Presys PCON-Y17 Series User manual

®
PCON-Y17
Pressure Controller
®
Technical Manual
EM0288-02
Q
Y
U
T
A
L
I
E
N
T
M
S
E
Y
G
S
A
T
N
E
A
M
M
presys

EC Declaration of Conformity
We declare under our sole responsability that the CE marked products, are in conformity
with the essential requirements of the following EC Directives when installed in
accordance with the installation instructions contained in the product documentation:
Series
PCON-Y17
Description
Pressure Controller
LVD
Low Voltage Directive
2006/95/EC of the European Parliament and of the Council of 12 December 2006 on
the harmonization of the laws of Member States relating to Electrical Equipment
designed for use within certain voltage limits.
EN 61010-1
Safety requirements for electrical equipment for measurement, control and laboratory
use –Part 1: General Requirements.
EN 61010-2:010
Safety requirements for electrical equipment for measurement, control and laboratory
use - Part 2-010: Particular requirements for laboratory equipment for the heating of
Materials.
EMC directive
2004/108/EC of the European Parliament and of the Council of 15 December 2004
on the approximation of the laws of the Member States relating to electromagnetic
compatibility and repealing Directive 89/336/EEC
EN 61326-1
Electrical equipment for measurement, control and laboratory use - EMC requirements
–Part 1: General Requirements.
São Paulo, 8 September 2015
Vinicius José Gomes Nunes
Antonio Rafael Sito Antunes
CEO
Engineering Manager
presys

PRESYS Instruments PCON-Y17
WARNING!
The instruments described in this technical manual are equipments for
use in specialized technical area. The user is responsible for the
configuration and selection of values of the parameters of the
instruments. The manufacturer warns against the risk of incidents with
injuries to both persons and property, resulting from the incorrect use of
the instrument.
Use the instrument only according to this technical
manual. Any operation not described here is not allowed.
WARNING!
Do not apply pressure above 130% of full scale of the control range (for
the range of 3000 psi, do not apply pressure above 110% of full scale
control range).
WARNING!
Be careful with the pressure connections. High pressures with a large
volume can cause damage to both persons and property.
Apply to the positive supply port pressure between 110% and 130% of
the full scale of the control module (for the 3000 psi range, the positive
pressure supply should be between 104% and 110%).
The pressure supply must be pneumatic, dry and clean air, nitrogen or
inert gases.
WARNING!
When a vacuum pump is attached to the negative supply port of the
pressure control module, it is strongly recommended to use a protection
valve to atmosphere in the vacuum pump. When controlling from a high
pressure to a low pressure the gas is exhausted through
the negative
supply port and can cause damage to the vacuum pump. Before going to
a low pressure, it is recommended to make a VENT operation.
WARNING!
Other pressure equipments and accessories (such as hose, reservoirs,
protection devices) connected to the calibrator must be appropriate to the
working pressure.
WARNING!
During startup, do not connect any instrument to the output port of the
Pressure Control Module. Before making
connections to the Pressure
Control Module, enter the CALIBRATOR menu and wait for the automatic
VENT operation.
presys

PRESYS Instruments PCON-Y17
WARNING!
The calibrator discharges gas to the atmosphere through the supply (-)
and vent ports. Leave these ports free.
The calibrator may produce high sound levels when discharging gas. Use
a muffling device at the vent port.
WARNING!
Before connecting the instrument under test to the output port, make a
VENT operation to discharge any gas inside the controller and go to
MEASURE mode. Before turning the calibrator off, it is recommended to
vent the gas inside the controller.
WARNING!
Before connecting the instrument under test to the output port it is
recommended to configure the output range parameters. These
parameters limit the controller output ran
ge, in order to protect the
instrument under test.
WARNING!
Avoid electric shock risk when touching the equipment. Use only suitable
power cable with earth connection and never power up the equipment to
the mains socket without an earth connection.
WARNING!
High voltage is present inside these equipments. It can cause great
damages and injuries. Do not make any repair service inside the
equipment without removing the plug from the supply.
NOTICE!
High level of electromagnetic noise can cause instability to the
equipment. The equipment is provided with electromagnetic interference
filters that protect not only the mains but also the equipment itself against
noise. These filters have no function if the unit is not earthed properly.
IMPORTANT!
All pressure equipment and accessories (such as hoses, connections,
adapters, etc.) connected to the calibrator must be clean, free of residues
such as oil, dirt, dust etc. These residues can cause irreversible damage
to the internal system of the pressure control module. It is mandatory to
use filters in both the pressure feed (SUPPLY (+)/ SUPPLY(-)
) and out
(OUT).
CAUTION: Damage caused by failure to observe the above recommendations
results in total or partial loss of the equipment warranty.
presys

PRESYS Instruments PCON-Y17
Table of Contents
1 - Introduction ...............................................................................................................1
1.1. General Description ...............................................................................................1
1.2. Technical Specifications.........................................................................................2
1.3. Special Software Features.....................................................................................4
1.4. Order Code ............................................................................................................5
1.5. Parts Identification..................................................................................................6
2 - Calibrator Operation..................................................................................................7
2.1. Calibrator Menu......................................................................................................8
2.1.1. Pressure Control Module –Connections .........................................................9
2.1.2. Measure Mode...............................................................................................14
2.1.3. Control Mode.................................................................................................16
a) Range Parameters...........................................................................................17
b) Changing the setpoint......................................................................................18
c) Stabilization Settings........................................................................................19
d) Control Settings...............................................................................................19
e) Preset Setpoints ..............................................................................................21
f) Step Function ...................................................................................................22
g) Nudge Function ...............................................................................................23
2.1.4. Vent Mode.....................................................................................................24
2.1.5. Absolute Mode...............................................................................................27
2.1.6. Changing the Pressure Unit...........................................................................28
2.1.7. Filter Intensity................................................................................................29
2.1.8. Auxiliar Input..................................................................................................30
2.1.9. Transmitter Power Supply (TPS) ...................................................................33
2.1.10. Auxiliary Input Scale Function......................................................................33
2.1.11. Calibration Examples...................................................................................35
a) Pressure Transmitter Calibration .....................................................................35
b) Pressure Switch Verification ............................................................................36
2.1.12. Leakage / Stability Test ...............................................................................38
2.2. HART®.................................................................................................................40
2.2.1. HART® Connections.......................................................................................40
2.2.2. Starting Communication ................................................................................42
2.2.3. Adjusting the Measurement Range of a HART® Transmitter..........................43
2.2.4. Adjusting the Measurement Range of a HART® Transmitter with Reference
(CH Option).............................................................................................................44
2.2.5. Checking / Adjusting HART® Transmitter mA Output.....................................45
presys

PRESYS Instruments PCON-Y17
2.2.6. Full-Hart Configurator (FH Option).................................................................46
2.3. Automatic Tasks ..................................................................................................48
2.3.1. Creating Tasks ..............................................................................................48
2.3.2. Performing Tasks ..........................................................................................51
2.3.3. Viewing Results.............................................................................................52
2.4. Data Logger .........................................................................................................53
2.5. Videos..................................................................................................................55
2.6. Settings................................................................................................................56
2.7. Built-in Web Server..............................................................................................57
2.8. SCPI Commands Set...........................................................................................58
3 - Maintenance.............................................................................................................60
3.1. Replacing the Current Input Fuse.........................................................................60
3.2. Replacing the Power Supply Fuse........................................................................60
4 - Pressure Units Conversion.....................................................................................61
presys

PRESYS Instruments PCON-Y17
Page 1
1 - Introduction
1.1. General Description
The PCON-Y17 Pressure Controller enables the control of pneumatic pressure
up to 3000 psi (210 bar), including gauge and absolute pressure (optional barometric
reference). Provides all features required for easier calibration and adjustment services
on process instruments such as pressure transmitters, pressure switches, and
manometers. It has a high level of accuracy, including aspects relating to changes in
room temperature, and it maintains the specifications over long periods of time.
Besides providing high accuracy pressure values, it also allows the measurement
of signals generated by the instrument under test which is being calibrated. This is
possible due to an embedded calibrator specific for these types of signal, including
current 4-20 mA. Thus, PCON-Y17 incorporates the functions of pressure controller,
digital pressure standard and calibrator for mA, V, mV, Ohms and RTD.
It is also a Hart®Communicator (optional) with a configurable internal
resistance, transmitter power supply and latest DD as optional, in order to easily
configure and calibrate Hart®pressure transmitters.
A highly visible 5.7” touchscreen display allows easy configuration and user-
friendly operation.
PCON-Y17 features full advanced documenting capabilities. It enables the
creation and execution of automatic calibration procedures (Tasks). After completing a
task, a Calibration Report is issued and it can be printed directly on a USB connected
printer or a PDF document can be generated.
Communication with the computer through SCPI commands for on-line data
acquisition and control of the calibrator.
presys

PRESYS Instruments PCON-Y17
Page 2
1.2. Technical Specifications
1.2.1. Pressure Control Module
Choose one range for the Pressure Control Module between 250 mmH2O and
3,000 psi.
Ranges *
Resolution
Control
Stability
Accuracy
(0) 0 –250 mmH2O
0.001
0.04 % FS**
0.025 % FS**
(1) 0 –1 psi
0.0001
0.02 % FS
0.025 % FS
(2) 0 –5 psi
0.0001
0.004 % FS
0.012 % FS
(3) 0 –15 psi
0.0001
0.002 % FS
0.012 % FS
(4) 0 –30 psi
0.0001
0.002 % FS
0.012 % FS
(5) 0 –100 psi
0.001
0.002 % FS
0.012 % FS
(6) 0 –250 psi
0.001
0.002 % FS
0.012 % FS
(7) 0 –500 psi
0.01
0.004 % FS
0.012 % FS
(8) 0 –1,000 psi
0.01
0.004 % FS
0.012 % FS
(9) 0 –3,000 psi
0.01
0.004 % FS
0.012 % FS
(*) Gage pressure, vacuum (only for range 3), or compound (from range 3 to 8). Absolute pressure is obtained with the
optional barometric reference.
(**) FS = Full Scale.
Accuracy values are valid within one year and temperature range from 20 to 26 °C.
Outside these limits add 0.005 % FS / C, taking 23 °C as the reference temperature.
These values are obtained through algorithms of temperature compensation on pressure
measurements.
1.2.2. Auxiliary Input
Input Ranges
Resolution
Accuracy
Remarks
millivolt
-150 mV to 150 mV
0.001 mV
0.01 % FS*
Rinput > 10 M
auto-ranging
-500 mV to -150 mV
0.01 mV
0.02 % FS
150 mV to 2450 mV
0.01 mV
0.02 % FS
volt
-10 V to 11 V
11 V to 45 V
0.0001 V
0.0001 V
0.02 % FS
0.02 % FS
Rinput > 1 M
mA
-5 mA to 24.5 mA
0.0001 mA
0.02 % FS
Rinput < 120
resistance
0 to 400
400 to 2500
0.01
0.01
0.01 % FS
0.03 % FS
Excitation current
0.85 mA, auto-ranging
Pt-100
-200 to 850 C
-328 to 1562 F
0.01 C
0.01 F
0.1 C
0.2 F
IEC 60751
Pt-1000
-200 to 400 C
-328 to 752 F
0.1 C
0.1 F
0.1 C
0.2 F
IEC 60751
Cu-10
-200 to 260 C
-328 to 500 F
0.1 C
0.1 F
2.0 C
4.0 F
Minco 16-9
Ni-100
-60 to 250 C
-76 to 482 F
0.1 C
0.1 F
0.2 C
0.4 F
DIN-43760
(*) FS = Full Scale.
presys

PRESYS Instruments PCON-Y17
Page 3
Accuracy values are valid within one year and temperature range from 20 to 26 °C.
Outside these limits add 0.001 % FS / C, taking 23 °C as the reference temperature.
1.2.3. General Specifications
Power supply from 100 to 240 Vac, 50/60Hz.
Thirty minutes warm-up time.
Transmitter power supply (TPS): 24 Vdc, with protection from short circuit (30 mA).
Contact input for calibration of pressure switches.
Operating temperature range: 0 to 50 C.
Relative Humidity: 0 to 90 % RH.
Engineering units –psi, bar, mbar, MPa, kPa, Pa, atm, at, mmH2O, cmH2O, ftH2O,
inH2O, inH2O@60°F, torr, mmHg, cmHg, inHg, inHg@60°F, gf/cm2, kgf/cm2, kgf/m2(see
section 4–Pressure Units Conversion). Temperature: °C, °F, K, °R.
Built in Web Server,
Ethernet communication. USB port for software/firmware upgrade.
HART®Communication Protocol (optional).
Includes technical manual and test leads.
Calibration Certificate (optional).
Pneumatic Connection: 1/8” Female BSPP.
Overpressure: For the range of 3000 psi, up to 110% full scale of the pressure control
module. For the other ranges, up to 130% of the full scale of the pressure control
module.
Power Supply: 100 to 240 Vac 50/60Hz (Time Delay Type Fuse 1 A / 250 V / 5x20 mm).
Dimensions: 135 mm x 350 mm x 270 mm (HxWxD) - DT Version –Desktop.
132 mm x 483 mm x 255 mm (HxWxD) - RM Version –Rack Mounting.
Weight: 5.0 kg approx. (DT Version - Desktop)
9.5 kg approx. (RM Version –Rack Mounting)
One-year warranty.
Notes:
* Changes can be introduced in the instrument, altering specifications in this manual.
* HART®is a FieldComm Group trademark.
presys

PRESYS Instruments PCON-Y17
Page 4
1.3. Special Software Features
- PRESET POINTS: edit your most frequently used setpoints and access them quickly.
- STEP: steps or setpoints with configurable time.
- STABILITY/LEAK TEST: measures the variation of the signal (be it Pressure or an
Auxiliary Input signal) within a configurable period.
- PRESSURE SWITCH TEST: automatic testing of pressure switches.
- AUTOMATED TASKS: create calibration work orders and automatic execution of
calibration services, storage of data and generation of calibration report to a directly
connected USB printer or PDF file.
- DATA LOGGER: monitoring of input or output signals, storage and visualization of data
in chart or table.
- VIDEOS: storage and viewing videos on the calibrator screen.
- MEMORY MANAGER: stores configuration types predefined by the user.
presys

PRESYS Instruments PCON-Y17
Page 5
1.4. Order Code
presys

PRESYS Instruments PCON-Y17
Page 6
1.5. Parts Identification
DT Version - Desktop
Front Panel
Fig. 01- Front Panel
Rear Panel
Fig. 02- Rear Panel
presys

PRESYS Instruments PCON-Y17
Page 7
RM Version –Rack Mounting
Fig. 03 –Front Panel PCON-Y17-RM
Note:
For the 3000 psi Pressure Control range there is no VENT PORT.
On the PCON-Y17-RM, the VENT and SUPPLY ports of the pressure control module
are located on the back of the module, as well as the power supply (100 to 240 Vac) and
Ethernet port.
2 - Calibrator Operation
Turn on the PCON-Y17 by pressing the power switch (located on the rear panel
for the DT version, and on the front panel for the RM version).
When powered on, the calibrator goes through a self-test routine. In case of
failure, it displays a message to indicate the error; if that occurs, please contact the
Presys Technical Assistance department.
After the self-test is completed, the display shows the main menu, as showed
below.
Fig. 04- Main Menu
WARNING!
During startup, do not connect any instrument to the output port of the
Pressure Control Module.
Before making connections to the Pressure
Control Module, enter the CALIBRATOR menu and wait for the automatic
VENT operation.
The main menu is divided in the following functions:
presys

PRESYS Instruments PCON-Y17
Page 8
CALIBRATOR –selects the input/output functions (Pressure Control Module, Auxiliary
Input Signals and Pressure Measurement Module), see section 2.1.
HART®–optional module that allows communication with devices that have Hart®
Communication Protocol, see section 2.2.
TASKS –performs calibrations automatically, see section 2.3.
DATA LOGGER –record measurements, enabling visualization on chart or table, see
section 2.4.
VIDEOS –features videos made by Presys to assist in the use of the calibrator, and can
also store videos made by the user, see section 2.5.
SETTINGS –general instrument settings, see section 2.6.
2.1. Calibrator Menu
To select the Pressure Control Module and the Auxiliary Input functions press
the CALIBRATOR button from the main menu.
When first entering the CALIBRATOR menu, the PCON-Y17 executes a VENT
operation and an AUTO RESET of the pressure read in the Pressure Control Module.
When it is done, it goes to MEASURE mode and opens the output isolation valve (see
the module layout in section 2.1.2). The following screen is displayed.
Fig. 05- Calibrator initialization
presys

PRESYS Instruments PCON-Y17
Page 9
The Pressure Control Module has three operation modes: MEASURE, VENT and
CONTROL (see sections 2.1.2,2.1.3 and 2.1.4). To select an operation mode, press one
of the buttons indicated below. The current state is indicated by the symbol.
Fig. 06- Pressure Control Module - Operation Modes
2.1.1. Pressure Control Module –Connections
PCON-Y17-DT controls the pressure in the output port with high precision and
stability. To control the pressure, a pressure between 110% and 130% of the full range of
the control range is applied to the SUPPLY (+) power port, except for the 3,000 psi range
which has a reduced range of 104% to 110% of the full scale. The pressure supply must
be pneumatic, dry and clean air, nitrogen or inert gases.
If it is not possible to supply the PCON-Y17 with a pressure between 110% and
130% of the full range of the control range (and between 104% and 110% FS for the
range of 3,000 psi), a lower supply pressure may be used, but the instrument control
range will also be reduced. The PCON-Y17 measures the supply pressure and
automatically sets its control parameters for better performance. If the pressure is not
sufficient, a warning is displayed on the screen.
The presence of water, oil or particles in the pressure control module can cause
irreparable damage to the system. Equipment and connections must be clean and dry
when connected to both pressure supply and output. The use of impurities filters is
necessary to protect the system. The use of 10 micron (m) filters is recommended.
presys

PRESYS Instruments PCON-Y17
Page 10
- For the DT version, all pneumatic connections are located on the rear panel of the PCON-Y17-DT.
- For the RM version, the pneumatic connections VENT, SUPPLY (+) and SUPPLY (-) are located on the
rear panel, pneumatic connections OUT and REF are located on the front panel of the PCON-Y17-RM.
Fig. 07- Pressure Control Module
If you do not need to control negative pressure, connect to the supply (-) port a
muffling device to reduce the sound levels.
To use the PCON-Y17 with negative supply (vacuum), remove the damper (air
diffuser) from the SUPPLY (-) port and connect the PCON-Y17 vacuum kit and the
vacuum pump (see Fig. 08). The negative supply does not need to be regulated. It is
recommended that the vacuum pump has less than 70 mbar absolute (-13.5 psig)
output.
When installing a vacuum feed, the pump must be protected against positive
pressure discharge by the controller, which may damage and/or reduce the performance
of the vacuum pump. This occurs in set point changes since the PCON-Y17 releases
positive pressure from the system to the atmosphere through the SUPPLY (-) port.
presys

PRESYS Instruments PCON-Y17
Page 11
The reservoir volume of the vacuum kit slows down and attenuates any sudden
increase in positive pressure, thereby protecting the output of the vacuum pump.
Fig. 08–Connections for Vacuum
Precautions must be taken to prevent the transfer of oil from the vacuum pump to
the PCON-Y17. Before turning off the vacuum pump supply, it is recommended to
disconnect the hose that connects the vacuum pump to the reservoir, allowing
atmospheric pressure to enter directly into the pump and not through the instrument.
Without this procedure, the oil present in the vacuum pump can progressively rise to the
PCON-Y17 through the hose.
presys

PRESYS Instruments PCON-Y17
Page 12
WARNING!
The instruments described in this technical manual are equipments for
use in specialized technical area. The user is responsible for the
configuration and selection of values of the parameters of the
instruments. The manufacturer warns against the risk of incidents with
injuries to both persons and property, resulting from the incorrect use of
the instrument. Use the instrument only according to this technical
manual. Any operation not described here is not allowed.
WARNING!
Do not apply pressure above 130% of full scale of the control range (for
the range of 3000 psi, do not apply pressure above 110% of full scale
control range).
WARNING!
Be careful with the pressure connections. High pressures with a large
volume can cause damage to both persons and property.
Apply to the positive supply port pressure between 110% and 130% of
the full scale of the control module (for the 3000 psi range, the positive
pressure supply should be between 104% and 110%).
The pressure supply must be pneumatic, dry and clean air, nitrogen or
inert gases
WARNING!
When a vacuum pump is attached to the negative supply port of the
pressure control module, it is strongly recommended to use a protection
valve to atmosphere in the vacuum pump. When controlling from a high
pressure to a low pressure the gas is exhausted through
the negative
supply port and can cause damage to the vacuum pump. Before going to
a low pressure, it is recommended to make a VENT operation.
WARNING!
Other pressure equipments and accessories (such as hose, reservoirs,
protection devices) connected to the calibrator must be appropriate to the
working pressure.
presys

PRESYS Instruments PCON-Y17
Page 13
WARNING!
The calibrator discharges gas to the atmosphere through the supply (-)
and vent ports. Leave these ports free.
The calibrator may produce high sound levels when discharging gas. Use
a muffling device at the vent port.
WARNING!
Before connecting the instrument under test to the output port, make a
VENT operation to discharge any gas
inside the controller and go to
MEASURE mode. Before turning the calibrator off, it is recommended to
vent the gas inside the controller.
IMPORTANT!
All pressure equipment and accessories (such as hoses, connections,
adapters, etc.) connected to the calibrator must be clean, free of residues
such as oil, dirt, dust etc. These residues can cause irreversible damage
to the internal system of the pressure control module. It is mandatory to
use filters in both the pressure feed (SUPPLY (+)/ SUPPLY(-)) a
nd out
(OUT).
presys

PRESYS Instruments PCON-Y17
Page 14
2.1.2. Measure Mode
In Measure mode, the calibrator shows the pressure measurement in the Control
Module. In this state, the control mode is disabled.
Fig. 09- Measure mode –screen
Note: The pressure controller of 3000 psi range does not have the output isolation valves (module connection to
the outlet is always open) or VENT (no VENT port to the atmosphere).
Fig. 10 - Measure mode –state of the valves
presys
This manual suits for next models
3