TECHNICAL
INSTRUCTIONS
Product: NewTom VGi – 5G – VGi evo – 5G XL
Code:
96600894 – GR MAIN WS VGi-5G-VGI EVO - UK
Object: MAIN WS REPLACEMENT
the guidelines for the replacement of the computer directly connected to the scanner
The provided code 9660089 includes the following parts:
979016 8 - WORKSTATION HP (UK)
97521890 - IMP.EL.CANBUS TERMINAT.9P F-F
97661231 - LAN PCI-E CARD
97661 66 – GRAPHIC ADAPTER FOR HP WORKSTATION
97661625 - CANBUS PCI-E KVASER CARD
In case the original workstation is still accessible, perform the following backups::
•Configuration data and calibration of the device (Menu File → Settings Backup).
•Patient exam.
•Network configuration (IP Address, Workgroup, Domain Name…).
•Potential personal files and documents, which needs to be restored in the new workstation.
Mount the LAN card included in the package inside the
workstation. For the complete and detailed installation
procedure, refer to the corresponding documentation (
“Service Manual” (VGi – 5G) or “Workstation Setup
Annex” (VGi evo – 5G XL).
Insert the CANBUS terminator on the CANBUS PCI card
and connect the CAN cable that comes from the scanner
(see image on the side).
RESTORE OF SOFTWARE COMPONENTS