2ERROR CODE WARNING
ERROR ‘HLb’: when the battery voltage is in warning voltage (3.1 V)and screen displays ‘HLb’
words , which means current electricity is about to ran out and can last a very short time .
ERROR ‘LLb’:when the battery voltage is in power off voltage (2.9V, scale will power off
itself automatically when screen displays ‘LLb’words.
When the scale is in power on status and screen displays ‘LLb’words accompanied by tick
sound and then power off automatically. which means scale is out of work already and is needed to
change battery immediately.
3SEALING OPERATION METHOD and ACCELERATION of
GRAVITYADJUSTMENT
Gravity acceleration compensation value can’t be adjusted directly after the PS1X price
computing scale adopts sealing protection. If needs to adjust the gravity acceleration, you must
destroy the sealing.
Our company adopts environmental label sealing, only needs to destroy the fragile label
sealing to startup the amending mode, do not need to disassemble the cover.
The operating steps of adjusting the acceleration of gravity A/D value and Zero full weight
learning are as follows:
3.1 Destroy the sealing
1) Destroy the fragile label sealing
2) Open the plastic sealing cover.
3) Only use the sharp-pointed instrument such as pen point to touch the adjustable switch lightly
can adjust the acceleration of gravityA/D value and zero full weight learning.
4) What you need to do is breaking the label based on the adjustment and do not need to open case.
The label sealing must be broke if you want to open the case. Please take the below pictures as
you references.