
Assembly, Operation, Maintenance of
MAGNUM
®
Membrane Tube Diffusers
and Membranes
7-2006
OTT GmbH
Tel: +49 (0)511 / 78631 0
Frankenring 21 Fax: +49 (0)511 / 78631 40
D 30855 Langenhagen Email: info@ott-system.com
Germany www.ott-system.com
page 6 of 6
5. Maintenance and cleaning of
membrane tube diffusers
OTT membrane tube diffusers operate reliable and
do not need intensive maintenance. Still regular
control and usual maintenance is strongly
recommended.
Wastewater contains various components which to
a certain extend do settle on the membrane. Some
are mentioned below:
carbonate (water hardness),
precipitants
polymeres (sludge conditioning)
Deposits on the diffuser are usually not harmful to
the membrane but might affect the pressure loss
and bubble pattern.
Certain solvents and oils can have a negative
impact on the membrane. In case of questions
address to you OTT representative.
Biological films are known to reduce the life time
on EPDM rubber membranes. If problems related
to biological films must be expected OTT
recommends using FLEXSIL membranes.
5.1 Lifetime
The lifetime of a diffuser is mainly depending on
the actual process and waste water conditions.
Therefore the suitability and quality of the chosen
membrane material is important. Please discuss
the given situation at the WWTP with the supplier
or your OTT representative to make a sustainable
choice.
OTT membranes are suitable for a waste water
quality as defined in the German „ATV
Arbeitsblattes A 115“ latest edition. This means the
amount of industrial wastewater may not exceed
20 % of the total wastewater flow. In case of
questions please contact your OTT representative.
In order to maintain an economically operation and
a long life cycle follow the assembly, operation and
maintenance instructions for OTT diffusers.
5.2 Performance control
The best indication for layers and crusts on the
membrane is the pressure loss. Monitor the
development of your blower pressure over time.
Define a critical pressure where measures should
be taken.
5.2.1 Flushing with air
First measure should be the flushing with air. This
should be done if possible preemptively on a daily
basis as described in chapter 3.
5.2.2 Cleaning
The diffusers membrane can be cleaned with
water and a high pressure cleaner. The distance
between the membrane and nozzle of the high
pressure cleaner should be at least 0.5 meters.
Alternatively a brush can be used to clean smaller
numbers of membranes. If possible put air through
the diffusers during the cleaning process.
Important: Mechanical or chemical impact can
cause serious damage of the membranes, handle
the diffusers and the cleaning equipment with care!
The membrane may not be damaged, folded or
wrenched while cleaning the diffuser.
Do not use any detergents, solvents, alcohol or
acids for cleaning without having discussed the
case with your OTT representative.
5.2.3 Flushing formic acid
From a certain point on flushing with air may not
be sufficient. The reason can be calcium in the
perforation holes of the diffuser. In this case OTT
recommends flushing with formic acid. Do not use
other acids. The formic acid should be injected into
the air supply or drop pipe. Use an acid
concentration of 80% and 250 ml/hr per meter
installed diffuser length.
Handling acids is dangerous. Wear appropriate
protective clothing. Choose a dosing mechanism
that does not imperil your health. Contact the acid
supplier and follow the prescribed directions for
handling acids.
6. Service
It is possible to send a sample diffuser / membrane
to OTT GmbH. Deposits and the quality of the
membrane will be analysed and an appropriate
advise for cleaning and further measures will be
worked out. Please clean the diffuser before
sending it. Use water (especially clean the flooded
pipe) but do NOT brush or clean away the deposits
on the membrane. Please take care that the
membrane does not dry during transportation - that
would affect the test.