
15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference
when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate
radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause
harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause
harmful interference in which case the user will be required to correct the interference at his own expense.
WARNING!
The Digital Sensor XCell sensor module utilizes thread locker suitable to an ambient temperature of -55°C to +74°C. If the
Digital Sensor is exposed to temperatures outside of its listed ratings, reapplication of thread locker when changing out the
XCell sensor module may be required.
Failure to follow this warning can result in serious personal injury or death.
NOTICE
This is a Class A product in accordance with CISPR 22. In a domestic environment, this product may cause radio
interference, in which case the user may be required to take adequate -measures.
NOTICE
The XCell sensor refers to the sensor portion of the digital sensor throughout this manual.
FCC Warning Statements
Changes or modifications not expressly approved by the manufacturer could void the user's -authority to operate the
equipment.
Industry Canada (IC) Warning Statements
The installer of this radio equipment must ensure that the antenna is located or pointed such that it does not emit RF field in
excess of Health Canada limits for the general population; consult -Safety Code 6, obtainable from Health Canada's website
www.hc-sc.gc.ca.
1.2 Product Warranty
The warranties made by GM with respect to the product are voided if the product is not installed, used and serviced in
accordance with the instructions in this manual. Please protect yourself and your employees by following the instructions.
6
US 1 Safety Regulations

ITEM WARRANTY PERIOD
S5000 Gas Monitor MSA warrants that this product will be free from mechanical defects and faulty workmanship for the
period specified in this table for each component, provided it is maintained and used in accordance
with MSA’s instructions and/or recommendations. Warranty shall not exceed.
Main Transmitter
Housing and PCBA
2 years from date of shipment.
Shall not exceed 2 years and 6 months from date of manufacture.
XCell Sensors 3 years from date of shipment.
Shall not exceed 3 years and 6 months from date of manufacture.
IR Sensor 2 years from date of shipment.
Shall not exceed 2 years and 6 months from date of manufacture.
Passive Catalytic
Bead
2 years from date of shipment.
Shall not exceed 2 years and 6 months from date of manufacture.
Passive MOS 2 years from date of shipment.
Shall not exceed 2 years and 6 months from date of manufacture.
Electrochemical
Sensors
12 months from date of shipment. Shall not exceed 18 months from date of manufacture.
This warranty does not cover filters, fuses, etc. Certain other accessories not specifically listed here may have different
warranty periods. This warranty is valid only if the product is maintained and used in accordance with Seller's instructions
and/or recommendations. The Seller shall be released from all obligations under this warranty in the event repairs or
modifications are made by persons other than its own or authorized service personnel or if the warranty claim results from
physical abuse or misuse of the product. No agent, employee or representative of the Seller has any authority to bind the Seller
to any affirmation, representation or warranty concerning this -product. Seller makes no warranty concerning components or
accessories not manufactured by the Seller, but will pass on to the Purchaser all warranties of manufacturers of such
components.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED, IMPLIED OR STATUTORY, AND IS
STRICTLY LIMITED TO THE TERMS HEREOF. SELLER SPECIFICALLY DISCLAIMS ANY WARRANTY OF
MERCHANTABILITY OR OF FITNESS FOR A PARTICULAR PURPOSE.
Exclusive Remedy
It is expressly agreed that Purchaser's sole and exclusive remedy for breach of the above warranty, for any tortious conduct of
Seller, or for any other cause of action, shall be the replacement at Seller's option, of any equipment or parts thereof, which
after examination by Seller is proven to be defective. Replacement equipment and/or parts will be provided at no cost to
Purchaser, F.O.B. Seller's Plant. Failure of Seller to successfully replace any nonconforming equipment or parts shall not
cause the remedy established hereby to fail of its essential purpose.
Exclusion of Consequential Damage
Purchaser specifically understands and agrees that under no circumstances will seller be liable to purchaser for economic,
special, incidental or consequential damages or losses of any kind whatsoever, including but not limited to, loss of anticipated
profits and any other loss caused by reason of non-operation of the goods. This exclusion is applicable to claims for breach of
warranty, tortious conduct or any other cause of action against seller.
7
1 Safety Regulations US

2 Description
2.1 Display
The S5000 utilizes a dot matrix LED display, capable of displaying four alphanumeric characters at a time. The display will
scroll words that exceed four letters. Most of the messages scroll twice across the screen before moving onto next selection.
Figure 1 S5000 Main Display
In addition to the red LED display, the S5000 uses six icons to indicate status. Green LED indicates power supply status. A
yellow triangle and red bell indicate fault and warning or alarm conditions respectively. The Bluetooth icon indicates that the
Bluetooth wireless technology is enabled on the device. Yellow "1" and "2" icon indicate which sensor gas reading is being
displayed, or during configuration which sensor's options are being accessed.
2.2 No Tool Interface
The S5000 does not require any tools or third party devices to change settings, reset alarms or perform any maintenance
operation. The EZtouchbutton works through the glass and does not require opening the explosion proof enclosure. The
EZtouchbutton works with bare fingers or with gloved hands, so long as the gloves are not black. See 4 Operation for more
information on navigating the menu with the EZtouchbutton.
Figure 2 Interface
The user menu can also be accessed using the round GM magnet on the General Monitors logo.
2.3 Bluetooth®Wireless Technology
The S5000 comes by default with Bluetooth communication. Using the X/SConnectApp on an appropriate smart phone or
tablet, users are able to interface with the S5000 menu options in a larger and more user friendly setting. The Bluetooth
8
2 Description US

communication can interact with the device within a maximum transmission distance of 70feet (21m).
WARNING!
Bluetooth operation is dependent upon signal availability of the wireless service(s) necessary to maintain the communication
link. Loss of wireless signal will prevent communication of alarms and other information to linked devices. Take appropriate
precautions in the event a loss of wireless signal occurs.
Failure to follow this warning can result in serious personal injury or death.
If ordered with Bluetooth, the device will be shipped with Bluetooth enabled. See 4 Operation for instructions on
disabling Bluetooth.
If the device was not ordered with Bluetooth it cannot be upgraded.
The S5000 and user provided communication device will need to be paired. This requires both devices to be in range and for a
pairing sequence inputting a 6digitpairing code. The instructions will be displayed on both the S5000 and communication
device.
There are communication devices capable of being used in classified areas. Please contact your MSA
representative for additional information.
2.4 Dual Sensing
The S5000 supports two digital sensors, or one IRsensor point IR detector and one digital sensor simultaneously with four
wire connections. However, the device will only support one passive sensor, either combustible catalytic bead or metal oxide
semiconductor (MOS) sensors, based on the ATO configuration. Passive catalytic bead uses three wires, passive MOS uses
four wires.
The S5000 Gas Monitor generates two independent analog outputs; one for each sensor connected to the transmitter. The
analog output associated with Sensor1 also has the digital HART (Highway Addressable Remote Transducer) communication
superimposed on the analog signal. If two sensors are connected, the digital HART communication carries information for both
sensors.
2.5 Retrofit Installations
The S5000 has the conduit entries located in the exact same orientation and distance from the wall and the mounting holes for
attaching to a wall are identical to the S4000Series detectors. Users will be able to re-use the existing wiring for the remotely
mounted sensors.
2.6 XCell Sensors Optimized for Fixed Gas Applications
XCell toxic and combustible catalytic bead sensors are developed and manufactured by MSA. Now optimized for fixed gas
applications, the XCell sensor platform is available in the S5000 and provides multiple benefits, including a standard 3-year
warranty on all XCellsensors.
One important optimization for fixed gas was incorporating the GM catalytic bead into the XCell sensor.
The XCell oxygen sensor does not use lead, but rather a non-consuming reaction chemistry. The XCell oxygen sensor is
expected to last well over 3years and can be safely stored on the shelf for at least 1 year without sensor performance
degradation. Changes in barometric pressure across the range of 86 kPA - 108 kPA have a negligible effect on the operation of
the sensor.
9
US 2 Description

2.7 TruCal Sensing Technology for CO and H2S Electrochemical Sensors
Using patented pulse check technology and proprietary Adaptive Environmental Compensation (AEC) algorithms, all XCell
sensors with TruCal verify operation by actively adjusting the sensor output for changes in sensitivity. Some XCell sensors
with TruCal also include Diffusion Supervision, which monitors the sensor inlet for obstructions that could prevent gas from
reaching the sensor.
Every six hours, an electrical pulse stimulates the XCell sensor similar to having actual calibration gas applied, providing a
snapshot of the sensor's sensitivity at the time of the pulse. Using this sensitivity snapshot, the sensor can diagnose sensor
failures like electrode poisoning, electrolyte leaking, or electrical connectivity issues.
AEC uses the sensitivity snapshots provided by the pulse check to adjust sensor output, compensating for environmental
impacts on sensor accuracy. If the AEC adjustment is greater than expected based on typical environmental impact
variations, the transmitter LED’s will slowly flash GREEN, alerting users that the sensor should be calibrated to reset the AEC
cycle. Users can also enable a Calibration Alert function that will send a milli-amp signal on the analog output to the control
room. The result is a sensor that actively self-monitors for operation and accuracy, with far fewer manual calibrations.
Diffusion supervision actively monitors the sensor inlet for obstructions. If an obstruction is detected, the sensor will go into a
fault mode to alert users and the control room that it is not seeing gas due to an obstruction. Objects residing directly on or in
the sensor inlet that result in a significant impact to the gas path are very likely to be detected by Diffusion Supervision.
Examples include paint, tape, water, and dirt. Small amounts of these materials can be visible on the inlet while not impacting
the gas path enough to trigger a Diffusion Supervision Fault. A fault signal will only be sent out when the system determines
that the amount of material that has accumulated on or inside the sensor inlet is negatively affecting the gas path.
Actual TruCal sensor performance will depend on the application, background gas exposure, and environment. To validate
XCell sensors with TruCal, it is recommended that users follow their regular calibration cycle and record the "as found" and "as
left" values. This data can be used to extend the time between calibrations depending on the required specification of the
application.
2.8 SafeSwap
The S5000 comes with patented SafeSwap technology, which allows users to change or replace XCellsensors without
needing to power down the instrument. Swap delay is enabled on the S5000 by default; a feature that gives users a 2minute
window to change sensors without triggering a fault condition. SafeSwap and Swap Delay are only applicable for
XCellsensors. For more information on SafeSwap and Swap Delay, see 6 Maintenance.
WARNING!
• As part of the product certification, it was verified that optional communication functions of this gas detection instrument
while operating at the maximum transaction rate do not adversely affect the gas detection operation and functions of the
instrument. The product certification, however, does not include or imply approval of the SafeSwap feature,
communications protocol or functions provided by the software of this instrument or of the communications apparatus
and software connected to this instrument.
• Follow the warnings below when removing or replacing sensors. Reference Figure 3 Exploded View for component
overview.
○ Never remove or replace a sensor body assembly or an IRSensor while under power or when explosive hazards
are present.
○ Confirm that the area is free of explosive hazards before removing or replacing an XCell Sensor under power.
○ To remove an XCell Sensor, unscrew XCell Sensor three full turns, wait 10 seconds, and then remove the XCell
Sensor completely.
Failure to follow these warnings can result in serious personal injury or death.
10
2 Description US

2.9 Housing
The S5000 comes in 316StainlessSteel for the highest corrosion resistance. All housings have ¾”NPT conduit entries.
Custom tags are available and easily attach to an integral ring. The JB5000 junction box comes in 316 Stainless Steel for the
highest corrosion resistance. The housing is offered in ¾" NPTof M25 conduit entries.
2.10 Component Overview
Figure 3 Exploded View
11
US 2 Description

2.11 Label Overview
Figure 4 Transmitter - Position of Labels Figure 5 Board stack - Position of Labels
Figure 6 Digital Sensor - Position of Labels
12
2 Description US

Figure 7 IRSensor - Position of Labels
Figure 8 Passive Sensor - Position of Labels
13
US 2 Description

3 Installation
3.1 Installation Warnings - Read Before Installation
WARNING!
• Refer to 9 Appendix: General Certification Information before installation and operation.
• When utilizing the P+F Wireless HART Bullet accessory, read and understand the Quick Start guides (P/N 10218978 &
10218979) and P+F Wireless HART Bullet instruction manual (WHA-BLT-F9D0-N-A0-*) before installing and using. The
P+F Wireless HART Bullet shall be installed as close to the S5000 transmitter, Junction Box or JB5000 as possible.
• Some digital sensors are provided in a fritless sensor housing. The fritless sensor housing is labeled as Div 2 or Zone 2
and is approved for Div 2 or Zone 2 installations only. The protection method is Nonincendive or Type n respectively.
Ensure all components are approved for the wiring method being used and in accordance with the National Electric Code
of the country of use, any applicable local regulations, and this manual.
• As part of the product certification, it was verified that optional communication functions of this gas detection instrument
while operating at the maximum transaction rate do not adversely affect the gas detection operation and functions of the
instrument. The product certification, however, does not include or imply approval of the SafeSwap feature,
communications protocol or functions provided by the software of this instrument or of the communications apparatus
and software connected to this instrument.
• Follow the warnings below when removing or replacing sensors. Reference Figure 3 Exploded View for component
overview.
○ Never remove or replace a sensor body assembly or an IRSensor while under power or when explosive hazards
are present.
○ Confirm that the area is free of explosive hazards before removing or replacing an XCell Sensor under power.
○ To remove an XCell Sensor, unscrew XCell Sensor three full turns, wait 10 seconds, and then remove the XCell
Sensor completely.
• Exposure to continuous levels of HCN gas affects the sensitivity of the HCN sensor and can result in erroneously low
gas readings. This can occur in as little as 30 minutes of exposure. Do not operate in the presence of continuous low
level background HCN gas. Use latching alarms to avoid an erroneous alarm clearance. Sensor sensitivity will recover
over the course of several hours once HCN gas is removed. Use an independent instrument to confirm the absence of
HCN gas and allow the sensor to return to zero prior to restarting operations.
• Plug all unused conduit entries with a suitably certified blanking/stopping plug.
• Do not paint the device. Avoid painting in areas where the S5000 and remote sensor junction box are located. If painting
is required in an area where an S5000 or remote sensor has been installed, exercise caution to ensure paint is not
deposited on the sensor inlet fitting. Paint solvents can also cause an alarm condition to occur or potentially poison
electrochemical sensors.
• Protect the device from extreme vibration.
• Do not mount the sensing head in direct sunlight without a sunshield (P/N 10180254).
• IR Sensors contain no user- or field-serviceable parts and must be returned to the factory for repair. Any attempt to open
the monitor will damage the unit and void the warranty.
Failure to follow these warnings can result in serious personal injury or death.
NOTICE
When installing the IRSensor, under no circumstances should a pry-bar be applied to the two legs that support the unit's
reflectors during installation or removal of the sensor. Applying force to the legs can permanently damage the IRSensor.
14
3 Installation US

3.2 Reviewing Shipment and Identifying Product Model
To determine the sensor type and options, check the shipping carton.
Figure 9 Shipping Label
Passive, Digital, and IR sensors are shipped attached to the S5000. Passive and IR Sensors are one-piece sensors. Digital
sensors comprise two parts: the sensor body assembly and the XCell Sensor. Sensor body assemblies must be installed and
tightened using a strap wrench.
Check the sensor details before attaching to the S5000 housing. The sensor details are listed on the inside of the XCell
Sensor. Unscrew the XCell Sensor from the sensor body assembly and check the label on the inside for gas type, range,
configuration ordered, serial number, and firmware revision number.
3.3 Product Installation Check List
Before Installation
• Review national electrical codes
• Review local procedural and building codes
• Determine optimum transmitter placement
• Determine wire requirements
• Determine mounting hardware requirements
• Review approvals and ensure suitability for installation
Mounting
• Attach appropriate sensor to housing or junction box (see 3.4.3 Sensor Orientation for proper sensor orientation)
• Mount transmitter or junction box using appropriate mounting hardware
• Confirm free air flow around the sensor
3.4 Mounting
WARNING!
Refer to 9 Appendix: General Certification Information before installation.
15
US 3 Installation

Some toxic gases are provided in a fritless sensor housing. The fritless sensor housing is labeled as Div 2 or Zone 2 and is
approved for Div 2 or Zone 2 installations only. The protection method is Nonincendive or Type n respectively. Ensure all
components are approved for the wiring method being used and in accordance with the National Electrical Code of the country
of use, any applicable local regulations, and this manual.
Failure to follow this warning can result in serious personal injury or death.
3.4.1 Sensor Mounting Location
The best location for the transmitter and the sensor may not be the same location. Sensors should be placed in a location
where a gas leak is most likely to be detected. When the best sensor placement would not allow the transmitter display to be
easily viewed or accessed, a remote junction box can be used to mount the sensor remotely from the transmitter, allowing both
to be installed in the optimum location.
Two main factors should be considered when choosing a sensor location. The first is the density of the target gas relative to
the air. Gases, such as propane, that are heavier than air should be placed near ground level while gases that are lighter than
air should be placed above potential leak sources.
Optimum sensor placement will depend on the surrounding processing equipment, such as pipes, valves, or turbines. MSA
offers a gas and flame mapping service that systematically evaluates potential sources of leaks and recommends detector
quantity and placement to create the most effective detection system.
3.4.2 Transmitter Mounting Location
The transmitter display should be mounted so that the screen is visible and easily accessed after installation. The electronics
assembly inside the enclosure have one orientation inside the cylindrical housing. Take care to position the conduit entries and
display so that the display is read in the correct orientation.
3.4.3 Sensor Orientation
WARNING!
Mount the IR Sensor with the sensor inlet fitting extended horizontally from the main enclosure (see Figure 10 Correct and
Incorrect Mounting Orientations for IR Sensor) to prevent the build-up of particulate or liquid matter on the monitor's optical
surfaces.
Mount the digital sensor with the sensor inlet fitting (see Figure 11 Correct and Incorrect Mounting Orientation for Digital
Sensors) pointed downward; otherwise, the inlet may become clogged with particulate matter or liquids.
Failure to follow this warning can result in serious personal injury or death.
Sensor orientation will depend on the sensor type. If mounting an IRSensor, whether locally on the transmitter or via remote
junction box, the sensor should be mounted horizontally. If the IRSensor is not mounted horizontally, the sensor will be prone
to more frequent beam blocking issues due to accumulated dust and condensation on the surface of the IRSensor. Figure 10
Correct and Incorrect Mounting Orientations for IR Sensor shows the correct and incorrect mounting orientations for the
IRSensor.
16
3 Installation US

Figure 10 Correct and Incorrect Mounting Orientations for IR Sensor
All other sensors, including electrochemical, catalytic bead, oxygen, passive catalytic bead, and passive MOS should be
mounted vertically with the gas inlet pointed downward. If the sensor is not mounted with the gas inlet facing down, it is more
likely to become clogged with particulate matter or liquids. Figure 11 Correct and Incorrect Mounting Orientation for Digital
Sensors shows the correct and incorrect mounting orientation for digital sensors. Passive catalytic bead and MOS sensors
come already installed on the transmitter housing.
Figure 11 Correct and Incorrect Mounting Orientation for Digital Sensors
3.4.4 Connecting Sensor to Transmitter Housing or Remote Junction Box
Digital and IRSensors are shipped attached to the main enclosure. The main sensor input is provided via a four-terminal
connection that provides a digital interface for all sensor modules. Up to two sensors (excluding passive sensors) can be
connected to a single transmitter with two analog outputs capable of representing the readings of the individual sensors.
Passive sensors are shipped already attached and electrically wired to the device. Only one passive sensor can be used on a
single S5000, and they are not interchangeable with other passive sensors or digital sensors.
Consider the sensor dimensions when choosing a mounting location for the transmitter or junction box.
To connect the sensor:
1. Loosen the set screw located on the lid using a 1.5mm Allen wrench.
2. Turn the lid counter-clockwise to remove.
3. Pull out display module to expose terminal connections.
17
US 3 Installation

Figure 12 Terminal Connections
4. Route the cable from the sensor through a conduit entry hole in the enclosure so that the sensor is oriented in the correct
position (see 3.6 Electrical Power Connections for details).
(Repeat to attach a second sensor to the S5000 transmitter).
5. Connect the sensor to the "Sensor 1" position on the electronics assembly.
a. If using a second sensor, connect it to the second sensor position.
NOTICE
If only using one sensor and it is connected to “Sensor2”position, the S5000 will enter Sensor Missing fault. See Disable
Sensor in 4.2.2 Sensor Setup for details on how to clear this fault.
Figure 13 Connecting Sensor to the Stack and
Grounding Terminal Figure 14 SOURCE/SINK Switch Location
18
3 Installation US

6. Verify the sensor connector is firmly seated on the terminal board.
7. Attach the sensor's ground to either of the grounding screws inside the S5000 housing.
8. Set the analog output to SOURCE or SINK using tweezers, flat head screwdriver or similar tool.
a. With the display board removed, locate “S1” switch on the main board (see Figure 14 SOURCE/SINK Switch
Location).
b. Set “S1” switch in the required position:
For SINK, set switch to the right side.
For SOURCE, set switch to the left side.
9. Replace the display module. Push firmly on the board stack.
NOTICE
Ensure that the electronics assembly is fully engaged in the mounting holes. If not fully seated, the touch interface
performance can be negatively affected.
10. Replace the cover by turning clockwise.
11. Tighten the set screw located on the lid using a 1.5mm Allen wrench.
NOTE: The JB5000 junction box is not compatible with IR400 and passive sensors.
3.4.5 Integrated Mounting Points
The S5000 transmitter can be surface mounted without any additional brackets using the integrated mounting tabs.
Figure 15 Integral Mounting Tabs
3.4.6 Mounting Points JB5000 Junction Box
The JB5000 junction box can be mounted directly using the 4 integrated 10-32 threaded holes on the back of the enclosure or
with the use of a mounting bracket (P/N 10206570).
19
US 3 Installation

3.4.7 Adjustable Pipe Mount
A Universal Pipe Mount Kit (P/N 10176946) can be used to mount the S5000 on pipes ranging from 20-150mm in diameter.
Two brackets are mounted over the top of the integrated mounting tabs and fitted with an adjustable pipe band (not included).
Figure 16 Adjustable Pipe Mount
3.4.8 Duct Mount
Duct mount kits are available for monitoring atmosphere inside flat or round ducts. Round duct mount kits are available for
small ducts 12-20" in diameter (P/N10179323) and large ducts 20-40" in diameter (P/N 10179324). The flat duct mount
(P/N10179322) is universal for flat ducts.
NOTICE
Consider the sensor type before choosing a duct mount location. IRSensors should be mounted horizontally and all other
sensors should be mounted vertically.
NOTICE
Air flow in the duct must be zero to ensure proper calibration.
Figure 17 Flat Duct Mount
20
3 Installation US
Other manuals for S5000
1
Table of contents
Other MSA Monitor manuals

MSA
MSA ICM Tx Unit NightFighter Troubleshooting guide

MSA
MSA Ultima User manual

MSA
MSA GASGARD 8 User manual

MSA
MSA Ultima X Series User manual

MSA
MSA Toxgard II User manual

MSA
MSA ULTIMA X5000 User manual

MSA
MSA ULTIMA X-Series User manual

MSA
MSA PrimaX IR User manual

MSA
MSA Chillgard 5000 User manual

MSA
MSA MV-1901F User manual