K3MA-LK3MA-L
Specifications
■Ratings
Input/Output Ratings
Relay Contact Output
K3MA-L 100-240VAC, K3MA-L-C 100-240VAC K3MA-L 24VAC/VDC, K3MA-L-C 24VAC/VDC
Supply voltage 100 to 240 VAC 24 VAC (50/60 Hz), 24 VDC
Operating voltage range 85% to 110% of the rated supply voltage
Power consumption (under
maximum load) 6 VA max. 4.5 VA max. (24 VAC)
4.5 W max. (24 VDC)
Insulation resistance 20 M
W
min. (at 500 VDC) between external terminal and case.
Insulation provided between inputs, outputs, and power supply.
Dielectric withstand voltage 2,000 VAC for 1 min between external terminal and case.
Insulation provided between inputs, outputs, and power supply.
Noise immunity
1,500 V on power supply terminals in normal or
common mode.
1
m
s, or 100 ns for square-wave noise with 1 ns.
480 V on power supply terminals in normal mode.
1,500 V in common mode.
1
m
s, or 100 ns for square-wave noise with 1 ns.
Vibration resistance Vibration: 10 to 55 Hz, Acceleration: 50 m/s2
5 min each in X, Y, and Z directions for 10 sweeps.
Shock resistance 150 m/s2(100 m/s2for relay contact outputs) 3 times each on 3 axes, 6 directions.
Ambient temperature Operating:
-
10
C to 55
C (with no condensation or icing)
Storage:
-
25
C to 65
C (with no condensation or icing)
Ambient humidity Operating: 25% to 85% (with no condensation)
Ambient atmosphere Must be free of corrosive gas.
Approved safety standards UL3121, conforms to EN61010-1 (Pollution degree 2/overvoltage category II)
Conforms to VDE0106/P100 (finger protection)
EMC (EMI) EN61326+A1 Industry
Emission Enclosure: CISPR 11 Group 1 class A: CISRP16-1/-2
Emission AC Mains: CISPR 11 Group 1 class A: CISRP16-1/-2
(EMS) EN61326+A1 Industry
Immunity ESD: EN61000-4-2: 4-kV contact discharge
8-kV air discharge
Immunity RF-interference: EN61000-4-3: 10 V/m (amplitude-modulated, 80 MHz to 1 GHz)
Electrical Fast Transient Noise: EN61000-4-4: 2 kV (power line)
Immunity Burst Noise: 1 kV line to line (I/O signal line)
Immunity Surge: EN61000-4-5: 1 kV (power line)
2-kV line to ground (power line)
Immunity Conducted Disturbance: EN61000-4-6: 3 V (0.15 to 80 MHz)
Immunity Voltage Dip/Interrupting: EN61000-4-11: 0.5 cycle, 0, 180
,
100% (rated voltage)
Weight Approx. 200 g
Item Resistive load (cos
f
= 1) Inductive load (cos
f
= 0.4, L/R = 7 ms)
Rated load 5 A at 250 VAC, 5 A at 30 VDC 1.5 A at 250 VAC, 1.5 A at 30 VDC
Rated carry current 5 A max. (at COM terminal)
Max. contact voltage 400 VAC, 150 VDC
Max. contact current 5 A (at COM terminal)
Max. switching capacity 2,000 VA, 192 W 375 VA, 36 W
Min. permissible load
(P level, reference value) 10 mA at 5 VDC
Mechanical life 20,000,000 times min. (at a switching frequency of 1,200 time/min)
Electrical life
(at an ambient temperature of 20
C) 100,000 times min. (at a rated load switching frequency of 10 time/min)