Pharma Test PTB 111E User manual

Manual Tablet Hardness Testing Instrument - Type PTB 111E
29-02840_01-2011E Page 1 of 4
PHARMA TEST AG +49 6182 9532-600
Siemensstrasse 5 +49 6182 9532-650
www.pharma-test.com
The Manual Tablet Testing Instrument PTB 111E is a dual force mode hardness test
apparatus as it can be used for either linear force or linear speed increase while Tablet
Hardness is tested. It offers a multiple point validation procedure for the built-in digital load
cell.
The instrument is made in strict compliance with the EP <2.9.8> and USP <1217>
Pharmacopoeia.
Enter the nominal test information for Hardness via the PTB 111E keyboard. Select the unit
to measure, Kp, N or Sc, now place the sample onto the Sample Dish and start the test. The
driven jaw will now run forward and touch the tablet to measure the hardness (tablet breaking
force).
The result is immediately displayed and printed to a connected Dot-Matrix or suitable PCL
Type Laser- or DeskJet printer. Repeat this until your series has been tested, get a full print
including each individual result, mean value and deviations.
The flexibility and the reproducibility of the results have made this and other models, like the
PTB311E series to become one of the most sold hardness testers.

Manual Tablet Hardness Testing Instrument - Type PTB 111E
29-02840_01-2011E Page 2 of 4
PHARMA TEST AG +49 6182 9532-600
Siemensstrasse 5 +49 6182 9532-650
www.pharma-test.com
Operating Principle…
Even in the existing monographs from USP and EP is no standard force setting or force
increase mode established. This usually is causing problems comparing results received of
different suppliers instruments of the same tablet. The hardness result is directly influenced
by the contact speed and force increase rate. Faster operated test jaw means lower
reproducibility and often higher results. In order to offer the possibility to select an operating
mode which will offer you similar results as the instruments you may already use, we allow to
select the force mode, linear force increase or linear speed increase as we do for the rate.
The only parameter we do not alter is the force touching the sample which is for all Pharma
Test Hardness Test Instruments very low as if you would try to break by hand. When touched
we start to increase.
Which force mode to select ?
Since more than 15 years all PHARMA TEST Hardness Test instruments offer the possibility
to select either linear Force or linear Speed increase.
Linear force increase certainly offers the most accurate control, as the rate of increase is
directly controlled by the electronically load cell used to read the force. Also it is quite simple
to validate the correct and linear operation as a Tablet of 100 Newton hardness will be
broken within 5 seconds if 20N/s had been set.
Linear speed increase can also be used. Here the driving speed of the motor is kept linear.
Actually if the touching force is kept low there is not too much difference in results between
the 2 systems.
Calibration and Validation
Built-in calibration and validation program for the
hardness station. To validate the hardness test station the
PT-MT magnetic tablet or different certified weights are
used. Use the PT-MT to qualify the correct breakpoint
detection, select a force, for example 50, 85 or 130 N
and run a test series, the resolution of the results should
be within 1.0N. The PT-MT instrument works like a tablet,
it withstands force and than “breaks”. For the 2 point
calibration of the hardness station a certified reference
weight of 10 kg is used for validation use 5, 10 and 15kg
weights. All calibration and validation results can be printed and countersigned. The PT-MT
will be used to qualify both, breakpoint detection as well as correctness of the set force
increase rate.
To prove the linearity of the instrument, the operator can program a print-out of the force
curve recorded during a test. This will show the linear increase of the adjusted force mode.
Also different weights, like the PTB-CAL15 which includes 5, 10, and 15kg, may be placed
onto the load cell or the PT-MT shall be used to validate the linearity. Using the parallel Port
a Matrix or PCL5 printer can be added and using the RS-232 COM port, all results can be
transmitted to a software program running on a computer system.

Manual Tablet Hardness Testing Instrument - Type PTB 111E
29-02840_01-2011E Page 3 of 4
PHARMA TEST AG +49 6182 9532-600
Siemensstrasse 5 +49 6182 9532-650
www.pharma-test.com
The PTB 111E Tablet Test Instrument offers:
¾Use automatic re-start facility to speed up the testing sequence
¾Documentation of all results using a separate Matrix or PCL Printer
¾OQ and Calibration program for the measurement station
¾Dual Point Calibration for the hardness test stations
¾Multiple Point Validation for the hardness test stations
¾Programmable print-out of force increase curve
¾Data transfer via RS-232 interface
¾Adjustable force control for hardness testing, select either linear force or linear speed
increase
¾Hardness testing in compliance with the EP <2.9.8> and USP <1217>
Pharmacopoeia
¾Test program for soft gelatine capsule testing
¾Only hardness testing instrument in the market which offers unique adjustment
facilities of force increase and breaking detection force rates to meet any upcoming
monograph.
¾Use calibration and validation program and report results.
Testing Tension Strength of Oblong and Caplets
The specially designed sample holder and force jaw is
suitable to be used in the PTB 111E Testing Instrument. It
is designed to test the tension strength at the break line of
shaped and also round tablets
Technical Data
Display: LED Display for No. of samples and hardness results
Keyboard: Numerical and function keys
Hardness: 5.0 - approx. 300 N (Newton), also 500N option
available
Accuracy: better 1N
Resolution: 300N = 0.0741N - 500N = 0.1482N
Measuring units: Thickness and diameter selectable in either mm
(Millimetre)or IN (Inches)
Hardness Result: Print-out and display selectable in either Newton (N),
kilopond (kp) or Strong Cobb (Sc)
Force rate: Adjustable for linear force increase or linear speed
increase
Range Linear Force Increase: 5.00 - 99.99 N/sec.
Accuracy: < 1%
Range Linear Speed Increase: 5 - 200.0 mm/minute

Manual Tablet Hardness Testing Instrument - Type PTB 111E
29-02840_01-2011E Page 4 of 4
PHARMA TEST AG +49 6182 9532-600
Siemensstrasse 5 +49 6182 9532-650
www.pharma-test.com
Accuracy: < 0.1%
Number of tests for statistics: Up to 250
Calibration Procedure: 10 kg reference weight (certified)
Validation breaking detection: PT-MT magnetic tablet
Validation force setting: PT-MT3 magnetic tablet
Validation linearity of the load cell: PTB-CAL15 certified weight 5, 10, 15 kg
Interface: RS-232 serial port
Parallel printer port for DOT Matrix or PCL Printer
Connection
Instrument Housing : Stainless steel to meet GLP requirements
Weights and Dimensions
Net weight: 10 kg
Gross weight: 15 kg
Packaging: 450 mm x 450 mm x 640 mm
Options
¾Extended hardness range up to approx. 550N
¾Oblong tension strength test jaws for different shapes
¾PTB32 software to enter manually individual weight of sample previous to hardness
test and get calculated batch statistics.
We reserve the right to make technical changes without any prior notice
Table of contents
Other Pharma Test Test Equipment manuals
Popular Test Equipment manuals by other brands

Time Electronics
Time Electronics 5051 Plus user manual

Elcometer
Elcometer 456 operating instructions

PCB Piezotronics
PCB Piezotronics ICP 481A Series Installation and operating manual

Hanna Instruments
Hanna Instruments Pool Line HI9310014 instruction manual

Vermason
Vermason 223002 Installation, operation and maintenance

Maschinenfabrik Reinhausen
Maschinenfabrik Reinhausen TAPCON 260 operating instructions