NANOE Zetamix ZetaSinter User manual

2
ZetaMix System®
Thank you for purchasing the Zetasinter furnace. To avoid any misuse and damage, please read the operations manual
carefully before using.
Content
1. Furnace Description .................................................................................................................................. 3
2. Technical Specifications........................................................................................................................... 3
3. Safety .......................................................................................................................................................... 4
a. Intended use
b. Assembling
c. Operation
d. Maintenance
4. Assembling ................................................................................................................................................. 6
5. Temperature controller ............................................................................................................................. 8
6. MoSi2 HeatinG
elements………………………………………………………………………………………………………………………………………….…. 13
7.
Troubleshooting……………………………………………………………………………………………………………
……………………….14
8. Electrical schematic diagram................................................................................................................. 15
9. Quick start procedure ...................................................................................................................... 16
10. EU Declaration of conformity (No 2022-01-a) ..................................................................................... 17

3
ZetaMix System®
1. Furnace Description
The Zetasinter Chamber Furnace is a chamber furnace using MoSi2 heating elements, working
temperatures are between 800°C and 1550°C. Samples are to be placed inside the chamber, on
ceramic plates or crucibles.
2. Technical Specifications
Heating Power
9kW
Voltage / Frequency / Current
3 phases + neutral 380V / 50Hz / 32A
Maximum temp.
1600°C
Working Temp.
1550°C
Suggested heating rate
≤5°C/min
Temperature controller accuracy
±1°C
Thermocouple
B type
Heating elements
U type MoSi2
Size
860x605x1120mm
Chamber size
250*250*250 mm

4
ZetaMix System®
3. Safety
a. Intended use
Only parts printed with Zetamix by Nanoe® filaments can be heat treated in Zetasinter. Reference
names of Zetasinter by Nanoe® filaments are:
- Zetamix White Zirconia®
- Zetamix Black Zirconia®
- Zetamix Alumine®
Installation must be done according the Zetasinter chamber furnace facility guide.
Installation, use and maintenance must be done according to this operation manual.
b. Assembling
Assembling must be done according to this operation manual.
Danger due to electrical Current.
During installation and maintenance
work, the electric supply to the furnace
must be switched off.
Work on electrical equipments may only
be performed by qualified personnel.
The set-up instructions and safety
regulations must be observed,
otherwise the furnace will be deemed
to have been used incorrectly,
effectively cancelling any claims
against Nanoe.

5
ZetaMix System®
c. Operation
Temperature controller
For each material reference, the heat program must be defined by the operator according to
guidelines provided by Nanoe (www.zetamix.fr).
The maximum safety temperature is set at 1600°C. Above, the temperature controller automatically
stops the heat treatment.
The recommended maximum working temperature is 1550°C.
Heating and cooling rates may not exceed 5oC/min (except emergency stop).
Do not turn off the power supply of the equipment if the furnace temperature is above 500°C.
After replacing the temperature controller, its settings must be adjusted before use.
Security warning
Do not touch the inner or outer surface of the furnace during or immediately after use.
Do not place any objects on the furnace body.
d. Maintenance
The power supply must be switched off before any maintenance work.
Danger due to electrical current.
During installation and maintenance
work, the electric supply to the furnace
must be switched off.
Work on the electrical equipment may
only be performed by qualified
personnel.
Hot surfaces, danger of burning.
You may not always realize that surfaces,
such as the furnace walls are hot. Do not
touch the surface.

6
ZetaMix System®
4. Assembling
-Install the furnace according to the zetasinter chamber furnace facility guide.
-Remove the plastic cover from the heating elements under the top cover (remove the screws
to do this).
-Install the alumina tube. Put the refractory blocks on both sides of the tube (cf. Alumina tube
chapter).
-Close the circuit breaker, turn on the power →the green light comes on and the cooling fan
starts to work.
-Turn the “lock”button clockwise →the temperature control unit is now switched on.
-Press the “Turn On”button (green light is now on).
Temprature controller
Turn ON
Turn ON
Lock
Cycle Left, Down, Up arrows
Selection screen
T/time screen
Danger due to electrical current.
During installation and maintenance
work, the electric supply to the furnace
must be switched off.
Work on the electrical equipment may
only be performed by qualified
personnel.

7
ZetaMix System®
-For the first heating, the furnace chamber must be “baked”for 2 hours at 300℃. Using the
temperature controller, program and run the "baking" cycle (cf. Temperature controller
chapter):
(C01=50; t01=85; C02=300; t02=120; C03=300; t03=65; C04=100; t04=-121)
-For the first high temperature heating (above 400°C), heating element must be “passivated”
for 2 hours at 1200°C (cf. heating element chapter). Using the temperature controller,
program and run the "passivation" cycle:
(C01=50 ; t01=385 ; C02=1200; t02=120; C03=1200; t03=365; C04=100; t04=-121)
-Install the vacuum pump and the flowmeter (cf. Vacuum pump & flowmeter chapter).
-Install the bubbler (cf. Bubbler chapter).

8
ZetaMix System®
5. Temperature controller
Heating program setting
Program state symbols
Stop state
When you turn on the “Lock”, the temperature controller displays the model and version. After a few
seconds, it switches to the off-state.
-PV : “TC temperature”(50°C at room temperature).
-SV : “StoP” is alternately displayed on the lower window.
①Process value (Thermo couple temperature °C)
(P V)
②Setting value (temperature °C- Program state*)
(S V)
③Setting key (confirm KEY)
④Data shift key (and program setup entry)
⑤Data reduction key (and program RUN/HOLD)
⑥Data add key (and program STOP)
⑦Function indicator :
-PRG controller in running state
-OP1 power output
-OP2 power output level
-COM Communication interface
-Other lamps not use

9
ZetaMix System®
Heating program setting
①From stop state, Press to go to the setup program state.
②The set points of the step StEPXX are displayed (C XX temperature in °C, and t XX time in min).
③Press , and to modify the value.
④Press to move to the next parameter. The program parameters are displayed in the following
order: setpoint1, time1, setpoint2, time2.
-Press and hold to return to the previous setting. The program step can be changed at
any time, even if the program is running.
⑤Set t××= -121 to indicate the end of the program (automatic switch-off).
⑥Press and simultaneously to exit the settings. If no key is pressed within 25 seconds, it also
exits to stop state.
Example:

10
ZetaMix System®
Specific settings:
Time Setting

11
ZetaMix System®
Heating programs saving
Run/Hold states
Run state:
In Stop state, press and hold until the SV window displays the “run” symbol. The controller runs
the defined program.
Hold state:
In Run state, press and hold until the SV window displays the “HoLd” symbol. Controller holds the

12
ZetaMix System®
given temperature and stop the timer.
In Hold state, press the until the SV window displays the “run” symbol. Controller starts running
again.
Step number and timer:
In Run or Hold states, press to display the current step number and timer (PV: current step time
setting and SV: current step elapsed time).
Return to Stop state:
In Run or Hold states, press, and hold until the SV window displays the “StoP” symbol. The
controller will stop (free cooling), the timer will be reset, and the step number will be set to 1.
6. MoSi2 Heating elements
In order to form a protective layer against oxidation, new heating elements must be heated to
1200°C for 2 hours with a ramp of 3°C/min (cf 4. Assembling). This process should be done when using
a Zetasinter for the first time or after replacing the heating elements.
The resistance of MoSi2 heating elements increases with use and the performance of the furnace
may decrease. Used MoSi2 heating elements are very brittle, especially after heat treatment at
1200°C.
New and used MoSi2 heating elements can be used together.
Replacement
The power must be turned off before maintenance or inspection.
①Unscrew the top cover and remove it.
②Unscrew of the clip of the connection bridge (for each rods of the U type heating element) and
remove it.
③Take off the connection bridges.

13
ZetaMix System®
1) 2) 3)
④Unscrew the clip from the ceramic jaws that hold the heating element.
⑤Remove the ceramic jaws and the refractory block.
⑥Remove the heating element which needs to be replaces
⑦Replace it with a new heating element. Replace the refractory block and ceramic jaws, make
sure they all fit in the same position as before.
4) 5) 6) 7)
⑧When tightening the clip screw, make sure that the bottom of the heating element does not touch
the bottom of the furnace chamber.
⑨Then follow the procedure④③②①to tighten the screw and complete the heating element
replacement.
7. Troubleshooting
Failure
Code 1
Code 2
Explanation
Solution
No power
/
/
The furnace is not
connected to the
power supply.
Check the power supply
and the electrical
connection.
Check and close the 32A
circuit breaker on the
furnace body (left).
Lock key is
turned
clockwise
(right) but the
temperature
controller is not
/
/
The control circuit
is not connected
to the power
supply because its
2A circuit breaker
is open.
Identify (if possible) the
cause of the failure.
Check and close the 2A
circuit breaker inside the
furnace (behind the control
panel).

14
ZetaMix System®
switched on.
No heating
power.
Heating
elements
voltage and
current are 0
while the
temperature
controller is on
run state.
The green
"power" light
is off, and
the red
"open circuit"
light is on.
The heater power
circuit is not
connected to the
power supply
because its 32A
circuit breaker is
open.
Identify (if possible) the
cause of the failure.
Check and close the 32A
circuit breaker inside the
furnace (behind the control
panel).
The green
"power" light
is off and the
red "open
circuit" light is
off.
Same as the
previous failure
and
the red light is
broken.
Same as the previous failure
and
change the red light.
The current of
the heating
elements is 0
while their
voltage is
maximum.
/
A heating element
is broken.
Identify which heating
element is broken and
change it.
Check first under the top
cover and then inside the
heating chamber.
Controller
panel SV
indicate “Oral”
/
/
The thermocouple
circuit is broken.
Change the thermocouple.
Controller
panel SV
indicate “Hial”
/
/
Furnace
temperature >Max
Let the furnace cool down.
Check the temperature
controller and the
thermocouple.

15
ZetaMix System®
8. Electrical schematic diagram
(380V Electrical schematic diagram)

16
ZetaMix System®
9. Quick start procedure
HEATING CURVE SETTING
1. Turn on the temperature controller by turning the LOCK button.
2. The temperature controller is initializing and then displays the initial screen.
3. Press the LETF ARROW button to display the heating curve setting.
4. The screen is now displaying the first segment (C 01) and the associated temperature. This
temperature can be modified by pressing the UP or DOWN ARROWS. After pressing the UP or
DOWN ARROWS for a few second, the decimal point will be displaced to the left in order to
change the temperature faster. With the side arrows, you can choose to change the number
of units, tens, hundreds...
5. Once the first temperature is set, press the CYCLE button to move to the first segment time
setting (T 01). You can change the time displayed in minutes by pressing the UP or DOWN
ARROWS. With the side arrow, you can choose to change the number of units, tens,
hundreds...
6. Then you can cycle through the segment by pressing the CYCLE button to build the desired
heating curve. The number displayed by the letter C or T corresponds to the number of the
segment.
7. Note that a plateau can be added by choosing the same temperature as the temperature of
the previous segment. If so, the time setting of that segment corresponds to the time of the
plateau.
8. Finally, when all segments are built, you must add a final step by choosing -121 as the
temperature of the last segment and not assigning any time control to this step.
9. In order to exit the heating curve setting, you can wait 20s and the temperature controller will
automatically return to the initial screen. You can check your heating curve by entering the
heating curve setting again (back to point 3).
RUN A HEATING CURVE
1. To run the selected heating curve, you must first press the TURN ON button to close the circuit
breaker. The green button will light up.
2. Press the DOWN ARROW (run) for more than one second to start the heating curve.
3. You can hold the heating curve by pressing the DOWN ARROW for a little longer than one
second. To resume the heating curve, press the DOWN ARROW again for more than one
second. You can stop the heating curve by pressing the UP ARROW for more than one
second.

17
ZetaMix System®
10. EU DECLARATION OF CONFORMITY (No 2022-01-a)
1.
Product model: Zetasinter Chamber Furnace
2.
Name and address of the manufacturer:
Nanoe SAS, 6 rue des frenes, 91160 Ballainvilliers
+339 81 98 33 64
3.
This declaration of conformity is issued under the sole responsibility of the manufacturer. It is based
on evaluation on a sampling of the aboved mentionned model.
4.
Object of the declaration : Tube furnace model Zetasinter Chamber Furnace
5.
The object of the declaration described above is in conformity with the relevant Union
harmonisation legislation:
EU directive on low voltage electrical equipments 2014/35/UE
EU directive on machinery 2006/42/EC
6.
Applicable standards for this declaration are :
EN 60519-1/2015 : Safety in installations for electroheating and electromagnetic processing - Part 1:
General requirements
EN 60519-2/2015 : Safety in electroheat installations - Part 2: Particular requirements for resistance
heating equipment.
EN 50156-1/2015 : Electrical equipment for furnaces and ancillary equipment - Part 1: Requirements
for application design and installation.
Signed for and on behalf of: Nanoe SAS
Ballainvilliers, on the 25/01/2022:
Guillaume de Calan, CEO
This manual suits for next models
1
Table of contents
Popular Laboratory Equipment manuals by other brands

Uvitron
Uvitron PORTA-RAY instruction manual

CORNING
CORNING LSE 6709 instruction manual

MGI
MGI Complete Genomics DNBSEQ-G800RS user manual

MRC
MRC SPECTRO-V11D Operation manual

Carver
Carver Mini C Operation, installation, and maintenance manual

Waters
Waters Xevo G2 QTof Operator's, overview and maintenance guide