NOVOTEST MF-1M User manual

Operating manual
Magnetometer
NOVOTEST MF-1M
NOVOTEST
Phone: +38-067-593-59-77
Email: sales@novotest.biz
Web: http://novotest.biz

RDC.OD.MF-1M.000 OM
Operating manual
Magnetometer NOVOTEST MF-1M
Page 2
CONTENTS
1 Description and operation of the device and its components.................................................................5
1.1 Purpose............................................................................................................................................5
1.2 Specifications..................................................................................................................................5
1.3 Package ...........................................................................................................................................5
1.4 Composition....................................................................................................................................6
1.5 Design and operation ......................................................................................................................7
1.6 Hall probe........................................................................................................................................7
1.7 Measurement instruments and accessories .....................................................................................8
1.8 Labeling and sealing .......................................................................................................................8
1.9 Packaging........................................................................................................................................8
2 Intended use ...........................................................................................................................................9
2.1 Operating restrictions......................................................................................................................9
2.2 Preparation for the use ....................................................................................................................9
2.2.1 Visual inspection......................................................................................................................9
2.2.2 Installation of batteries.............................................................................................................9
2.2.3 Battery charging.......................................................................................................................9
2.2.4 Hall probe connection............................................................................................................10
2.3 Operation.......................................................................................................................................11
2.3.1 Turning on..............................................................................................................................11
2.3.2 “Measurement” section..........................................................................................................11
2.3.2.1 Selection of measurement units ..........................................................................................11
2.3.2.2 Measurement modes ...........................................................................................................12
2.3.3 Measurement..........................................................................................................................13
2.3.4 Calibration..............................................................................................................................13
2.3.5 “Archive” section...................................................................................................................14
2.3.6 “Settings” section...................................................................................................................15
2.3.7 Memory card..........................................................................................................................15
2.3.8 Information.............................................................................................................................17
2.3.9 Turning off.............................................................................................................................17
3 Maintenance of the device and its components ...................................................................................18
3.1 Safety precautions.........................................................................................................................18
3.2 Verification ...................................................................................................................................18
3.2.1 Preconditions for verification.................................................................................................18
3.2.2 Steps and accessories of verification .....................................................................................18

RDC.OD.MF-1M.000 OM
Operating manual
Magnetometer NOVOTEST MF-1M
Page 3
3.2.3 Visual inspection....................................................................................................................19
3.2.4 Testing....................................................................................................................................19
3.2.5 Measurement error definition.................................................................................................19
3.2.6 Documentation of verification...............................................................................................19
3.3 Warranty obligations.....................................................................................................................19
3.3.1 Basic warranty........................................................................................................................19
3.3.2 Extended warranty .................................................................................................................20
3.3.3 Warranty for repaired or replaced parts.................................................................................20
3.3.4 Wearing parts.........................................................................................................................20
3.3.5 Owner obligations..................................................................................................................20
3.3.6 Warranty restrictions..............................................................................................................21
3.3.7 Cases uncovered by warranty ................................................................................................21
3.3.8 Warrants and consumer law...................................................................................................21
3.4 Maintenance..................................................................................................................................22
4 Running repairs....................................................................................................................................23
5 Storage .................................................................................................................................................23
6 Transportation......................................................................................................................................23
7 Disposal................................................................................................................................................23

RDC.OD.MF-1M.000 OM
Operating manual
Magnetometer NOVOTEST MF-1M
Page 4
Caution!
Please read this operating manual carefully before using the Magnetometer NOVOTEST MF-1M.
This operating manual (hereinafter referred to as OM) includes the information required to inform
user about the performance and operating procedure of the product, magnetometer NOVOTEST MF-
1M (hereinafter referred to as the device or magnetometer). The document contains the specification,
description of the design and operating principle, as well as the information necessary for the right
product operation. Before using the device, user must read this manual, as the device must be operated
by a person aware of the operating principle and design of the device.
The right and efficient operation of the device provides for the following:
−measurement technique;
−conditions of measurement that meet the measurement technique;
−a trained user who has read this operating manual.
The manufacturer reserves the right to make minor changes that do not impair the technical
specification of the product. These changes may not be mentioned in this document.
The delivery set of the device includes the operation documents that are a part of this manual and
the device registration certificate.
This operating manual applies to all product modifications.

RDC.OD.MF-1M.000 OM
Operating manual
Magnetometer NOVOTEST MF-1M
Page 5
1 DESCRIPTION AND OPERATION OF THE DEVICE AND ITS COMPONENTS
1.1 Purpose
Magnetometer (teslameter) NOVOTEST MF-1M is designed to measure magnetic field strength
(magnetic induction). The device is used in industrial and construction fields.
The magnetometer is designed to test the following parameters:
−checking of equipment for compliance with the required technical specifications;
−magnetic field measurement during testing of ferromagnetic products by the magnetic particle
inspection;
−definition of residual magnetization;
−monitoring of permissible values of industrial interference;
−research and testing of magnetic field levels.
1.2 Specifications
A magnetometer is a portable device containing a circuit board with electronic components and a
battery. The magnetic field transducer (Hall probe) is designed to measure the intensity of a constant
magnetic field.
The main specifications of the device are shown in tab. 1.1.
Table 1.1 – The main characteristics of the device
Measurement range, G
±100
±1000
±2500
Measurement accuracy, G
±(1+0.05H) where Н –
module of the numerical
value of readings of the
magnetometer expressed in
Gauss
Operating conditions:
– ambient temperature for the electronic unit, оС
– ambient temperature for the probe, оС
– relative air humidity, %
– air pressure, kPa
from -10 to +40
from -20 to +40
up to 98 at +25 оC
from 84 to 106.7
Display resolution, G
0.1
Setting time, s, max
1
Dimensions:
– electronic unit, mm
– probe, mm
120х75х36
Ø20х115
Weight:
– electronic unit, kg, max
– probe, kg, max
0.25
0.08
Powering from two Ni-MH AA batteries (or alkali batteries)
1.2 (1.5) V
Operating time, h, min
20
Measurement units
G, mT, A/cm
1.3 Package
Electronic unit of magnetometer.............................................................................................1 pc.
Hall probe................................................................................................................................1 pc.

RDC.OD.MF-1M.000 OM
Operating manual
Magnetometer NOVOTEST MF-1M
Page 6
АА battery...............................................................................................................................2 pc.
Protective silicone case ...........................................................................................................1 pc.
Charger....................................................................................................................................1 pc.
Operating manual RDC.OD.MF-1M.000 OM........................................................................1 pc.
Transport packaging................................................................................................................1 pc.
* By request of the customer, the package can be added by extra equipment or parts. The exact
information about the scope of delivery is specified in the datasheet.
1.4 Composition
The magnetometer consists of the electronic unit and a Hall probe connected through a socket on
the top part of the electronic unit (fig. 1.1).
The body of the device is made of a shockproof ABS plastic. The front panel of the electronic unit
features a graphic display and control buttons. The graphic display shows the measured values of the
magnetic field and the state of charge of the battery. The bottom part of the rear panel features a
battery compartment.
1 – socket for the probe connection; 2 – protective silicone case; 3 – graphic display; 4 – keyboard;
5 – serial number of the device; 6 – Hall probe; 7 – the probe pin; 8 – serial number of the probe; 9 –
battery compartment; 10 – АА battery; 11 – battery compartment cover.
Figure 1.1 – Magnetometer NOVOTEST MF-1M
Control buttons and their functions:
– turning off and on;
– selection of measurement units;
– selection of measurement mode;
– right functional button;
– navigation up in the menu/increase of the input value;

RDC.OD.MF-1M.000 OM
Operating manual
Magnetometer NOVOTEST MF-1M
Page 7
1.5 Design and operation
The operating principle of a magnetometer is based on the physical Hall effect using the Hall
linear elements (transducers) of PH type.
The structural scheme of the magnetometer is shown on the fig.1.2.
Х – linear Hall element; ADC – analog-to-digital converter;
M – microprocessor controller; G – graphic display; PS – power source.
Figure 1.2 – Schematic stricture of magnetometer NOVOTEST MF-1M
The Hall element converts the magnitude of the magnetic field into a voltage which is fed to the
input of analog-to-digital converter “ADC”. The synchronous analog-to-digital conversion of the
signal leads to a cyclic sequence of digital codes, based on which the microprocessor controller “M”
calculates the exact value of Hall electromotive force and converts it into measurement unit of
induction or magnetic field strength and displays the reading in digital or graphic form on the graphic
display “G”. The microprocessor controller “M” also monitors the status of the “PS” power system of
the magnetometer.
1.6 Hall probe
The working surface of the probe features a hole with the Hall element. The sensitive area of the
Hall element is arranged along the working surface of the magnetic field transducer.
The magnetometer readings correspond to the magnitude of the constant magnetic field induction
vector component normal to the working surface of the magnetic field transducer Вithat is numerically
equal to the scalar product of the magnetic field induction vector B and the unit normal vector n (fig.
1.3).
– navigation down in the menu/decrease of the input value.
M
ADC
G
Х
PS

RDC.OD.MF-1M.000 OM
Operating manual
Magnetometer NOVOTEST MF-1M
Page 8
(a) (b)
Figure 1.3 – The readings depending on the passing of the magnetic induction vector through the
Hall probe.
The positive reading of the magnetometer means that the normal component of the induction
vector of a constant magnetic field at the measurement point goes outside the working surface of the
magnetic field transducer. (fig. 1.3а).
A negative reading of the magnetometer means that the normal component of the induction vector
of a constant magnetic field at the measurement point is within the working surface of the magnetic
field transducer (fig. 1.3б).
1.7 Measurement instruments and accessories
The adjustment and setting of the device in case of detection of malfunctions should be performed
by the manufacturer.
1.8 Labeling and sealing
The front panel of the device features the designation of the device with the trademark of the
manufacturer.
The rear panel of the device features the factory number.
1.9 Packaging
The electronic unit and the converter are delivered in packaging containers that eliminate their
damage during transportation.
To avoid damage to the cable and the connectors of the device, it is necessary to disconnect the
converter from the device before placing it in the packaging container.

RDC.OD.MF-1M.000 OM
Operating manual
Magnetometer NOVOTEST MF-1M
Page 9
2 INTENDED USE
2.1 Operating restrictions
The device should be operated away from direct exposure to dust and aggressive environments
with consideration of the parameters of the test objects under the agreed technical requirements and
used within its specifications.
The device must be operated by a user who have read this operating manual.
After transporting the device to the place of operation at a negative ambient temperature and
introducing it into a room with a positive temperature, the product must be kept in the package for at
least 6 hours to avoid damage to the device due to moisture buildup.
2.2 Preparation for the use
2.2.1 Visual inspection
The device must be visually inspected and checked for damages of the processing unit, Hall probe,
socket, and connecting cable.
2.2.2 Installation of batteries
Install the batteries in the battery compartment by removing protective silicone case and opening
the compartment cover by gently pressing and pulling down until the cover is completely detached.
The batteries or battery are installed according to the polarity indicated on the device. Close the battery
cover until it clicks into place and put on the protective silicone case. (fig. 2.1).
Figure 2.1 – Battery installation
2.2.3 Battery charging
Install the batteries in the battery compartment, having previously removed the protective silicone
case and opened the compartment cover by gently pressing and pulling it down until the cover is
completely detached. The batteries or a battery should be installed according to the polarity indicated
on the device. Close the battery cover until it clicks and put a protective silicone case back on.
To charge the battery, user should to the following steps:
−Connect the battery to the charger terminals;
−Plug the charger back in.
The battery gets fully charged within 14 hours. The charger shall not be left unattended during the
charge.
To avoid the battery failure during prolonged storage, the battery should be recharged with an
interval of at least 2 months, even if it has not been used.

RDC.OD.MF-1M.000 OM
Operating manual
Magnetometer NOVOTEST MF-1M
Page 10
2.2.4 Hall probe connection
Connect the Hall probe via the connecting cable to the probe socket on the processing unit.
Connect the connecting cable so that the red dot on the plug and the connector are aligned (fig. 2.2).
Figure 2.2 – The probe connection
Caution!
To prevent damage to the connectors and cables, follow the instructions for working with these
connectors below.
The connectors used in the device (fig. 2.3) consist of two parts: the socket of the device and pin (plug)
of the cable.
Figure 2.3 – Connectors of the device
The connection and disconnection between socket and pin are shown in fig. 2.4.
Caution!
Disconnecting the pin from the socket, hold it by the rubbed part of the body, do not pull the cable!
Figure 2.4 – Handling of connectors
Socket of the device
Pin of the cable
Connection
Wrong disconnection
Right disconnection

RDC.OD.MF-1M.000 OM
Operating manual
Magnetometer NOVOTEST MF-1M
Page 11
Note – Connection and disconnection from the processing unit are allowed only with the
magnetometer powered on.
2.3 Operation
2.3.1 Turning on
Turning on is performed by pressing “ ” until a short screensaver appears on the graphic
display (fig. 2.5).
Figure 2.5 – Screensaver after turning on
The device then goes to the main menu that can be navigated by quick pressing of “ ” and
“ ”, selection of the section is made by pressing “ ”.
2.3.2 “Measurement” section
When the “Measurement” section is selected (fig. 2.6), the device goes to the measurement
depending on the chosen parameters (fig. 2.7).
Figure 2.6 – “Measurement” section
Figure 2.7 – Displaying of information during the operation
2.3.2.1 Selection of measurement units
Selection of measurement units is made by a quick pressing “ ”. The device will offer the
following measurement units (fig. 2.8):
– Gauss;
– Millitesla;
– Ampere/cm;
–CODE.
Measurement unit
Battery charge status
The measured magnetic
induction value

RDC.OD.MF-1M.000 OM
Operating manual
Magnetometer NOVOTEST MF-1M
Page 12
Note – The CODE scale is used for calibration by the manufacturer.
Figure 2.8 – Selection of measurement units
Select the needed measurement unit by “ ” and “ ” and press “ ”. The selected
measurement unit is shown on the graphic display of the device.
After turning on, the magnetometer will measure in Gauss.
2.3.2.2 Measurement modes
To select the measurement mode, press “ ”. The device will offer the following measurement
modes (fig. 2.9):
– Normal – normal operation mode;
– Graphical – graphical mode.
Figure 2.9 – Selection of measurement modes
In the “Normal” mode the device will display the current measurement value (fig. 2.10).
Figure 2.10 – Displaying of the reading in the “Normal” mode
In “Graphical” mode, the graphic display of the magnetometer will show the following
information (fig. 2.11).
The numeric value of the set range
The current measured value
The current position of the
measured value on the
graph
Graphic value of the
threshold

RDC.OD.MF-1M.000 OM
Operating manual
Magnetometer NOVOTEST MF-1M
Page 13
Figure 2.11 – The graphic scheme of the device operation
The threshold level and range are set in the “Settings” section. If the threshold of the measured
value is exceeded, the sound signal is generated (the signal can be disabled in the “Settings” section).
2.3.3 Measurement
To measure the normal component of the induction vector of the constant magnetic field on the
surface of a metal product or a permanent magnet, install the Hall probe with the working surface
against the needed place of the test object (fig. 2.12).
Figure 2.12 – Measurement of the normal component of the induction vector of a constant magnetic
field on the end and side surface of the test object
To measure the tangential (contacting) component of the induction vector of a constant magnetic
field on the surface of a metal product or a permanent magnet, install the probe with the edge of the
working place against the needed place of the test object (fig. 2.13).
Figure 2.13 – Measurement of the tangential component of the magnetic field induction vector on the
end and side surfaces of the test object
The value of the measured value will be shown in the graphic display of the device (fig. 2.14).
Figure 2.14 – Displaying of the measured values of a magnetic induction
2.3.4 Calibration
Caution!
Calibration shall be performed only by a highly qualified specialist using the magnets with a valid
verification certificate, and only if required operationally!
When the “CALIBRATION” section is entered (fig. 2.15), the display will show the information
as on the fig. 2.16.

RDC.OD.MF-1M.000 OM
Operating manual
Magnetometer NOVOTEST MF-1M
Page 14
Figure 2.15 – “Calibration” section
Figure 2.16 – “Calibration” section
A magnetometer shall be calibrated by four points: 0 G, + 999 G, -999 G, and 500 G with the
magnetic inductometer SH1-1:
1. Calibrating by the first point, user must set the zero, for that the probe shall be put in a
horizontal position away from sources of magnetic fields (fig. 2.17), and then set 0 G on in
the «reference value» section of the display
Figure 2.17 – The probe position when zeroing
2. For a calibration by the second point, circuit the magnetometer with magnetic induction meter
SH1-1 under the instruction manual of the magnetic inductometer.
3. Turn SH1-1 on and set the reference value of magnetic induction as +999 G. Set +999 G on the
display of the magnetometer in the “reference value” cell.
4. Find the maximum code value at one of the points of the measured circuit and save the value in
the probe memory by pressing the “edit” button.
5. For calibration by the third point, turn the probe by 180° and set -999 G in the “nominal value”
cell of the magnetometer and perform operations described in c.4.
6. For calibration by the fourth point set the magnetic induction value as 500 G in SH1-1, and set
500 G in the “reference value” cell of a magnetometer and do the steps described in c.4.
2.3.5 “Archive” section
When the “Archive” mode is selected (fig. 2.18), the device goes to the list of the saved
measurements. The device can store up to 256 measurements.
Figure 2.18 – “Archive” section
The current value of
the code
The calculated value of the
function (polynomial)

RDC.OD.MF-1M.000 OM
Operating manual
Magnetometer NOVOTEST MF-1M
Page 15
2.3.6 “Settings” section
When the “Setting” section is selected, the device goes to its settings (fig. 2.19).
Figure 2.19 – “Settings” section
2.3.6.2 The device settings
This menu features the setting of the following parameters (fig. 2.20):
−Language: selection of the language of the device (Russian, English, Spanish);
−Brightness: change of the display brightness (10. 20. …, 100%);
−Contrast: change of the display contrast (10. 20. …, 100%);
−Auto turning off: setting the automatic turnoff when the device is not in use (OFF, 1 minute, 5
minutes, 10 minutes, 30 the minutes).
−Keyboard sound: setting the sound signal when pressing keys (OFF, ON.);
−Scale: Selection of measurement units (Gauss, mT, A/cm, COD);
−Range: measurement of the magnetic induction is performed only within the set range (±1000
G);
−А. scale: (М1, М2, OFF);
−Signal: turning on and off of the sound signal (OFF, ON). If the measured value exceeds the
threshold, the sound signal will be generated.
−Level: setting the measurement threshold of the magnetic induction.
Figure 2.20 – The device settings
2.3.7 Memory card
The device features the memory card for storage of the measurement values in the memory.
2.3.7.1 Calibration backup
Entering the “Memory card” section (fig. 2.21), the device goes to the memory menu (fig. 2.22).
Figure 2.21 – “Memory card” section

RDC.OD.MF-1M.000 OM
Operating manual
Magnetometer NOVOTEST MF-1M
Page 16
Figure 2.22 – Calibration backup
It is recommended to back up the calibration before calibrating the factory settings of the probe.
This option allows for recovering the calibration after incorrect settings.
2.3.7.2 Downloading backups
The saved calibrations can be downloaded from the Hall probe memory. This function is needed
to recover the right calibration in the event of incorrect settings of the probe. (fig. 2.23).
Figure 2.23 – Downloading the calibration backups
2.3.7.3 Memory card clearance
Selecting this section (fig. 2.24), the data stored in the archive of the device will be cleared. After
clearing the SD card, the Archive will be empty.
Figure 2.24 – Menu of the memory card clearance
2.3.7.4 Memory state
When the “Memory state” section is selected (fig. 2.25), the table will appear showing the
allowable number of saved measurements, the already saved measurements, and the measurements to
be saved (fig. 2.26).
Figure 2.25 – The memory state menu

RDC.OD.MF-1M.000 OM
Operating manual
Magnetometer NOVOTEST MF-1M
Page 17
Figure 2.26 – Viewing of the memory stare
2.3.8 Information
In this menu section (fig. 2.27) user can view the information about the device, manufacturer and
representative offices around the world. The “INFORMATION” mode also allows viewing the serial
numbers of the device and the probe, the firmware version of the magnetometer and the number of
measurements, for this the needed bookmark should be opened using the keys “ ” and “ ”.
Figure 2.27 – “Information” section
2.3.9 Turning off
The device turns off by a continuous pressing of “ ” button.

RDC.OD.MF-1M.000 OM
Operating manual
Magnetometer NOVOTEST MF-1M
Page 18
3 MAINTENANCE OF THE DEVICE AND ITS COMPONENTS
3.1 Safety precautions
After commissioning of the device, it is recommended to inspect it periodically to check the
following:
−operability;
−compliance with the operating requirements;
−battery charge level;
−damage to the device components.
When using the charger connected to 220V at 50 Hz, user should observe the regulations in
“Safety procedure when operating consumer electronic units”.
If the device is not in use for a long time, the batteries must be removed from the battery
compartment.
The device can be used only by the persons instructed and certified by the ІІ qualification category
of safety engineering for the operation of measurement devices.
3.2 Verification
The verification is recommended to be done at least once a year.
The verification technique hereinafter referred to as verification, applies to the Magnetometer
NOVOTEST MF-1M and stipulates the methods and accessories of primary and periodic verification.
The device shall be verified by the State metrology service or other authorized authorities and
organizations with the right of verification.
Note – Verification of the device and the certificate of verification is executed only under customer
order.
3.2.1 Preconditions for verification
Verifications can be performed in the following conditions:
– ambient temperature: +20 ±2оС;
– relative air humidity: 65 ±15%;
– air pressure: 100 ±8 kPa.
3.2.2 Steps and accessories of verification
During verification, user shall do the steps and use the accessories specified in table 3.1.
Table 3.1 – Steps and accessories of verification
Step of verification Number
of points Verification
accessories
Compulsoriness
after production
and repairs
operation and
storage
1. Visual inspection
3.2.3
Yes
Yes
2. Testing
3.2.4
Yes
Yes
3. Measurement error
definition 3.2.5
Magnetic
inductometer
SH1-1
Yes
Yes
4. Documentation of
verification
3.2.6
Yes
Yes

RDC.OD.MF-1M.000 OM
Operating manual
Magnetometer NOVOTEST MF-1M
Page 19
3.2.3 Visual inspection
The device shall be packaged in line with to c. 1.3 of RDC.OD.MF-1M.000.OM.
The user should visually check the processing unit, the connecting cable and the probe; the
insulation and marking must be in normal condition and no damages shall be found. User should also
check the connectors for cleanness and the absence of disconnected or loosely fixed components (it
can be checked by a slight shaking the device), as well as tightness of fixing of the probe pin in the
socket on the electronic unit.
3.2.4 Testing
The device shall be prepared for operation in line with c. 2.2 of RDC.OD.MF-1M.000.OM.
The control buttons of the device and switching of the measurement modes must be checked.
3.2.5 Measurement error definition
The permissible measurement error should be defined with magnetic inductometer SH1-1 by the
following procedure:
1. Circuit the magnetometer and magnetic inductometer SH1-1 under the operating manual of the
magnetic inductometer.
2. Turn on the magnetic inductometer SH1-1 and set the magnetic field to 1 mT, and the device
display will show +1 mT; the reading may vary by not more than 0.5 %.
3. Reverse the current flow and the display of magnetometer will show the change of readings.
4. Measure the magnetic field during supply of magnetic strength by magnetic inductometer: 5
mT, 10 mT, 50 mT, 100 mT;
5. By the readings define relative measurement error of magnetic field strength in percent by the
formula below:
100%
Д
Д
мТл мТл
мТл
δ
−
= ⋅
(1)
Where mT (
мТл
) – the reading of magnetometer, mT;
Д
мТл
– the actual value of magnetic induction displayed by magnetic inductometer, mT.
The permissible measurement error shall not exceed ±0.1+0.05 Н.
3.2.6 Documentation of verification
The verification of a device recognized as suitable for use is documented with a verification
certificate of the approved form and its marking.
The negative verification is documented by a certificate of inoperability specifying the reason and
eliminating the stamp of the previous verification.
3.3 Warranty obligations
The below-mentioned information of the warranty service applies to all the NOVOTEST products.
The manufacturer warrants the compliance of the device with the technical specifications and
compliance with the terms of transportation, storage, and operation, and timely maintenance at least
once a year.
3.3.1 Basic warranty
The device purchased from NOVOTEST or an authorized dealer is covered by the basic 3-year
warranty.
If any part of the device fails due to the defect of the material or a manufacturer, it shall be
repaired or replaced by the manufacturer, or any authorized dealer of NOVOTEST, regardless of the
transferring of the ownership right to a third party during the warranty period.

RDC.OD.MF-1M.000 OM
Operating manual
Magnetometer NOVOTEST MF-1M
Page 20
The warranty on batteries and chargers is provided by the manufacturers of batteries and chargers,
and therefore they are not covered by the NOVOTEST warranty. However, the authorized dealer of
NOVOTEST will assist you in making warranty claims regarding batteries, batteries, and chargers.
The warranty applies from the date of purchase of the device, normally from the shipment date. If
the device is purchased by an intermediate, the warranty period starts from the date of receipt of the
device by an intermediate.
3.3.2 Extended warranty
There is a special program of extension of the basic warranty from 3 to 5 years. To take advantage
of this program, user needs to pay the certificate with the purchasing. The terms and conditions of the
extended warranty are specified in the datasheet.
3.3.3 Warranty for repaired or replaced parts
All branded parts of NOVOTEST installed during the warranty repair apply to the NOVOTEST
warranty (up to the end of the warranty period). The spare parts replaced during the warranty service,
shall not be returned to the device owner.
3.3.4 Wearing parts
The parts wearing during the device operation are divided into two main categories. The first part
includes the parts that need to be replaced or adjusted with frequency stipulated by the maintenance
schedule of the device, and the second part includes the wearing parts, which frequency of replacement
or adjustment depends on the conditions of the product operation.
3.3.4.1 The parts replaced during the maintenance
The parts listed above have a limited service life and need replacement or adjustment with the
frequency stipulated by the device maintenance schedule of the. For these parts, the basic warranty
applies before the first replacement or adjustment. The warranty period for each part shall not exceed
the limitations (by the time of operation of the device or working hours), specified in the conditions of
the basic warranty.
−built-in batteries;
−oil and working fluids.
3.3.4.2 Wearing parts
The parts listed below that have limited service life or need to be replaced (adjusted) due to the
damage. However, these parts are subject basic warranty of NOVOTEST for 12 months:
−transducer and their components;
−connecting cables;
−parts and mechanisms exposed to mechanical impact during operation.
Note: The parts wearing due to friction (such as movable parts of measuring transducers,
ultrasonic piezoelectric transducers, support arm etc.) are not covered by the basic NOVOTEST
warranty, if these parts fail due to wear and tear during the device operation. However, if in the
warranty period these parts fail due to defect of material or workmanship, they shall be repaired or
replaced under the basic warranty.
3.3.5 Owner obligations
The operating manual and datasheet of the device state the information about the proper operation
and maintenance of the device.
Proper operation and maintenance of the device will help avoid costly repairs caused by incorrect
operation, negligence or improper maintenance. Besides, following our recommendations will extend
the service life of the device. Therefore, the device owner should:
Table of contents
Other NOVOTEST Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Bosch
Bosch GIM 120 Professional Original instructions

Pfeiffer Vacuum
Pfeiffer Vacuum PPT 200 operating instructions

Seaward
Seaward PowerPlus 1557 operating instructions

TSI Instruments
TSI Instruments ALNOR MicroManometer AXD 550 owner's manual

Samoa
Samoa 365601 Parts and technical service guide

Belimo
Belimo EXT-KA-SE100 user manual

Beko
Beko BEKOMAT 12 KW Instructions for installation and operation

Autonics
Autonics MX4W Series product manual

LEM
LEM Analyst 2050 operating instructions

AEMC
AEMC PowerPad III 8336 user manual

LCR
LCR Pro1 quick start guide

Kirchner und Tochter
Kirchner und Tochter RA 65 Installation and operating instructions