Olympus DIEGO ELITE User manual

Popular Ultrasonic Jewelry Cleaner manuals by other brands

Probuilder 16937 instruction manual

Probuilder

Probuilder 16937 instruction manual

Kränzle Round Cleaner 300 INOX Operating manual with spare parts list

Kränzle

Kränzle Round Cleaner 300 INOX Operating manual with spare parts list

RIDGID 36023 manual

RIDGID

RIDGID 36023 manual

JOCCA 1249 instruction manual

JOCCA

JOCCA 1249 instruction manual

Kärcher FC 7 Cordless quick start guide

Kärcher

Kärcher FC 7 Cordless quick start guide

Connoisseurs SafeWave LaSonic quick start guide

Connoisseurs

Connoisseurs SafeWave LaSonic quick start guide

Kärcher SB OB manual

Kärcher

Kärcher SB OB manual

Laser 6166 instruction manual

Laser

Laser 6166 instruction manual

poolstar NetSpa Super Vac user manual

poolstar

poolstar NetSpa Super Vac user manual

Hammerhead DCFSCP18 Operator's manual

Hammerhead

Hammerhead DCFSCP18 Operator's manual

Rothenberger RODIA 1600 Instructions for use

Rothenberger

Rothenberger RODIA 1600 Instructions for use

Hedson DRESTER X Series Operation manual

Hedson

Hedson DRESTER X Series Operation manual

ETEKCITY CD-2800 user manual

ETEKCITY

ETEKCITY CD-2800 user manual

MaxVac DV15 MB instruction manual

MaxVac

MaxVac DV15 MB instruction manual

Makita DVC560 instruction manual

Makita

Makita DVC560 instruction manual

Forch INOX CLEANER 350 W manual

Forch

Forch INOX CLEANER 350 W manual

IPC ML 15.150 TST manual

IPC

IPC ML 15.150 TST manual

Maximex 9935390500 Instructions for use

Maximex

Maximex 9935390500 Instructions for use

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

HANDPIECE CLEANING
QUICK-REFERENCE GUIDE
Wipe the external surfaces of the handpiece, cord, and plug with enzymatic detergent, sterile water, or tap water
immediately after use.
MANUAL CLEANING
PRE-CLEANING
Fully immerse the handpiece, cord, and plug in the
cleaning solution. Allow them to soak for ten
minutes. Note the manufacturer’s instructions
regarding the temperature and dilution factor for the
detergent.
While immersed, gently scrub the areas shown below with the appropriate brushes (handpiece cleaning kit item EG70338009)
1
2
This is designed as a quick-reference guide only. This guide is not a substitute for reading and learning
the entire operator’s manual and/or instructions for use (IFU) provided the DIEGO ELITE System. Olympus
cannot encourage or recommend any actions that deviate from or are not covered by the IFU.
The gear shaft
The three-pin receptacle pocket and plug
The interior of the blade retention mechanism
The handpiece plug
E048xxxx · X.000 · 11/13 · PR
Specifications, design, and accessories are subject to change without any notice or obligation on the part of the manufacturer.
Postbox 10 49 08, 20034 Hamburg, Germany
Wendenstrasse 14–18, 20097 Hamburg, Germany
Phone: +49 (0)40 237 730, Fax: +49 (0)40 230 761
www.olympus-europa.com
Using a nylon-bristle brush, scrub the shelf that
contains the Olympus logo and part number.
Flush the interior of the handpiece with cool distilled,
deionized, or potable water. Actuate the retention
mechanism while flushing. Ensure water flows through
all drainage ports.
Use 70% isopropyl alcohol to wipe and flush the
handpiece, power cord, and plug. Flush the exterior to
remove any remaining residue.
Dry completely with a lint-free cloth. Visibly inspect the entire handpiece for any remaining contamination and repeat
the cleaning procedure if needed.
Actuate the blade retention mechanism.
3
7
4
5
6
Automated cleaning can be performed in an automatic
washer-disinfector as an alternative to manual cleaning.
The handpiece is designed for sterilization by steam autoclave.
Handpieces can be sterilized wrapped or unwrapped.
Place the handpiece and cord, on the washer rack or
place them in the approved tray (Item EGMDHPTRAY).
The washer should be set to the Vario TD cycle. After
automated cleaning, visually inspect the handpiece for
any remaining residue and repeat the cleaning procedure
if needed.
AUTOMATED CLEANING STERILIZATION
Steam sterilization
Cycle Gravity prevacuum
Temperature 132° C 132° C
time 15 minutes 4 minutes
Drying Time 60 minutes 30 minutes
To prevent damage, the handpiece must be completely vacuum dried before storage.
STORAGE
HANDPIECE CLEANING – QUICK-REFERENCE GUIDE
1
2
Note: To satisfy international, regional, governmental, or health-
care facility requirements, the handpiece may be processed at
temperatures up to 137° C without causing damage.