PrimeCare™ PCB901 Rotating Half Rail
Intended use: The rotating half rails are intended to help patients with exit from and entry to
the PCB901 bed. These devices are NOT intended to be used for restraint purposes.
Intended application environment: This device is intended to be used in a long term care
facility, application environment 3. The environmental conditions for use and storage are as
follows.
Use Environment: Ambient temperature: 10°C ~ 40°C (50°F~104°F)
Relative humidity range: 30% ~ 75%
Transport/storage Ambient temperature: -10°C ~ 50°C (14°F~122°F)
environment: Relative humidity range: 10% ~ 93%
Maximum patient weight/safe working load: This device is not intended to support the
entire weight of a patient. This device has been tested to allow a downward load of 750 N (168
lbf) and a side load of 500 N (112 lbf).
Description of device function: This device has a single upright locked position. When the
upright locked position is not desired, the lock may be released by lifting the handle to unlock
and then rotate the device down to its lowered position.
Frequently used functions:
Lift the release handle to rotate the device.
The device locks only in the upright position.
Contraindications to use: This device shall not be used with patients who do not have the
cognitive ability to understand the operation of the device or have the strength to release the
device from its locked position.

DOC-040-2350 PCB901 Rotating Half Rail Instructions| REV 001
The information in this document is subject to change without notice.
Page3
Contents of package:
Description QTY
Right side rotating assist bar 1
Left side rotating assist bar 1
3/8” x 2.5” square neck bolt 4
3/8” knob 4
Left/right reference graphic:
STANDARD CONVENTIONS USED
DANGER
Information that appears under the DANGER description concerns the protection of
personnel from direct and pending hazards that, if not avoided, will result in immediate,
serious personal injury or death in addition to damage of the equipment.
WARNING
Information that appears under the WARNING description concerns the protection of
personnel from possible hazards that can result in injury or death in addition to damage
of the equipment.
CAUTION
Information that appears under the CAUTION description concerns the protection of
personnel from possible hazards that may result in minor injury or damage of the
equipment.
NOTE:
Information that appears under the NOTE description gives added information, which
helps in understanding the item being described.

DOC-040-2350 PCB901 Rotating Half Rail Instructions| REV 001
The information in this document is subject to change without notice.
Page4
Safety information:
DANGER
These devices are only to be mounted in the positions and orientations shown in this
documentation. Failure to comply with these instructions could result in patient/user/operator
harm and or death.
WARNING
This device is only to be used with the Primus Medical PrimeCare™ B901 bed frame.
WARNING
This device rotates. When routing cables from other equipment in the medical bed, precautions
shall be taken to avoid squeezing those between parts of the medical bed.
WARNING
No modification of this equipment is allowed.
WARNING
Do NOT use assist bars as handles for moving the bed.
WARNING
Keep all moving parts free of obstructions.
WARNING
NEVER allow anyone under the bed at any time.
WARNING
Supervision is required when this product is operated by or near children or people with
disabilities.
WARNING
DO NOT let any body parts protrude over the side or between parts, especially when
the bed is being moved or operated.

DOC-040-2350 PCB901 Rotating Half Rail Instructions| REV 001
The information in this document is subject to change without notice.
Page5
WARNING
Bed should be at lowest suitable height for entry and exit.
ENTRAPMENT WARNING
Accurate assessment of the patient and monitoring of correct maintenance and use of
equipment are required to prevent entrapment. For additional information on product
and safety issues for bed frames and rails review the manual specific to the products
you are using. If bed frames have been serviced or any other adjustments have been
made, make sure all parts are securely back in place before operating the bed frame.
ENTRAPMENT WARNING
The risks associated with these devices used in conjunction with specialty mattresses and
mattress overlays have not been evaluated. When these specialty mattresses or mattress
overlays are used the end use facility must take steps to ensure that the therapeutic benefit
outweighs the risk of entrapment.
ASSIST RAIL WARNING
Other manufacturers assist rails/bars may not be compatible and can lead to
entrapment issues or harm to patients. Only compatible Primus Medical assist bars and
rails may be used on this bed. Make sure mattress is the correct size for bed frame and
the assist bars are secured to frame to decrease the risk of entrapment.
Installation:
1. Remove the half rails from the packaging.
2. Remove the 2 knobs from the mounting bolts for one half rail.
3. Install the left or right side half rail by lowering the half rail over the top of the bed deck
lining up the holes in the half rail mounting bracket with the holes in the square tube on
the deck.
NOTE: There are alignment pins that must be paid attention to. These pins help with the
identification of the appropriate mounting positions.

DOC-040-2350 PCB901 Rotating Half Rail Instructions| REV 001
The information in this document is subject to change without notice.
Page6
4. Insert the 3/8”x2.5” square neck bolts from the outside of the bed frame and install the
knobs on the ends of the bolts.
WARNING
The knobs must be installed tightly to ensure proper mounting of the
device.
5. Repeat the procedure for the opposite side.
Assist bar mounting positions and orientations:
Head deck only

DOC-040-2350 PCB901 Rotating Half Rail Instructions| REV 001
The information in this document is subject to change without notice.
Page7
Cleaning Instructions:
WARNING
If cleaning assist bars while still mounted to the bed frame, all bed frame cleaning
instructions must be followed.
•Prior to cleaning unplug power supply cords
•Sanitize and wash all components. DO NOT submerge the assist bars.
•Use standard water pressure. DO NOT power wash or steam clean any parts
•Rinse completely with water (Maximum temperature 110oF or 43oC). Solvents,
alcohol or petroleum should not be used on the bed.
•Make sure all parts are dry before using or storing
NOTE: THIS DEVICE IS NOT COMPATIBLE WITH WASHDOWN TUNNELS
Inspections:
WARNING
Failure to take care of your device may decrease the life of your product and increase
maintenance.
Always inspect and service the device at the required intervals.
INITIAL INSPECTION
•Inspect All Components.
•Check components for obvious damage.
•Verify proper function.
Quarterly Maintenance Check
•Inspect all welds for cracks.
•Ensure proper function of the device.
•Inspect all fasteners for looseness and wear. Replace or tighten as necessary.
•Ensure no tubes are bent.
ADVICE TO RESPONSIBLE ORGANIZATION
•Carry out all adjustment and cleaning procedures specified.
•Assembly of this device and modifications made during the actual service life
require evaluation to the requirements of IEC 60601-1 and IEC 60601-2-52.
•Dispose of this device following all local rules and regulations.

DOC-040-2350 PCB901 Rotating Half Rail Instructions| REV 001
The information in this document is subject to change without notice.
Page8
Table of contents
Other Primus Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Softneta
Softneta MedDream Install manual

LINET
LINET Eleganza 3 User manual and technical description

Invacare
Invacare 6631 Assembly, installation and operating instructions

AREQUIPMENT
AREQUIPMENT Monobloc 4SW instruction manual

ECOPOSTURAL
ECOPOSTURAL C7944 instructions

Vermeiren
Vermeiren Albatros instruction manual

Braun
Braun Oral-B Plak Control MD 9000 use instructions

Centromed
Centromed Mattress Variator instruction manual

DDD
DDD SoloMobile User Intro Leaflet

Fresenius Medical Care
Fresenius Medical Care 2008K2 manual

Invacare
Invacare Carroll Assist Bar installation instructions

Mettler Electronics
Mettler Electronics Sonicator Plus 930 instruction manual