nal von minden Reactif Touch User manual

new art laboratories
Reac f Touch
Manual
Rapid Tests
Laboratory Diagnos cs
Laboratory Service
Consul ng & Service

nal von minden GmbH · Manual
Reac f Touch

1
1
nal von minden GmbH
Carl-Zeiss-Strasse 12
47445 Moers
Germany
Friedenstrasse 32
93053 Regensburg
Germany
Tel.: +49 941 29010-0
Fax: +49 941 29010-50
info@nal-vonminden.com
www.nal-vonminden.com
Version 1.0
Januar 2017

nal von minden GmbH · Manual
Reac f Touch

Table of Contents
Sec on 1 Introduc on ........................................................................................................................................ 2
Intended Use.................................................................................................................................................. 2
Sec on 2 Analyzer Components.......................................................................................................................... 3
Port Func ons................................................................................................................................................ 5
Op onal Accessories...................................................................................................................................... 5
Sec on 3 Ini al Startup ...................................................................................................................................... 6
Plugging Analyzer In....................................................................................................................................... 6
Installing Ba eries (Op onal) ........................................................................................................................ 6
Interfacing to a Computer.............................................................................................................................. 7
Connec vity................................................................................................................................................... 7
Analyzer So ware Updates............................................................................................................................ 7
Loading the Printer Paper .............................................................................................................................. 7
Powering Up .................................................................................................................................................. 8
Powering Down.............................................................................................................................................. 8
Sec on 4 Touchscreen LCD Naviga on ................................................................................................................ 9
Sec on 5 Start-Up Wizard................................................................................................................................. 15
Sec on 6 Specimen Test ................................................................................................................................... 16
Entering the Canister Code .......................................................................................................................... 16
Specimen/Strip Prepara on ........................................................................................................................ 17
Strip Test - Quick Test................................................................................................................................... 17
Strip Test - Full Test ..................................................................................................................................... 22
Strip Test - Customized Test ........................................................................................................................ 23
Strip Test - Work List ................................................................................................................................... 23
Sec on 7 Analyzer Setup .................................................................................................................................. 29
Test Se ngs ................................................................................................................................................. 29
Analyzer Confi gura on ................................................................................................................................ 34
Setup Review ............................................................................................................................................... 40
Lockout ........................................................................................................................................................ 40
Diagnos c Tests ........................................................................................................................................... 48
Service ......................................................................................................................................................... 49
Sec on 8 Records Reviewg ............................................................................................................................... 50
Clear All Data ............................................................................................................................................... 50
Data Search.................................................................................................................................................. 50
Review All Data .......................................................................................................................................... 51
Sec on 9 QC Test.............................................................................................................................................. 53
QC Lockout................................................................................................................................................... 53
Strip Prepara on.......................................................................................................................................... 53
QC Test Procedures...................................................................................................................................... 53
Sec on 10 Cleaning and Maintenance .............................................................................................................. 57
Cleaning ....................................................................................................................................................... 57
Changing Ba eries....................................................................................................................................... 61
Touch Screen LCD Calibra on ...................................................................................................................... 62
Sec on 11 Troubleshoo ng .............................................................................................................................. 64
Appendix 1 Urine Analyzer Specifi ca ons ......................................................................................................... 68
Appendix 2 Performance Characteris cs of Urinalysis Reagent Strips................................................................ 69
Appendix 3 URS Parameter Table...................................................................................................................... 71
Appendix 4 Result Print-Out ............................................................................................................................. 73
Appendix 5 Barcode reader .............................................................................................................................. 75
Appendix 6 Safety Informa on ......................................................................................................................... 76
Appendix 7 Catalog........................................................................................................................................... 78
Appendix 8 Index of Symbols............................................................................................................................ 79
Appendix 9 Warranty........................................................................................................................................ 80

nal von minden GmbH · Manual
Reac f Touch
nal von minden GmbH · Manual
Reac f Touch
2nal von minden GmbH · Manual
Reac f Touch
Sec on 1 Introduc on
The Reac f Touch Urine Analyzer is a portable, easy to use analyzer. It is designed to read
only Reac f Urinalysis Reagent Strips.
The analyzer can be plugged into an electrical outlet and used on a bench top. It can also
be powered by ba eries to be able to freely move the analyzer from one tes ng site to
another.
The instrument performs a system test each me it is turned on. It automa cally calibrates
each me a test is run.
Results are displayed on the touchscreen and can be printed (if desired). The results can
also be transferred to a computer using an RS232 serial port or USB port on the back of the
instrument. The analyzer stores results from 2000 pa ent tests and is able to recall past
pa ent test results using the Records Review func on.
Note: Keys on the touchscreen are listed in bold. Display items on the screen and set-
ngs are listed in bold italics in this manual.
Note: Contains useful ps on using the analyzer.
Cau on: Should be followed carefully to ensure analyzer operates correctly and is
not damaged.
Intended Use
The Reac f Touch Urine Analyzer is intended for use in conjunc on with the Reac f
Urinalysis Reagent Strips for the semi-quan ta ve detec on of the following analytes in
urine: Glucose, Bilirubin, Ketone (Acetoace c acid), Specifi c Gravity, Blood, pH, Protein,
Urobilinogen, Nitrite, Leukocytes, Ascorbic Acid, Albumin, Crea nine, Calcium, Pro-
tein-to-Crea nine Ra o, and Albumin-to-Crea nine Ra o. The instrument is intended for
professional, in vitro diagnos c use only.
This Instruc on Manual contains direc ons to unpack the analyzer, for Reac f Urinalysis
Reagent Strip tes ng, and how to keep it in good working condi on. Throughout the
Operator’s Manual, these symbols can be seen:

3
Sec on 2 Analyzer Components
1. Analyzer
2. Power Cord and Supply Adapter
3. Printer Paper
4. Test Table
5. Test Table Insert

1. Printer Cover
2. LCD
3. Bu on
4. Printer Roller
5. Printer Paper
6. Printer
7. Test Table
8. Test Table Insert
9. SN Label
10. RS232-Port
11. USB Port 1
12. USB Port 2
13. DC Socket
14. SD card socket
15. Ba ery compartment
4nal von minden GmbH · Manual
Reac f Touch

Port Func ons
No. Port/ Socket Func on
10. RS232 Port Data transfer or Barcode reader or
External printer
11. USB Port 1 Data transfer
12. USB Port 2 So ware update or external printer
13. DC Socket DC adapter
14. SD card socket Install SD card for so ware update
15. Ba ery compartment Install 6 AA Alkaline ba eries
Op onal Accessories
1. RS232C Barcode Readerr
2. USB or RS232C Data Transfer Cable
3. Bluetooth adapter
4. Ethernet via USB to RJ45 Adapter
5. USB wireless net-card
6. SD card
7. USB Flash Drive
5

Sec on 3 Ini al Startup
Carefully remove the contents of the shipping carton. Check the carton and instrument
for visible signs of damage. If damaged, immediately contact the carrier. Remove wrap-
pings and check for all items (refer to Sec on 2 Analyzer components).
Place the instrument on a level work surface where temperature and humidity are fairly
constant.
Sec on 3 Ini al Startup
6nal von minden GmbH · Manual
Reac f Touch
Interfacing to a Computer
The instrument can send results to a computer via the serial port or the USB port located
Plugging Analyzer In
Plug the appropriate end of the power cord into the DC socket located on the back of
the analyzer. Plug the other end of the power cord into an AC electrical wall outlet.
Installing Ba eries (Op onal)
Place the analyzer on its side. Open the ba ery cover by sliding the tab away from the
cover. Place the 6 AA Alkaline ba eries into the analyzer. Close the ba ery cover and
turn the instrument back onto its base.
Cau on: Do not use the unit or the strips outside of the opera ng temperature
ranges listed below.
Analyzer: 0-40ºC (32-104ºF)
Strips: 15-30ºC (59-86ºF)
Do not place the analyzer outside or near windows, ovens, hot plates, or radiators.

7
on the back of the analyzer. This requires a 9-pin null modem serial cable that can be
purchased separately at an electronics store or from your local representa ve.
Connec vity
Reac f Touch allows for Bluetooth, Ethernet or wireless network connec vity, and use of
op onal barcode reader to eliminate manual steps and reduce transcrip on errors. Refer
to Sec on 7 Analyzer Setup: Analyzer Confi gura on: Connec vity and Reac f Touch
Data Transfer Insert.
Analyzer So ware Updates
From me to me nal von minden GmbH will add new features and make improvements
to the Reac f Touch urine analyzer so ware.
These so ware updates will be available on an electronic memory card (SD card or USB
fl ash drive) which is inserted into the so ware update socket (by the USB port or SD card
socket). The USB port is located on the back of the instrument, and the SD card socket is
located under the ba ery compartment cover and on the right-hand side of the ba ery
compartment when you face the bo om of the instrument.
Upda ng the so ware is a simple procedure. Instruc ons for upda ng the so ware
please refer to the Reac f Touch So ware Update Instruc on.
Loading the Printer Paper
Open the printer cover by pulling up on the tab. Open the paper roll compartment cover
by pulling forward and li ing up. Place the new paper roll into the printer paper compart-
ment with the paper unrolling from underneath and toward the compartment wall. Pull
about 4 inches (or 10 cm) of paper through. Close the printer and paper roll covers by
clicking them into posi on.
Note: The analyzer is enabled to automa cally print the results (to turn off the auto-
ma c print func on refer to Sec on 7 Analyzer Setup: Analyzer Confi gura on:
Printer Setup). The analyzer uses regular thermal paper or s cker label paper.
Powering Up
Press the On/Off bu on located on the front of the instrument. The analyzer will run
an automa c system diagnos c test each me it is turned on. This is the fi rst screen
displayed.

8nal von minden GmbH · Manual
Reac f Touch
Powering Down
Before turning the analyzer off , always ensure that there is no strip on the test table and
that the table and insert are clean. Press the On/Off bu on for at least 5 seconds. The
test table will retract into the analyzer. If there is no strip on the test table, the door will
close and the analyzer will switch off .
When the analyzer is turned on for the fi rst me, the Start-Up Wizard will be displayed
for a quick setup procedure. Refer to Sec on 5 Start-Up Wizard for further instruc ons
regarding the Start-Up Wizard.

9
Sec on 4 Touchscreen LCD Naviga on
Analyzer Confi gura on is performed by using the LCD touchscreen. Selected icons, which
are shown below, and text can be pressed with the fi nger to change se ngs or to enter
or exit screens. If the analyzer does not respond, press the symbol or text for a slightly
longer me or with slightly higher pressure to ac vate the touchscreen area. If the ana-
lyzer s ll does not respond, refer to Sec on 11 Troubleshoo ng.
Screens
The touchscreen serves as a guide for how to operate the Reac f Touch Analyzer. The
screen will display messages, instruc ons and op ons, which the user can respond to by
pressing the appropriate selec on on the screen.
The fi rst main screen is the ini al screen. It displays the operator’s name, me/date,
the strip quan ty that can be used, power source used, and connec vity method. This
screen indicates the instrument is ready to perform any of the four ac ons or tasks
shown below.
Each screen that follows the ini al screen has an icon, tle bar, and touch-sensi ve
ac ve areas. In some cases, the screen will also display instruc ons, messages, or error
messages. The icon indicates the main sec on working (1 of the 4 sec ons listed on
Note: The screen on the le is the ini al screen for analyzers with the Strip Lockout
and the screen on the right is the ini al screen for analyzers without the Strip
Lockout.
The screen will not display the strip quan ty if Strip Lockout is not included on
the analyzer.

10 nal von minden GmbH · Manual
Reac f Touch
Round Bu ons
These bu ons typically appear on screens that require a selec on among several items.
The bu on with a fi lled circle is the current selec on. To change your selec on, press an
unfi lled circle. The newly selected circle (bu on) will now be marked as . Press
to return to the previous screen.
the ini al screen). The screen needs to be pressed lightly in the touch-sensi ve area to
ac vate a response.
Note: If a pressed area does not respond as expected, slide fi nger across the approp-
riate selec on area.
There are three types of areas that respond to pressing the screen: Round Bu ons,
Boxed Areas, and Scroll Arrows.
Cau on: Use of anything hard or pointed on the touchscreen may cause damage.

11
Boxed Areas
These are areas on the screen enclosed in boxes. Press any area within the box to ac -
vate that func on. Boxed areas vary in size. The boxes located on the ini al screen are
examples of larger areas. Smaller box selec ons include and found at the
bo om of the screen.
Note: When an op on can be selected, it is displayed with a thick black frame and
will respond when pressed. If an op on is not available, it is displayed with a
gray frame and will not respond when pressed.
Scroll Arrows
Press the up and down Scroll Arrows on the right side of the screen to scroll through the list
of informa on on the le side of the screen. Once informa on on the le side of the screen
is highlighted, press the relevant op on to confi rm or change se ngs.

12 nal von minden GmbH · Manual
Reac f Touch
To switch between keyboards, press
AB
C
to use the lowercase keyboard. Press to use the
upper case keyboard.
To type in a name, number, birth date, etc., press the appropriate bu on. Selec ons will
appear in the data entry box. When switching between keyboards, all values will be retained
in the entry box on both keyboard screens.
The maximum number of characters allowed is 25. An audible tone will sound when the
maximum number of characters has been exceeded.
A er entering the informa on, press from keyboard screen.
Symbol Name Descrip on
Setup Menu Shows the Setup Menu screen
Records Review Shows the Records Review screen
QC Test Shows the QC Test screen
Specimen Test Shows the Specimen Test screen
Test Se ng Shows the Test Se ng screen
Analyzer Confi gura on Shows the Analyzer Confi gura on screen
Keyboards
When the screen prompts the user to enter informa on for operator ID, pa ent name, and/or
pa ent iden fi ca on, a keyboard will appear on the screen.
Symbols

13
Setup Review Shows the Setup Review screen
Lockout Shows the Lockout screen
Diagnosis Shows the Diagnosis screen
Service Shows the Service screen
Bluetooth Transfers data by Bluetooth
LAN Transfers data by LAN
WLAN Transfers data by WLAN
Ba ery Indicates the unit is opera ng from ba ery
RS232 Transfers data by RS232
Printer paper Indicates printer paper is out when fl ashing
Exit Returns to the previous screen
Up Moves the selec on up
Down Moves the selec on down
Start Begin tes ng
Checkmark Confi rm the selec on and con nue to the next
step.
Checkmark Cancel the selec on and return back to the
previous screen
Checkmark Saves the new selec on and returns to the previ-
ous screen
Cancel Cancels any changes made and returns to the previous
screen
Print Print the content on the screen
Enter Saves the new selec on and returns to the previous
screen
Con nue Confi rms the current display on the screen and con nu-
es to the next step
Disable Disables the current func on

14 nal von minden GmbH · Manual
Reac f Touch
Delete Deletes the entered informa on
Add Adds new item
Set On Operator setup menu, sets the operator
Select On Operator setup menu, selects the operator
Restore Resets to the default se ngs
Send Data If the se ngs are OK, press to transfer the selec-
ted data to a PC
View View the data selected
Previous View previous data set
Next View the next data set
Help Displays the Help screen
Up Moves the selected parameter up in order
Down Moves the selected parameter down in order
Enter Saves the new input and returns to the previous
screen
Cancel Cancels the current input
Capital Switches to the upper case keyboard
Small le er Switches to the lower case keyboard
Period Period key
Blank Space key
Hyphen hyphen key

15
Sec on 5 Start-Up Wizard
If it is the fi rst me the Reac f Touch Urine Analyzer is turned on (following an automa c
system diagnos c test), the analyzer will run through a quick setup procedure. This pro-
cedure selects se ngs and basic func ons of the analyzer.
The Start-Up Wizard allows the user to select the following se ngs:
• Language
• Date and me
• Test Number
• Type of strip
• Results format
Once selected, the analyzer will display a confi rma on screen to check that preferences are
correct.

16 nal von minden GmbH · Manual
Reac f Touch
Sec on 6 Specimen Test
A en on: For the analyzers without the Strip Lockout func on, skip the sec on Ente-
ring the Canister Code and go directly to the next sec on Specimen/Strip
Prepara on on the next page.
For analyzers with Strip lockout func on, begin with the sec on Entering
the Canister Code.
For instruc ons on how to iden fy if the analyzer has the Strip Lockout
func on, refer to the Ini al Screen on page 8. Analyzers without the
lockout func on will not display Strip Quan ty.
Entering the Canister Code
When a new canister of strips is required, the analyzer will request that a canister code
be entered from the new canister. The code can be entered manually from the keypad or
scanned with the Barcode reader, if installed.
A er self-tes ng or pressing on the ini al screen, the following screen will
be shown.
Scan or manually enter the canister code from the strip canister by pressing the data
entry box. A barcode reader can also be used to scan the canister code into the analyzer
if it is turned On. Press the data entry box to display the Numeric Keypad to enter the
new canister code manually. Press to cancel and exit the screen.
If the canister code is incorrect, the following screen will be shown.