Omron K3MA-L User manual

1
K3MA-LTemperature Meter
Highly Visible LCD Display with 2-color
(Red and Green) LEDs
■Wide input range - select from two types of plati-
num-resistance thermometers and ten types of
thermocouples.
■Front-panel key operation for easy setting.
■Average processing function suppresses flicker.
■Temperature input shift and temperature unit
selection functions.
■Easy confirmation of max/min display.
■Short 80-mm depth (measured from edge of face
plate).
■Finger protective cover (standard equipment) pro-
tects against electric shock.
■Water- and dust-proof NEMA4X (IP66 equivalent)
front panel.
■Recognized to U.S. and Canadian requirements
under the Component Recognition Program of UL
with CE marking.
Ordering Information
Model Number Legend:
1. Input Type
L: Platinum-resistance thermometer or thermocouple
2. Output Type
None: No output
C: With relay contact output (SPDT)
3. Supply Voltage
100-240VAC: 100 to 240 VAC
24VAC/VDC: 24 VAC/VDC
CUS
Input type Supply voltage Output Model
Platinum-resistance thermometer or
thermocouple 100 to 240 VAC None K3MA-L 100-240VAC
1 relay contact output (SPDT) K3MA-L-C 100-240VAC
24 VAC/VDC None K3MA-L 24VAC/VDC
1 relay contact output (SPDT) K3MA-L-C 24VAC/VDC
K3MA-L-@
1 2 3

2
K3MA-LK3MA-L
Specifications
■Ratings
Input/Output Ratings
Relay Contact Output
K3MA-L 100-240VAC, K3MA-L-C 100-240VAC K3MA-L 24VAC/VDC, K3MA-L-C 24VAC/VDC
Supply voltage 100 to 240 VAC 24 VAC (50/60 Hz), 24 VDC
Operating voltage range 85% to 110% of the rated supply voltage
Power consumption (under
maximum load) 6 VA max. 4.5 VA max. (24 VAC)
4.5 W max. (24 VDC)
Insulation resistance 20 M
W
min. (at 500 VDC) between external terminal and case.
Insulation provided between inputs, outputs, and power supply.
Dielectric withstand voltage 2,000 VAC for 1 min between external terminal and case.
Insulation provided between inputs, outputs, and power supply.
Noise immunity
1,500 V on power supply terminals in normal or
common mode.
1
m
s, or 100 ns for square-wave noise with 1 ns.
480 V on power supply terminals in normal mode.
1,500 V in common mode.
1
m
s, or 100 ns for square-wave noise with 1 ns.
Vibration resistance Vibration: 10 to 55 Hz, Acceleration: 50 m/s2
5 min each in X, Y, and Z directions for 10 sweeps.
Shock resistance 150 m/s2(100 m/s2for relay contact outputs) 3 times each on 3 axes, 6 directions.
Ambient temperature Operating:
-
10
C to 55
C (with no condensation or icing)
Storage:
-
25
C to 65
C (with no condensation or icing)
Ambient humidity Operating: 25% to 85% (with no condensation)
Ambient atmosphere Must be free of corrosive gas.
Approved safety standards UL3121, conforms to EN61010-1 (Pollution degree 2/overvoltage category II)
Conforms to VDE0106/P100 (finger protection)
EMC (EMI) EN61326+A1 Industry
Emission Enclosure: CISPR 11 Group 1 class A: CISRP16-1/-2
Emission AC Mains: CISPR 11 Group 1 class A: CISRP16-1/-2
(EMS) EN61326+A1 Industry
Immunity ESD: EN61000-4-2: 4-kV contact discharge
8-kV air discharge
Immunity RF-interference: EN61000-4-3: 10 V/m (amplitude-modulated, 80 MHz to 1 GHz)
Electrical Fast Transient Noise: EN61000-4-4: 2 kV (power line)
Immunity Burst Noise: 1 kV line to line (I/O signal line)
Immunity Surge: EN61000-4-5: 1 kV (power line)
2-kV line to ground (power line)
Immunity Conducted Disturbance: EN61000-4-6: 3 V (0.15 to 80 MHz)
Immunity Voltage Dip/Interrupting: EN61000-4-11: 0.5 cycle, 0, 180
,
100% (rated voltage)
Weight Approx. 200 g
Item Resistive load (cos
f
= 1) Inductive load (cos
f
= 0.4, L/R = 7 ms)
Rated load 5 A at 250 VAC, 5 A at 30 VDC 1.5 A at 250 VAC, 1.5 A at 30 VDC
Rated carry current 5 A max. (at COM terminal)
Max. contact voltage 400 VAC, 150 VDC
Max. contact current 5 A (at COM terminal)
Max. switching capacity 2,000 VA, 192 W 375 VA, 36 W
Min. permissible load
(P level, reference value) 10 mA at 5 VDC
Mechanical life 20,000,000 times min. (at a switching frequency of 1,200 time/min)
Electrical life
(at an ambient temperature of 20
C) 100,000 times min. (at a rated load switching frequency of 10 time/min)

3
K3MA-LK3MA-L
■Measuring Ranges
Platinum-resistance Thermometer
Thermocouple
■Characteristics
Note: The indication accuracy of the K thermocouple at a temperature of
-
200 to 1300
C is
2
C
1 digit maximum.
The indication accuracy of the T and N thermocouples at a temperature of
-
100
C or less is
2
C
1 digit maximum.
The indicator accuracy of the U and L thermocouples at any temperature is
2
C
1 digit maximum.
The indication accuracy of the B thermocouple at a temperature of 400
C or less is unrestricted.
The indication accuracy of the R and S thermocouples at a temperature of 200
C or less is
3
C
1 digit maximum.
Input Pt100 JPt100
Range
C
-
200 to 850
-
199.9 to 500.0 0.0 to 100.0
-
199.9 to 500.0 0.0 to 100.0
F
-
300 to 1500
-
199.9 to 900.0 0.0 to 210.0
-
199.9 to 900.0 0.0 to 210.0
Parameter
Input K J T E L U N R S B
Range
C
-
200
to
1300
-
20.0
to
500.0
-
100
to
850
-
20.0
to
400.0
-
200
to
400
-
199.9
to
400.0
0
to
600
-
100
to
850
-
200
to
400
-
199.9
to
400.0
-
200
to
1300
0
to
1700
0
to
1700
100
to
1800
F
-
300
to
2300
0.0
to
900.0
-
100
to
1500
0.0
to
750
-
300
to
700
-
199.9
to
700.0
0
to
1100
-
100
to
1500
-
300
to
700
-
199.9
to
700.0
-
300
to
2300
0
to
3000
0
to
3000
300
to
3200
Parameter
Indication accuracy (at 23
5
C)
(See note.) Thermocouple:
(
0.5% of indication value or
1
C, whichever greater)
1 digit max.
Platinum-resistance thermometer:
(
0.5% of indication value or
1
C, whichever greater)
1 digit max.
Input Thermocouple: K, J, T, E, L, U, N, R, S, B
Platinum-resistance thermometer: JPt100, Pt100
Measurement method Double integral method
Sampling period 500 ms
Display refresh period Sampling period (sampling times multiplied by number of averaging times if average processing
is selected.)
Max. displayed digits 4 digits (
-
1999 to 9999)
Display 7-segment digital display, Character height: 14.2 mm
Polarity display “
-
” is displayed automatically with a negative input signal.
Zero display Leading zeros are not displayed.
Input shift Input shift equivalent to the setting value supported for all points within the sensor measurement
range.
Hold function Max hold (maximum value), Min hold (minimum value)
Hysteresis setting Programmable with front-panel key inputs (0001 to 9999).
Other functions Display color change (green (red), green, red (green), red)
Average processing (simple average OFF/2/4/8 operations)
Setting change lockout
Parameter initialization
Output Relay contact (SPDT)
Delay in comparative outputs 1 s max.
Enclosure ratings Front panel: NEMA4X for indoor use (equivalent to IP66)
Rear case: IEC standard IP20
Terminals: IEC standard IP00 + finger protection (VDE0106/100)
Memory protection Non-volatile memory (EEPROM) (possible to rewrite 100,000 times)

4
K3MA-LK3MA-L
Nomenclature
Name Functions
1. Main indicator Displays current values, parameters, and set values.
2. Opera-
tion indi-
cators
1 Lit when output 1 is ON.
SV Lit when a set value is being displayed or changed.
Max Lit when the main indicator is showing the MAX value.
Min Lit when the main indicator is showing the MIN value.
3. Level indicator Displays the current level that the K3MA-L is in. (See below for details.)
4. MAX/MIN Key Used to display the MAX and MIN values when a measurement value is being displayed.
5. Level Key Used to change the level.
6. Mode Key Used to allow the main indicator to indicate parameters sequentially.
7. Shift Key Used to enable a set value to be changed. When changing a set value, this key is used to move along the digits.
8. Up Key Used to change a set value. Used to set or clear a forced-zero function when a measurement value is being dis-
played.
Level indicator Level
/Protect
Not lit Operation
Adjustment
2Initial setting
%Advanced-function setting
MAX/MIN LEVEL SHIFTMODE UP
1
SV
Max
Min
2. Operation indicators
1. Main indicator
8. Up key
6. Mode key 7. Shift key
5. Level key
4. MAX/MIN key
3. Level indicator

5
K3MA-LK3MA-L
Operation
■Main Functions
Input Types and Ranges
Note: The initial value is “
thermocouple K (
-
200 to 1300
C/
-
300 to 2300
F).”
Temperature Unit Selection
Either centigrade (
C) or fahrenheit (
F) can be selected as the
temperature unit.
OUT Types (Comparative Output Models Only)
OUT 1 can be set to operate in one of the three following modes
in accordance with the compared values:
• Upper limit (High Acting):
The output is turned ON when the measurement value is greater
than its set value.
• Lower limit (Low Acting):
The output is turned ON when the measurement value is less
than its set value.
• Upper and lower limits (Outside Band Acting):
An upper limit (H set value) and lower limit (L set value) can be
set independently.
The output is turned ON when the measurement value is greater
than the upper-limit set value or less than the lower-limit set val-
ue.
Parameter Setting Input type Meaning
(-3
Platinum-resistance
thermometer Pt100
-
200 to 850
C
-
300 to 1500
F
-
199.9 to 500.0
C
-
1999 to 900.0
F
0.0 to 100.0
C 0.0 to 210.0
F
JPt100
-
199.9 to 500.0
C
-
199.9 to 900.0
F
0.0 to 100.0
C 0.0 to 210.0
F
Thermocouple K
-
200 to 1300
C
-
300 to 2300
F
-
20.0 to 500.0
C 0.0 to 900.0
F
J
-
100 to 850
C
-
100 to 1500
F
-
20.0 to 400.0
C 0.0 to 750.0
F
T
-
200 to 400
C
-
300 to 700
F
-
199.9 to 400.0
C
-
199.9 to 700.0
F
E 0 to 600
C 0 to 1100
F
L
-
100 to 850
C
-
100 to 1500
F
U
-
200 to 400
C
-
300 to 700
F
-
199.9 to 400.0
C
-
199.9 to 700.0
F
N
-
200 to 1300
C
-
300 to 2300
F
R 0 to 1700
C 0 to 3000
F
S 0 to 1700
C 0 to 3000
F
B 100 to 1800
C 300 to 3200
F
Parameter Setting Meaning
#4 "
Display in
C.
%
Display in
F. Parameter Setting Meaning
.433 '(
Upper limit: Alarm
operates at upper
limit.
+.
Lower limit: Alarm
operates at lower
limit.
'(+.
Upper and lower
limits: Alarm oper-
ates at upper and
lower limits.
ON
OFF
ON
OFF
ON
OFF
Upper Limit (High Acting) Lower Limit (Low Acting) Upper and Lower Limits
(Outside Band Acting)
Hysteresis
Hysteresis
Hysteresis
OUT1 value
Measurement
value
Output
Measurement
value
OUT1 value
Output
OUT1 upper-limit
value
Measurement
value
OUT1lower-limit
value
Output
Hysteresis

6
K3MA-LK3MA-L
Temperature Input Shift
Input shift equivalent to the setting value supported for all points
within the sensor measurement range.
Parameter Initialization
This function returns all of the parameters to their initial values.
Use this to reset the K3MA-L after returning it to its factory-set
condition.
Average Processing
Average processing stabilizes displayed values to minimize flicker
by averaging the fluctuating input signals. Average processing
can be performed for the measurement values in either of four
steps (OFF, 2 times, 4 times, or 8 times).
This is useful for ignoring rapid fluctuations, e.g., eliminating spike
noise.
Hysteresis (Comparative Output Models Only)
The hysteresis of comparative outputs can be set to prevent chat-
tering in the output when the measurement value fluctuates finely
near the OUT value.
Changing the Display Color
The color of the value displayed can be set to either red or green.
For comparative output models, the display color can be set to
change from green to red, or from red to green, according to the
status of the comparison criterion.
Display Auto-return Time
This function automatically returns the display to the operation
level’s current value if no keys are pressed for a preset time
(called the display auto-return time).
Move-to-Protect-Level Time
The time required to shift to the protect level can be set as
desired.
MAX/MIN Display
The maximum and minimum measurement (display) values from
the time the power is turned ON until the current time can be
stored and displayed. This is useful, for example, when measur-
ing the maximum value.
Parameter Setting
(-2
to
Parameter Setting Meaning
(-(3 .%%
---
.-
Initializes all param-
eters.
Temperature
Before shift
Temperature input
shift value
After shift
Input
Upper limit (high acting)
Set value
Output ON
OFF
Hysteresis
Measurement value
192.0
258.9
OUT1 value Green
Red
MAX/MIN LEVEL SHIFTMODE UP
MAX/MIN
MAX/MIN
MAX/MIN
MAX/MIN
Max
LEVEL SHIFTMODE UP
MAX/MIN
Min
LEVEL SHIFTMODE UP
Current value
MAX value
MIN value

7
K3MA-LK3MA-L
■External Connections
Terminal Arrangement
Output
terminals
Input
terminals
Power
supply
E1
E2
A1
A2
E3
E4
E5
E6
+
−
E4
E5
E6
A
B
B
Models with
comparative
output
For platinum-
resistance
thermometer input
For thermocouple
input
100- to 240-VAC type or
24-VAC/VDC type
(No polarity for 24-VDC
connection.)
Terminal No. Name Description
Operation power Connects the operation power supply.
Thermocouple or platinum-resistance ther-
mometer input Connects the thermocouple or platinum-re-
sistance thermometer input.
Outputs Outputs the relay outputs.
A1 A2
-
E6 E5
-
E4 -
E2 E3
-
E1 ,

8
K3MA-LK3MA-L
■Block Diagram
Note:Relay output models only.
X
EEPROM
5 V 12 V
Input Input
Micro-
computer
Key
Display
Output circuit
Constant voltage circuit
Power supply circuit
Contact
output (See
note.)

9
K3MA-LK3MA-L
■Levels
“Level” refers to a grouping of parameters. The following table lists the operations that are possible in each of the levels, and the diagram
tells how to move between levels. There are some parameters that are not displayed for certain models.
Note: The move-to-protect-level time can be set in the advanced-function setting level.
Level name Function Measurement
Protect Setting lockouts. Continue
Operation Displaying current values, and setting OUT 1 value. Continue
Adjustment Setting communications writing control. Continue
Initial setting Making initial settings of input type, output operating action, and
other parameters. Stopped
Advanced-function setting Setting average processing, display color settings, and other ad-
vanced function parameters. Stopped
Protect level
Operation level
Power ON
+
Time set
by user
1 s min. - 123.4
Flashing stops if key
is released.
1 s min.
Initial setting level
Password "−169"
1 s min.
Advanced-function
setting level
Continue to press the key for 2 s min.
Indicates change of level.
1 s min.
(See note.)
Adjustment level
Less than 1 s
+

10
K3MA-LK3MA-L
■Parameters
Operation level
Set one of these.
Current value
OUT1
value
OUT1
upper-limit
value
OUT1
lower-limit
value
Note: 1. Some parameters are not displayed for certain models.
2. The K3MA-L will stop measurement if the level is changed to the initial setting level or
the advanced-function setting level.
3. If the input range is changed, some parameters are set to default values.Therefore,
set the input range first.
4. Settings displayed in reversed colors are initial settings.
Power ON
∼
∼
∼
∼∼
0
Adjustment level
Press Level Key
Press Level Key
for less than 1 s.
for less than 1 s.
Temperature input shift
For models with the
comparative output function
MODE
MODE
MODE
MODE

11
K3MA-LK3MA-L
Enter
password
"−169"
Unit: times
Unit: s
Input type
OUT1 type
Move to
advanced-
function
setting level
Parameter
initialization
Average
processing
OUT1
Hysteresis
Display
color
change
Display
auto-return
time
Move-to-
protect-
level time
Green (red)
Green
Red (green)
Red
Upper Limit
Lower Limit
Upper/Lower
Limits
Press Level
Key for
more than 1 s.
Press Level Key for less than 1 s.
Initial setting level Advanced-function
setting level
Press Level
Key for
more than 3 s.
∼
∼
∼ ∼
∼∼
/
/
//
∼∼
∼∼
Unit: s
MODE
MODE
MODE
MODE
MODE
Temperature
unit
MODE
MODE
MODE
MODE
MODE
Temperature unit
OUT1 type
Password:
-
169
Models
with the
compara-
tive out-
put
function
Models
with the
compar-
ative out-
put
function
OUT1 hysteresis
Settings displayed in reversed colors are initial settings.

12
K3MA-LK3MA-L
Operation/Adjustment Lockouts
Restricts key operations for operation level and adjustment level.
• Initial setting is 0.
• This cannot be displayed on models not equipped with the com-
parative output function.
Setting Level Lockouts
Restricts shifting to initial setting level or advanced-function set-
ting level.
Setting Change Lockout
Restricts setting changes by key operation. When this lockout is
set, it is no longer possible to shift to a setting change mode.
However, all protect level parameters can still be changed.
■Initial Settings
Operation/adjustment
lockouts
Setting level lockouts
Setting change lockout
//
//
/
Protect level
MODE
MODE
MODE
Operation Level
Press Level Key + Mode Key for more than 1 s.
Press Level Key + Mode Key
for more than preset time.
Param-
eter Setting Operation level Moving to
adjustment
level
Process value
display Set value
display
. /3
Allowed Allowed Allowed
Allowed Allowed Prohibited
Allowed Prohibited Prohibited
Parameter Setting Shift to initial
setting level Shift to
advanced-
function
setting level
("/3
Allowed Allowed
Allowed Prohibited
Prohibited Prohibited
Parameter Setting Setting change by key
operation
53/3 .%%
Allowed
.-
Prohibited
Press the Level Key for 3 s min.to move to the initial setting
level.
Select the input type.
Press the Level Key for less than 1 s min.to return to the
operation level.
Measurement starts.
Power ON
If desired, shift to the adjustment level to set the temperature
input shift value.
If required, shift to the advanced-function setting level to set the
number of measurements for averaging, hysteresis values,
auto-zero time, display color change, display auto-return time,
or move-to-protect-level time.
Set the OUT values.

13
K3MA-LK3MA-L
■Setting Example
Initial Settings
The settings for the following example are shown here.
Example: Monitoring the temperature of an industrial furnace
Here, the temperature inside the furnace is to be displayed in cen-
tigrade (
C).
Temperature sensor: E52-PR Thermocouple, Measurement
range: 0 to 1,400
C.
1. Set the K3MA-L input type to the thermocouple R input
range.
Parameter:
(-3
(input type), Setting value:
2. Select centigrade (
C) as the temperature unit.
Parameter:
#4
(temperature unit), Setting value:
"
If you are using a comparative output model, make the setting as
desired.
■Application Examples
MAX/MIN LEVEL SHIFTMODE UP
1
SV
Max
Min
Industrial furnace
K3MA-L-C AC100-240V
E52-PR Thermocouple
K3MA-L
K3MA-L
E5EN
K3MA-L
Monitoring the temperature of
an industrial furnace Sending a temperature alarm
for molding equipment Monitoring the bearing
temperature for a generator
motor
Industrial furnace
PID control
Molding equipment Alarm output
Generator motor
Alarm output
• Monitoring the temperature of an indus-
trial furnace/sintering furnace.
• Monitoring/alarm function for disinfect-
ing equipment.
• Monitoring (failsafe checking) abnormal
temperatures in molding equipment.
• Monitoring the liquid temperature for
cleaning devices.
• Monitoring temperature rises in electric
power generating facilities.
• Inspecting temperatures in machines
and devices.

14
K3MA-LK3MA-L
Dimensions
Optional Parts (Order Separately)
Installation
1. Insert the K3MA-L into the panel cut-out hole.
2. For a waterproof installation, insert the rubber gasket onto
the body of the K3MA-L.
3. Fit the adaptor into the grooves on the left and right sides of
the rear case, then push it until it contacts the panel to
secure the K3MA-L.
■Wiring Precautions
• Use crimp terminals.
• Tighten the terminal screws to a torque of approximately 0.5
N
¼
m.
• To avoid the influence of noise, route signal lines and power lines
separately.
■Wiring
• Use the following M3 crimp terminals.
■Unit Markings (Provided)
• The unit markings are not attached to the K3MA-L. Select the de-
sired markings from the provided sheet.
Note: For scales and gauges, use the unit markings that are
specified by the relevant laws or regulations.
Name Shape Model
Splash-proof Soft Cover K32-49SC
Hard Cover K32-49HC
ABCDE
UPSHIFTMODELEVELMAX/MIN
80
44.8
48
96
1.3
12
85 97
101.2
91 120 min.
75 min.
45
+0.5
0
92+0.5
0
14.2 mm
7.6 mm
Finger protective
cover (provided) Main indicator
character size Panel cut-out
5.8 mm max.
5.8 mm max.

15
K3MA-LK3MA-L
Precautions
!Caution
Do not touch the terminals while the power is being supplied.
Doing so may result in electric shock.
!Caution
Do not disassemble the product or touch the internal compo-
nents of the product while the power is being supplied. Doing
so may result in electric shock.
!Caution
Do not allow pieces of metal or wire clippings to enter the prod-
uct. Doing so may result in electric shock, fire, or malfunction.
!Caution
Perform correct settings for the product according to the control
application. Failure to do so may cause unexpected operation,
resulting in damage to the product or injury.
!Caution
Take safety measures, such as installing a separate monitoring
system, to ensure safety even if the product fails. Product fail-
ure may prevent comparative outputs from being generated,
resulting in serious accidents.
Observe the following precautions to ensure safety.
1. Maintain the power supply voltage within the range speci-
fied in the specifications.
2. Maintain the load within the ratings specified in the specifi-
cations.
3. Check each terminal for correct number and polarity before
connecting it. Incorrect or reverse connections may dam-
age or burn out internal components in the product.
4. Tighten the terminal screws securely. The recommended
tightening torque is 0.43 to 0.58 N¼m. Loose screws may
cause fire or malfunction.
5. Do not connect anything to unused terminals.
6. Provide a switch or circuit breaker so that operators can
easily turn OFF the power supply when necessary. Also
provide appropriate indications of such devices.
7. Do not attempt to disassemble, repair, or modify the prod-
uct.
8. Do not use the product where flammable or combustible
gases are present.
Application
General Precautions
1. Do not use the product in the following locations:
• Locations subject to direct radiant heat from heating
equipment.
• Locations subject to exposure to water, oil, or chemicals.
• Locations subject to direct sunlight.
• Locations subject to dust or corrosive gases (particularly,
sulfuric gas or ammonia gas).
• Locations subject to severe changes in temperature.
• Locations subject to icing or condensation.
• Locations subject to shock or vibration.
2. Do not block heat dissipation around the product, i.e., pro-
vide sufficient space for heat dissipation.
3. Ensure that the rated voltage is reached within two seconds
after the power is turned ON.
4. Conduct aging for 15 minutes min. after power is turned ON
for correct measurement.
5. Do not touch the slit sections or terminals while the power
is being supplied to prevent the product from being affected
by static electricity.
6. Do not lay heavy objects on the product during use or stor-
age. Doing so may deform or deteriorate the product.
7. Do not use paint thinner for cleaning. Use commercially
available alcohol.
Mounting
• Mount the product to a panel that is 1 to 8 mm thick.
• Install the product in a horizontal position.
• Use crimp terminals that match screw sizes.
Noise Prevention
• Install the product as far as possible from devices that generate
strong, high-frequency fields (such as high-frequency welders or
sewing machines) or surges.
• Install surge absorbers or noise filters on nearby devices that
generate noise (particularly motors, transformers, solenoids,
magnet coils, and other devices that have a high inductance
component). Do not connect a surge absorber to the tempera-
ture sensor input section of the K3MA-L.
• To prevent inductive noise, separate the terminal block wiring for
the product from high-voltage or high-current power lines. Do not
route the wiring for the product in parallel with or tie it in a bundle
with power lines.
Take the following countermeasures against inductive noise in
input lines.
Temperature Inputs
Separate the lead wire that connects the product with a temperature
sensor from the load line to prevent the product from being affected by
inductive noise.
• When using a noise filter for the power supply, check for the volt-
age and current and install it as close as possible to the Temper-
ature Meter.
• Do not install the product near radios, television sets, or wireless
devices. Doing so may cause reception interference.
Increasing Service Life
• Do not use the product in locations where the temperature or hu-
midity exceeds the ratings or where condensation may occur.
When installing the product in a panel, be sure that the temper-
ature around the product (not the temperature around the panel)
does not exceed the ratings. The product service life depends on
the ambient temperature. The higher the ambient temperature,
the shorter the service life. To extend the product service life,
lower the temperature inside the Temperature Meter.
• Use and store the product within the temperature and humidity
ranges given in the specifications. When gang-mounting Tem-
perature Meters or arranging them vertically, heat generated by
the Temperature Meters will cause the internal temperature to
rise, reducing the service life. In such cases, consider forced
cooling methods, such as using a fan to circulate air around the
Temperature Meters. Do not, however, allow only the terminals
to be cooled. Doing so will increase measurement error.
• The life of the output relays are greatly affected by the switching
capacity and switching conditions. Use these relays within their
rated load and electrical life. The contacts may fuse or burn if
they are used past their electrical life.
+
−
Line filter
Tempera-
ture Meter
Surge absorber
Signal
input
Power
supply
input
Power
supply
input Tempera-
ture Meter

16
K3MA-LK3MA-L
■Troubleshooting
When an error occurs, error details will be displayed on the main indicator. Confirm the error from the main indicator and take the appropri-
ate countermeasures.
Level display Main indicator Error contents Countermeasures
Not lit
$
RAM memory error Repair is necessary.
Consult your OMRON sales rep-
resentative.
$
EEPROM memory error When this error is displayed,
press the Level Key for 3 sec-
onds, and the settings will be re-
stored to the factory settings.
If the error cannot be recovered,
repair is necessary.
Consult your OMRON sales rep-
resentative.
Not lit Flashes
2$11
Input error Confirm that the temperature
sensor is correctly connected,
and that there are no broken sig-
nal lines to the temperature sen-
sor.
If the condition does not return to
normal, repair is necessary.
Consult your OMRON sales rep-
resentative.
Not lit Flashes
The measurement value after
temperature input correction ex-
ceeds 9999.
Thetemperatureinput correction
value may be inappropriate.
Use the adjustment level to re-
view the temperature input cor-
rection value.
Not lit Flashes
The measurement value after
temperature input correction is
lower than -1999.
Thetemperatureinput correction
value may be inappropriate.
Use the adjustment level to re-
view the temperature input cor-
rection value.
In the interest of product improvement, specifications are subject to change without notice.
ALL DIMENSIONS SHOWN ARE IN MILLIMETERS.
To convert millimeters into inches, multiply by 0.03937. To convert grams into ounces, multiply by 0.03527.
Cat. No. N109-E1-01
OMRON Corporation
Industrial Automation Company
Measuring and Supervisory Controls Department
Shiokoji Horikawa, Shimogyo-ku
Kyoto, 600-8530 Japan
Tel: (81)75-344-7108/Fax: (81)75-344-7189 Printed in Japan
1001-1M (1001) (B)
Other manuals for K3MA-L
2
Table of contents
Other Omron Monitor manuals

Omron
Omron NS - User manual

Omron
Omron K3MA-F - User manual

Omron
Omron F3UV Series User manual

Omron
Omron HCG-801 User manual

Omron
Omron SMART ACTIVE PARTS User manual

Omron
Omron FZ-M08 User manual

Omron
Omron EQUO Series User guide

Omron
Omron XPECTIA FZD User manual

Omron
Omron FZ5-110 User manual

Omron
Omron K3HB-C Series User manual

Omron
Omron HBF-306 User manual

Omron
Omron AX-OP05-E User manual

Omron
Omron K3HB-H User manual

Omron
Omron Sysmac NA Series User manual

Omron
Omron HBF-306 User manual

Omron
Omron fz5-l35 series User manual

Omron
Omron K3HB User manual

Omron
Omron K3HB User manual

Omron
Omron K3NP - 1998 User manual

Omron
Omron ns series User manual