Powerex PX-SLP-1 User manual

Sleep Lab Panel
Please read and save these instrucons. Read carefully before aempng to assemble, install, operate or maintain the product described.
Protect yourself and others by observing all safety informaon. Failure to comply with instrucons could result in personal injury and/or
property damage! Retain instrucons for future reference.
Descripon
Powerex • 150 Producon Drive • Harrison, OH 45030 •USA
P 1.888.769.7979 • F 513.367.3125 • www.powerexinc.com
PX-99-0534 •03/2017
Page 1 of 8
Powerex Sleep Lab Panel components are cleaned for use with
oxygen. Each inlet and outlet DISS assembly is tested for leakage.
Each unit is designed and built in accordance with the Naonal Fire
Protecon Associaon and Compressed Gas Associaon guidelines.
The installaon and maintenance should be conducted in
accordance with NFPA 99 or CSA Z7396.1.
Features & Benets
•Five year parts and one year labor warranty
• Provides 50 psi (3.4 bar) – Oxygen Source
• Check Valve Risers – allow easy service
• Sleep Lab Panels meet or comply with NFPA 99 Standards –
simplifying connecons and reducing cost
• Easy to service layout/design
• Flowmeter control panel incorporates dial adjustment
• Ease of maintenance
• DISS gas specic connecons
• Single owmeter panel
• Dual owmeter panels
PX-SLP-1 Sleep Lab Panel PX-SLP-2 Sleep Lab Panel
Contents
Features & Benets 1
Introducon 2
Major components 3
Installaon 5
Installaon of Inlet & Outlet DISS 5
Sleep Lab Panel Installaon 5
Maintenance 6
Flowmeters 7
Operang Instrucon 7
Cleaning/Maintenance 7
Troubleshoong Guide 8
Appendix A Technical Specicaons 10
Terms & Warranty 11
Technical Assistance 1-888-769-7979
Safety Guidelines
This manual contains informaon that is very important to know
and understand. This informaon is provided for SAFETY and to
PREVENT EQUIPMENT PROBLEMS. To help recognize this informaon,
observe the following symbols.
Danger indicates an imminently hazardous
situaon which, if not avoided, WILL result in death
or serious injury.
Warning indicates a potenally hazardous
situaon which, if not avoided, COULD result in
death or serious injury.
Cauon indicates a potenally hazardous situaon
which, if not avoided, MAY result in minor or
moderate injury.
Noce indicates important informaon, that if not
followed, may cause damage to equipment.
NOTE: Note indicates informaon that requires special aenon.

Sleep Lab Panel
Powerex • 150 Producon Drive • Harrison, OH 45030 •USA
P 1.888.769.7979 • F 513.367.3125 • www.powerexinc.com
PX-99-0534 •03/2017
Page 2 of 8
Installaon
Installaon of the Inlet & Outlet DISS Assemblies
Inlet and outlet DISS assemblies must be installed in accordance
with NFPA 99 or CSA Z7396.1. An internal nitrogen purge must be
used during the brazing operaon. The purge gas should ow away
from the DISS body. Brazing alloys per appropriate standards must
be used. Before brazing, wet rags must be wrapped around the tube
extensions next to the DISS body to prevent overheang and possible
damage to the DISS O-ring seals. Direct the ame away from the DISS
body. The DISS body temperature must not exceed 300°F to prevent
damaging the O-ring seals.
The system must be tested (per appropriate standards) to ensure
that no cross-connecons have been made.
The system must be tested (per appropriate standards) for leaks.
Note: Pressure in the system will increase or decrease with
temperature rise or fall.
Miss-connecon of the gases could lead to serious
or fatal injury to paents. Following installaon,
DISS connecons must be tested for cross-connecon (per appropriate
standards) to ensure that the intended services are correctly connected to
the appropriate service lines.
5. The center of the rough-in box should be 60” in height above
the nished oor. Fasten the back box to horizontal braces
installed between the studs so that the front edge of the rough
in box will be ush or slightly recessed with the nished wall
covering.
6. Before brazing, remove the plasc tube caps from the DISS
assembly (risers) copper extensions. DISS assemblies must be
installed in accordance with “Installaon of Inlet and Outlet
DISS Assemblies” above.
Sleep Lab Panel Installaon
1. Powerex Sleep Lab Panels may be plumbed in either direcon
– with the paent (use) side on the le or the right. This saves
the plumber me and money.
2. The rough-in (back box) is shipped with a cardboard dust cover
installed. You will need to remove the dust cover to install
the rough-in and perform the pressure test. The dust cover
should be re-installed aer the pressure test to protect the
DISS connecons unl the wall covering (drywall, paint etc.)
is complete.
3. The DISS risers are shipped loose and must be installed in the
back box.
4. Adjustable depth mounng anges (top & boom) should be
ulized to align front edge of box to be ush with the drywall
surface.
Plasc Caps
Adjustable
Flanges
X
Dust Cover
7. The system must be tested (per appropriate standards) to
ensure that no cross-connecons have been made. The
system must be tested (per appropriate standards) for leaks.
Note: Pressure in the system will increase or decrease with
temperature rise or fall.
8. Aer the system passes the leak test, install the labels provided
in the back box paying careful aenon to the direcon of the
gas ow (le to right, or right to le). The direcon of ow
aects where you place Inlet and Outlet Labels in the back box.
Inlet &
Outlet Labels
9. Aer the wall covering is complete, the dust cover may be
removed from the rough-in box and the hose assemblies and
front panel may be installed.
10. Make sure the hose assemblies are aached to the
appropriate inlet/ow meter connecon point, and
outlet/ow meter connecon point. Failing to do so
may aect the paent and the sleep lab test results.
Note: Sleep Lab Panels may be plumbed in either direcon –
with the paent (use) side on le or the right.

Sleep Lab Panel
Powerex • 150 Producon Drive • Harrison, OH 45030 •USA
P 1.888.769.7979 • F 513.367.3125 • www.powerexinc.com
PX-99-0534 •03/2017
Page 3 of 8
11. When all connecons are complete and leak tested with an
oxygen compable leak test soluon, fasten the sleep lab front
panel to the back box with the four #8-32 x ¾” screws provided.
12. On the front panel, label each owmeter in the space provided
with the sleep room it controls (label not provided). See
Appendix A.
Mounng Screws
Miss-connecon of the gases could lead to serious
or fatal injury to paents. Following installaon,
DISS connecons must be tested for cross-connecon (per appropriate
standards) to ensure that the intended services are correctly connected to
the appropriate service lines.
Make certain the labeling coincides with the gas
service, and areas controlled by the Sleep Lab
Panel and that it is easily read.
Maintenance
1. Clean the exterior of the stainless steel Sleep Lab Panel with a
glass cleaner. Strong solvents will damage the etched screened
prinng on the sleep lab panel.
2. Never use glass cleaner on the owmeters.
3. If it is necessary to clean the owmeters wipe them clean with
a cloth dampened with water
4. Wipe dry with a clean dry cloth.
Flowmeters
Accuracy: ± 10% of the indicated ow.
Operang Temperature: 69°F (20.6°C) to 71F° (21.6° C)
Storage Temperature: -40°F (40°C) to 140°F (60° C)
Max Humidity: 95 %
Note: Storage / Transport outside the specied range may cause
damage to the ow meter.
Specicaons are subject to change without noce.
Operang Instrucons
Note: Tri-Tech Medical, Inc. recommends the use of kink proof
paent tubing.
1. Be sure Dial Flowmeter is in the “OFF” posion.
2. Connect the Dial Flowmeter to a 50 psi (3.4 bar) Oxygen Source.
3. Set the ow by aligning the Indicang Pointer ▼to the desired
ow.
Note: The accuracy of ow will be adversely aected by any variaon
of inlet pressure, ambient temperature or outlet resistance from
rated values.
Read this user manual before installing or
operang the Dial Flowmeter.
There is NO FLOW between sengs. The Indicang Pointer ▼ must point
to a specic number on the dial to obtain ow.
ONLY use appropriate indexed ngs to connect
the Dial Flowmeter to appropriate source. Use
Oxygen connecons for oxygen owmeters and use Air connecons for Air
owmeters.
Cleaning/Maintenance
1. Disconnect all connecons before cleaning.
2. Clean exterior surfaces of the Dial Flowmeter with a cloth
dampened with water.
3. Wipe dry with a clean cloth.
4. Store the Dial Flowmeter in a clean area free from grease, oil
and other sources of contaminaon.
DO NOT immerse the Dial Flowmeter in any kind
of liquid. This will cause damage to the Dial
Flowmeter and will void the warranty.
This Dial Flowmeter is not eld serviceable.

Sleep Lab Panel
Powerex • 150 Producon Drive • Harrison, OH 45030 •USA
P 1.888.769.7979 • F 513.367.3125 • www.powerexinc.com
PX-99-0534 •03/2017
Page 4 of 8
Troubleshoong
If the Dial Flowmeter fails to funcon, consult the Trouble Shoong
Guide below. If the problem cannot be solved, consult your provider.
Problem Probable Cause Remedy
No Flow •Flowmeter set in OFF
posion
•Dial set between sengs
•Internal failure
•Hoses are not properly
connected between the
DISS inlet and Flowmeter
•Turn to desired seng
•Move Dial to ow
seng
•Return to Powerex
•Remove Front Panels
and check connecons
Dial will not turn /
Hard to turn
•Dial seized •Return to Powerex
Will not shut o •Flowmeter not in OFF
posion
•Turn Dial to the OFF
posion
To prevent risk of burns, re, or injury to person(s):
• ALWAYS conrm prescribed ow before administering to paent and
monitor ow on a frequent basis.
• Always follow ANSI and CGA standards for Medical Gas Products and
Flowmeters and Oxygen Handling.
• DO NOT use oils, greases, organic lubricants or any combusble
materials on or near this Dial Flowmeter.
• DO NOT use near any type of ame or ammable/explosive substances,
vapors or atmosphere.
• DO NOT smoke in an area where oxygen is being administered.
• This Dial Flowmeter contains magnec, ferrous material that may aect
the results of an MRI.
• NO FLOW is delivered when the Indicang Pointer ▼ is poinng to OFF.
• Flow is available only at the listed increments. There is NO FLOW
between increments.
• Only personnel instructed and trained in its use
should operate this Dial Flowmeter.
• Be sure all connecons are ght and leak free.
• Only use oxygen-safe leak detector to test for leaks.
• DO NOT autoclave.
• DO NOT gas sterilize with (EtO) Ethylene Oxide.
• DO NOT clean with aromac hydrocarbons.
• DO NOT immerse Dial Flowmeter in any kind of liquid. This will void the
warranty.
• DO NOT aempt to repair the Dial Flowmeter.
• Store the Dial Flowmeter in a clean area when not in use.
NOTE: The medical oxygen outlet you are connecng to in the
paent room must be labeled “Not For Use With A Flowmeter”.
Flow regulaon is controlled by the Dial Flowmeter that is on the
Sleep Lab Panel.

Sleep Lab Panel
Powerex • 150 Producon Drive • Harrison, OH 45030 •USA
P 1.888.769.7979 • F 513.367.3125 • www.powerexinc.com
PX-99-0534 •03/2017
Page 5 of 8
Appendix A
Technical Specicaons
NOTE: Sleep Lab Panels me be plumbed in either direcon – with the paent (use) side on le or the right.

Sleep Lab Panel
Powerex • 150 Producon Drive • Harrison, OH 45030 •USA
P 1.888.769.7979 • F 513.367.3125 • www.powerexinc.com
PX-99-0534 •03/2017
Page 6 of 8
Notes

Sleep Lab Panel
Powerex • 150 Producon Drive • Harrison, OH 45030 •USA
P 1.888.769.7979 • F 513.367.3125 • www.powerexinc.com
PX-99-0534 •03/2017
Page 7 of 8
Powerex Limited Warranty
Warranty and Remedies. (a) Standard Period of Warranty – Parts and Labor. Powerex warrants and represents all Products shall be free from
Defects for the rst twelve (12) months from the date of shipment by Powerex. During such warranty period, Powerex shall be fully liable for all
Defects in the Products (the “Product Defects”), i.e., all costs of repair or replacement, which may include “in and out” charges, so long as the
Products are located in the United States or Canada, and the Products are reasonably located and accessible by service personnel for removal.
“In and out” charges include the costs of removing a Product from buyer’s equipment for repair or replacement.
(b) Addional Period of Warranty – Parts Only (No Labor). In addion to the above, Powerex warrants the products described herein to be
free from defects in material and workmanship for forty-eight (48) months from end of Standard Period of Warranty with the excepon of
any components which are recommended to be replaced in less than sixty months in our Installaon/Operaon manuals. Within said period
Powerex will repair or replace any part or component which is proven to be defecve in either material or workmanship. This warranty covers
parts only. Labor is not included. This warranty is valid only when the product has been properly installed according to Powerex specicaons,
used in a normal manner and serviced according to factory recommendaons. This warranty does not cover failures due to damage which
occurs in shipment or failures which result from accidents, misuse, abuse, neglect, mishandling, alteraon, misapplicaon or damage due to
acts of nature.
(c) General. Powerex warrants each Powerex branded Pipeline Accessory (collecvely “Products”, individually each a “Product”) to be
free from defects in material and workmanship (“Defects”) at the date of shipment. EXCEPT AS SET FORTH BELOW, NO OTHER WARRANTY,
WHETHER EXPRESS OR IMPLIED, INCLUDING ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE, SHALL EXIST
IN CONNECTION WITH THE SALE OR USE OF SUCH PRODUCTS. TO THE EXTENT PERMITTED BY LAW, ANY AND ALL IMPLIED WARRANTIES ARE
EXCLUDED. All warranty claims must be made in wring and delivered to Powerex in accordance with the procedures set forth on its website
(www.powerexinc.com), or such claim shall be barred. Upon mely receipt of a warranty claim, Powerex shall inspect the Product claimed to
have a Defect, and Powerex shall replace any Product which it determines to have had a Defect; provided, however, that Powerex may elect,
upon return of the Product, to refund to buyer any part of the purchase price of such Products paid to Powerex. Freight for returning Products
to Powerex for inspecon or for shipping warranty parts shall be paid by buyer where permied by applicable law. Powerex is not responsible
for any import fees, taxes, dues, licenses or other fees imposed by any governmental authority upon the producon, sale, shipment and/or use
of Products covered hereunder. The warranes and remedies herein are the sole and exclusive remedy for any breach of warranty or for any
other claim based on any Defect, or non-performance of the Products, whether based upon contract, warranty or negligence.
(d) Coverage. The warranty provided herein applies to Powerex pipeline products only.
(e) Excepons. Notwithstanding anything to the contrary herein, Powerex shall have no warranty obligaons with respect to Products:
(i) That have not been installed in accordance with Powerex’s wrien specicaons and instrucons;
(ii) That have not been maintained in accordance with Powerex’s wrien instrucons;
(iii) That have been materially modied without the prior wrien approval of Powerex; or
(iv) That experience failures resulng from operaon, either intenonal or otherwise, in excess of rated capacies or in an otherwise improper
manner.
The warranty provided herein shall not apply to: (i) any defects arising from corrosion, abrasion, use of insoluble lubricants, or negligent
aendance to or faulty operaon of the Products; (ii) ordinary wear and tear of the Products; or (iii) defects arising from abnormal condions of
temperature, dirt or corrosive maer; (iv) any OEM component which is shipped by Powerex with the original manufacturer’s warranty, which
shall be the sole applicable warranty for such component.
Limitaon of Liability. NOTWITHSTANDING ANYTHING TO THE CONTRARY HEREIN, TO THE EXTENT ALLOWABLE UNDER APPLICABLE LAW,
UNDER NO CIRCUMSTANCES SHALL POWEREX BE LIABLE FOR ANY INCIDENTAL, CONSEQUENTAL, PUNITIVE, SPECULATIVE OR INDIRECT LOSSES
OR DAMAGES WHATSOEVER ARISING OUT OF OR IN ANY WAY RELATED TO ANY OF THE PRODUCTS OR GOODS SOLD OR AGREED TO BE SOLD
BY POWEREX TO BUYER. TO THE EXTENT ALLOWABLE UNDER APPLICABLE LAW, POWEREX’S LIABILITY IN ALL EVENTS IS LIMITED TO, AND SHALL
NOT EXCEED, THE PURCHASE PRICE PAID. In the event of breach of any warranty hereunder, Powerex’s sole and exclusive liability shall be
at its opon either to repair or to replace any defecve product, or to accept return, transportaon prepaid, of such product and refund the
purchase price; in either case provided that wrien noce of such defect is given to Powerex within twelve (12) months from date of shipment
to Buyer, that the product is found by Powerex to have been defecve at the me of such shipment, that the product has been installed and/or
operated in accordance with Powerex’s instrucons, that no repairs, alteraons or replacements have been made by others without Powerex’s
wrien approval, and that Buyer noes Powerex in wring within een (15) days aer the defect becomes apparent and promptly furnishes
full parculars in connecon therewith; and provided further that in no event shall the aggregate liability of Powerex in connecon with breach
of any warranty or warranes exceed the purchase price paid for the product purchased hereunder. Powerex may, at its opon, require the
return of any product, transportaon and dues prepaid, to establish any claim of defect made by Buyer. Unless otherwise agreed in wring

Sleep Lab Panel
Powerex • 150 Producon Drive • Harrison, OH 45030 •USA
P 1.888.769.7979 • F 513.367.3125 • www.powerexinc.com
PX-99-0534 •03/2017
Page 8 of 8
(a) Powerex will not accept and shall have no responsibility for products returned without its prior wrien consent, and (b) Powerex will not
assume any expense or liability for repairs to products made outside of its plant by third pares. In the event Powerex elects to replace a
defecve product, costs of installaon, labor, service, and all other costs to replace the product shall be the responsibility of Buyer.
Powerex shall not, except as set forth above, be otherwise liable to Buyer or to any person who shall purchase from Buyer, or use, any products
supplied hereunder for damages of any kind, including, but not limited to, indirect, special or consequenal damages or loss of producon of
loss of prots resulng from any cause whatsoever, including, but not limited to, any delay, act, error or omission of Powerex. Supplier’s repair
or replacement of any Product shall not extend the period of any warranty of any Product.
Warranty Disclaimer. Powerex has made a diligent eort to illustrate and describe the Products in its literature, including its Price Book,
accurately; however, such illustraons and descripons are for the sole purpose of idencaon, and do not express or imply a warranty that
the Products are merchantable, or t for a parcular purpose, or that the Products will necessarily conform to the illustraons or descripons.
Product Suitability. Many jurisdicons have codes and regulaons governing sales, construcon, installaon, and/or use of Products for
certain purposes, which may vary from those in neighboring areas. While Powerex aempts to assure that its Products comply with such codes,
it cannot guarantee compliance, and cannot be responsible for how the product is installed or used. Before purchase and use of a Product,
please review the Product applicaons, and naonal and local codes and regulaons, and be sure that the Product, installaon, and use will
comply with them.
Claims. Any non-warranty claims pertaining to the Products must be led with Powerex within (6) months of the invoice date, or they will not
be honored. Prices, discounts, and terms are subject to change without noce or as spulated in specic Product quotaons. Powerex shall not
be liable for any delay or failure arising out of acts of the public enemy, re, ood, or any disaster, labor trouble, riot or disorder, delay in the
supply of materials or any other cause, whether similar or dissimilar, beyond the control of Company. All shipments are carefully inspected and
counted before leaving the factory. Please inspect carefully any receipt of Products nong any discrepancy or damage on the carrier’s freight bill
at the me of delivery. Discrepancies or damage which obviously occurred in transit are the carrier’s responsibility and related claims should be
made promptly directly to the carrier. Returned Products will not be accepted without prior wrien authorizaon by Powerex and deducons
from invoices for shortage or damage claims will not be allowed. UNLESS OTHERWISE AGREED TO IN WRITING, THE TERMS AND CONDITIONS
CONTAINED IN THIS LIMITED WARRANTY WILL CONTROL IN ANY TRANSACTION WITH POWEREX. Any dierent or conicng terms as may
appear on any order form now or later submied by the buyer will not control. All orders are subject to acceptance by Powerex.