PCB Piezotronics ICP 353B77 User manual

Model 353B77
ICP® Accelerometer
Installation and Operating Manual
For assistance with the operation of this product,
contact PCB Piezotronics, Inc.
Toll-free: 800-828-8840
24-hour SensorLine: 716-684-0001
Fax: 716-684-0987
E-mail: [email protected]
Web: www.pcb.com
The information contained in this document supersedes all similar information that
may be found elsewhere in this manual.
Total Customer Satisfaction – PCB
Piezotronics guarantees Total Customer
Satisfaction. If, at any time, for any
reason, you are not completely satisfied
with any PCB product, PCB will repair,
replace, or exchange it at no charge. You
may also choose to have your purchase
price refunded in lieu of the repair,
replacement, or exchange of the product.
Service – Due to the sophisticated nature
of the sensors and associated
instrumentation provided by PCB
Piezotronics, user servicing or repair is
not recommended and, if attempted, may
void the factory warranty. Routine
maintenance, such as the cleaning of
electrical connectors, housings, and
mounting surfaces with solutions and
techniques that will not harm the
physical material of construction, is
acceptable. Caution should be observed
to insure that liquids are not permitted to
migrate into devices that are not
hermetically sealed. Such devices should
only be wiped with a dampened cloth
and never submerged or have liquids
poured upon them.
Repair – In the event that equipment
becomes damaged or ceases to operate,
arrangements should be made to return
the equipment to PCB Piezotronics for
repair. User servicing or repair is not
recommended and, if attempted, may
void the factory warranty.
Calibration – Routine calibration of
sensors and associated instrumentation is
recommended as this helps build
confidence in measurement accuracy and
acquired data. Equipment calibration
cycles are typically established by the
users own quality regimen. When in
doubt about a calibration cycle, a good
“rule of thumb” is to recalibrate on an
annual basis. It is also good practice to
recalibrate after exposure to any severe
temperature extreme, shock, load, or
other environmental influence, or prior
to any critical test.
PCB Piezotronics maintains an ISO-
9001 certified metrology laboratory and
offers calibration services, which are
accredited by A2LA to ISO/IEC 17025,
with full traceablility to N.I.S.T. In
addition to the normally supplied
calibration, special testing is also
available, such as: sensitivity at elevated
or cryogenic temperatures, phase
response, extended high or low
frequency response, extended range, leak
testing, hydrostatic pressure testing, and
others. For information on standard
recalibration services or special testing,
contact your local PCB Piezotronics
distributor, sales representative, or
factory customer service representative.
Returning Equipment – Following
these procedures will insure that your
returned materials are handled in the
most expedient manner. Before returning
any equipment to PCB Piezotronics,
contact your local distributor, sales
representative, or factory customer
service representative to obtain a Return
Warranty, Service, Repair, and
Return Policies and Instructions
Materials Authorization (RMA)
Number. This RMA number should be
clearly marked on the outside of all
package(s) and on the packing list(s)
accompanying the shipment. A detailed
account of the nature of the problem(s)
being experienced with the equipment
should also be included inside the
package(s) containing any returned
materials.
A Purchase Order, included with the
returned materials, will expedite the
turn-around of serviced equipment. It is
recommended to include authorization
on the Purchase Order for PCB to
proceed with any repairs, as long as they
do not exceed 50% of the replacement
cost of the returned item(s). PCB will
provide a price quotation or replacement
recommendation for any item whose
repair costs would exceed 50% of
replacement cost, or any item that is not
economically feasible to repair. For
routine calibration services, the Purchase
Order should include authorization to
proceed and return at current pricing,
which can be obtained from a factory
customer service representative.
Warranty – All equipment and repair
services provided by PCB Piezotronics,
Inc. are covered by a limited warranty
against defective material and
workmanship for a period of one year
from date of original purchase. Contact
PCB for a complete statement of our
warranty. Expendable items, such as
batteries and mounting hardware, are not
covered by warranty. Mechanical
damage to equipment due to improper
use is not covered by warranty.
Electronic circuitry failure caused by the
introduction of unregulated or improper
excitation power or electrostatic
discharge is not covered by warranty.
Contact Information – International
customers should direct all inquiries to
their local distributor or sales office. A
complete list of distributors and offices
can be found at www.pcb.com.
Customers within the United States may
contact their local sales representative or
a factory customer service
representative. A complete list of sales
representatives can be found at
www.pcb.com. Toll-free telephone
numbers for a factory customer service
representative, in the division
responsible for this product, can be
found on the title page at the front of this
manual. Our ship to address and general
contact numbers are:
PCB Piezotronics, Inc.
3425 Walden Ave.
Depew, NY 14043 USA
Toll-free: (800) 828-8840
24-hour SensorLineSM: (716) 684-0001
Website: www.pcb.com
E-mail: [email protected]
DOCUMENT NUMBER: 21354
DOCUMENT REVISION: B
ECN: 17900
MOUNTNG THREAD SEE DRAWING
5-40 A
M3 X 0.50
B
10-32 C
M5 X 0.80 D
1/4-28
E
M6 X 1.00
F
BOLT THREAD SEE DRAWING
10-32 C
M5 X 0.80 D
1/4-28
E
M6 X 1.00
F
M8 X 1.25
F
1
1
2
2
3
3
4
4
A A
B B
CODE
IDENT. NO.
52681
DWG. NO.
SCALE: SHEET
DRAWN CHECKED ENGINEER
TITLE
UNLESS OTHERWISE SPECIFIED TOLERANCES ARE:
DIMENSIONS IN MILLIMETERS
[ IN BRACKETS ]
XX ±0.13
ANGLES 2 DEGREES
3425 WALDEN AVE. DEPEW, NY 14043
(716) 684-0001 E-MAIL: [email protected]
DIMENSIONS IN INCHES
DECIMALS XX ±.01
ANGLES 2 DEGREES
FILLETS AND RADII
.003 - .005
DECIMALS X ± 0.3
FILLETS AND RADII
0.07 - 0.13
INSTALLATION DRAWING 081-XXXX-90
1 OF 1N.T.S.
FOR STANDARD
081 SERIES MOUNTING
XXX ±.005
JDM 3/9/07 ECB 3/9/07 JJD 3/9/07
"A"
5-40
MOUNTING INSTRUCTIONS
(METRIC DIMENSIONS IN BRACKETS)
MOUNTING HOLE PREPARATION:
.101[2.57]
X .20[5.1]MIN.
5-40 UNC-2B
X .15[3.8]MIN.
4.) RECOMMENDED MOUNTING TORQUE,
4-5 INCH POUNDS
[45-55 NEWTON CENTIMETERS].
"B"
M3 X 0.50
MOUNTING INSTRUCTIONS
(ENGLISH DIMENSIONS IN BRACKETS)
MOUNTING HOLE PREPARATION:
2.5[.099]
X 4.6 [1.8]MIN.
M3 X 0.50-6H
X 3.3[.13]MIN.
4.) RECOMMENDED MOUNTING TORQUE,
45-55 NEWTON CENTIMETERS
[4-5 INCH POUNDS].
"C"
10-32
MOUNTING INSTRUCTIONS
(METRIC DIMENSIONS IN BRACKETS)
4.) RECOMMENDED MOUNTING TORQUE,
10-20 INCH POUNDS
[113-225 NEWTON CENTIMETERS].
MOUNTING HOLE PREPARATION:
.159[4.04]
X .23[5.8]MIN.
10-32 UNF-2B
X .15[3.8]MIN.
"D"
M5 X 0.80
MOUNTING INSTRUCTIONS
(ENGLISH DIMENSIONS IN BRACKETS)
4.) RECOMMENDED MOUNTING TORQUE,
113-225 NEWTON CENTIMETERS
[10-20 INCH POUNDS].
MOUNTING HOLE PREPARATION:
4.22[.166]
X 7.62 [.300]MIN.
M5 X 0.8-6H
X 5.08[.200]MIN.
"E"
1/4-28
MOUNTING INSTRUCTIONS
(METRIC DIMENSIONS IN BRACKETS)
4.) RECOMMENDED MOUNTING TORQUE,
2-5 FOOT POUNDS
[3-7 NEWTON METERS].
MOUNTING HOLE PREPARATION:
.218[5.54]
X .300[7.62]MIN.
1/4-28 UNF-2B
X .200[5.08]MIN.
"F"
M6 X 0.75, M6 X 1.00, M8 X 1.25
MOUNTING INSTRUCTIONS
(ENGLISH DIMENSIONS IN BRACKETS)
4.) RECOMMENDED MOUNTING TORQUE,
3-7 NEWTON METERS [2-5 FT POUNDS].
M8 X 1.25
MOUNTING HOLE PREPARATION:
6.75[.266]
X 8.64 [.340]MIN.
M8 X 1.25-6H
X 5.00[.197]MIN.
M6 X 0.75
MOUNTING HOLE PREPARATION:
5.31[.209]
X 7.62 [.300]MIN.
M6 X 0.75-6H
X 5.08[.200]MIN.
M6 X 1.0
MOUNTING HOLE PREPARATION:
5.05[.199]
X 8.10 [.320]MIN.
M6X 1.0-6H
X 6.35[.250]MIN.
"B"
"C"
"G"
MOUNTING INSTRUCTIONS
FOR SPECIAL THREAD LENGTHS
(METRIC DIMENSIONS IN BRACKETS)
THREAD DEPTH : B= X + 1 THREAD PITCH
DRILL DEPTH : C= B + 3 THREAD PITCH
SEE A-F FOR APPROPRIATE DRILL AND TAP
THREAD PITCH= 1/TPI [P]
3.) FOR BEST RESULTS, PLACE A THIN LAYER OF SILICONE GREASE (OR EQUIVALENT) ON INTERFACE PRIOR
TO MOUNTING.
MOUNTING SURFACE SHOULD BE FLAT TO WITHIN .001(0.03) TIR OVER DIM 'A' WITH A OR BETTER FINISH FOR BEST RESULTS.
DRILL PERPENDICULAR TO MOUNTING SURFACE TO WITHIN 1.
SENSOR
SENSOR
"THRU-BOLT"
"THRU-BOLT" STUD MOUNT
INTEGRAL STUD MOUNT
SENSOR
THREAD
MOUNTING
THREAD
SENSOR
THREAD
MOUNTING
THREAD
STANDARD STUD MOUNT
"X"
"X"
SENSOR THREAD
"X"
1111
1
1
1
1
2222
2
2
2
MOUNTING HOLE PREPARATION:
DRILL DIA.
X "C"MIN.
TAP
X "B"MIN.
1
"A"
"A"
"A"
"P"
1 THREAD PITCH
SHOWN
"X"
SENSOR
INTEGRAL
MOUNTING
STUD
63[1.61]
REVISIONS
REV DESCRIPTION DIN
P UPDATE DRAWING 25686
2
1
MOUNTING THREAD SEE DRAWING
5-40 A
M3 X 0.50 B
10-32 C
M5 X 0.80 D
1/4-28 E
M6 X 1.00 F
081-XXXX-90
PCB Piezotronics Inc. claims proprietary rights in
the information disclosed hereon. Neither it nor any
reproduction thereof will be disclosed to others
without the written consent of PCB Piezotronics Inc.

This manual suits for next models

1

Other PCB Piezotronics Measuring Instrument manuals

PCB Piezotronics IMI Sensors 608A11 User manual

PCB Piezotronics

PCB Piezotronics IMI Sensors 608A11 User manual

PCB Piezotronics IMI SENSORS 646B11 User manual

PCB Piezotronics

PCB Piezotronics IMI SENSORS 646B11 User manual

PCB Piezotronics IMI SENSORS TOM607A11 User manual

PCB Piezotronics

PCB Piezotronics IMI SENSORS TOM607A11 User manual

PCB Piezotronics 353B04 User manual

PCB Piezotronics

PCB Piezotronics 353B04 User manual

PCB Piezotronics 333B55 User manual

PCB Piezotronics

PCB Piezotronics 333B55 User manual

PCB Piezotronics M639A91 User manual

PCB Piezotronics

PCB Piezotronics M639A91 User manual

PCB Piezotronics ICP 350C23 User manual

PCB Piezotronics

PCB Piezotronics ICP 350C23 User manual

PCB Piezotronics P357A09/NC User manual

PCB Piezotronics

PCB Piezotronics P357A09/NC User manual

PCB Piezotronics 3503A1160KG/-0002 User manual

PCB Piezotronics

PCB Piezotronics 3503A1160KG/-0002 User manual

PCB Piezotronics 052M94A User manual

PCB Piezotronics

PCB Piezotronics 052M94A User manual

PCB Piezotronics IMI Sensors M623C00 User manual

PCB Piezotronics

PCB Piezotronics IMI Sensors M623C00 User manual

PCB Piezotronics IMI SENSORS 604B11/050BZ User manual

PCB Piezotronics

PCB Piezotronics IMI SENSORS 604B11/050BZ User manual

PCB Piezotronics 356M57 User manual

PCB Piezotronics

PCB Piezotronics 356M57 User manual

PCB Piezotronics ICP IMI SENSORS MXM622A11 User manual

PCB Piezotronics

PCB Piezotronics ICP IMI SENSORS MXM622A11 User manual

PCB Piezotronics J357B11 User manual

PCB Piezotronics

PCB Piezotronics J357B11 User manual

PCB Piezotronics J353B11 User manual

PCB Piezotronics

PCB Piezotronics J353B11 User manual

PCB Piezotronics W138A10/M038CY010EB User manual

PCB Piezotronics

PCB Piezotronics W138A10/M038CY010EB User manual

PCB Piezotronics Tourmaline ICP 134A24 User manual

PCB Piezotronics

PCB Piezotronics Tourmaline ICP 134A24 User manual

PCB Piezotronics IMI SENSORS M623C10 User manual

PCB Piezotronics

PCB Piezotronics IMI SENSORS M623C10 User manual

PCB Piezotronics Triaxial ICP PCB-339A31/NC User manual

PCB Piezotronics

PCB Piezotronics Triaxial ICP PCB-339A31/NC User manual

PCB Piezotronics IMI SENSORS 66102APZ1 User manual

PCB Piezotronics

PCB Piezotronics IMI SENSORS 66102APZ1 User manual

PCB Piezotronics EX357A94 User manual

PCB Piezotronics

PCB Piezotronics EX357A94 User manual

PCB Piezotronics IMI SENSORS 646B01 User manual

PCB Piezotronics

PCB Piezotronics IMI SENSORS 646B01 User manual

PCB Piezotronics 356A61 User manual

PCB Piezotronics

PCB Piezotronics 356A61 User manual

Popular Measuring Instrument manuals by other brands

Pro-Tecs PNA 4500 quick guide

Pro-Tecs

Pro-Tecs PNA 4500 quick guide

Acrel ACR10R Series Installation and operation instruction

Acrel

Acrel ACR10R Series Installation and operation instruction

Idexx VetTest Operator's manual

Idexx

Idexx VetTest Operator's manual

GE Sievers 500 RL Operation and maintenance manual

GE

GE Sievers 500 RL Operation and maintenance manual

Hach FH950 user manual

Hach

Hach FH950 user manual

Hioki PW3390 Measurement guide

Hioki

Hioki PW3390 Measurement guide

Sper scientific 840010 quick start guide

Sper scientific

Sper scientific 840010 quick start guide

Trotec BI15 operating manual

Trotec

Trotec BI15 operating manual

ADInstruments DTVLINK-5 user manual

ADInstruments

ADInstruments DTVLINK-5 user manual

DNT Grabstar Audio user manual

DNT

DNT Grabstar Audio user manual

GESTRA VK 14 Original Installation Instructions

GESTRA

GESTRA VK 14 Original Installation Instructions

C-Tech 16-T0012/P product manual

C-Tech

C-Tech 16-T0012/P product manual

Agilent Technologies 3588A datasheet

Agilent Technologies

Agilent Technologies 3588A datasheet

Time Electronics 7006 user manual

Time Electronics

Time Electronics 7006 user manual

AccuEnergy AcuRev 1310 Series user manual

AccuEnergy

AccuEnergy AcuRev 1310 Series user manual

MONARCH INSTRUMENT Pocket Tach 99 instruction manual

MONARCH INSTRUMENT

MONARCH INSTRUMENT Pocket Tach 99 instruction manual

Omron K3NH Operation manual

Omron

Omron K3NH Operation manual

Worldcast Systems ecreso NEXT FM user manual

Worldcast Systems

Worldcast Systems ecreso NEXT FM user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.