8. Maintenance
8.1 Maintenance considerations
Maintenance Considerations
Attention
1. Only the dealers authorized by the manufacturer or the qualified personnel
after special training can conduct preventive maintenance or performance
adjustment for medical oxygen concentrator.
2. In case of medical oxygen concentrator failure, contact the dealer or
manufacturer for repair. Do not disassemble for repair by yourself.
3. Clean the humidifier bottle every day. Add pure water to humidifier bottle and
the water level shall be between the maximum and minimum water levels.
4. Regularly clean air suction filtering foam and filtering cotton according to
requirements. Clean in advance in case of too much dust. oxygen concentrator shall
not be started before filtering cotton and foam are installed.
5. Power cord is provided according to the power of oxygen concentrator, please do
not replace arbitrarily.
8.2 Cleaning of humidifier bottle
To clean, first pull out the silicone rubber tube from humidifier bottle, take out
the humidifier bottle. Humidifier bottle is generally cleaned with clean water, in case
of any water scale, rinse with clean water after descaling. During cleaning, pay
attention to cleaning the small air inlet hole at the bottom of the core tube in the bottle,
so as to keep oxygen unobstructed.
After cleaning, sterilize (soak in disinfectant containing available chlorine of
500mg/L) and rinse with pure water. Add proper amount of pure water to the bottle,
the water level shall be between the maximum (MAX) and minimum (MIN) water
lines, and then tighten the bottle cap, put into the transparent cover and insert the
silicone rubber tube.
8.3 Cleaning of atomizer (atomization models only)
After using the atomizer, pour out all residual liquid from atomizing cup, rinse
with clean water.
After cleaning, sterilize (soak in disinfectant containing available chlorine of
500mg/L) and rinse with pure water.
8.4 Cleaning of primary air suction filter (filtering sponge) and secondary
filter
Timely cleaning and replacement of primary air suction filter and secondary filter
are very important for protecting air compressor and molecular sieve, and extending
the service life of medical oxygen concentrator, please timely clean and replace
according to requirements.
Primary air suction filter shall be cleaned once a week. During cleaning, draw out
filtering sponge and rinse with clean water, it can be used after natural drying.