16. If any of the following situations arises, get the equipment
checked by service personnel:
a. The power cord or plug is damaged.
b. Liquid has penetrated into the equipment.
c. The equipment has been exposed to moisture.
d. The equipment does not work well, or you cannot get it to
work according to the users manual.
e. The equipment has been dropped and damaged.
f. The equipment has obvious signs of breakage.
17. DO NOT LEAVE THIS EQUIPMENT IN AN UNCONTROLLED
ENVIRONMENT WHERE THE STORAGE TEMPERATURE IS
BELOW -20°C (-4°F) OR ABOVE 60°C (140°F). IT MAY DAMAGE
THE EQUIPMENT.
18. External equipment intended for connection to signal
input/output or other connectors, shall comply with relevant UL
/ IEC standard (e.g. UL 1950 for IT equipment and ANSI/AAMI ES
60601-1: 2005 AND CAN/CSA-C22.2 No. 60601-1:08 / IEC 60601
series for systems –shall comply with the standard IEC
60601-1-1, Safety requirements for medical electrical systems.
Equipment not complying with UL 60601-1 shall be kept outside
the patient environment, as defined in the standard.
19. Unplug the power cord from the power adapter jack to
disconnect the device.
Caution:
It may cause the danger of explosion if battery is incorrectly
replaced. Replace only with same or equivalent type
recommended by the manufacturer.