8
The MyGlucoHealth Wireless Meter and
strip system is one of the most accurate
personal diabetic care products on the
market today. When tested against cali-
brated laboratory instruments, the MyGlu-
coHealth Meter meets higher performance
standards when compared to many off-the-
shelf glucometers.
In glucose concentrations of less than 4,2
mmol/L, MyGlucoHealth tested within plus
or minus 6.5% of laboratory calibration,
89% of the time and within 7.5%, 100% of
the time. When testing concentrations of
greater than 4.2 mmol/L, MyGlucoHealth was within 5%, 65% of
the time, within 10%, 92% of the time and within 15%, 98% of
the time.
The MyGlucoHealth Wireless meter is plasma calibrated. This
makes results directly comparable to HbA1c test results from a
laboratory. Plasma results are approximately 12% higher than
whole blood results. Keep this in mind when comparing MyGlu-
coHealth test readings to those of whole blood calibrated meters.
A few high-end or professional quality blood glucose meters, in-
cluding the MyGlucoHealth Wireless meter, are already calibrated
or set to give a “plasma” test result, making this conversion un-
necessary. For an accurate comparison of your blood glucose
meter to a laboratory result, you must rst know how the meter
is calibrated.
METER
ACCURACY
note:
When comparing readings between an older meter and MyGluco-
Health you may see differences due to the higher accuracy standards
and plasma calibration of MyGlucoHealth.