NF IV-202F4 User manual

NF Corporation
TRANSIMPEDANCE AMPLIFIER
IV-202F4 / IV-204F3
INSTRUCTION MANUAL

DA00062341-002 1 / 11 IV-202F4 / IV-204F3
――― Preface ―――
Thank you for purchasing the IV-200 series transimpedance amplifier.
Please read the "Safety Information" on the next page first to ensure the product is used in a
correct and safe manner.
● Alert symbols in this manual
This manual uses the following alert symbols. The instructions next to these symbols must be
carefully obeyed to ensure the equipment operator’s safety and prevent damage to equipment.
This contains information for avoiding electric shock and other risks when handling
the equipment that could result in death or bodily injury to the operator.
This contains information for avoiding risks when handling the equipment that could
result in minor or moderate injury to the operator or damage to property.
CAUTION
WARNING

DA00062341-002 2 / 11 IV-202F4 / IV-204F3
────Safety Information────
To ensure safe use, be sure to observe the following safety precautions.
NF Corporation shall not be held liable for damages that arise from a failure to observe these safety
precautions or warnings or cautions in the instruction manual.
Be sure to observe the information of the instruction manual.
The instruction manual contains information for the safe operation of the product.
Be sure to read the information first before using the product.
All the warnings in the instruction manual are intended for preventing risks that may lead to serious
accidents. Ensure to obey them.
Be sure to ground the product.
To prevent electric shock, the enclosure must be connected to an electrical ground.
Check the power supply voltage.
This product operates on the power supply voltage indicated in the instruction manual.
Prior to connecting the power supply, check that the voltage of the power supply matches the rated
power supply of the product.
If you notice anything strange
If this product emits smoke, an unusual smell, or abnormal noise, immediately power it off and stop
using it.
If such an anomaly occurs, prevent anyone from using this product until it has been repaired, and
immediately report the problem to NF Corporation or one of our representatives.
Do not operate in an explosive atmosphere.
An explosion or other such hazard may result.
Do not remove the cover.
The product may contain high-voltage parts. Never remove the cover.
Even when the inside of this product needs to be inspected, do not touch the inside. All such
inspections are to be performed by service technicians designated by NF Corporation.
Do not modify the product.
Never modify the product. Modification to the product may pose a new risk. We may refuse the
repair of a modified product.
Ensure that water does not get into this product.
Using the product in wet condition may cause electric shock and fire. When water etc. get into the
product, immediately power it off, and contact NF Corporation or one of our representatives.
If lightning occurs, power off and disconnect this product.
A lightning may cause an electric shock, a fire and a failure.

Safety Information
DA00062341-002 3 / 11 IV-202F4 / IV-204F3
WARNING
Electromagnetic compatibility
This product is a CISPR 11 Group 1 Class A compliant device. This product may cause interference
if used in residential areas. Such use must be avoided unless the user takes special measures to reduce
electromagnetic emissions to prevent interference to the reception of radio and television broadcasts.
●Safety symbols
Caution, refer to instruction manual
This notifies the user of potential hazards and indicates that he/she must refer to the
instruction manual.
Caution, possibility of electric shock
This indicates that an electric shock may occur under specific conditions.
Protective earth ground terminal
This indicates that the terminal needs to be grounded in order to prevent electric shock
accidents.
Before operating the product, ensure to connect this terminal to an earth ground.
Warning
This contains information to avoid risks in equipment handling that could result
in loss of life or bodily injury.
Caution
This contains information to avoid risks in equipment handling that could result
in minor or moderate injury to person or damage to property or the equipment.
●Other symbols
This indicates that the terminal or the outer conductor of the connector is connected to the
enclosure.
This indicates that the terminal or the outer conductor of the connector is connected to the
signal ground.
This indicates that the terminal or the outer conductor of the connector is isolated from the
enclosure.
It indicates, however, that the difference of the potential from the ground potential is
restricted to 42 Vpk or less for safety (this product is used grounded so the enclosure
potential and ground potential are equal).
Waste disposal
This product does not include batteries and is designed in accordance with the RoHS Directive (EU).
Dispose of the product in accordance with the laws and regulations of the area where you use it.
CAUTION

1. Before Using Product
DA00062341-002 4 / 11 IV-202F4 / IV-204F3
1. Before Using Product
1.1 About Power Connection
1.1.1 Connecting to Dedicated LP Series Power Supply
To obtain the best noise characteristics, we recommend the dedicated low noise DC power supply
LP series.
To connect this product to this series of power supply, use the “Output cable D”(PA-001-3029)
as shown below.
Figure 1-1 Connecting to LP series power supply
Figure 1-2 Pin assignment
This product will be damaged by reversely connecting the power supply. Supplying a voltage
greater than ± 16.5 V will damage the internal circuits of this product.
There are concave and convex parts for the power connector, as shown in Figure 1-2, to prevent the
wrong connection. If the orientation of the connector is correct, excessive force is not required.
Please be careful of the orientation of the connector.
Securely connect the cable to the power connector of the main unit.
Attention
Convex
Concave
Main unit
LP series power supply
(±15V)

1. Before Using Product
DA00062341-002 5 / 11 IV-202F4 / IV-204F3
1.1.2 Connecting to Power Supply other than LP Series
To connect this product to a power supply other than one of the LP series, use the supplied power
supply cable as shown below.
Connect the connector of the cable to the power connector of the main unit.
Connect the split end of the power supply cable to the ±15 V power supply
(split end: red to +15 V, black to GND, blue to -15 V).
Figure 1-3 Connecting to power supply other than LP series
Figure 1-4 Pin assignment
This product will be damaged by reversely connecting the power supply. Supplying a voltage
greater than ± 16.5 V will damage the internal circuits of this product.
There are concave and convex parts in the power connector, as shown in Figure 1-4, to prevent the
wrong connection. If the orientation of the connector is correct, excessive force is not required.
Please be careful of the orientation of the connector.
Securely connect the cable to the power connector of the main unit.
This product does not operate on a single power supply (+30 V / GND). Be sure to use a dual
power supply (+15 V / GND / −15 V).
Attention
Main unit
Power supply (±15V)
Power supply cable (supplied)
Convex
Concave

1. Before Using Product
DA00062341-002 6 / 11 IV-202F4 / IV-204F3
1.2 Signal Input and Output Connection
An SMB connector is used for signal input and output. However, there is a restriction in the
connector portion due to the product’s outer shape. Use a connector with an external diameter of
φ6.2 mm or smaller for connection.
If a large parallel capacitance is connected to the input of this product, overshooting, ringing, or
oscillation may occur. Make the input wiring as short as possible. We recommend 50-Ω coaxial
cables with lengths of approximately 10 cm and 40 cm for the IV-202F4 and IV-204F3,
respectively.
The absolute maximum rating of the input current is ±20 mA. Be aware of this value to avoid
causing damage.
Be careful of input short-circuits that could damage devices connected to the output.
1.3 Adjuster for Frequency Response (FREQ ADJ)
If a large input parallel capacitance ※
1-1 is connected to this product, overshooting, ringing, or
oscillation may occur. In this case, the FREQ ADJ adjuster may be able to improve measurement
stability.
To compensate for input parallel capacitance, make adjustments by turning the adjuster clockwise.
However, the upper cut-off frequency becomes lower as you turn the adjuster clockwise.
※1-1:Approximately 20 pF or higher (IV-202F4).Approximately 50 pF or higher (IV-204F3).

2. Reference Data
DA00062341-002 7 / 11 IV-202F4 / IV-204F3
10f
100f
1p
10p
100p
10 100 1k 10k 100k 1M 10M
Frequency[Hz]
雑音電流密度[V/√Hz]
450fA/√Hz typ. @10kHz
1f
10f
100f
1p
10p
10 100 1k 10k 100k 1M
Frequency[Hz]
雑音電流密度[V/√Hz]
70fA/√Hz typ. @1kHz
-50
-40
-30
-20
-10
0
10
110 100 1k 10k 100k 1M 10M
Frequency[Hz]
Gain[dB]
遮断周波数=1.5MHz
0dB=4MV/A
-50
-40
-30
-20
-10
0
10
110 100 1k 10k 100k 1M 10M 100M
Frequency[Hz]
Gain[dB]
遮断周波数=13MHz
0dB=100kV/A
2. Reference Data
This chapter provides reference data.Although the performance of this product may not achieve
the level of these data, please understand that each item of all products has been strictly tested and
confirmed to meet the specifications before shipment.
2.1 Frequency Characteristics
The frequency characteristics shown below represent the gain based on 10 kHz.
Figure 2-1 Frequency characteristic (IV-202F4) Figure 2-2 Frequency characteristic (IV-204F3)
2.2 Equivalent Input Noise Current Density
The characteristics shown below are calculated by dividing the output noise voltage density by
the current gain.
Figure 2-3 Noise characteristic (IV-202F4) Figure 2-4 Noise characteristic (IV-204F3)
fc=
fc
Equivalent input noise current density [A/√Hz]
Equivalent input noise current density [A/√Hz]

3. Specifications
DA00062341-002 8 / 11 IV-202F4 / IV-204F3
3. Specifications
3.1 Overview
The IV-202F4 and IV-204F3 are compact current-to-voltage conversion amplifiers with low noise
characteristics. Wide bandwidths of up to 10 MHz and 1 MHz are realized for the IV-202F4 and
IV-204F3, respectively.
These products stably operate for up to an input parallel capacitance of approximately 20 pF
(IV-202F4) or 50 pF (IV-204F3) without adjustment.
3.2 Configuration
Main unit ×1
Power supply cable ×1
Instruction manual ×1
3.3 Absolute Maximum Rating
Power supply voltage
(± VS)
±18 V
Input current
± 20 mA
3.4 Electrical Characteristics
Unless otherwise specified, power supply voltage Vs = ± 15 V, temperature 23 ± 5 °C, and input
parallel capacitance Cs = 5 pF.
3.4.1 Input Section
Item/Model
IV-202F4
IV-204F3
Note
Input form
DC coupling unbalanced single ended input
SMB connector (male)
Input impedance
30 Ω or lower
1.2 kΩor lower
at DC
Input bias current
5 pA typical
—
Maximum input current
±100 μA
±2.5 μA
—
Equivalent input noise
current density
600 fA/ Hz or lower
450 fA/ Hz typical
90 fA/ Hz or lower
70 fA/ Hz typical
Input opened
IV-202F4: at 10 kHz
IV-204F3: at 1 kHz
Equivalent input noise current
20 nArms
or lower typical
1.1 nArms
or lower typical
Input opened
BW: 3 Hz to 20 MHz
Recommended signal
source impedance
25 kΩor higher,
20 pF or lower
1 MΩor higher,
50 pF or lower
—
3.4.2 Output Section
Item/Model
IV-202F4
IV-204F3
Note
Output form
DC coupling unbalanced single ended output
SMB connector (male)
Phase between input and output
Inverting
—
Output impedance
50 Ω± 10 %
—
Maximum output voltage
±10 V
—
Maximum output current
±5 mA
—
Output offset voltage
±10 mV
Input opened

3. Specifications
DA00062341-002 9 / 11 IV-202F4 / IV-204F3
3.4.3 Amplifier Section
Item/Model
IV-202F4
IV-204F3
Note
Gain
1 × 105V/A ± 5%
4 × 106V/A ± 5%
—
Frequency bandwidth
DC to 10 MHz
DC to 1 MHz
Based on 10 kHz
+1 to −3 dB
3.5 Power Supply
Item/Model
IV-202F4
IV-204F3
Note
Operating supply voltage range
±15 ± 1 V
—
Consumption current
±40 mA or lower
±25 mA typical
No signal
Vs = ± 15 V
Connector
DF11-4DP-2DS(52) , 4 pins, male
by Hirose Electric Co., Ltd.
—
Pin Assignment
Pin number
1
2
3
4
Function
+Vs
GND
GND
−Vs
Figure 3-1 Pin assignment diagram
3.6 General
Performance guarantee temperature range 23 ± 5°C
Operating temperature and humidity ranges 0 to 40°C, 10 to 90 %RH (non-condensation)
Storage temperature and humidity ranges −10 to 60°C, 10 to 80 %RH (non-condensation)
External dimensions 80 × 15 × 21 mm (without protrusions)
Weight Approximately 25 g
RoHS Directive 2011/65/EU
3.7 Adjuster for Frequency Response (FREQ ADJ)
If a large input parallel capacitance*3-1 is connected to this product, overshooting, ringing, or
oscillation may occur. In this case, the FREQ ADJ adjuster may be able to improve measurement
stability.
To compensate for input parallel capacitance, make adjustments by turning the adjuster clockwise.
However, the upper cut-off frequency becomes lower as you turn the adjuster clockwise.
* 3-1: approximately 20 pF or higher (IV-202F4). Approximately 50 pF or higher (IV-204F3).
3 4
1 2

3. Specifications
DA00062341-002 10 / 11 IV-202F4 / IV-204F3
3.8 Power Supply Cable (Supplied)
Length (including connector) Approximately 50 cm
IV-202F4/IV-204F3 side DF11-4DS-2C(52), 4 pins, female
Hirose Electric Co., Ltd.
Power supply side Split end, 4 wires (0.2 mm2)
Red to +Vs, Blue to −Vs, Black (2 lines) to GND
3.9 External View
* For the SMB connector used as the input and output connector, use a connector with a diameter
of φ6.2 mm or smaller.
* The external view shown above is that of the IV-202F4. The external dimensions of the IV-204F3
are the same; only the model name is different.
Figure 3-2 External view
Adjuster for frequency response
Material and Surface Treatment Unit: mm
Panel: Aluminum, Alumite treatment
Case: Aluminum, non-chromium chemical conversion coating
Power supply connector
Output connector
Input connector

3. Specifications
DA00062341-002 11 / 11 IV-202F4 / IV-204F3
3.10 Block Diagram
Figure 3-3 Block Diagram

ーーーーWARRANTY ーーーー
The NF Corporation certifies that this product was thoroughly tested and inspected and found
to meet its published specifications when it was shipped from our factory.
All NF products are warranted against defects in materials and workmanship for a period of one
year from the date of shipment. During the warranty period, NF will repair a defective product
without any charge for the parts and labor. For repair service under warranty, the product must be
returned to either NF or an agent designated by NF. The purchaser shall prepay all shipping
charges, duties, and taxes for the product to either NF or an agent from another country, and the
shipping charge for returning the product to the purchaser shall be paid by NF.
This warranty shall not apply to any defect, failure, or damage caused by a) improper use; b)
improper or inadequate maintenance and care; or c) modifications made by the purchaser or
personnel other than NF representatives.
If there are any misplaced or missing pages, we will replace the manual. Please contact a sales
representative.
NOTES
Reproduction of the contents of this manual is forbidden by applicable laws.
The contents of this manual may be revised without notice.
Information provided in this manual is intended to be accurate and reliable. However, we
assume no responsibility for any damage regarding the contents of this manual.
We assume no responsibility for influences resulting from the operations in this manual.
Copyright 2017, NF Corporation
IV-202F4/IV-204F3 INSTRUCTION MANUAL
NF Corporation
6-3-20, Tsunashima Higashi, Kohoku-ku, Yokohama
223-8508 JAPAN
Phone: +81-45-545-8111 Fax: +81-45-545-8191

http://www.nfcorp.co.jp/ NF Corporation
6-3-20, Tsunashima Higashi, Kohoku-ku, Yokohama 223-8508 JAPAN
This manual suits for next models
1
Table of contents
Other NF Amplifier manuals
Popular Amplifier manuals by other brands

Soundstream
Soundstream SA.120 Owner's manual and installation guide

FMR Audio
FMR Audio R-390B quick start guide

Marshall Amplification
Marshall Amplification VDA-104-3GS-2 user manual

Audio Note
Audio Note KAGEKI BALANCED Owner's Information

Roland
Roland SA-1000 Connection guide

Onkyo
Onkyo TX-SV515PRO II instruction manual