OSAKA VACUUM TC163H User manual

OSAKA VACUUM, LTD.
Document No. : ND00338-R06
INSTRUCTION MANUAL
POWER SUPPLY
For the pump
MODEL: TC163H
TC223
TC523H
TC203
TC353
TC1104
IMPORTANT
Read rule carefully for safe installation, operation and instruction.
Please keep this instruction manual by your side.
OSAKAVACUUM, LTD.

OSAKA VACUUM, LTD.
Document No. : ND00338-R06
1
-INTRODUCTION-
The power supply provides the pump with power for operation.
Prior to operating the power supply, we recommend you to read this instruction manual
carefully to understand various functions and correct handling thoroughly.
If you keep this manual by your side, it will be helpful when you are not sure of something
during operation or you have something wrong.
-WARRANTY-
Our company warrants the quality of power supply in accordance with the conditions of
warranty prescribed in “STANDARD WARRANTY CONDITIONS”.
-PRECAUTIONS-
Please read carefully the precautionary instruction on the operation and installation given on
from page 2 to page 3.
Special care should be taken for the parts with the following heading, “Caution” and
“Warning” signs throughout in this manual.
! CAUTION: An important consideration is described in order to operate the pump
and the power supply safely.
! WARNING: An important consideration is described in order to avoid any injuries or
a risk of electric shock when operating the pump and the power supply.

OSAKA VACUUM, LTD.
Document No. : ND00338-R06
2
1) Input power
! CAUTION
Input power voltage isAC90~121V orAC180~253V.
Set the input voltage select switch of the rear panel of the power supply according to the
input voltage.
Supply the power excluding noise, surge and voltage fluctuation to the power supply.
2) Combinations of the power supply and the pumps.
! CAUTION
Match the power supply with the correct pump.
Prior to operation, supply the input power and press the F1 button, and then check the
model indicated on the top in display.
Correct combinations of the power supply and the pumps, refer to page 6, table 1.
The power supply is set to the model matched to the pump at shipping from our company.
The nameplates are stuck on the side and bottom panels of the power supply at
shipping from our company.
When requesting our company for repair or making other inquiries on the product, please
do not forget to tell us the serial number.
3) Fan mode selection
! CAUTION
Set the “function for detecting fan disconnected” to “ON” or “OFF” depending on
air-cooled type or water-cooled type of the pump.
The fan mode select switch is set to “FAN” at shipping from our company.
4) Earth
! WARNING
Be sure to connect the earth line of the INPUT connector to EARTH/GROUND.
5) Connections of cables
! CAUTION
DO NOT disconnect the cables or connectors of the power supply in operation.
6) Disassembly and modification
! CAUTION
DO NOT disassemble or modify the pump, the power supply and the cable.
It may result in failures or accidents and fails safety.

OSAKA VACUUM, LTD.
Document No. : ND00338-R06
3
7) Electric shock
! WARNING
DO NOT touch the inside of the alive power supply.
DO NOT touch the inside of the power supply within 5 minutes after turning the input
power off, or you may get electric shock.
8) Ambient temperature/humidity
! CAUTION
Avoid locations where the power supply is exposed to high temperature and humidity.
9) Water drop
! CAUTION
Avoid locations where the water drops wet the power supply.
10) Dust and flammable/corrosive gas
! CAUTION
Avoid locations where dust, flammable gas or corrosive gas may affect the power supply.
11) Electric/magnetic field
! CAUTION
Avoid locations where strong electric field or magnetic field exists.
12) Vibration
! CAUTION
Avoid locations where excessive vibration exists.
13) Space
! CAUTION
Be sure to provide sufficient space around the cooling fan on the rear of the power supply.
-OPENING THE CRATE-
Check that the power supply and attachments are not damaged caused during transportation. There
are spare parts and attachments below in a standard set of the power supply.
Input power connector 1 pc.
Remote connector 1 pc.
Serial communication connector 1 pc.
Instruction manual of power supply 1 copy
Instruction manual of serial communication 1 copy
“AC100/110V” seal 1 sheet

OSAKA VACUUM, LTD.
Document No. : ND00338-R06
4
CONTENTS
Introduction ------------------------------------------------------------------------------------------1
Warranty ------------------------------------------------------------------------------------------1
Precautions ------------------------------------------------------------------------------------------1
Opening the crate ------------------------------------------------------------------------------------------3
1. Feature ---------------------------------------------------------------------------------------------------6
2. Specifications ------------------------------------------------------------------------------------------7
3. Buttons and display ---------------------------------------------------------------------------------8
3.1 Input voltage select switch ---------------------------------------------------------------8
3.2 FUNCTION buttons (F1,F2,F3) ---------------------------------------------------------------8
3.3 START button ------------------------------------------------------------------------8
3.4 STOP/RESET button ------------------------------------------------------------------------8
3.5 Display ---------------------------------------------------------------------------------9
3.5.1 Indication of each operation status
3.5.2 Indication by Function buttons
4. MAINTENANCE MODE ----------------------------------------------------------------------- 10
4.1 Function for no-load test ------------------------------------------------------------------------10
4.2 Function for resetting total operation time ------------------------------------------------------10
4.3 Function for canceling protection signal ----------------------------------------------------- 11
4.4 Function for detecting fan disconnected ------------------------------------------------------- 11
4.5 Function for selecting LOCAL/REMOTE/SERIAL COMMUNICATION ----------------11
4.5.1 LOCAL mode
4.5.2 REMOTE mode
4.5.3 SERIAL COMMUNICATION mode
4.6 Function for using CRC code ---------------------------------------------------------------------12
4.7 Function for error (alarm) log ---------------------------------------------------------------------12
5. Protective functions ---------------------------------------------------------------------------------14
5.1 Protective function ---------------------------------------------------------------------------------14
5.2 Warning function ---------------------------------------------------------------------------------16
6. Reset of error ---------------------------------------------------------------------------------17
6.1 Over frequency and System error ---------------------------------------------------------17
6.2 Theothers -----------------------------------------------------------------------------------17
7. Installation and electrical connections ---------------------------------------------------------------17
7.1 Installation ---------------------------------------------------------------------------------17
7.2 Electrical connections ---------------------------------------------------------------18
7.2.1 Connection of input power cable
7.2.2 Connection of motor cable
7.2.3 Connection of fan cable
7.2.4 Connection of remote cable

OSAKA VACUUM, LTD.
Document No. : ND00338-R06
5
8. Maintenance and inspection ------------------------------------------------------------------------22
8.1 Tuning ---------------------------------------------------------------------------------22
8.2 Adjusting the depth of character in display (LCD contrast) ------------------------------- 22
8.3 No-load test of power supply ---------------------------------------------------------------22
9. Others -------------------------------------------------------------------------------------------------- 23
Appendix
Attached drawings
Outside view drawing of the power supply ---------- drawing number : X3-00080
Input/output Communications of the power supply --------- drawing number : X3-00081
Standard warranty conditions

OSAKA VACUUM, LTD.
Document No. : ND00338-R06
6
1. Feature
This power supply is a high-frequency inverter to drive the pump. The power supply controls
motor current to a fixed value and set the output voltage proportional to output frequency. Thus
the power supply gives full play to its ability and can start the pump in a short time.
The power supply shows the status of the pump and power supply on the display of the front
panel, for example, output frequency, operation status and so on.
! CAUTION
Table 1 shows the combinations of the power supply and the pumps. Check the model and
match the power supply with the correct pump.
Table 1. Combinations of power supply and the pumps
Power supply model ⇒Pump models
TC163H ⇒TH162/163
TC203 ⇒TG200/203
TS50
TC223 ⇒TG220F
TC353 ⇒TG350F
TG450F
TC523H ⇒TH520/522
TC1104 ⇒TG800F
TG1100F

OSAKA VACUUM, LTD.
Document No. : ND00338-R06
7
2. Specifications
Table 2 shows the specifications of the power supplies.
Table 2. Specifications of the power supplies
Power supply model TC163H TC203 TC223 TC353 TC523H TC1104
Applicable the model of pumps TH162
TH163
TG200
TG203
TS50
TG220F TG350F
TG450F
TH520
TH522
TG800F
TG1100F
Temperature ‐5 - 40℃
Humidity 35 - 85%
Ambient
Pollution degree 2
Phase Single
Voltage 1) AC90 - 110 or AC180 - 253[V]
Frequency 50/60[Hz]±2[%]
Max Power 490[VA] 610[VA] 590[VA] 610[VA] 590[VA] 610[VA]
Rating Power 210[VA] 210[VA] 96[VA] 92[VA] 210[VA] 86[VA]
Input
Over voltage
category Ⅱ
Phase 3 phases
Voltage 2) 45[V] 60[V] 60[V] 60[V] 60[V] 60[V]
Output
Frequency 2) 800[Hz] 800[Hz] 800[Hz] 750[Hz] 400[Hz] 560[Hz]
Brake method DC constant current
Brake operation time 4[min] 7[min] 4[min] 9[min] 14[min] 30[min]
Weight Approx. 2.7[kg]
Max. EARTH/GROUND leakage
current 3.5[mA]
IP Rating IP20 (Service area or operation area)
Withstand Voltage 3) Viso. INPUT(supply)-EARTH/GROUND
Viso. INPUT(supply)-REMOTE(interface)
Viso. INPUT(supply)-SERIAL(interface)
1.5[kV]
2.3[kV] (SELV)
2.3[kV] (SELV)
1) Set the input voltage switch of the rear panel of the power supply according to the input
voltage.
2) Value of normal operation
3) Don’t perform the withstand voltage test. When testing, the inner parts must remove.
! CAUTION
DO NOT perform the withstand voltage test.

OSAKA VACUUM, LTD.
Document No. : ND00338-R06
8
3. Buttons and display
This section explains buttons and display of the power supply.
Display
FUNCTION buttons (F1,F2,F3)
STOP/RESET button
START button
Input voltage select switch
Figure 1. Front and rear panel of the power supply
3.1 Input voltage select switch
! CAUTION
Input power voltage is AC90~121V or AC180~253V.
Set the input voltage select switch of the rear panel of the power supply according to the input
voltage.
Supply the power excluding noise, surge and voltage fluctuation to the power supply.
Input voltage select switch is set to “200” at shipping.
3.2 FUNCTION buttons (F1,F2,F3)
These buttons are for entering into MAINTENANCE MODE, confirming or altering value of
parameters on each functional items in MAINTENANCE MODE. Besides, confirming or altering
value of parameters for serial communication. How to operate MAINTENANCE MODE is
shown at section 4. MAINTENANCE MODE. How to operate serial communication is separately
shown at INSTRUCTION MANUAL SERIAL COMMUNICATION.
3.3 START button
Press to start driving when the operation mode is “LOCAL mode”.
3.4 STOP/RESET button
Press to stop driving when the operation mode is “LOCAL mode”.
Also, press to reset the power supply stopped with error with all Operation Modes.

OSAKA VACUUM, LTD.
Document No. : ND00338-R06
9
3.5 Display
3.5.1 Indication of each operation status
The display indicates rotating speed, operation status, total operation time, operation mode, error
message and others during operation of the pump. Details of display are shown at following Table
3.
Table 3. The indication of display
Operation Status Contents Display Note.3
Power ON Top: model of the power supply (for 5 seconds)
Bottom: revision number of the power supply (for
5 seconds)
TC1104
Rev.A
Under Standby or Under
rotating at Free Run Top: operation status and operation mode
Bottom: rotating speed and total operation time
Note.1
Standby L
0rpm 12345h
Under Acceleration Top: operation status and operation mode
Bottom: rotating speed and total operation time
Note.1
Acceleration L
3000rpm 12345h
Under rotating at
Normal Top: operation status and operation mode
Bottom: rotating speed and total operation time
Note.1
Normal L
33600rpm 12345h
Under Acceleration
from normal operation
Note.2
Top: operation status and operation mode
Bottom: rotating speed and total operation time
Note.1
Acceleration L
18000rpm 12345h
Under deceleration Top: operation status and operation mode
Bottom: the rest of braking time and total
operation time
Brake L
30:00 12345h
Under
MAINTENANCE
MODE
Top: parameter
Bottom: value of parameter Shown at section 4
Under Working
Protective Functions Top: error message
Bottom: total operation time Shown at section 5
Note.1) rotating speed means output frequency times 60.
Note.2) Acceleration means blinking.
Note.3) “L” in display means operation mode is set to Local Mode. Each “R”, “S” means
operation mode is set to REMOTE, SERIAL COMMUNICATION. “Operation mode”
means the way for operating the power supply.

OSAKA VACUUM, LTD.
Document No. : ND00338-R06
10
3.5.2 Indication by FUNCTION buttons
The display is indicated as follows by pressing the FUNCTION buttons under except
MAINTENANCE MODE or working protective functions.
The model and revision number of the power supply are indicated as follows by pressing the F1
button.
TC1104
Rev.A
The value of parameters for serial communication is indicated as follows by pressing the F1
button again. How to alter the value of parameters is shown at INSTRUCTION MANUAL
SERIAL COMMUNICATION.
RS232C Echo:OFF
9600 8bit N 1sp
“LCD Contrast” is indicated as follows by pressing F1 button again.
LCD Contrast
5
You can adjust the depth of character on this page. Details of “LCD Contrast” is shown at section
8.2. The display returned to the first indication by pressing the F1 button again.
4. MAINTENANCE MODE
You can select value of parameter in some functional items at MAINTENANCE MODE.
At first, press F3 button and F1 button (press F3 button before F1 button) on front panel for 5
seconds when “Standby” is indicated on display, and “Maintenance Mode” is indicated on display.
Next, press F2 button, and the present value of each parameters are indicated every time F1
button is pressed. In case of altering the value of parameter, select the value of parameter to alter
by pressing F2 button, and press F3 button on the value to be set. Refer to Figure.2
4.1 Function for no-load test
ON: Able to operate with no load (Refer to section 8.3)
OFF: Unable to operate with no load
Function for no-load test is set to “OFF” at shipping.
! CAUTION
Be sure to set the function for no-load test to “OFF” usually.
4.2 Function for resetting total operation time
ON: The total operation time indicated on the display is reset to zero. When this functional item
is left ON, the total operation time will be reset to zero every time input power is supplied.
OFF: The total operation time is not reset to zero.
Function for resetting total operation time is set to “OFF” at shipping.

OSAKA VACUUM, LTD.
Document No. : ND00338-R06
11
4.3 Function for canceling protection signal
ON: Protection signal is effective.
OFF: Protection signal is cancelled.
The protection signal means interlock signal connected to the remote connector on the rear panel.
This signal works as interlock for the purposes below:
zTo stop the power supply when cooling water for the pump is cut off.
zTo stop the power supply when backing pressure of the pump rises up.
zTo stop the power supply for other reason.
Set this function to “OFF” when operating the power supply without closing this protection
signal.
Function for canceling protection signal is set to “ON” at shipping.
! CAUTION
Be sure to set the function to “ON” in closing protection signal in the remote connector on
the rear panel so as to interlock.
4.4 Function for detecting fan disconnected
ON: Detection of Fan trouble is effective, and the power (DC12V) for fan is supplied from
the fan connector on the rear panel while input power is supplied.
OFF: Detection of Fan trouble is cancelled, and the power for fan is not supplied.
Set to “ON” or “OFF” according to cooled way of the pump.
“Function for detecting fan disconnected” is set to “ON” at shipping without cooled way
of the pump.
4.5 Function for selecting operation mode LOCAL/REMOTE/SERIAL COMMUNICATION
Function is to select the way for operating the power supply.
4.5.1 LOCAL MODE
You can use START/STOP buttons on the front panel in LOCAL mode.
You cannot use START/STOP signals of the remote connector and serial communication
connector on the rear panel because the signal is ignored.
4.5.2 REMOTE MODE
You can use START/STOP signals of the remote connector on the rear panel.
You cannot use START/STOP buttons on the front panel and START/STOP signals of the
serial communication connector.

OSAKA VACUUM, LTD.
Document No. : ND00338-R06
12
4.5.3 SERIAL COMMUNICATION MODE
You can use START/STOP signals of the serial communication connector on the rear panel.
You cannot use START/STOP buttons on the front panel and START/STOP signals of the
remote connector.
Refer to the INSTRUCTION MANUAL SERIAL COMMUNICATION that is a separate
volume.
Function for selecting operation mode is set to “LOCAL” at shipping.
4.6 Function for using CRC code
ON: CRC code is effective.
OFF: CRC code is not effective.
Details of CRC code are shown at INSTRUCTION MANUAL SERIAL COMMUNICATION.
Function for using CRC code set to “OFF” at shipping.
4.7 Function for error (alarm) log
You can look at error (alarm) log in the past 8 times. If you press F3 button, error (alarm) log will
be cleared to “No Failure”

OSAKA VACUUM, LTD.
Document No. : ND00338-R06
13
Stand by L
0rpm 12345h
F1 F3
a part of switches
F1 Next display
F2 Alteration
F3 Decision or clear
F2
Maintenance Mode
(F1:Next F2:In)
F1 F3F2
F1 F3F2
No Load Test
OFF
F1 F3F2F1 F3F2
Operation Time
Reset OFF
F1 F3F2F1 F3F2
Protection Signal
ON
F1 F3F2F1 F3F2
F1 F3F2
F1 F3F2 No Load Test
ON
F1 F3F2F1 F3F2
F1 F3F2F1 F3F2
Operation Time
Reset ON
Protection Signal
OFF
Fan Alarm
ON
F1 F3F2F1 F3F2
Operation Mode
Select LOCAL
F1 F3F2
F1 F3F2
F1 F3F2F1 F3F2
F1 F3F2F1 F3F2
Fan Alarm
OFF
Operation Mode
Select REMOTE
5 seconds
F1 F2F1 F2
F1 F3F2F1 F3F2
F1 F3F2F1 F3F2
F1 F3F2F1 F3F2
F1 F3F2
F1 F3F2
F1 F3F2F1 F3F2 Operation Mode
Select SERIAL
F1 F3F2
F1 F3F2
F1 F3F2F1 F3F2
F1 F3F2F1 F3F2
F3F3
(Water cooled type cannot be set to ON.)
CRC
OFF
F1 F3F2F1 F3F2
F1 F3F2
F1 F3F2
CRC
ON
F1 F3F2F1 F3F2
F1 F3F2
F1 F3F2
F2F2 F3F3 No Load Test
0rpm 12345h
Failure Log
F2:Disp F3:Clear
F1 F3F2F1 F3F2
F1 F3F2F1 F3F2
No Failure
(Last Time)
F1 F3F2F1 F3F2
F1 F3F2F1 F3F2
No Failure
(2times before)
F1 F3F2F1 F3F2
F1 F3F2F1 F3F2
No Failure
(8times before)
F1 F3F2F1 F3F2
F3 button can clear failure log.
"No Load Test" blinks
Figure 2. MAINTENANCE MODE

OSAKA VACUUM, LTD.
Document No. : ND00338-R06
14
5. Protective functions
5.1 Protective function
If following error happens on the power supply or the pump, protective function works, suitable
error message is indicated and blinking on the display. The pump stops with brake when errors
shown in Table.4-1 occur, and the pump stops without brake when errors shown in Table.4-2
occur. The power supply holds indication of the error message and the FAILURE signal from the
remote connector.
Table 4-1. Protective functions
Kind of errors
Display Contents Measures
Acceleration
overtime
When the power supply does not
become in NORMAL condition within
the specified time in spite of start button
was pressed.
* As for limit of acceleration time, see
table 5.
Check the following on the pump:
・Backing pressure of the pump is lower than
the specified value.
・The system does not have leakage.
・Gas flow to the pump does not exceed.
Re-acceleration
overtime
When the power supply does not
become in NORMAL condition within a
specified time after the indication of the
display changes from in NORMAL
condition to inACC condition. (When
the power supply becomes in ACC.
condition again, “acceleration”
indication blinks.)
* As for limit of re-acceleration time,
see table 5.
Check the following on the pump:
・Backing pressure of the pump is lower than
specified value.
・The system does not have leakage.
・Gas flow to the pump does not exceed.
Protection signal
open
When protection signal connected to the
pin No. 2 and 10 of the remote
connector on the rear panel of the power
supply is open.
Note.1
Check the interlock signal.
Fan disconnected
When the fan cable is not connected.
Note.1 Check the following:
・Air-cooled type pump
Check the connection of fan cable is correctly
connected.
・Water-cooled type pump
Set the function for detecting fan
disconnected to “OFF”. (Refer to section 4.4)
Acc. Overtime1
30:00 12345h
Acc. Overtime2
30:00 12345h
ProtectionSignal
30:00 12345h
Fan Disconnected
30:00 12345h

OSAKA VACUUM, LTD.
Document No. : ND00338-R06
15
When output current from the power
supply to the motor of the pump is
extremely low.
Check the following on the power supply.
・The lubricating oil is not less than the
specified amount. If the lubricating oil is less
than the specified amount, the output current
decreases and the power supply will stop with
this trouble.
! CAUTION
However, DO NOT check and control the
amount of the lubricating oil by the indication
of this trouble code.As for proper amount of
oil, refer to instruction manual of the pump.
Light load
When the motor cable is broken. (In this
case, the pump stops by free run.) Check the motor cable.
Note.1) The power supply does not indicate these error messages before pressing the START button.
“Stand-by” indication changes to one of these error messages indication only after the
START button was pressed.
Table 4-2. Protective functions
Kind of errors
Display Contents Measures
Output over-current Abnormal over-current flows because of
short in output or others. Operate the power supply in no-load
operation mode. Check whether the error is
caused by the power supply or by the pump
and contact OSAKA VACUUM, LTD.
* Procedure for no-load test of the power
supply is shown at section 8.3.
Pump overheat
When the thermal protector built in the
pump works.
Note.1
Check the following on the pump:
・The system does not have leakage.
・Gas flow to the pump does not exceed.
・Water-cooled type pump
Cooling water fall off.
P/S overheat
Power supply is overheated. Check the following on the power supply:
・The cooling fan inside of the power supply
works normally.
・The vent of the power supply do not be
blocked.
・The ambient temperature of the power
supply is under 40℃.
Input voltage
dropping When input voltage to the power supply
is lower than the specified value.
This trouble message is indicated in the
range of the input voltage in which the
control circuit of the power supply can
work. Also, if the input power is cut and
the power recovers within 1 second, the
power supply continues to operate. The
power supply stops if the input power
does not recover within 1 second, and
indicates this trouble message.
Note.1
Check the input voltage.
Over Current
12345h
Motor Overheat
12345h
No Load
30:00 12345h
P/S Overheat
12345h
InputVoltageLow
12345h

OSAKA VACUUM, LTD.
Document No. : ND00338-R06
16
Over frequency
When output frequency from the power
supply is higher than about 105% of
rated output frequency.
Contact OSAKAVACUUM,LTD
System error
When there is some trouble(s) inside
power supply. Contact OSAKAVACUUM,LTD
Note.1) The power supply does not indicate these error messages before pressing the START button.
“Stand-by” indication changes to one of these error messages indication only after the
START button was pressed.
Table.5 Detected time for Acceleration / re-acceleration overtime (unit: minute)
Power supply model TC163H TC203 TC223 TC353 TC523H TC1104
Acceleration overtime 10 10 2 4 20 25
Re-acceleration overtime 5 5 1 2 10 10
5.2 Warning function
Concerning the pump which uses grease type bearings (TG220F / TG350F / TG450F / TG800F /
TG1100F ), ”Change Bearing” is displayed at front panel when operation time reaches 90% of
bearing lifetime. “Operation status (Standby, Acceleration, Normal, Brake)” and “Change
Bearing” are displayed alternately in the 1.5 seconds cycle.
Bearing lifetime is shown at following table7.
Table 6. Warning function
Kind of alarm
Display Contents Measures
Change Bearing
↕1.5 seconds cycle
It is necessary to change bearings
built in the pump.
! WARNING
Contact OSAKA VACUUM,
LTD because of necessary to
change bearings.
After bearings exchange, please
press STOP/RESET button on
flont panel for 5 seconds. So The
alarm is reset.
Table 7. Durability of bearing
Power supply model TC223 TC353 TC1104
Durability of bearing 20000[h] 20000[h] 30000[h]
System Error
12345h
Over Speed
12345h
Normal L
20400rpm 12345h
Change Bearing
20400rpm 12345h

OSAKA VACUUM, LTD.
Document No. : ND00338-R06
17
6 Reset of error
6.1 Over frequency and System error
Turn off the power supply and then turn on again.
6.2 The others
Press the STOP button on the front panel of the power supply to reset the power supply, whether
it is in LOCAL operation or in REMOTE operation or in SERIAL COMMUNICATION
operation. Prior to reset, verify the error and resolve the error. The power supply does not restart
when the power supply was in REMOTE operation and the START signal was left in closed even
if the power supply was reset. Make the START signal opened and then make it closed to restart
the power supply.
7. Installation and electrical connections
7.1 Installation
The power supply is suitable for installation in rack or on floor.
Install the power supply in rack as follows.
1, Remove the rubber legs on the bottom of the power supply.
2, Support the power supply’s weight with either rails or plates.
3, Screw front panel of the power supply on rack at 2 points.

OSAKA VACUUM, LTD.
Document No. : ND00338-R06
18
! CAUTION
In case of installing the power supply in rack, be sure to support its weight with either rails
or plates. Weight of power supply is approximately 2.7kg.
Be sure to provide space more than 10mm around both sides of the power supply.
Be sure to provide 150mm space for cable connections around the rear of the power supply.
DO NOT install the power supply such locations as below:
1. Location in high temperature and/or high humidity
2. Location where the water drops wet the power supply
3. Location where is dusty
4. Location where is harmful gas such as acid, alkali, corrosive gas and so on.
5. Location where is flammable/explosive gas.
6. Location where is strong electric/magnetic field.
7. Location in big and/or much vibration
7.2 Electrical connections
②
①
Rear of the
Power Supply
Molecular pump
motor
connector
④
③
⑤
FAN
connector
Connection of the cables between the power supply and the pump is shown at Figure 3.
Figure 3. Electrical connections
Refer to table.8 as for ①to ⑤in Figure 3.

OSAKA VACUUM, LTD.
Document No. : ND00338-R06
19
Table 8. Attached connectors and cables
Name Connector type of the
power supply side Connector type of the
pump side Remarks
① Input power
connector YAMATE
AP-400(B) accessory
② Remote
connector DDK
17JE-23150-02(D8A) accessory
TC223,TC353
Bendix
PT06A-12-8S(SR) Exclusive cable
TC163H,TC203
TC523H,TC1104
③ Motor cable Nanaboshi
NJC-207-PM-UL
Nanaboshi
NJC-308-P
Exclusive cable
④ Serial
communication
connector
DDK
17JE-13090-02(D8A) accessory
⑤
Fan cable
(only to air
cooled type
pump)
Nanaboshi
NJC-165-PM-UL AMP
172159-1 Exclusive cable
7.2.1 Connection of input power cable
Connect the cable to the input power connector (attachment).
1) Select the cable size from the Table9.
2) Strip sheath of the cable about 30mm from the end.
3) Set the solderless terminals at the end of the cable, and cover the part of the cable which is
clamped by a protective tube.
4) After thread the cable through a bushing, screw the solderless terminals at each pin of the
connector, and screw the cable clamp.
N (R)
AC90~121V or AC180~253V, 50/60Hz, single phase Note.1
L (S)
E Earth Note.2
! Warning
Note.1) Prior to supplying power, be sure to set the input power select switch according
to the input voltage.
Note.2) Be sure to earth the power supply.
5) Connect the input power cable to the “INPUT” connector on the rear panel of the power
supply.