8
blood pressure monitor be kept 30 cm away from other wireless devices,
such as WLAN unit, cell phone, microwave oven, etc. It can’t be used near
acve HF SURGICAL EQUIPMENT and the RF shielded room of an
ME SYSTEM for magnecresonance imaging, where the intensity of EM
DISTURBANCES is high.
12. If Irregular Heartbeat (IHB) brought on by common arrhythmias is detected
in the procedure of blood pressure measurement, a signal of will be
displayed. Under this condion, the Electronic Sphygmomanometer can
keep funcon, but the results may not be accurate, it’s suggested that you
consult with your physician for accurate assessment.
There are 2 condions under which the signal of IHB will be
displayed:
1) The coecient of variaon (CV) of pulse period >25%.
2) The dierence of adjacent pulse period ≥0.14s, and the number of such
pulse takes more than 53 percentage of the total number of pulse.
13. Please do not use a cu other than supplied by the manufacturer, otherwise
it may bring biocompable hazard and might result in measurement errors.
14. The monitor might not meet its performance specicaons or cause safety
hazard if stored or used outside the specied temperature and humidity
ranges in specicaons.
15. Please do not share the cu with other infecve person to avoid cross
infecon.
16. This equipment has been tested and found to comply with the limits for a
Class B digital device, pursuant to part 15 of the FCC Rules. These limits are
designed to provide reasonable protecon against harmful interference in a
residenal installaon. This equipment generates, uses and can radiate
radio frequency energy and, if not installed and used in accordance with the
instrucons, may cause harmful interference to radio communicaons.
However, there is no guarantee that interference will not occur in a
parcular installaon. If this equipment does cause harmful interference to
radio or television recepon, which can be determined by turning the
equipment o and on, the user is encouraged to try to correct the
interference by one or more of the following measures:
— Reorient or relocate the receiving antenna.
MX86007 - IFU - 01.2021.indd 8MX86007 - IFU - 01.2021.indd 8 2021/01/06 13:222021/01/06 13:22