MTEC PAC300 User manual

2
Table of Contents
1. First and Foremost: Three Key Precautions to Protect PAC300 from Damage
1.1 Protecting Microphone during Purging ………………………………............3
1.2 Caution to Avoid Breaking Window ……………………………………………….3
1.3 Protecting Window from Moisture Damage …………………………………..3
2. Unpacking
2.1 Unsealing the Hermetically Sealed Shipping/Storage Case ……………..3
2.2 Packing List …………………………………………………………………………………...3
2.3 Damage Inspection …………………………………………………………………………4
3. PAC300 Description
3.1 Introduction to the PAC300 ………………………………………………………………….4
4. Set-Up Directions
4.1 Insert Rear Lever …………………………………………………………………………….5
4.2 Mounting and Alignment of the PAC300 in the FTIR ……………………..…6
4.3 Electrical Connections …………………………………………………………………….9
4.4 Purge Gas Connections …………………………………………………………………..10
5. Operation
5.1 Preamplifier Gain Control …………………………………………………………..….10
5.2 Rear Lever Operation …………………………………………………………………….11
5.3 Sample Cup Loading ……………………………………………………………………..11
5.4 Sample Cup Insertion …………………………………………………………………..…13
5.5 Purging ………………………………………………………………………………………....14
5.6 FTIR Spectrum Acquisition Parameters ………………………………………..…14
5.7 Test Procedure ……………………………………………………………………………….15
5.8 Acquiring and Normalizing a Sample Spectrum …………………………..…15
5.9 Advanced Methods of Photoacoustic Spectroscopy ……………………...16
5.10 Storage of the PAC30 ……………………………………………………………………16
6. User Adjustments and Servicing
6.1 Setting the Sample and Reference Cup O-ring Seal Compression …..17
6.2 Window Replacement ……………………………………………………………………17
7. Service Assistance
7.1 Returning Units for Factory Servicing ……………………………………………..19
7.2 Foreign Returns ……………………………………………………………………………….19
8. Product Warranty and Disclaimer …………………………………………………………………..….20

3
1. First and Foremost: Three Key Precautions to Protect PAC300 from Damage
1.1 Protecting Microphone during Purging
Purging with zero grade helium gas is used to enhance the sensitivity of photoacoustic
measurements and to reduce the moisture level in the sample chamber which
superimposes a water vapor spectrum on the sample spectrum and also interferes with
condensed sample signal generation. However, a too high gas flow rate can easily damage
the PAC300’s sensitive microphone. Accidental high flow rates may occur when gas valves
are being opened or adjusted. The PAC300 microphone is protected from gas flow surges
as long as the rear lever is in the SEAL position.
1.2 Caution to Avoid Breaking Window
Do not place samples in the brass sample cup that extend above 1 mm from the top of the
cup. Samples that extend above the top of the cup will break the PAC300’s window.
1.3 Protecting Window from Moisture Damage
The PAC300’s KBr window is moisture sensitive. The window can be protected from
moisture damage by storing the unit in its hermetically sealed case with a 10 g packet of
indicating silica gel ( http://silicagelpackets.com/indicating-silica-gel-packets/indicating-
silica-gel-packets/10-gram-indicating.html ).
2. Unpacking
2.1 Unsealing the Hermetically Sealed Shipping/Storage Case
In cold weather, it is advisable to let the case warm up for 24 hrs. before opening to
prevent moisture condensation on the KBr window.
2.2 Packing List
All PAC300 photoacoustic accessories include:
1. PAC300 with focusing mirror and KBr window.
2. Black handled sample holder (inserted in PAC300) with brass sample nest, five
small and five large removable sample cups, five nest spacers to elevate samples,
sample cup fixture and funnel, tweezers and Allen wrench to adjust cup o-ring
compression.
3. Carbon reference sample mounted in black handle.
4. Liquid crystal for imaging IR beam position.
5. Helium flow meter.
6. Hermetically sealed shipping/storage case

4
Some accessories may also include one or more of the following:
1. FTIR specific baseplate mounting.
2. Desk-top power supply.
3. Glassy carbon phase reference.
4. FTIR specific connection cable(s).
5. PAC300 elevation adjustment shims.
6. FTIR purge coupling.
7. SH003 and/or SH004 options.
2.3 Damage Inspection
Please examine items for evidence of shipping damage. Report any damage to the carrier
and to MTEC for direct purchases or the reseller that shipped the accessory.
3. PAC300 Description
3.1 Introduction to the PAC300
The PAC300
enables
direct
measurement by photoacoustic detection
of
optical
absorption spectra
of
s
o
lid
s and semisolids and is primarily used
in
FTIR
spectrometers:
http://www.mtecpas.com/Docs/Photoacoustic%20Technology%20Overview.p
df .
The
detector
can be
used
for
a wide
range
of
measurements
as
described
in
the
applications
liter
atur
e and literature bibliography:
http://www.mtecpas.com/applicationslibrary.html ,
http://www.mtecpas.com/Docs/FTIR%20PAS%20Bibliography%20Draft%205.pdf .
The PAC300 can also be used with nonvolatile liquid and noncorrosive gas samples which
are not toxic.
The PAC300 is
designed for
ease of use.
It mounts
in the
s
t
a
n
d
a
r
d
FTIR slide-
mount fixture or on FTIR specific baseplates. Sample changing, purging,
and
sealing
ar
e controlled
by a
single lever, detector preamplifier
gain is
controlled
by a
12
position switch
on the
detector,
and
the detector
is
powered
either
by the FTIR or by a
115/230 V,
60/50 Hz desk top power
supply
with
an international standard IEC 320
input
connector. This
allows the power
supply to be
used with appropriate
cord sets for any country
of the world
.
Inserts into FTIR Slide Mount
Inserts into FTIR Sample compartment
PAC300 for slide mount fixture (left)
or baseplate mounting (right).

5
4. Set-Up Directions
4.1 Insert Rear Lever
The stainless steel rear lever is packed separately from the PAC300 in the shipping case
and should be inserted into one of the three slots on the back of the PAC300. Use the slot
that allows the lever to rotate most conveniently in your FTIR through the 90 degrees of
rotation necessary to purge and seal the sample. Note that the sample holder can be
inserted from either side of the PAC300 to accommodate both right-to-left and left-to-
right beam directions in different FTIRs.
Purge Inlet
Sample Holder
Rear Lever
FTIR Purge
Coupler
Baseplate

6
4.2 Mounting and Alignment of the PAC300 in the FTIR
The PAC300 is either mounted on an FTIR specific baseplate or in the standard sample
slide mount found in all FTIRs. Baseplate and slide mounted PAC300s are prealigned and
usually will not require the adjustments described below.
Alignment can be checked using the
i
nfrared heat
sensitive liquid crystals supplied
with the
PAC300.
Lever
Slots
Color Change Indicates
FTIR beam focal spot. If the
spot is centered in the cup,
the PAC300 is aligned.
PAC300 needs to be moved
horizontally perpendicular
to the IR beam to center the
spot in sample cup.
IR Beam
1
2

7
In
order
to check
alignment, place a liquid crystal in the sample cup following
these directions in its package:
Heat Sensitive Liquid Crystals (HSLCs)
The HSLCsare usedto alignthe photoacoustic detector in the FTIRby imaging the focal spot of the IR beam in the sample
cup. Each HSLC covers a slightly different ambient temperature range. The one best suited for a given ambient
temperature should be selected by trying each. Place a HSLC in the large sample cup supported on the foam spacer. A
brass spacer should be used under the foam to elevate the HSLC to a point near the cup rim. The heat sensitive side of
the HSCL can be identified by finger touch induced heating. The detector position should be adjusted to center the
focal spot image in the sample cup. See the Instruction Manual for additional information.
Then
block
the infrared beam
and
insert the sample holder with the
li
qu
id
crystal
into the PAC300.
Fully
open the
aperture
of the
FTIR. Elevate
the
crystal
by
moving the rear lever
to the
CLOSED PURGE position. Remov
e
the
beam
block, wait
several seconds, remove
the
sample
cup, and
examine
the
image
pattern
before
it fades
as
the
temperature falls.
The
types
of images
that
may be
observed
are
shown
above. Image
1
indicates ideal alignment. Image
2
shows
s
ide
-
t
o-
s
ide
misalignment.
This can
be
corrected
by
moving
the PAC300
perpendicular
to the
beam
in the
direction opposite
to the
misal
i
gnment
direction. The side-to-side
adjustment
can be
made as shown below either
by
loosening
the
screws
on the
baseplate and
sliding the
PAC300
side-to-side
or by adjusting the
slide
plate
relative
to
its mounting
yoke.
Image
3
indicatesthat
the PAC300
needs
to be
raised. The height of
baseplate mounted PAC300s can be adjusted with the screws that support
the accessory on the baseplate as shown below.
PAC300 needs to be
moved vertically to
center the spot in sample
cup.
3
Shims for height adjustment
Side-to-side adjustment
Side-to-side adjustment

8
In the case of slide mounted PAC300s, some
FTI
R
slide
fixtures have provision for
height adjustments
while
others
do
not.
If
th
e
r
e
is no
adjustment provided
on the
FTIR, height adjustments
are
made by placing, as shown above, small
shims provided with
the
detector between
the PAC300
s
lide
mount
and
the
yoke
coupling plate. Adding shims lowers the detector
and moves
th
e image
in the
direction
of the front of the PAC300
.
In some cases it may be desirable with slide mount fixtures to support the
PAC300 directly on the FTIR’s or other instrument’s baseplate. Four 5-40
thread screws can be used as adjustable legs as shown below.
If slide mount fixture’s position is designed to be adjustable in the direction of the IR beam
, it may be observed that the
infrared beam image on the liquid crystal
is
large
and
weak. This may be caused by
the PAC300 not being positioned properly relative
the IR beam’s focal spot which is normally at the midpoint of the beam’s path through the
FTIR's
s
a
mple compartment. The PAC300’s optics require that the PAC300’s
slide plate be coincident with the FTIR’s focal plane with the hole in its slide
plate centered on the focal spot of the FTIR. To achieve this, the slide mount’s
position can be adjusted while watching the PAC300’s interferogram signal to
maximize it.
This nut is one of four used to
raise or lower the PAC300 on
a baseplate mounted unit.
Four 5-40 Screws Support
the PAC300 and allow for
height adjustment.
FTIR’s focal spot

9
4.3 Electrical Connections
The PAC300 is powered either by the FTIR or an MTEC desktop power supply. The power
and signal cables are connected to the vibration isolated part of the accessory after first
being secured to the stationary part to avoid transmitting mechanical vibrations to the
PAC300’s sensitive microphone.
The desktop power supply has +15 and -15 Volt outputs and connects to either 115V 60 Hz
or 230V 50 Hz power lines. It has an IEC 320 power cord connector and an RJ9 output
connector.
The PAC300’s preamplifier gain is adjustable in 12 steps from 2X to 10,000X to
accommodate samples with different thermal properties and FTIR mirror velocities.
Power In
Signal Out
Gain Switch
On-Off switch
120/230 Volt
switch

10
4.4 Purge Gas Connections
Purging with Zero Grade Helium increases the signal by a factor of 2 to 3 and helps to exhaust
moisture and CO2 from the sample chamber. A two stage gas regulator valve should be
connected to a helium tank with its output connected to the MTEC flow meter. The
flowmeter output is connected to the inlet of the stationary part of the PAC300. Its outlet is
connected to the vibration isolated sample chamber. Consult Section 5.5 before purging the
PAC300.
5. Operation
5.1 Preamplifier Gain Control
A
twelve position switch controls
the gain as fo
llows
:
step# 1
2
3
4
5 6 7 8 9
10
11
12
gain
2
5
10
20
50 100 200 500 1000
2000
5000
10,000
factor
The gain
should
be set high
enough
to give an
inter
f
er
ogr
a
m
amplitude of
several volts
but not too high to
either cause preamplifier clipping
(<20V
pe
ak
-
to-peak)
or to
overloadthe FTIR's
A to D
converter.
Higher gains
are
used
for
higher
FTIR
mirror velocities because
the
signal
amplitude decreases
as
modulation frequency increases. Higher gai
ns
are
also
used when sample spectra are
being
acquired relative
to c
a
r
bon
reference
background spectra because totally absorbing carbon
gener
a
tes
higher signal amplitudes than partially absorbing samples whichusually
a
l
s
o
have higher thermal mass.
To Gas
Cylinder
Line Pressure
Tank Pressure
Control Flow Rate
Only with this Valve
Leave This Valve
Always Open
Rear Lever Always In the SEAL
Position When Adjusting Valves
Flow Rate 10cc/s or Less

11
5.2 Rear Lever Operation
The
rear
lever
rotates
over an
angle
of 90 degrees
to elevate
the
sample
hol
d
er
into the
detector's sample chamber
and
to compress
the holder’s o-ring to
seal
the
chamber.
The
lever
also
automatically controls
the
purg
e
valve
during
the 90 degree
rotation.
A
detent snap
device
indexes
four
rota
ti
onal positions
over the 90 degree
rotation. These
are the
OPEN, OPEN PU
RGE, CLOSED
PURGE,
and SEAL
positions
which
are shown schematically
on
the
detector's
sides for
easy
reference during
operati
o
n.
In the
OPEN position
the
sample holder
may be
withdrawn
or
i
nserted
for
sample cha
n
g
e
.
In the
OPEN PURGE position purge
gas flows
through
the
sample chamber
and
out the
bottom because
the
sample holder o-ring
is
not s
e
ale
d
.
In the
CLOSED PURGE position purge
gas flows
past
a hole
leading to
the
sample chamber
but not
through
it.
Consequently, purging
is
slower
but
there
is no flow over
the sample whichmight
blow
powders
out of the
cup.
In the SEAL
position
the
sample chamber
is
sealed
and
spectra
may
be
acquired.
The
purge
gas
still
can flow but is
sealed
from the
sampl
e
chamber so the microphone is protected from any surges in the gas flow that
might occur when gas valves are being opened, adjusted or closed.
5.3 Sample Cup Loading
The PAC300 is
supplied with
five
small
and five
large samp
le
cups and five
brass spacer inserts.
The cups and
inserts fit
into a
b
r
a
ss nest
which
is
attached
to
the
black
sample holder handle.
The slot in the
brass part
should
be
kept oriented perpendicular
to
the
handle length.
The o-ring
seals
the
detector when
the lever is
rotated
fully
to
the SEAL
location
and
must
be
kept clean.
v
Brass nest rim
Sample cups
Sample cup
elevation spacers

12
To
avoid breaking the detector window,
do
not place anyth
i
n
g which
extends higher than the rim
of
the brass nest because when the rear
lever is rotated
to
the SEAL
location, the
rim
is located just below the
detector window.
An
yth
i
ng extending above the rim will be forced into
the window and break
i
t
.
Samples, depending
on size,
are placed
in
the small sample
cups
or
in
the
large brass
cup
holder
(in the
latter
case
with
or
without
a
large sample
cup).
The
brass spacer inserts
ar
e
used
to
eliminate
excess
volume
in
the sample
cup. A
higher signal
level
is
consequently obtained because
the
signal
is
inversely proportional
to the
g
a
s
volume.
The
inserts
and
sample
levels in
cups
should
be
chosen
so
that th
e
sample
is
approximately
1 mm or more
below the
brass
cup
holder
rim.
T
h
i
s
distance allows
the
photoacoustic signal
to be
generated
in the gas
withou
t
thermal interference
by the
window
above the
sample.
Course samples
ma
y
be
placed
in cups
using tweezers. Fine powders
may
be
loaded with the
funnel
by
first placing
cups
either directly
in the
brass
nest
or in the
slotted
cup
fixture
and
then placing
the
funnel
in
position
.
An
alternative approach
is to scoop
powders from storage containers
with
the
sample
cups held by
tweezers
and
then
use the
slotted
cup
fixture
to
align the tweezers
in
order
to
facilitate
easy
insertion
of
the cu
p
into
the
brass
cup
holder.
Be
careful not
to spill
samples
onto
the o-ring.
On
ly
enough
sample
to cover the
bottom
of the
sample
cup is
necessar
y
.
Samples
(both
micro
and
macro)that
evolve H20
vapor should
be
r
un
with
a
desiccant
in the
sample chamber
to avoid
vapor bands
in
s
pe
c
tr
a.
Magnesium perchlorate
is a very
effective desiccant
and can often be
used
safely.
This
chemical
is,
however,
a
strong oxidizing agent
and may
create
a
hazard when used
in close
proximitywith certain samples. It
is
the operator's
responsibility
to
consult
and
observe appropriate safety
in
fo
r
m
at
io
n

13
The
desiccant
is
placed
in a
large sample
cup. This cup is
inserted in
to
the
brass holder
, a
slotted spacer insert
is
placed
over
the
desi
ccant
cup, and the
slots
are aligned
to
permit
gas
circulation between the
desi
cca
nt
and
sample
cups.
Finally, the sample
is
put
in a
second
cup
which
is inserted
above the
desiccant
cup.
Never
leave
desiccant
in the
sample cup when
the
detector
is not
sealed because
a
corrosive liquid
will form
as
moisture
is
collected
by the
desiccant
from the room
a
i
r. If the detector is stored for an extended
time, check the status of the desiccant periodically.
There
are
several effective approaches
in
addition
to the
desiccative
cup,
which
may be
used alone
or in
combination
to
reduce vapor
b
a
n
d
interference
from a
sample. These approaches include vacuum and/or oven
drying
of
samples prior
to
measurements, reduction
of the
amount
of
sampl
e
placed
in the cup,
dry
gas
purging (zero grade helium),
and
spectral
subtraction.
In
terferen
ce
bands
may also be
present
due to
residual vapors
in the
detector from
the
previous samples, degassing
of
internal
components,
or from the
storage
bo
x.
These bands
can be
readily reduced
to
an
acceptable
level by
purging.
Note
that bands
due to
vapor absorption
in the
detector are always
p
o
s
i
tive
pointing whereas vapor absorption
in
the spectrometer produces negati
ve
pointing
b
a
nds.
5.4 Sample Cup Insertion
The
sample
cup can be
inserted
from
either
side of the
detector si
nce
different sides
of the
detector
are
accessible with different FTIRs.
The
sample holder is ready
to be slid into place
for elevation
into
th
e sample
chamber
and
sealing by the rear lever.
Desiccant
Slotted Spacer

14
5.5 Purging
Purging increases the signal
level by
approximately
a
factor
of 2 to
3
and
reduces moisture
in the
sample chamber. Care must
be
taken
in
purg
i
n
g
the
detector, however,
to avoid
large pressure fluctuationsthat
could
damage
the
microphone.
Use
clean dry helium
gas such as "zero
grade".
Only
the primar
y
pressure regulator
valve on the gas
cylinder should
be
used
to
control the
flow
rate as indicated in Section 4.4.
If the
regulator
has a
second
valve
at
the
regulator output
set
this
valve to the full open
position
and leave
it
so
set. Keep the
detector
lever
i
n
the SEAL
position whenever
the gas is being
turned
on or off and
when the
flow is being
adjusted. These
are the
times
when pressure surges are most
likely to occur and the SEAL
position isolates
the
microphone.
Use a flow
rate
of 10-20 cc/s for most
sa
mples.
For
fine powders
just fill sample cups with a thin layer. To insure that the
powder cannot be blown out of the cup during purging, just before inserting the sample
into the PAC300, reduce the helium flow to zero as per Section 4.4, insert the sample cup
into the PAC300 and rotate the rear lever to the SEAL position. Now increase the helium
flow to 5 cc/s and rotate the rear lever just to the Closed Purge position and wait 30 s.
Next rotate the rear lever to the SEAL position and collect the spectrum of the sample.
Reduce the helium flow to zero, rotate the rear lever to the OPEN position, and remove
the sample. Rotate the rear lever to the seal position, set the helium flow to between 5
and 10 cc/s, and rotate the rear lever to the OPEN position. Prepare the next sample to be
loaded and repeat the sample insertion steps given above starting with reducing the
helium flow to zero.
Ten
seconds purge time
in each of the
purge locations
is
usually
suffi
ci
ent.
Longer times
may be
necessary
if
considerable moisture
is
p
r
e
s
e
n
t.
Leaving
the
detector sealed with desiccant under
the
sample
cup for
a
n
extended time during measurements
is the
best
way to
remove
the
last traces
of
moi
s
ture.
When
the
detector
is
initially purged with helium there
will be
a
gradual decrease
in
signal amplitude
as
helium exchanges with
air in
th
e
rear volume
of the
microphonethat
is
connected
to the
sample chamber
by
a
fine
capillary.
This
drift
will be
minimized
if
helium
is
kept sealed
in
th
e
detectorat
all
times.
The
signal
will also
gradually decrease
due to
di
ff
u
s
i
on
of
helium at
the
o-ring seals. Signal drift
can be
eliminated
by
purging with
dry
nitrogen
but the
signal enhancement
of
helium
is
lost.
5.6 FTIR Spectrum Acquisition Parameters
For general use, the followingare
usua
lly
appropriate:
1. Minimum mirror velocity
tha
t
is
stable
2.
Maximum source
ap
e
r
tu
r
e
3.
Resolution
of 8
cm
-1
4. Scan number depending on signal-to-noise ratio

15
5.7 Test Procedure
1.
Remove
the red
protective
cap, place the
carbon
black
reference (Fi
g
.
4.11) in the
detector
and
purge
the
detector,
if
desi
red.
CAUTION.
Do
not allow anything
to
contact the black
absorber
surface
of
the carbon reference.
It
is
e
a
sily
damaged. Keep the
red protective cap
on
the carbon bl
ac
k
reference when not in use.
Do
not leave the carbon black reference in the detector when
a
backgroundspectrum is not being
acquired.
2.
Adjust
the
detector
gain for
a peak-to-peak amplitude of several
volts. This will assure t ha t the
preamplifier is not clipping (clipping
occurs
at approximately
20 volts
ma
x
i
mum
peak-to-peak)
and that
the FTIR
input is
not overloading.
3. If
the PAC300detector initially
does not
produce
a
signal,
di
sc
onnect
and
then reconnect the telephone connector
on the
white
cable
located next to
the
detector's
gain
sw
itch.
4.
Acquire
two
single beam
spectra
of 8
scansat
8 cm-1
resolution
and an
OPD
mirror velocity
of 0.16
cm/
s
with
the MTEC
carbon
reference
.
If the FTIR
spectrometer
does not
have
an OPD
velocity
of 0.16 c
m
/s
substitute
the
closest
availa
ble
.
5.
Ratio
the two
single beam spectra
to
obtain
a 100% line.
P
e
ak
-
t
o-
peak
noise
should
be 0.4% or less i n t he r a ng e f ro m 19 0 0 t o 2 1 0 0 cm-1 for
most FTIR
i
nstruments
using
a
helium purge.
The
peak-to-peak
noise will be
found to
increase
if
a
higher mirror
vel
o
ci
t
y
is
used
due to the slow
time
response
of
photoacousti
c
signal gener
a
tion
.
5.8 Acquiring and Normalizing a Sample Spectrum
1. Load the sample (see 5.3 and 5.4) and purge (see 5.5) the detector, if desired.
2. Set the detector gain to provide a signal level which is less than or equal to what was
used for the carbon black spectrum (see 5.7) and acquire the sample spectrum (see 5.6).
3. Divide the sample spectrum by a carbon black spectrum in order to obtain an
absorbance like spectrum of the sample. Some FTIR’s have parameter settings to
automatically ratio the sample to reference spectra and display it in photoacoustic units.
Do not touch
black surface!

16
5.9 Advanced Methods of Photoacoustic Spectroscopy
Advanced photoacoustic spectroscopic methods utilize both the magnitude and phase of
the photoacoustic signal often to investigate depth varying composition in the form of
layers or gradients. Glassy carbon is the best phase reference material for such
investigations.
A general overview of some interesting advanced applications is available here:
http://www.mtecpas.com/Docs/Photoacoustic%20Technology%20Overview.pdf.
The PAC300 can be used to study samples with depth varying composition in the form of
layers and gradients over depth ranges of a few tens of microns. If deeper depths are of
interest, MTEC’s MicroLap device (http://www.mtecpas.com/microlap.html) can be used
on planar samples. For further information in general see:
http://www.mtecpas.com/Docs/MicroLap%20Technology.pdf
http://www.mtecpas.com/Docs/MicroLap3.pdf
http://www.mtecpas.com/Docs/Introduction%20to%20Photoacoustic%20Spectroscopy%2
0with%20Step%20Scan%20and%20Constant%20Velocity%20Scan%20FTIR%20Spectromet
ers.pdf
http://www.mtecpas.com/Docs/BCG.pdf
http://www.mtecpas.com/Docs/MicroLap1.pdf
http://www.mtecpas.com/Docs/MicroLap2.pdf
MTEC offers two options with sampling heads for sequential diffuse reflectance,
photoacoustic absorbance, and transmittance measurements (Option SH003) and for
microsamples in the form of single particles and fibers (Option SH004). For further
information see:
http://www.mtecpas.com/Docs/MTEC%20Sampling%20Head%20Options.pdf
http://www.mtecpas.com/Docs/Prac%20Guide%20to%20FTIR%20Photoacoustic.pdf
5.10 Storage of the PAC300
The PAC300’s KBr window is moisture sensitive. The window can be protected from
moisture damage by storing the unit in its hermetically sealed case with a 10 g silica gel
desiccant pack with indicator to tell when to change to fresh desiccant:
http://silicagelpackets.com/indicating-silica-gel-packets/indicating-silica-gel-packets/10-
gram-indicating.html

17
6. User Adjustments and Servicing
6.1 Setting the Sample and Reference Cup O-ring Seal Compression
Two set
screws control
the
compression
of
the sample
cup's o-ring
sea
l
.
These
screws must be
adjusted symmetrically
to
extend
the
same distance
in
order
to
avoid
cocki
n
g
the o-ring
flange relative
to the
detector
body.
When properly
adjusted,
the
sample holder handle should not
be
moveable side-to-side relative
to the
housing when
the rear lever
is in
the
SEAL
p
o
s
itio
n
.
6.2 Window Replacement
MTEC offers KBR, quartz, CsI, KRS5, and Polyethylene windows. It
is
recommendedthat
the
detector
be
returned
to MTEC for
w
i
ndow
replacement.
If the
user prefers
to
replace
the
window,
follow
the
fol
l
o
w
ing di
recti
o
ns.
Disconnect the power and signal cables and the purge line.
Loosen the top set screw, remove the two top slide plate screws and the attached
part which clamps the cables.
No side-to-side motion in
rear lever SEAL position
Loosen

18
Loosen the upper vibration isolator nuts and back off the screws as shown to
allow the detector to be lifted off the lower vibration isolation platform.
The detector’s mirror can now be removed by unscrewing its two attachment
screws to expose the four screws of the window’s o-ring compression plate.
Remove
Loosen

19
Once these screws are removed using a ball point Allan wrench, the plate can be
lifted off and the window replaced (diameter=18mm, thickness=3mm). Great care
must be taken in tightening the screws to compress the window o-ring. The
screws must be tightened in very small equal angle increments with a light touch
to avoid cracking the window. Carefully tighten
the
fou
r
compression screws to
compress
the o-ring while
keeping
an
equal
gap on
a
ll sides between
the
compression plate
and the
detector housing.
The detector should be
reassembled by reversing the disassembly steps.
7. Service Assistance
7.1 Returning Units for Factory Servicing
Please contact
MTEC for
authorization prior
to
shipping
any
items for
repair.
(MTEC email: mtec@mtecpas.com,
telephone: 1-515-292-7974; telefax: 1-515-292-
7974.) A
ll
transportation charges
and
insurance
are the
customer's
responsibility.
A
purchase order number must
be
provided before repairs
will be
made. (See
also
Section
8 for
Product
Wa
rra
nty
.
)
Package
the
return carefully,securing
all
parts
to
prevent dama
g
e
during transit. Seal the empty sample cup in the detector.
Send the
entire system including powersupply
and
ca
b
les.
Shipping address
for
returns:
MTEC
P
h
otoacousti
cs,
Inc.
3507
Oakland
St
r
eet
Ames, Iowa 50014
U.
S.A.
7.2 Foreign Returns
If
possible,
ship the
package
via a
courier,
such as
Federal Express,
who will act
on
behalf
of MTEC as a
customs broker.
This
simplifies
the
return, expedites
clearance through
U.S.
Customs,
and
saves extra
frei
g
h
t
forwarder
costs to
MTEC
which are
billed back to the
customer. Prepay
the
air waybill through to
Ames, Iowa.
Consign
to the
shipping address above,
and
list MTEC's telephone
number (515-292-7974).
To avoid co
mp
lic
ations
and
delays,
label
the shipment
"free domicile."
This
simplifies
U.S.
C
u
s
t
o
m
s
clearance.
A copy of an invoice
showing the total value
is
required fo
r
customs.
This invoice
must state
the
value, that
the
product
was "Made
i
n the
U.S.A." and is a
"Returned American
Product for Repair
Only."
Ple
a
s
e
follow
your
local
shipper's instructions
concerning
all
shipping
r
e
qu
ir
eme
n
ts
.
If the
return
is
made through
a
freight
forwarder,
be
sure the freight
i
s prepaid
to Ames, Iowa, if
possible,
or
at least
to Des
Moines,
Iowa. Use
the
"free domicile" declaration
to avoid
complications
and
de
la
ys
.

20
8. Product Warranty and Disclaimer
PRODUCT WARRANTY
MTEC Photoacoustics, Inc. warrants the Model 300 Photoacoustic Detector and its accessories
to be free from defects in material or workmanship and to operate to published specifications
under normal use, for a period of one year from the date of original shipment. No other
warranty is expressed or implied. If examination by MTEC Photoacoustics, Inc. discloses a
product defect, obligation is limited to repairing, replacing, or giving credit for the purchase
price, at our option. KBr windows, carbon black references, batteries, Option SH003 absorber
elements, and Option SH004 tungsten needles are not warranted. MTEC Photoacoustics is not
liable for any consequential damages or for any damages which might occur during vacuum
and/or bake out operations. Components other than those manufactured by MTEC
Photoacoustics, Inc., including microphones will carry the original manufacturer's warranty. All
transportation charges on items returned are the customer's responsibility. Contact MTEC
for authorization prior to returning any items for warranty claims.(See Section 7.)
PRODUCT
DIS
CLAIMER
MTEC
Photoacoustics,
Inc. will
not assume responsibility fo
r
any
damages
to
persons
or
to property
due to
the operation
of or
t
o
results obtained with
the
MTEC Model 300
photoacoustic
detector system.
Loosen
Table of contents
Other MTEC Security Sensor manuals
Popular Security Sensor manuals by other brands

Littfinski Daten Technik
Littfinski Daten Technik SX-GB-8-B Assembly instruction

ETC
ETC echoflex MOS-IR A Series installation guide

Saso
Saso SASO PB100A installation instructions

Interlogix
Interlogix EV1012 Series Installation sheet

GJD
GJD D-TECT Universal GJD630 manual

DELTA DORE
DELTA DORE Tyxal+ DME 180 manual