INSTRUCTION MANUAL
EC/TDS/SALT Testr
Large Screen Waterproof Multi Range
Conductivity/TDS/Salt Tester
with Temperature Display
Introduction
Thank you for selecting microprocessor-based waterproof EC / TDS / SALT tester with
large dual line display. You have one of the following models:
• ECTestr11 • ECTestr11+ • TDSTestr11 • TDSTestr11+ • SALTTestr11
Non-plus modelsNon-plus models
Non-plus modelsNon-plus models
Non-plus models (ECTestr11, TDSTestr11 & SALTTestr11) come with user-replaceable
two-pin type sensor and have many user friendly features such as Dual-range
measurement, the Hold function, Automatic Temperature Compensation (ATC) and
Self-Diagnostic Messaging capabilities.
Plus modelsPlus models
Plus modelsPlus models
Plus models (ECTestr11+ & TDSTestr11+) come with the user-replaceable cup type
sensor and have additional features such as Multi-range measurement, up to 3-point
calibration and higher resolution measurement.
Before You Begin
Remove the electrode’s protective cap. Soak the electrode for a few minutes in
alcohol to remove any oil stains on the electrodes which will affect the accuracy of
the tester. Rinse thoroughly with de-ionized water and shake off dry.
This Instruction Manual is also available for download at: www.eutechinst.com or www.4oakton.com Switching On
Press ON/OFFON/OFF
ON/OFFON/OFF
ON/OFF key to switch on the tester. The LCD shows the power-up sequence
as illustrated in Figure 2. When the tester is on, if you do not press a key for 8.5
minutes, the tester automatically switches off to conserve batteries.
Range Selection
Depending on the selected model, you can set the tester to limit its reading to a
particular measuring range (PU, LO or HI) or full scale (AUTO). The default setting
is AUTO. When you select a range other than AUTO, the tester can be calibrated
only for that particular range.
If you try to measure a sample which has a higher
conductivity/TDS value than that of the selected measuring range, the LCD shows
‘OR’ error message. Refer SpecifSpecif
SpecifSpecif
Specificaica
icaica
icationstions
tionstions
tions section for available ranges of the
selected model.
TT
TT
To select a ro select a r
o select a ro select a r
o select a rangang
angang
ange:e:
e:e:
e:
1. Switch off the tester. Press and hold °C/°F°C/°F
°C/°F°C/°F
°C/°F key and then switch on the tester
using ON/OFFON/OFF
ON/OFFON/OFF
ON/OFF key. Release °C/°F°C/°F
°C/°F°C/°F
°C/°F key.
2. The tester goes to range selection mode. The LCD shows the currently selected
Range (the default is AUTO) in the lower display. The upper display shows the
maximum possible reading for the selected range. Press HOLDHOLD
HOLDHOLD
HOLD key repeatedly
until you see the required range (PU, LO or HI).
Note:Note:
Note:Note:
Note: If no key is pressed within 5 seconds, LCD shows power-up sequence and
tester goes to measurement mode.
3. The tester automatically confirms the last selection if no key is pressed for 5
seconds. Upper display momentarily shows ‘CO’. The LCD shows power-up
sequence and the tester goes to measurement mode.
Figure 2: Power Up Sequence
Figure 3: Range selection sequence from AUTO to HI for ECTestr11+
Measurement
1. Press the ON/OFFON/OFF
ON/OFFON/OFF
ON/OFF key to switch on the tester. The ‘MEAS’ indicators appears
when the tester is in measurement mode.
2. Dip the electrode into the test solution making sure that it is fully immersed.
Stir to clear any trapped air bubbles from the electrode and let the reading
stabilize. For plus models, you can opt for the cup style measurement by filling
the electrode cup with sample of test solution.
Note:Note:
Note:Note:
Note: The LCD indicates ‘Or’ (over range) if the reading is outside the selected
range. If this occurs, select an appropriate range to suit the reading.
3. The upper display shows the main reading (conductivity/TDS/Salt) of the
solution, automatically temperature compensated (ATC) to normalized temperature
of 25°C. The lower display shows the temperature of the solution.
HOLD Function
This feature lets you freeze the display for a delayed observation.
1. Press HOLDHOLD
HOLDHOLD
HOLD key to freeze the measurement. The tester goes to hold mode and
‘HOLD’ indicator is displayed in LCD. The measurements are frozen and the ‘MEAS’
indicator disappears.
2. Press HOLDHOLD
HOLDHOLD
HOLD key again to release the measurement. The ‘HOLD’ indicator is no
longer displayed. The tester goes back to measurement mode.
Figure 4: HOLD Function
Temperature Unit of Measurement Selection
This feature lets you set the unit of measurement of temperature to either Celsius (°C)
or Fahrenheit (°F).
When the tester is in the measurement mode, press °C/°F°C/°F
°C/°F°C/°F
°C/°F button. The temperature
display toggles between the Celsius and Fahrenheit reading.
Figure 5: Temperature unit of measurement selection
KK
KK
Kee
ee
ey Functiony Function
y Functiony Function
y Function
- Power on and off the tester
(The tester automatically switches of, if no button is pressed for 8.5 seconds)
- In measurement mode, temperature reading switches between Celsius &
Fahrenheit
- In calibration mode, switches the tester to temperature calibration mode
- In temperature calibration mode, exits calibration mode without confirming
calibrated values
- In measurement mode, switches to hold mode freezing the display.
- In hold mode, switches back to measurement mode
- In manual calibration and temperature calibration modes, exits calibration mode
without confirming calibrated values
- In range selection mode, selects a range
- In measurement mode, enters calibration mode
- In calibration mode, adjusts the calibration values
- In hold mode, enters TDS factor setting mode
- In TDS factor setting mode, adjusts TDS factor
°F°C/
Key Functions
Figure 1: Battery compartment
Note:Note:
Note:Note:
Note: INCINC
INCINC
INC & DECDEC
DECDEC
DEC keys are located inside the
battery compartment.....
Refer figure 1.
Note:Note:
Note:Note:
Note:
For ECTestr11 & ECTestr11+ models, the
caption of HOLD key is ‘HOLD/ENT’‘HOLD/ENT’
‘HOLD/ENT’‘HOLD/ENT’
‘HOLD/ENT’
°C
MEAS MEAS
ATC
6.00
25.5
pp
t
pp
t
°F
ATC
6.00
78.0
°F
°C/
HOLD
ON/OF F
2000
us
°F°C/
Press & hold
key and switch on
the tester
°C/°F
20.00
Au o
ms
ENT
HOLD
200.0
us
PU LO
20.00
ms
HI
CO
HI
Power-up
Sequence
Auto confirmation occurs fo r
the last selected range
if no key is pressed for 5 sec.
ENT
HOLD
ENT
HOLD
Tes t e r
Off
1
2
3
2 2 2
5 seconds
+
-
°C
MEAS
ATC
1500
26.8
ppm
°C
ATC
1500
26.8
ppm
°C
MEAS
ATC
1535
26.2
ppm
HOLD
HOLD modeMeas urement
mode
Measurement
mode
HOLD HOL D
-1- -2- -3-
ON/O FF
°C°F
°C
°F
MEAS pp m
HOLD
pH% R.mV
S
CAL AT C
20.00Ec.p
Au o
°C
MEAS
ATC
0.0
26.8
uS
Model number &
software revision number
Selected ra nge &
upper limit of the range
Measurement
mode
ms
TT
TT
To beo be
o beo be
o begin automagin automa
gin automagin automa
gin automatic calibrtic calibr
tic calibrtic calibr
tic calibraa
aa
ation:tion:
tion:tion:
tion:
1. Switch on the tester. Make sure the tester is in measuring mode. Press INCINC
INCINC
INC or
DECDEC
DECDEC
DEC key to enter conductivity calibration mode.
2. ‘CAL’ indicator appears in LCD. The display briefly shows ‘CAL’ and the number
of points the tester will be calibrated.
3. The upper display shows the conductivity reading and the lower display
sequentially shows calibration standard values 1413 uS & 12.88 mS (for
ECTestr11) or 84 uS, 1413 uS & 12.88 mS (for ECTestr11+) if the
measuring range of the tester is set to AUTO.
Figure 7: Two-point automatic calibration sequence for ECTestr11
TT
TT
To enao ena
o enao ena
o enabb
bb
ble/disale/disa
le/disale/disa
le/disabb
bb
ble auto calibrle auto calibr
le auto calibrle auto calibr
le auto calibraa
aa
ation and mtion and m
tion and mtion and m
tion and multi-point calibrulti-point calibr
ulti-point calibrulti-point calibr
ulti-point calibraa
aa
ation:tion:
tion:tion:
tion:
1. Switch off the tester. Press and hold INCINC
INCINC
INC key and then switch on the tester using
ON/OFFON/OFF
ON/OFFON/OFF
ON/OFF key.
2. The tester goes to auto calibration selection mode. The lower display shows ‘A.CAL’
and the upper display blinks the current choice (‘Yes’ or ‘No’).
Note:Note:
Note:Note:
Note:
This mode is available only for ECTestr11 & ECTestr11+ models. For other
models, go directly to 1-point selection mode, described in step 4 below.
Press INCINC
INCINC
INC or DECDEC
DECDEC
DEC key to select ‘Yes’ (to enable auto calibration) or ‘No’ (to disable
auto calibration)
Note:Note:
Note:Note:
Note:
Press °C/°F°C/°F
°C/°F°C/°F
°C/°F key, if you wish to skip this setting without confirming changes.
NoteNote
NoteNote
Note: Press °C/°F°C/°F
°C/°F°C/°F
°C/°F key twice, if you wish to return to measurement mode without
confirming changes.
3. Press HOLD/ENTHOLD/ENT
HOLD/ENTHOLD/ENT
HOLD/ENT key to confirm the selection. The display shows ‘CO’.
4. The tester goes to 1-point calibration selection mode. The lower display shows
‘1.Pnt’ and the upper display blinks the current choice (‘Yes’ or ‘No’).
Press INCINC
INCINC
INC or DECDEC
DECDEC
DEC key to select ‘Yes’ (to enable 1-point calibration) or ‘No’ (to
disable 1-point calibration, i.e. to enable multi-point calibration).
Note:Note:
Note:Note:
Note: Press °C/°F°C/°F
°C/°F°C/°F
°C/°F key if you wish to skip this setting without confirming.
5. Press HOLD/ENTHOLD/ENT
HOLD/ENTHOLD/ENT
HOLD/ENT key to confirm the selection. The display shows ‘CO’ for few
seconds and then shows power-up sequence The tester goes to measurement
mode.
Figure 6: Selecting auto calibrtaion & 1-point calibration for ECTestr11+
Auto Calibration
Auto calibration feature is available only for conductivity in ECTestr11 &
ECTestr11+ models. Make sure that ‘auto calibration’ is enabled as described in
the previous section. Auto calibration is suitable if you use standard conductivity
solutions for calibration process.
If you selected 1-point calibration, you need to choose a calibration standard
corresponds to the selected measuring range of the tester listed below. If you
have selected multi-point calibration & AUTO measuring range, you can choose any
of the calibration standards listed below. During auto calibration, the tester
recognizes the calibration standard if its value is within 50% tolerance. For multi-
point calibration (with AUTO measuring range), the tester automatically scans
through all possible calibration points until all of them are calibrated.
TT
TT
To pro pr
o pro pr
o pree
ee
eparpar
parpar
pare calibre calibr
e calibre calibr
e calibraa
aa
ation standartion standar
tion standartion standar
tion standards:ds:
ds:ds:
ds:
Use fresh calibration standard solutions listed in the above table. Prepare each
solution in two beakers - one for rinsing and the other for calibration. Rinse the
electrode in de-ionized water before calibration.
Selected Measuring Range Calibration Standard
PU (0 to 200.0 uS/cm) 84 us
LO (0 to 2000 uS/cm) 1413 uS
HI (0 to 20.00 mS/cm) 12.88 mS
AUTO 84 us, 1413 uS, 12.88 mS
Note:Note:
Note:Note:
Note:
If you have selected a specific measuring range for the tester, the lower display
shows the corresponding calibration standard value that matches the selected
measuring range.
4. Rinse the electrode with the calibration standard that you intend to calibrate and
then dip the electrode in the other beaker with same calibration standard. Swirl
gently to create a homogenous sample and allow time for the reading to stabilize.
Note:Note:
Note:Note:
Note: For multi-point calibration, the lower display automatically locks at the
calibration standard value that closely matches. The tolerance range is ± 50% of
the calibration standard. The tester shows error message ‘Er.1’ if you try to calibrate
with a solution whose conductivity is outside the tolerance range.
Note:Note:
Note:Note:
Note:
Press INCINC
INCINC
INC or DECDEC
DECDEC
DEC key if you wish to exit from auto calibration, during any of
the above steps.
5. Press HOLD/ENTHOLD/ENT
HOLD/ENTHOLD/ENT
HOLD/ENT key to confirm the calibration. LCD shows ‘CO’ for 2 seconds. The
calibration is complete and the tester returns to measurement mode, if this is a 1-
point calibration.
6. For multi-point calibration, the tester goes to the next calibration point, lower
display showing next calibration standard values. Rinse the electrode in de-ionized
water and repeat step 4 & 5 to continue calibrating with next calibration standard
solution.
Note:Note:
Note:Note:
Note: The tester shows error message ‘Er.0’ and returns to measuring mode if the
temperature of the calibration solution is not within 0°C to 50°C.
Note:Note:
Note:Note:
Note: The tester shows error message ‘Er.1’ if you press HOLD/ENTHOLD/ENT
HOLD/ENTHOLD/ENT
HOLD/ENT key before the
tester recognizes the calibration standard.
About Calibration
To ensure higher accuracy, the tester must be calibrated on a regular basis.
Calibration can be manual or automatic (only for ECTestr11 & ECTestr11+); calibration
can be 1-point or multi-point. You can choose any combination of the above two
options for calibration. If you calibrate the tester for 1-point, the calibration is applied
for all the measuring ranges. In applications where you need higher accuracy, and
when you intend to measure values in more than one range, it is recommended to
select multi-point calibration.
Selection of Automatic or Manual Calibration
ECTestr11 & ECTestr11+ models support both automatic & manual calibration while
all other models have to be calibrated manually. In automatic calibration, the tester
automatically detects and verifies known conductivity standard solutions (84uS,
1413uS & 12.88mS) . In manual calibration, you can use non-standard solutions
which may be specific for your application.
The factory default is Automatic calibration (Auto). You can enable or disable
automatic calibration as described below (for only ECTestr11 & ECTestr11+ models).
Selection of 1-point or Multi-point Calibration
The factory default is 1-point Calibration. For higher accuracy, it is recommended that
you calibrate the tester for multiple ranges if you intend to measure values in multiple
ranges. You can enable or disable multi-point calibration as described below.
Note:Note:
Note:Note:
Note:
If you have selected a specific measuring range for the tester, selecting multi-
point calibration has no meaning, as the tester can only be calibrated for 1-point
for the selected range. Set the tester to ‘AUTO’ measuring range if you wish to
calibrate multi-points.
-4- -5- -6-
ON/OF F
Press & hold INC
key and switch on
the tester
YES
ENT
HOLD
Power-up
Sequence
A.CAL
A.CAL
A.CAL
CO
ENT
HOLD
YES
ENT
HOLD
ENT
HOLD
CO
After
1 sec.
Tes t e r
Off
1
2
2
3 4
4
5
ENT
HOLD
Measurement
Mode
OR
CAL ATC
0.0
1413
uS
CAL ATC
0.0
12.88
uS
CAL
CAL ATC
The two calibration stds . values
displayed sequentially u ntil
tester finds a matching so lution
within 50% tolerance .±
CAL ATC
1410
1413
uS
Tester dipped in
1413 uS std. solution
and automatically detects
the calibration std.
CAL ATC
1410
CO
uS
Confirmation of 1st point
displayed for
2 seconds
ENT
HOLD
CAL AT C
1410
12.88
mS
CAL ATC
12.80
12.88
mS
CAL ATC
12.80
CO
The next calibration s td.
displayed until tester fi nds
a close match within
50%
to
l
e
r
a
n
ce
±
mS
Confirmation of 2nd point
displayed for
2 seconds
Tester dipped in
12.88 mS std. solution
& automatically detects
the calibrati on std.
Measurement
Mode
3
1
2 54
6
3
Displaye d for
2 seconds and
tester goes to
calibrati on mode
3. The upper display shows the measured temperature reading based on the last set
offset (if any) and the lower display shows the default (uncalibrated) temperature
reading based on factory settings. Use INCINC
INCINC
INC and DECDEC
DECDEC
DEC keys to adjust the upper
temperature reading to the known temperature value of the solution.
Notes:Notes:
Notes:Notes:
Notes:
The temperature adjustment window is ± 5°C (± 9°F) from the default reading.
4. Wait for 5 seconds for the tester to automatically confirm the temperature
calibration value by displaying ‘CO’ and return to the measurement mode.
Note:Note:
Note:Note:
Note:
To exit temperature calibration mode without confirming the calibration, press
°C/°F°C/°F
°C/°F°C/°F
°C/°F key or HOLD/ENTHOLD/ENT
HOLD/ENTHOLD/ENT
HOLD/ENT key before the automatic confirmation takes place.
Note:Note:
Note:Note:
Note:
The tester shows error message ‘Er.0’ and returns to measuring mode if the
temperature of the solution is not within 0°C to 50°C.
Selected Acceptable Calibration Standard Range
Measuring Range Conductivity TDS/Salt
PU 2.0 - 200.0 uS/cm 2.0 - 200.0 ppm
LO 200 - 2000 uS/cm 200 - 2000 ppm
HI 2.00 - 20.00 mS/cm 1.00 - 10.00 ppt
AUTO Select a calibration standard nearer to application sample
5. For multi-point calibration rinse the electrode in de-ionized water and repeat
step the above steps with another standard solution.
TDS Factor Setting
TDS factor is only applicable for TDSTestr11 & TDSTestr11+ models.
The factory default TDS factor is 0.71. You can adjust the TDS factor to suit
different samples of your applications.
TT
TT
To co c
o co c
o changhang
hanghang
hangee
ee
e TDS fTDS f
TDS fTDS f
TDS factoractor
actoractor
actor:::::
1. Switch on the tester. Make sure the tester in measurement mode. Press HOLDHOLD
HOLDHOLD
HOLD
key to bring the tester to the HOLD mode.
2. Press INCINC
INCINC
INC or DECDEC
DECDEC
DEC key to enter the TDS factor setting mode.
3. The upper & lower displays of LCD show the last configured TDS factor. The
upper display is adjustable. Use the INCINC
INCINC
INC or DECDEC
DECDEC
DEC key to adjust the TDS factor.
The adjustable range is 0.4 to 1.0
Note:Note:
Note:Note:
Note:
If you do not press INCINC
INCINC
INC or DECDEC
DECDEC
DEC key within 5 seconds, the tester shows the
confirmation ‘CO’ and returns to measurement mode.
4. Wait for 5 seconds for the tester to automatically confirm the new setting by
displaying ‘CO’ and return to the measurement mode.
Figure 9: TDS Factor setting
Temperature Calibration
Temperature calibration need not be performed every time, unless the temperature
reading differs from that of an accurate thermometer. If temperature calibration is
performed, Conductivity/TDS/Salt calibration is mandatory.
1. Switch on the tester. Make sure the tester is in measuring mode. If required,
press °C/°F°C/°F
°C/°F°C/°F
°C/°F key to select the desired unit of measurement for temperature
(Celsius or Fahrenheit). Dip the tester into a solution of known temperature
and allow time for the temperature reading to stabilize.
2. Press INCINC
INCINC
INC or DECDEC
DECDEC
DEC key to bring the tester to the calibration mode. CAL’ indicator
appears in LCD. Immediately press °C/°F°C/°F
°C/°F°C/°F
°C/°F key to switch to the temperature
calibration mode.
Note:Note:
Note:Note:
Note:
When you enter calibration mode, if the conductivity/TDS/salt reading is outside
the specified range (Or), the tester shows ‘Er.1’ error message. You can still proceed
to the temperature calibration mode by pressing °C/°F°C/°F
°C/°F°C/°F
°C/°F key immediately. If the °C/°F
key is not pressed within 2 seconds, the tester exits the calibration mode and returns
to the measurement mode.
Figure 10: Temperature calibration sequence
Figure 8: One-point manual calibration sequence for ECTestr11+
-7- -8- -9-
Manual Calibration
In manual calibration, the tester allows 1-point calibration for each measuring range.
You can use customized calibration solutions with known conductivity/TDS values to
calibrate the tester. . The following table shows acceptable conductivity/TDS ranges of
calibration solutions for each measuring range. Make sure your calibration solutions
are within the given ranges.
TT
TT
To pro pr
o pro pr
o pree
ee
eparpar
parpar
pare calibre calibr
e calibre calibr
e calibraa
aa
ation standartion standar
tion standartion standar
tion standards:ds:
ds:ds:
ds:
Use fresh calibration solutions. Measure conductivity/TDS values of the solution with a
meter known to be accurate. Prepare each solution in two beakers - one for rinsing
and the other for calibration. Rinse the electrode in de-ionized water before
calibration.
TT
TT
To beo be
o beo be
o begin mangin man
gin mangin man
gin manual calibrual calibr
ual calibrual calibr
ual calibraa
aa
ation:tion:
tion:tion:
tion:
1. Switch on the tester. Make sure the tester is in measuring mode. Rinse the
electrode with the calibration standard that you intend to calibrate and then dip
the electrode in the other beaker with same calibration standard. Swirl gently to
create a homogenous sample and allow time for the reading to stabilize.
2. Press INCINC
INCINC
INC or DECDEC
DECDEC
DEC key to enter calibration mode. The ‘CAL’ indicator appears in
LCD. The display briefly shows ‘CAL’ and the number of points the tester will be
calibrated.
3. The upper display shows the measured conductivity/TDS reading of the solution
based on previous calibration (if any) and the lower display shows the default
(uncalibrated) conductivity/TDS reading.
Note:Note:
Note:Note:
Note:
The tester shows error message ‘Er.1’ :
(a) If the reading is over range (Or) of selected measuring range of the tester ,
or
(b) If the default (uncalibrated) reading is not within the acceptable calibration
standard range.
Use INCINC
INCINC
INC and DECDEC
DECDEC
DEC keys to adjust the upper display to the correct conductivity/TDS
value of the calibration solution.
Note:Note:
Note:Note:
Note:
The calibration adjustment window is ± 50% from the default reading.
Note:Note:
Note:Note:
Note:
If you do not press INCINC
INCINC
INC or DECDEC
DECDEC
DEC key within 5 seconds, the tester shows the
confirmation ‘CO’ and returns to the measurement mode. However, the tester is
not calibrated to new values yet. The old calibration is still active. If this happens,
press INCINC
INCINC
INC or DECDEC
DECDEC
DEC key once again to enter calibration mode.
4. Wait for 5 seconds for the tester to automatically confirm the calibration by
displaying ‘CO’ and return to the measurement mode.
Note:Note:
Note:Note:
Note:
To exit calibration mode without confirming the calibration, press HOLD/ENTHOLD/ENT
HOLD/ENTHOLD/ENT
HOLD/ENT
key before the automatic confirmation takes place.
Note:Note:
Note:Note:
Note:
The tester shows error message ‘Er.0’ and returns to measuring mode if the
temperature of the calibration solution is not within 0°C to 50°C.
.
CAL
ATC
1400 1413
1410 1410
Displayed for
2 seconds and
Tester goes to
Calibration mode
Tester dipped in
1413 uS std. solution
Measurement
Mode
CO
Confirmation
displayed for
2 seconds
1
23 43
After
5 sec.
1413
OR
Upper display reading
Adjusted to the actual
Conductivity value of
Ca li bratio n std. s olu tio n
CAL ATC CAL ATC CAL ATC
uS uS uS
6.35
pp
t
°C
ATC
26.8
0.71 0.50
0.71 0.71
HOLD
HOLD mode TDS Factor setting mode
Measurement
Mode
OR
CO
Confirmation
displayed for
2
seco
n
ds
1
23 43
After
5 sec.
HOLD
0.50
CAL CAL CAL
°C
°C °C
°C °C
MEAS
AT C
7.50 7.50
26.8
mS
°C
MEAS
AT C
7.50
27.5
mS
ATC
26.8 27.5
7.50 26.8 26.8
mS
°F
°C/
OR
OR
Dipped in a solution
of known temperature
Calibration
mod e
Temperature Calibration mod e
Measurement
mode
Measurement
Mode
CAL
°C
27.5
Confirmation
displayed for
2 seconds
CO
1
23
4
2 3
Set required unit of
measurement for
temperature
3 Bars indicates Battery is full (100%)
2 Bars indicates 50% of the battery life is left
1 Bar indicates 25% of the battery life is left
Blinking battery casing indicates the need to replace
batteries with fresh ones as specified by manufacturer
The sensor electrodes short circuited
Replacement sensor is not connected properly to the
tester during sensor replacement
Measured value or temperature value exceeds the specified
maximum or minimum value
Blinking ‘ATC’, ‘Or’ or ‘Ur’ indicates that there is a short
or open circuit at the built in temperature sensor
Calibration error due to temperature value not within the
specified range
Calibration error due to Conductivity/TDS/Salt value not
within the specified calibration standard range
Self-Diagnostic Messages
ErEr
ErEr
Er
.1.1
.1.1
.1
Low battery
indicator
Or / UrOr / Ur
Or / UrOr / Ur
Or / Ur
(Still)
Over range /
Under range
signal
AA
AA
ATT
TT
TC / Or /C / Or /
C / Or /C / Or /
C / Or /
UrUr
UrUr
Ur (Blinking)
ErEr
ErEr
Er.0.0
.0.0
.0
Error
Message
Electrode Replacement
When the tester fails to calibrate or gives fluctuating readings in calibration
standards, you need to change the electrode module. You can replace the
electrode module at a fraction of the cost of a new tester.
1. With dry hands, grip the ribbed tester collar with electrode facing you. Twist the
collar counter clockwise (see Figure13-A). Save the ribbed tester collar and
O-ring for later use.
2. Pull the old electrode module away from the tester.
3. Align the four tabs of the new electrode module so that they match the four
slots on the tester (see Figure 13-B).
4. Gently push the module into the slots to sit it in position. Push the smaller
O-ring fully onto the new electrode module. Push the collar over the module
and thread it into place by firmly twisting clockwise.
Note:Note:
Note:Note:
Note: It is necessary that you recalibrate the tester prior to measurement after
an electrode replacement.
Accessories
Item Order Code
ECTESTR, TDSTESTR & SALTESTR TDSSENSOR
replacement sensor
ECTESTR+ & TDSTESTR+ TDSSENSORPLUS
replacement sensor
Warranty
The waterproof testers are warranted to be free from manufacturing defects for 1 year
and electrode module for 6 months, unless otherwise specified. If repair, adjustment or
replacement is necessary and has not been the result of abuse or misuse within the
time period specified, please return the tester - freight prepaid - and correction will be
made without charge. Out of warranty products will be repaired on a charge basis.
Return of Items
Authorization must be obtained from your distributor before returning items for any
reason.
When applying for authorization, please include information regarding the reason the
item(s) are to be returned.
Note:Note:
Note:Note:
Note: Eutech Instruments/Oakton Instruments reserve the right to make improvements
in design, construction and appearance of products without notice. Prices are subject
to change without notice.
ModelModel
ModelModel
Model ECTECT
ECTECT
ECTestr11estr11
estr11estr11
estr11 ECTECT
ECTECT
ECTestr11+estr11+
estr11+estr11+
estr11+ TDSTTDST
TDSTTDST
TDSTestr11estr11
estr11estr11
estr11 TDSTTDST
TDSTTDST
TDSTestr11+estr11+
estr11+estr11+
estr11+ SALSAL
SALSAL
SALTTTT
TTTT
TTestr11estr11
estr11estr11
estr11
Range: PU - 0 to 200.0 uS/cm - 0 to 200.0 ppm -
LO 0 to 2000 uS/cm 0 to 2000 uS/cm 0 to 2000 ppm 0 to 2000 ppm -
HI 0 to 20.00 mS/cm 0 to 20.00 mS/cm 0 to 10.00 ppt 0 to 10.00 ppt 0 to 10.00 ppt
Resolution: PU - 0.1 uS/cm - 0.1 ppm -
LO 10 uS/cm 1 uS/cm 10 ppm 1 ppm -
HI 0.10 mS/cm 0.01 mS/cm 0.10 ppt 0.01 ppm 0.10 ppt
Accuracy ± 1% of Full Scale
Calibration Type Auto or Manual Auto or Manual Manual Manual Manual
Calibration Points 1 or 2 points 1, 2 or 3 points 1 or 2 points 1, 2 or 3 points 1 point
Calibration Window ± 50% from each point
Calibration PU - 2.0 - 200.0 uS/cm - 2.0 - 200.0 ppm -
Standard Range:LO 200 - 2000 uS/cm 200 - 2000 uS/cm 200 - 2000 ppm 200 - 2000 ppm -
(Manual) HI 2.00 - 20.00 mS/cm 2.00 - 20.00 mS/cm 1.00 - 10.00 ppt 1.00 - 10.00 ppt 1.00 - 10.00 ppt
Sensor Type Two-pin Cup Two-pin Cup Two-pin
TDS Factor - - 0.4 to 1.0 (Default 0.71)-
Temperature
Range in °C 0.0 to 50.0°C
Range in °F 32.0 to 122°F
Resolution 0.1°C (0.1°F)
Accuracy ±0.5°C (±0.9°F)
Calibration point 1 point
Calibration Window ± 5°C (± 9°F)from factory default
ATC 0 to 50°C
Temp Coefficient 2% per °C
Normalization Temp 25.0°C
Auto Off 8.5 minutes after last key press
Operating Temp 0 to 50°C
Power Battery 4 X 1.5V”A76" micro alkaline battery
Battery Life >150 hrs
LCD Display Custom Dual Display27mm(H)X21mm(W)
Dimensions Tester: 16.5 cm X 3.8 cm; 90g
Weight Boxed: 22cm X 6cmX 5cm; 170 g
Specifications
Changing Batteries
Replace the batteries when the low battery indicator
starts blinking.
1. Open the battery compartment lid (with attached
lanyard loop).
2. Remove old batteries by pulling plastic ribbon.
Replace with fresh ones. Note polarity as shown in
figure 11.
Electrode Maintenance
1. Always keep the sensor electrodes clean. Rinse the electrodes with de-ionized
water and wipe them dry with clean cloth before storing with its protective cap.
For cup type electrodes, remove the white plastic cup and insert to thoroughly
clean viscous solutions. Never scratch electrodes with a hard substance.
2. For better performance, soak the electrode in alcohol for 10 to 15 minutes
and rinse with de-ionized water before starting any measurement process.
This is to remove dirt and oil stains on the electrode which may affect the
accuracy of the measurements.
Figure 13: Removing collar & inserting electrode
68X068056 Rev 2 May 06
+
-
Plastic
Ribbon
B
Rotate ribbed collar
away fr om you
Ribbed col lar Ribbed collar Small
O-Ring
Small Tab Larg e Tab
Large
O-Ring
Insert Electrode Module
A
Parameter Factory Default
User calibration (conductivity/TDS/salt) (Reset)
Temperature unit of measurement Celsius (°C)
Temperature offset 0
Auto calibration (for ECTestr11 & ECTester11+) Enable
1-point calibration Enable
Conductivity calibration factor (for ECTestr11 & ECTester11+) 1.0
TDS factor (for TDSTestr11 & TDSTester11+) 0.71
Figure 11: Resetting sequence
Reset
Reset option allows you to restore the calibration and other parameters back to
factory default settings.
1. Switch off the tester. Press and hold the HOLDHOLD
HOLDHOLD
HOLD key and then switch on the tester
using ON/OFFON/OFF
ON/OFFON/OFF
ON/OFF key. Release HOLDHOLD
HOLDHOLD
HOLD key.
2. The lower display shows ‘rSt’ (reset) and the upper display blinks ‘No’. Use INCINC
INCINC
INC
or DECDEC
DECDEC
DEC key to select ‘Yes’ (to proceed with resting) or ‘No’ (to quit without
resetting).
Note:Note:
Note:Note:
Note:
Press °C/°F°C/°F
°C/°F°C/°F
°C/°F key if you wish to skip to measurement mode without making any
selection.
3. Press HOLDHOLD
HOLDHOLD
HOLD key to confirm your selection. LCD shows ‘CO’. If ‘Yes’ is selected, the
tester resets to its factory default values as listed below. LCD shows power-up
sequence and tester goes to measurement mode.
Figure 11: Battery compartment
-10- -11-
ON/O FF
Press & hold HOLD
key and switch on
the tester
YES
Power-up
Sequence
CO
Te st e r
Off
1
2
2
3
HOL D HOL D
HOL D

This manual suits for next models

3

Other Oakton Multimeter manuals

Oakton pHTestr 1 User manual

Oakton

Oakton pHTestr 1 User manual

Oakton 35632 Series User manual

Oakton

Oakton 35632 Series User manual

Oakton 35632-Series User manual

Oakton

Oakton 35632-Series User manual

Oakton pH 110 User manual

Oakton

Oakton pH 110 User manual

Oakton pH 5+ User manual

Oakton

Oakton pH 5+ User manual

Oakton pH/CON 510 series User manual

Oakton

Oakton pH/CON 510 series User manual

Popular Multimeter manuals by other brands

Fluke 805 user manual

Fluke

Fluke 805 user manual

Testboy TB 315 operating instructions

Testboy

Testboy TB 315 operating instructions

Fluke 8012A instruction manual

Fluke

Fluke 8012A instruction manual

Owon HDS-N series user manual

Owon

Owon HDS-N series user manual

Maxwell 25401 user manual

Maxwell

Maxwell 25401 user manual

TPI 190 Series instruction manual

TPI

TPI 190 Series instruction manual

Benning CM 12 operating manual

Benning

Benning CM 12 operating manual

KNOVA KN 8059 manual

KNOVA

KNOVA KN 8059 manual

Truper MUT-60A manual

Truper

Truper MUT-60A manual

Fluke 106 user manual

Fluke

Fluke 106 user manual

Fluke 8050? instruction manual

Fluke

Fluke 8050? instruction manual

Tense EM-07E user manual

Tense

Tense EM-07E user manual

Klein Tools MM720 instruction manual

Klein Tools

Klein Tools MM720 instruction manual

Keithley 177 instruction manual

Keithley

Keithley 177 instruction manual

Amprobe DM73C user manual

Amprobe

Amprobe DM73C user manual

Agilent Technologies U1241B quick start guide

Agilent Technologies

Agilent Technologies U1241B quick start guide

Cropico DO7e Operating	 instruction

Cropico

Cropico DO7e Operating instruction

Sonel CMM-60 user manual

Sonel

Sonel CMM-60 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.