Prodigit 4016 Series User manual

4016 Series
Operation manual
S/N:900401602 REV:E

Material Contents Declaration
(材料含量宣称)
(Part Name)
零件名称
Hazardous Substance (
有毒有害物质或元素
)
铅(Pb) 汞(Hg)
镉(Cd)
六价铬
(Cr6+)
多溴联
苯(PBB)
多溴二苯醚
(PBDE)
PCBA
(印刷电路装配件)
x
○
x
○
○
○
Electrical part not on
PCBA’s
未在
PCBA
上的电子零件
x
○
x
○
○
○
Metal parts
金属零件
○
○
○
x
○
○
Plastic parts
塑料零件
○
○
○
○
x
x
Wiring
电线
x
○
○
○
○
○
Package
封装
x
○
○
○
○
○
对销售之日的所售产品
,
本表显示
, PRODIGIT
供应链的电子信息产品可能包含这些物质。注意
:
在所售产品中可能
会也可能不会含有所有所列的部件。This table shows where these substances may be found in the supply chain
of Prodigit electronic information products, as of the date of sale of the enclosed product. Note that some of the
component types listed above may or may not be a part of the enclosed product. ○:表示该有毒有害物质在该部
件所有均质材料中的含量均在SJ/T 11363-2006 标准规定的限量要求以下。○:Indicates that the concentration of
the hazardous substance in all homogeneous materials in the parts is below the relevant threshold of the SJ/T
113632006 standard. ×:表示该有毒有害物质至少在该部件的某一均质材料中的含量超出SJ/T 11363-2006 标准
规定的限量要求。×:Indicates that the concentration of the hazardous substance of at least one of all
homogeneous materials in the parts is above the relevant threshold of the SJ/T 11363-2006 standard.
Note(注释):
1.Prodigit has not fully transitioned to lead-free solder assembly at this moment;However, most of the
components used are RoHS compliant.
(此刻,Prodigit 并非完全过渡到无铅焊料组装;但是大部份的元器件一至于RoHS的规定。)
2. The product is labeled with an environment-friendly usage period in years.
The marked period is assumed under the operating environment specified in the product specifications.
(产品标注了环境友好的使用期限制(年)。所标注的环境使用期限假定是在此产品定义的使用环境之下。)
Example of a marking for a 10 year period:
(例如此标制环境使用期限为10年)

SAFETY SUMMARY
The following general safety precautions must be observed during all phases of operation, service,
and repair of this instrument. Failure to comply with these precautions or with specific warnings
elsewhere in this manual violates safety standards of design, manufacture, and intended use of the
instrument. PRODIGIT assumes no liability for the customer's failure to comply with these
requirements.
GENERAL
This product is a Safety Class 1 instrument (provided with a protective earth terminal). The protective
features of this product may be impaired if it is used in a manner not specified in the operation
instructions.
ENVIRONMENTAL CONDITIONS
This instrument is intended for indoor use in an installation category I, pollution degree 2 environments.
It is designed to operate at a maximum relative humidity of 80% and at altitudes of up to 2000 meters.
Refer to the specifications tables for the ac mains voltage requirements and ambient operating
temperature range.
BEFORE APPLYING POWER
Verify that the product is set to match the available line voltage and the correct fuse is installed.
GROUND THE INSTRUMENT
This product is a Safety Class 1 instrument (provided with a protective earth terminal). To minimize
shock hazard, the instrument chassis and cabinet must be connected to an electrical ground. The
instrument must be connected to the ac power supply mains through a three conductor
power cable, with the third wire firmly connected to an electrical ground (safety ground) at the power
outlet. Any interruption of the protective (grounding) conductor or disconnection of the protective earth
terminal will cause a potential shock hazard that could result in personal injury.
FUSES
Only fuses with the required rated current, voltage, and specified type (normal blow, time delay, etc.)
should be used. Do not use repaired
Fuses or short circuited fuse holder. To do so could cause a shock or fire hazard.
DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE.
Do not operate the instrument in the presence of flammable gases or fumes.
KEEP AWAY FROM LIVE CIRCUITS.
Operating personnel must not remove instrument covers. Component replacement and internal
adjustments must be made by qualified service personnel. Do not replace components with power
cable connected. Under certain conditions, dangerous voltages may exist even with the power cable
removed. To avoid injuries, always disconnect power, discharge circuits and remove external voltage
sources before touching components.
DO NOT SERVICE OR ADJUST ALONE.
Do not attempt internal service or adjustment unless another person, capable of rendering first aid and
resuscitation, is present.
DO NOT EXCEED INPUT RATINGS.
This instrument may be equipped with a line filter to reduce electromagnetic interference and must be
connected to a properly grounded receptacle to minimize electric shock hazard. Operation at line
voltages or frequencies in excess of those stated on the data plate may cause leakage currents in
excess of 5.0 mA peak.
DO NOT SUBSTITUTE PARTS OR MODIFY INSTRUMENT.
Because of the danger of introducing additional hazards, do not install substitute parts or perform any
unauthorized modification to the instrument. Return the instrument to a PRODIGIT ELECTRONICS
Sales and Service Office for service and repair to ensure that safety features
are maintained.
Instruments which appear damaged or defective should be made inoperative and secured against
unintended operation until they can be repaired by qualified service personnel.

DECLARATION OF CONFORMITY
Company Name: PRODIGIT ELECTRONICS CO., LTD
Address:8F, No.88, Baojhong Rd., Sindian District, New Taipei City,Taiwan.
Declares under sole responsibility that the product as originally delivered
Product Names:Digital Power Analyzer
Model Numbers:4016
(And other customized products based upon the above)
Product Options:
Safety and EMC Information:
This declaration covers all options and customized products based on the above products.
Complies with the essential requirements of the Low Voltage Directive 2014/35/EU and the EMC
Directive 2014/30/EUand carries the CE Marking accordingly.
Safety standard:
Safety standards following:
IEC 61010-1:2010
EMC standard:
EN 61326-1:2006
EN 61326-2-1:2006
EN 55011:2009+A1:2010
EN 61000-3-2:2006+A1:2009+A2:2009
EN 61000-3-3:2008
EN 61000-4-2:2009
EN 61000-4-3:2006+A1:2008+A2:2010
EN 61000-4-4:2004+A1:2010
EN 61000-4-5:2006
EN 61000-4-6:2009
EN 61000-4-8:2010
EN 61000-4-11:2004
January, 18, 2019
Date
The holder of the verification is authorized to use this verification in connection with the EC declaration
of conformity according to the Directives. The CE marking may only be used if all releveant and
effective EC Directives are complied with. Together with the manufacturer’s own documented
production control, The manufacturer (or his European authorized representative) can in his EC
Declaration of Conformity Verify compliance with the directives.

SAFETY SYMBOLS
Direct current (DC)
Alternating current (AC)
Both direct and alternating
Three-phase alternating current
Off (Supply)
On (Supply)
Protective earth (ground)
Caution!Refer to this manual before using the meter.
Caution, risk of electric shock
CAT IV – Is for measurements performed at the source of the low-voltage
installation.
CAT III – Is for measurements performed in the building installation.
CAT II – Is for measurements performed on circuits directly connected to the low-
voltage installation.
CAT I – Is for measurements performed on circuits not directly connected to Mains.
Fuse

4016 series Digital Power Analyzer operation manual
Table of Contents
Chapter 1 Introduction ·························································································1
1-1. GENERAL DESCRIPTION ·············································································1
1-2. FEATURES ····························································································1
1-3. STANDARD ACCESSORIES ··········································································1
1-4. OPTION ·······························································································1
1-5. SPECIFICATIONS 1 ···················································································2
1-6. SPECIFICATIONS 2 ···················································································3
Chapter 2 Installation···························································································5
2-1 INSPECTION···························································································5
2-2 CHECK LINE VOLTAGE ··············································································5
2-3 FUSE EXCHANGE ····················································································5
2-4 GROUNDING REQUIREMENTS·······································································6
2-5 ENVIRONMENTAL REQUIREMENTS ·································································7
2-6 REPAIR································································································7
2-7 CLEANING ····························································································7
2-8 POWER UP····························································································7
2-9 CONNECTION TO THE LOAD INPUT TERMINAL ON THE REAR PANEL···························8
2-10 RS232 INTERFACE OPTION··········································································8
2-11 GPIB INTERFACE OPTION ···········································································8
2-12 USB INTERFACE OPTION ············································································8
Chapter 3 Operation ····························································································9
3-1. 4016 SERIES DIMENSION DESCRIPTION ····························································9
3-2. 4016 I/O INTERFACE FUNCTION ·································································· 10
3-3. OPERATING INSTRUCTIONS: ······································································ 11
3-4. EXTERNAL CURRENT MEASUREMENT WIRING DIAGRAM ······································ 40
Chapter 4 Communication Interface programming operation ·································· 44
4-1. INTRODUCTION ···················································································· 44
4-2. GPIB ADDRESS MODIFICATION INSTRUCTIONS ················································· 44
4-3. RS232 PROTOCOL ················································································· 44
4-4. 4016 SERIES COMMUNICATION INTERFACE PROGRAMMING COMMAND LIST ··············· 46
4-5. THE DESCRIPTION OF ABBREVIATION···························································· 49
4-6. COMMUNICATION INTERFACE PROGRAMMING COMMAND DESCRIPTION ···················· 49
4-7. COMMUNICATION INTERFACE PROGRAMMING COMMAND SYNTAX DESCRIPTION ········· 57
Appendix 1 4016 series USB Instruction ······························································ 58
Appendix 2 4016 series LAN Instruction······························································· 60

Figures
Fig 2-1 4016 Series AC Input Connection································································5
Fig 2-2 4016Series fuse holder ··············································································6
Fig 2-3 4016 Series RS232 interface·······································································8
Fig 2-4 4016 Series GPIB interface·········································································8
Fig 2-5 4016 Series USB interface··········································································8
Fig 3-1 4016 Digital Power Analyzer dimension ····················································· 10
Fig 3-2 4016 Series I/O interface Connection························································· 10
Fig 3-3 4016 Display panel·················································································· 11
Fig 3-4 AC Standby power wiring diagram ···························································· 25
Fig 3-5 DC power accumulation wiring diagram ···················································· 29
Fig 3-6 Series resistance partial pressure wiring diagram······································· 40
Fig 3-7 CT measurement current wiring diagram ··················································· 40
Fig 3-8 4016 three-phase four-wire Y connection 3 sets of electricity meter
wiring diagram·································································································· 41
Fig 3-9 4016 three-phase three-wire △connected to three sets of electricity
meter wiring diagram························································································· 42
Fig 3-10 4016 three-phase three-wire △connected to two sets of meter diagram ······· 43
Fig 4-1 RS232 INTERFACE CONNECTION OF REAR PANEL ··································· 45

Tables
Table 1-1 4016 Series Specifications ········································································2
Table 1-1A 4016 Series Specification ········································································4
Table 4-1 Communication Interface programming setting command summary ················· 48
Table 4-2 GPIB COMMAND TERMINATOR ····························································· 57


4016Series Operation Manual 1
Chapter 1 Introduction
1-1. General description
5 digit V/A/W Meter,display the Voltage(Vrms, Vpeak, Vmax., Vmin)、Current (Irms,
Ipeak, Imax, Imin.),Watt, Voltampere (VA), Frequency Crest Factor、Power Factor,Total
Harmonic Distortion of Voltage(VTHD), Voltage Harmonic(VH)、Total Harmonic
Distortion of Current(ITHD), Current Harmonic(IH)
1-2. Features
The main features of the 4016 series of digital Power Analyzer are highlighted below.
1.2.1. Less than 0.001W standby power measurements meet IEC62301 Energy Star 0.03W
standby power.
1.2.2. Harmonic analysis to the 50thorder
1.2.3. 6 voltage range and 18 current range
1.2.4. Up to 800Vpeak/200Apeak input
1.2.5. Color graphical display
1.2.6. Data log with Vrms, Arms, Watt, PF,VTHD and ITHD
1.2.7. The built-in Power switch can be Controlled the ON/OFF degree angle (0~359˚)
according to of the Voltage source phase angle
1.2.8. RS-232, GPIB, USB, Ethernet Interface (O p tio n )
1.2.9. 9942 Measuring fixture box (O p tio n )
1.2.10. Supports the Inrush Current of the power supply at startup and the Surge Current test
When the load is suddenly plugged in (Hot Plug-in).
1-3. Standard Accessories
a 4016 Series operation manual...................................1PCs
1-4. Option
1.4.1. GPIB interface
1.4.2. RS232 interface
1.4.3. GPIB interface
1.4.4. USB interface + USB & LAN DRIVER CD
1.4.5. LAN interface+ USB & LAN DRIVER CD
1.4.6. 9942 Measuring fixture box
1.4.7. GPIB cable 1 M
1.4.8. GPIB cable 2 M
1.4.9. USB TYPE A TO TYPE B cable 1.8 M.

2PRODIGIT
1-5. Specifications 1
Table 1-1 4016 Series Specifications
AC INPUT
LINE
100Vac~230Vac ± 10%
FREQUENCY
50/60 Hz ±3Hz
PROTECT FUSE
0.5A/250V (5*20mm)
MAX.POWER
CONSUMPTION
50VA
Model
Voltage
Current
Dimension(HxWxD)
WEIGHT
4016 500V 20A 99.4 mm x 213 mm x 304 mm

4016 Series Operation Manual 3
1-6. Specifications 2
4016 Specification
Item Specification
ACV & DCV
Vrms, Vpk+/Vpk-, Vmax/Vmin,
V Harmonic
Range
20Vpeak/0.001V
40Vpeak/0.001V
80Vpeak/0.01V
200Vpeak/0.01V
400Vpeak/0.01V
800Vpeak/0.1V
Max. Input 800Vpeak/500Vrms
Accuracy
±0.1% of (Reading + Range)
±0.5% of (Reading + Range, For Peak)
ACA & DCV
Arms, Apk+/Apk-
,
Amax/Amin,
A Harmonic
Shunt 0.05A
(10Ω)
Range
0.002Apeak/0.1uA
0.004Apeak/0.1uA
0.008Apeak/0.001mA
0.02Apeak/0.001mA
0.04Apeak/0.001mA
0.08Apeak/0.01mA
Max. Input 0.08Apeak/0.05Arms
Shunt 0.5A
(1Ω)
Range
0.2Apeak/0.01mA
0.4Apeak/0.01mA
0.8Apeak/0.1mA
Max. Input 0.8Apeak/0.5Arms
Shunt 5A
(0.04Ω)
Range
2Apeak/0.1mA
4Apeak/0.1mA
8Apeak/0.001A
Max. Input 8Apeak/5Arms
Shunt 20A
(0.005 Ω)
Range
10Apeak/0.001A
20Apeak/0.001A
40Apeak/0.001A
50Apeak/0.001A
100Apeak/0.01A
200Apeak/0.01A
Max. Input 200Apeak/20Arms
Accuracy
±0.1% of (Reading + Range)
±0.5% of (Reading + Range,For Peak)
VCF & ICF
Range
0.0000~9.9999
Accuracy
±0.5% of (Reading + Range)
AC Power & DC Power
Watt, VA, VAR
Range
Vrange*Arange
Accuracy
±0.2% of (Reading + Range)
PF
Range
±0.001~1.000/0.001
Accuracy
1% of (Reading + Range, Corresponds to V and A)
Frequency
Range
dc, 20~1000/0.1Hz
Accuracy
± 0.1 Hz
V/A Harmonic
Number
1~50 th/ Same as ACV, ACA meter
Accuracy
±0.5% of (Reading + Range)
V/A THD
Range
0%~255%/0.001%
Accuracy
±0.5% of (Reading + Range)

4PRODIGIT
Inrush V/A
Voltage
Range
Same as ACV & DCV
Max. Input
Accuracy
±2% of (Reading + Range)
Current
Shunt 20A (0.005 Ω)
Range
Same as ACA & DCA
Max. Input
Accuracy
±2% of (Reading + Range)
Measurement Wide
100mS
AC ON/OFF
Programmable
output switch
ON
0~359°/1°
OFF
0~359°/1°
Accuracy
Max. +/- 1° @50/60Hz
Standby Power
Accumulated Time
0D0H0M0S~9999D23H59M59S
WHr
0.000000~999.999999 WHr / 1.000~9999.999 KWHr
Counter
0H0M0S~99H59M59S
Accuracy
±0.2% of (Reading + Range)
Data Log
Item
Vrms
、
Arms
、
Watt
、
PF
、
VTHD
、
ITHD
Updata Rate
0.2
、
0.5
、
1
、
2
、
5
、
10 Second
ON/OFF Cycling
ON time
0M0.200S~ 10M0S
OFF Time
0M0.200S~ 10M0S
Repeat Cycle
0~9999
Low Pass Filter(V & A) 50KHz
Interface(Option) RS-232, GPIB, USB, Ethernet
Operating Theory
Rms Voltage (Vrms)
Rms Current (Arms)
+ or - Peak Value
(+/-Vpk, +/-Apk)
Max[Value(t)] or Min[Value(t)]
Max.or Min Value
(Vmax/Vmin, Amax/Amin,
Wmax/Wmin)
Max[Value] or Min[Value]
Crest Pactor (VCF, ICF)
Peak Value/Rms Value
Active Power (Watt)
Apparent Power (VA)
Reactive Power (VAR)
Power Factor (P.F.)
Harmonic
Tatal Harmonic Distortion
(%)
Rms Sampling Rate
4096 sample/Cycle @ 50/60 Hz
Inrush Sampling
<2.5us
V/A ADC
Dual 16-Bit, 500KSPS ADC with DSP
Power Input
110/220V 50/60Hz
Consumption
38VA
Protection(fuse)
Shunt 0.05A (10Ω) 3.6x11mm 250Vac 0.2A Fast
Shunt 0.5A (1Ω)
3.6x11mm 250Vac 1A Lag
Switch
6*30mm 250V/25A
Display
3.5" TFT LCD, 320 x RGB x 240
Dimensions
Height
99.4 mm with feet
Width
213 mm
Depth
304 mm
Weight
3.5 Kg
Storage temperature
-20 ºC to +60 ºC (-4 °F to 140 °F)
Operating temperature
0 ºC to 40 ºC (32 °F to 104 °F)
Maximum operating altitude
2000 M (6562 ft)
Maximum relative humidity
80% for temperatures up to 31 °C (88 °F) decreasing linearly to 50 % relative humidity at 40 °C (104 °F)
Table 1-1A 4016 Series Specification
∫
T
i
dtA
T
0
2
1
dtAV
T
i
T
i
∫×
0
1
1
2
50
2
3
2
2
... HHHH ÷++
ArmsVr ms
Wa t t
×
22 Hq
Hr +
22
WVA −

4016 Series Operation Manual 5
Chapter 2 Installation
2-1 Inspection
The 4016 Series AC/DC Digital Power Analyzer was carefully inspected before shipment. If
instrument damage has occurred during transport, please inform Prodigit's sales and service
office or representative.
You’re 4016 Series AC/DC Digital Power Analyzer was shipped with a power cord for the type
of Terminal blocks used at your location. If the appropriated cord was not included, please
contact your nearest Prodigit sales office to obtain the correct cord. Refer to “check line
voltage “to check the line voltage is 100V~230Vac.
2-2 Check line voltage
The 4016 Series AC/DC Digital Power Analyzer can operation with 100 Vac ~230Vac input as
indicated on the label on the rear panel.
Make sure that the factory check mark corresponds to your nominal line voltage. Skip this
procedure if the label is corrected marked.
2.2.1. With the 4016 Series AC/DC Digital Power Analyzer power OFF, disconnect the
Power cord.
2.2.2. Refer the drawing on the rear panel of 4016 Series high power Digital Power
Analyzer in Fig 2-1.
Fig 2-1 4016 Series AC Input Connection
2-3 Fuse Exchange
This product has the power fuse, and exchanges it according to the following procedure.
Never fail to turn off the power of this product, and disconnect the plug of the AC Power
cable.

6PRODIGIT
2.3.1.Check the rating of the line fuse and replace it with the correct fuse if necessary.
100V~230V use T0.5A/250V (5*20mm).
2.3.2.The AC line fuse is located below the AC line receptacle see Fig 2-2. Use a small
Screwdriver to extract the fuse holder, to change a new one. Change an appropriate
Specifications fuse which indicated in Table 1-1.
2.3.3.Reinstall fuse holder and connect the power cord.
Fig 2-2 4016Series fuse holder
2-4 Grounding requirements
SHOCK HAZARD
1. It is requested to use the 3Pin plug connector only for 4016 Series mainframe to out of
danger when electric leakage. And the complete and proper grounded is necessary.
2. The 4016 Series high power AC/DC Digital Power Analyzer is equipped with three
conductor cable which plugs in an appropriate receptacle to ground the instrument's
cover.
To avoid the fire or electronic shock, the Fuse that will be used in the product should have
the safety standard in the area of the region you use. Any use of improper Fuse or shorting
the Fuse holder would be extremely dangerous and would be strictly prohibited.
・Before exchanging the Fuse, if there are abnormal odor or abnormal noise,
・Please stop using immediately and ask for the repair.
T0.5A/250V (5*20mm)
T0.5A/250V (5*20mm)

4016 Series Operation Manual 7
2-5 Environmental requirements
•Indoor use.
•Measurement Category I.
•Pollution Degree 2.
•Relative Humidity 80% Max.
•Ambient Temperature 0 to +40°C
•Altitude up to 2000m.
•The equipment is not for measurements performed for CAT II, III and IV.
•Transient Overvoltage on the mains supply can be 2500V.
2-6 Repair
If the instrument is damaged, please attach a tag to the instrument to identify the owner and
indicated the require service or repairing. And inform the Prodigit sales and service office or
representative.
Fuses (F1, F2, Fuse 25A), overcurrent protection diodes (D1~4, D31~32) and switching
components (14016402) are consumables and are not covered by the warranty.
2-7 Cleaning
2-8 Power Up
The following procedure should be followed before applying mains power:
2.8.1. Turn off (O) the POWER switch
2.8.2. Check that the power cord is corrected.
2.8.3. Check that nothing is connected to the DC INPUT on the rear panels.
2.8.4. Turn on POWER switch.
Use a soft or slightly damp cloth to clean this product.
BEFORE you clean the unit, switch the mains power off and disconnect the input lead.
・Please do NOT use any organic solvent capable of changing the nature of the plastic
such as benzene or acetone.
・Please ensure that no liquid is allowed to penetrate this product.

8PRODIGIT
2-9 Connection to the load Input Terminal on the Rear Panel
Connection procedure of the load input terminal on the rear panel
2.9.1. Turn off POWER switch.
2.9.2. Check that the output of the equipment under test is off.
2.9.3. Check the polarity of the connection and connect the load wire to the output
Terminal of the equipment under test.
2-10 RS232 interface option
Fig 2-3 shows the RS232 interface (Female) on the rear panel connects 4016 Series
Digital Power Analyzer to RS232 port of computer in one by one configuration .The
RS232 BAUD-RATE can be set in the front panel
Fig 2-3 4016 Series RS232 interface
2-11 GPIB interface option
2.11.1 The maximum number of devices including the controller is no more than 15.
2.11.2 The maximum length of all cable is no more than 2 meters times the number
Of devices connected together, up to 20 meters maximum.
Fig 2-4 4016 Series GPIB interface
2-12 USB interface option
Fig 2-5 shows the USB interface in the rear panel of 4016 Series Digital Power
Analyzer . Please Refer Appendix B.
Fig 2-5 4016 Series USB interface

4016 Series Operation Manual 9
Chapter 3 Operation
This chapter describes the front panel function and operation of each 4016 Series Digital Power
Analyzer, The communication Interface programming is described in Chapter 4.
3-1. 4016 Series dimension description
213
287

10 PRODIGIT
Fig 3-1 4016 Digital Power Analyzer dimension
3-2. 4016 I/O interface function
The 4016 Series Digital Power Analyzer I/O interface includes Vsense, Load, Source,
SYNC TTL, and EXT.
10
9
11
8
2
1
3
4
5
6
7
Fig 3-2 4016 Series I/O interface Connection
1. Power Input 100~230V 50/60Hz
2. Voltage measurement contact positive terminal
3. Voltage measurement contact negative terminal
4. Positive end of the EUT
5. Negative end of the contact of the EUT
6. Measuring the positive input of the power supply
7. Measuring the negative input of the power supply
8. Internal Switch Synchronization Signal
9. External current signal input positive terminal
10. External current signal input negative terminal
11. Interface card (optional RS232 GPIB USB)

4016 Series Operation Manual 11
3-3. Operating instructions:
1
2
6
5
3
4
7
8
9
Fig 3-3 4016 Display panel
•Model number (1)
Refers to model number, of 4016 Series Digital Power Analyzer.
•4016 Power switch (2)
•Number keys (3)
•Key (4), Turn output on or off.
※The output switch has an angle control function. In the AC mode, VSource and
VSense need to input power supply to operate normally.
•Key (5), Enter or leave the edit page
•Key (6), clear data
•Key (7), Switch data and graphics pages.
•The arrow keys and Select keys (8) switch the screen and select functions.
•Key (9), Return to the Menu menu
Table of contents
Other Prodigit Measuring Instrument manuals