PTW MP3-PL Installation instructions

Assembly Instructions and User Manual
MP3-PL
Therapy Beam Analyzer T41050
Version 0 or higher
Contents
Operating Manual
Technical Manual
Service Manual
D850.131.05/01 en
2016-05 Bru

MP3-PL Therapy Beam Analyzer T41050
General Information
2
D850.131.05/01 en
General Information
The product bears the CE-mark "CE-0124" in
accordance with the Council Directive
93/42/EEC about Medical Devices and fulfills
the essential requirements of Annex I of this
directive. The product is a class IIb device
(MDD).
Electrical Safety
The product fully complies with the require-
ments of the standard IEC 61010-1.
Electromagnetic Immunity
The product fully complies with the require-
ments of the following standards:
IEC 61326-1
The user manual is an integral part of the
product. It should always be kept near the
product. Observance of the manual is a pre-
requisite for proper product performance and
correct operation.
Operator safety, specified measuring accura-
cy, and interference-free operation can only be
guaranteed if original products and parts are
used. Furthermore, only the accessories listed
in this manual are approved by PTW-Freiburg
and only they or other accessories whose use
has been expressly permitted by PTW-
Freiburg may be used in conjunction with the
product. Safe operation and proper product
performance are not guaranteed if accessories
or consumables from other manufacturers are
used.
PTW-Freiburg cannot be held liable for dam-
age resulting from the use of accessories or
consumables from other manufacturers or
when the user ignores the instructions and in-
formation given in this manual.
The warranty period is 1 (one) year and begins
on the day of delivery.
It is unaffected by repairs covered by the war-
ranty regulations.
PTW-Freiburg only considers itself responsible
for the safety, reliability, and performance of
the product if the assembly, extension, read-
justment, modification, or repair is carried out
by PTW-Freiburg or by persons authorized by
PTW-Freiburg, and if the product is used in
compliance with the technical documentation.
In case of any questions concerning the ser-
vice, support, or warranty, please contact your
supplier.
This manual is in conformity with the product
specifications and all applicable safety stand-
ards that are valid as at the printing date. All
rights are reserved for devices, circuits, tech-
niques, software, and names as referred to in
the manual.
PTW-Freiburg reserves the right to make mod-
ifications.
Please contact PTW or your local representa-
tive for the most current information concern-
ing the products of interest.
No part of the technical documentation may be
reproduced without written permission from
PTW-Freiburg.
PTW-Freiburg is a registered manufacturer
according to the ElektroG (Elektro- und Elek-
tronikgeräte-Gesetz).
Elektro-Altgeräte-Register (EAR) Registration
number DE15599992
PTW-Freiburg works in strict accordance with
a quality management system that is continu-
ously updated according to the national and in-
ternational standards.

MP3-PL Therapy Beam Analyzer T41050
General Information
D850.131.05/01 en
3
PTW-FREIBURG
Physikalisch-Technische Werkstätten
Dr. Pychlau GmbH
Lörracher Str. 7
79115 FREIBURG
GERMANY
Phone: +49 761 49055-0
Fax: +49 761 49055-70
info@ptw.de
www.ptw.de

MP3-PL Therapy Beam Analyzer T41050
Contents
4
D850.131.05/01 en
Contents
General Information 2
Contents 4
Illustrations 6
Intended Use 7
Safety Information 8
Operating Manual 15
1The Components of the MP3-PL System 15
1.1 Perspex Tank 17
1.2 Moving Mechanism 18
1.3 Positioning Device 19
1.4 SCANLIFT Carriage and Reservoir 20
1.5 Control Unit and Control Pendant 20
2Assembly of the System 21
2.1 Assembling SCANLIFT, Positioning Device, Chamber Cable Holding Device and Perspex Tank 21
2.2 Assembling the Junction Box and the Moving Mechanism 23
3Installation of the System for Measurement 27
3.1 Quickstart (Workflow) 27
3.2 Mounting Detectors on the Moving Mechanism 28
3.3 Connecting the Detectors 35
3.4 Mounting and Connecting the Reference Detector 37
3.5 Mounting the Inhomogeneous Phantom T43031 39
3.6 Mounting the Hemisphere T41029.3.231 39
3.7 Mounting the Wedge T41029.1.220 40
3.8 Mounting the Film Holding Device T40040 40
3.9 Connecting SCANLIFT 41
3.10 Filling with Water 42
3.11 Connecting the Control Unit and Control Pendant 44
4Installation Check 45
5Error Messages and Troubleshooting 46
Technical Manual 47
6Technical Specifications 47
7Order Information 49

MP3-PL Therapy Beam Analyzer T41050
Contents
D850.131.05/01 en
5
Service Manual 51
8Cleaning 51
9Preventive Maintenance 53
10 Storage 54
11 Disposal of the Product 54
Literature 55
Appendix: China Electronic Industry Standard Compliance 56
Index 58

MP3-PL Therapy Beam Analyzer T41050
Illustrations
6
D850.131.05/01 en
Illustrations
Figure 1: Definition of patient environment 13
Figure 2: MP3-PL System with accessories for measurements with 24 PinPoint chambers 15
Figure 3: MP3-PL System with accessories for measurements with a measuring and a reference detector 16
Figure 4: MP3-PL Perspex tank with junction box and moving mechanism (top: front view; below: rear view) 17
Figure 5: Connections at the junction box 18
Figure 6: Positioning device for tank alignment 19
Figure 7: Assembling SCANLIFT, positioning device, chamber cable holding device and Perspex tank 22
Figure 8: Mounting the A-axis 24
Figure 9: Mounting the B-axis 25
Figure 10: Mounting the C-axis 26
Figure 11: Inserting the 24 PinPoint chambers into the 3D-detectorblock 28
Figure 12: Vertical mounting of the 3D-detectorblock with 24 PinPoint chambers 29
Figure 13: Horizontal mounting of the 3D-detectorblock with 24 PinPoint chambers 29
Figure 14: Inserting the 24 PinPoint chambers into the linear detector holder 30
Figure 15: Vertical mounting of the linear detector holder with 24 PinPoint chambers 31
Figure 16: Horizontal mounting of the linear detector holder with 24 PinPoint chambers 32
Figure 17: Horizontal mounting of the Bragg Peak chamber 33
Figure 18: Vertical mounting of the Bragg Peak chamber 34
Figure 19: Laying and bundling of the chamber cables 35
Figure 20: Chamber cable holding device T20015.1.200 36
Figure 21: Mounting the reference detector 37
Figure 22: Mounting the reference detector holder 38
Figure 23: Mounting the inhomogeneous phantom T43031 39
Figure 24: Mounting the hemisphere T41029.3.231 39
Figure 25: Mounting the wedge T41029.1.220 40
Figure 26: Mounting the film holding device T40040 40
Figure 27: Coupling of the plastic hose of the water reservoir 41

MP3-PL Therapy Beam Analyzer T41050
Intended Use
D850.131.05/01 en
7
Intended Use
The MP3-PL Therapy Beam Analyzer system
T41050 is a computer-controlled measuring system
for recording dose distributions (depth dose curves
and profiles) in order to check the emitted irradia-
tion of particle accelerators (such as a cyclotron or
synchrotron).
For filling or emptying the MP3-PL Therapy Beam
Analyzers the SCANLIFT carriage and reservoir is
used.
The basic system consists of the following compo-
nents:
Perspex tank with moving mechanism
TBA CONTROL UNIT T41013
TBA CONTROL PENDANT T41031
SCANLIFT carriage and reservoir T41024
MP3-M positioning device T41022.1.021
Chamber cable holding device T20015.1.200
2 multi-channel dosemeters MULTIDOS T10004
2 detector connection boxes T16007
24 PinPoint chambers TM31015
3D-detectorblock T21003
COM-Server (RS232-TCP/IP Interface)
The basic system can be expanded with additional
accessories (see section 7, Order Information).
NOTE
Please observe the information given in the user
manuals supplied with the SCANLIFT carriage
and reservoir, the multi-channel dosemeter
MULTIDOS, the TBA CONTROL UNIT, the TBA
CONTROL PENDANT, the PC and the ionization
chambers.
NOTE
The corresponding software is the only means
for running and analyzing measurements. Follow
the instructions given in the corresponding user
manual.

MP3-PL Therapy Beam Analyzer T41050
Safety Information
8
D850.131.05/01 en
Safety Information
This is the safety alert symbol. It is used to
alert the user to potential hazards. Obey all
safety messages that follow this symbol to
avoid possible bodily injury or equipment
damage.
All safety messages consist of the following com-
ponents:
–Safety alert symbol and signal word
–Type of danger
–Source of danger
–Consequence
–Measures to prevent hazard
Signal Words
DANGER
Indicates an imminent hazard. If not avoided, the
hazard will result in death or serious injury.
WARNING
Indicates a hazard. If not avoided, the hazard
can result in death or serious injury.
CAUTION
Indicates a potential hazard. If not avoided, the
hazard could result in minor injury or product /
property damage.
NOTE
Provides useful information to assure that you
get the most from your equipment.
Safety Information
DANGER
Operation in areas where an explosion hazard
may occur or in oxygen-enriched atmospheres
Explosion Hazard!
The product is not suitable for operation in areas
of risk where an explosion hazard may occur.
Explosion hazards may be caused by the use of
combustible anaesthetics, skin-cleansing agents
and disinfectants.
Furthermore the product is not suitable for appli-
cation in oxygen-enriched atmospheres. The
atmosphere is considered to be oxygen-enriched
when more than 25 % of oxygen or nitrous oxide
is added to the ambient air.

MP3-PL Therapy Beam Analyzer T41050
Safety Information
D850.131.05/01 en
9
WARNING
Improper handling.
Patient Hazard!
The product is a medical electrical device and
must only be handled by persons who are trained
in the use of such equipment and are capable of
applying it properly. The operator must be trained
in the use of the device.
WARNING
Electricity is a source of risk, particularly when
the product is not in perfect operating condition
or when it is operated inappropriately.
Shock Hazard!
Strictly observe the following warnings. Failure to
do so may endanger the lives of the patient, the
user and other persons involved.
Before using the product, the user must as-
certain that it is in correct working order and
operating condition.
Before putting the device into operation, visu-
ally inspect all connection cables for signs of
damage. Damaged cables and connectors
must be replaced immediately.
When disconnecting the device from the pow-
er line, first remove the plug from the wall out-
let, then disconnect the power cord from the
device.
Devices on which moisture condensation has
developed as a result of temperature changes
must not be switched on unless completely
dry.
WARNING
Liquids must not enter the product. If liquids
have entered the product, it must be thor-
oughly inspected before being used again.
The devices are IEC 61140 protection class I
equipment. They may be put into operation
only when connected to a properly installed
power outlet with earthing contact.
All devices of a system must be connected to
the same electric circuit. Devices which are
not connected to the same circuit must be
electrically isolated (use RS232 interface with
galvanic insulation).
Exclusion of operation in the patient envi-
ronment: Neither the product nor any peripheral
devices may be operated in the patient environ-
ment (see Figure 1).
Exclusion of operation as device with patient
contact: The device is not intended for use in
direct contact with the patient. Neither the device
nor any peripheral device may have contact to
the patient.
Never touch the patient and open connectors of
the device at the same time.
Exclusion of operation as controlling instru-
ment: The device is only designed for use as a
measuring device. The device must not be used
to control radiodiagnostic equipment or radio-
therapy units.

MP3-PL Therapy Beam Analyzer T41050
Safety Information
10
D850.131.05/01 en
WARNING
Electricity is a source of risk, particularly when
the product is not in perfect operating condition
or when it is operated inappropriately.
Shock Hazard!
Strictly observe the following warnings. Failure to
do so may endanger the lives of the patient, the
user and other persons involved.
Use of peripheral devices: Peripheral devices
(PC, printer) may only be connected if they meet
the requirements of standard UL 60950-1 or
IEC 60950-1 or of standard IEC 61010-1.
Devices may be connected to other devices or to
parts of systems only if it has been ascertained
that this connection does not impair the safety of
the patient, the operator or the environment.
If the device specifications do not contain infor-
mation as to connecting the device to other
equipment, you must consult the manufacturer of
the other equipment or an expert about the ef-
fects of the connection on the patient, the opera-
tor or the environment. Always observe standard
IEC 61010-1.
WARNING
Operation in conjunction with particle beams that
activate the material of the product.
Risk to Persons!
Certain types and energies of particle beams
may activate the material of the product. In these
situations it must be ascertained that there are
no persons in the vicinity of the product until the
radiation of the activated material has decayed.
WARNING
Magnetic and electrical fields are capable of in-
terfering with the proper performance of the de-
vice.
Equipment Failure!
For this reason make sure that all external devic-
es operated in the vicinity of the product comply
with the relevant EMC requirements. X-ray
equipment, MRI devices, radio systems, cell
phones, etc. are possible sources of interference
as they may emit higher levels of electromagnet-
ic radiation. Keep the product away from these
devices and verify its performance before use.
WARNING
Hazards originating from other system compo-
nents.
Patient Hazard! Equipment Damage!
Observe the safety information provided in the
user manuals of the system components.

MP3-PL Therapy Beam Analyzer T41050
Safety Information
D850.131.05/01 en
11
CAUTION
Use of the product without observing the user
manual.
Bodily Injury! Equipment Damage!
Always use the product in compliance with the
user manual. Otherwise, the intended protection
can be reduced.
Use the product only in conjunction with the
products approved for this purpose by PTW
and/or the products listed in the section "Intend-
ed Use".
Handle the product with care to avoid equipment
damage.
CAUTION
Operation under inadequate ambient conditions.
Equipment Damage!
Before connecting the device to the power line,
make sure that line voltage and frequency are
equivalent to the ratings indicated on the name-
plate of the device.
Set up the device in a location which affords
sufficient ventilation. The air gaps of the device
must not be obstructed.
Always observe the ambient conditions as indi-
cated in the `Technical Specifications`.
CAUTION
Operation in conjunction with particle beams.
Equipment Damage!
Continuous irradiation with a pencil beam
through the entrance window at a fixed position
can cause a local discoloring of the material. If
such a discoloring appears, the position of the
beam traversing the entrance window should be
shifted (e.g., laterally by 15 mm). Continued irra-
diation through the original position can lead to
radiation damage of the entrance window mate-
rial. Cracks can occur.
CAUTION
Improper handling of cables and connectors.
Equipment Damage!
To avoid damage to the cables and connectors,
observe the following points.
Do not kink the cables. Observe the minimum
bend radius of the cables. Never exert pres-
sure on the cables. Do not step on the cables.
Do not roll trolleys over the cables. Do not pull
or twist the cables.
Disconnect the cables from the devices during
transport and when the devices are not in
use. Do not let the cables hang down unre-
strained.
The connectors must always be clean. Do not
allow the connectors to lie on the floor. Al-
ways protect connectors when pulling them
through cable conduits.

MP3-PL Therapy Beam Analyzer T41050
Safety Information
12
D850.131.05/01 en
CAUTION
Improper handling.
Risk of injury!
If the water phantom is lowered to the bottom
position by means of the SCANLIFT lift mecha-
nism, only very little space (approx. 10 –12 mm)
is left between the clamping screw of the cham-
ber cable holding device and the lower push bar
of the SCANLIFT.
While the lift element is moving, ensure that no
other persons are working on the SCANLIFT.
Otherwise, these persons are at risk of suffering
crush injuries at their hands and fingers.
CAUTION
Improper handling.
Risk of injury!
The chamber cable holding device is very heavy
(several kilograms).
Push the chamber cable holding device carefully
to its final position. Otherwise, there is a risk of
suffering crush injuries at the hands and fingers
at the end of the rails.
CAUTION
Improper handling.
Risk of injury!
When used with the SCANLIFT, rotating the
MP3-PL by means of the positioning device is
not allowed.
When the positioning device rotates, the base
plate of the chamber cable holding device may
collide with the higher push bar of the
SCANLIFT. This may damage the product.
While the lift element is moving, ensure that no
other person rotates the positioning device. Oth-
erwise, these persons are at risk of suffering
crush injuries at their hands and fingers.

MP3-PL Therapy Beam Analyzer T41050
Safety Information
D850.131.05/01 en
13
NOTE
Set up the device so that the operator has a
clear, unobstructed view of the indicators and
controls and can operate the controls easily.
NOTE
Do not position the device so that it is difficult to
operate the disconnecting device.
Figure 1: Definition of patient environment
Symbols on the Nameplate
Symbol
Description
Please refer to user manual!
The product bears the CE-mark.
Manufacturer and date of manufac-
ture
Reference number
Serial number
Separate collection of electrical and
electronic equipment!
(refer also to the section "Disposal
of the Product")
Labeling according to "Administra-
tion on the Control of the Pollution
caused by Electronic Information
Products (ACPEIP)" (China RoHS)
(refer also to the Appendix)
For a description of additional symbols, if appli-
cable, please refer to the equipment description.

MP3-PL Therapy Beam Analyzer T41050
Safety Information
14
D850.131.05/01 en

MP3-PL Therapy Beam Analyzer T41050
Operating Manual –The Components of the MP3-PL System
D850.131.05/01 en
15
Operating Manual
1 The Components of the MP3-PL System
Figure 2: MP3-PL System with accessories for measurements with 24 PinPoint chambers

MP3-PL Therapy Beam Analyzer T41050
Operating Manual –The Components of the MP3-PL System
16
D850.131.05/01 en
Figure 3: MP3-PL System with accessories for measurements with a measuring and a reference detector

MP3-PL Therapy Beam Analyzer T41050
Operating Manual –The Components of the MP3-PL System
D850.131.05/01 en
17
1.1 Perspex Tank
The moving mechanism as well as several ancillary
devices (refer to section 2) are attached to the Per-
spex tank. The tank (Figure 4) is filled with water. The
water level should remain approx. 20 mm below the
top edge of the tank. Horizontal and vertical lines on
the tank allow it to be easily leveled. As an alternative,
you can use the circular bubble level that is part of the
MP3-PL accessories.
Additional marks at the bottom and on the walls of the
tank allow it to be aligned, using the accelerator laser
system.
The Perspex plates are glued together. On the inside
of the tank, a special sealing agent protects the glue
from penetrating water.
Figure 4: MP3-PL Perspex tank with junction box
and moving mechanism (top: front view;
below: rear view)

MP3-PL Therapy Beam Analyzer T41050
Operating Manual –The Components of the MP3-PL System
18
D850.131.05/01 en
Figure 5: Connections at the junction box
1 without function
2 without function
3 Control cable to the control unit
4 Motor cable A
5 Motor cable B
6 Motor cable C
7 Motor cable to the control unit
8 Control pendant
1.2 Moving Mechanism
CAUTION
Improper handling.
Risk of injury!
Do not reach into the Perspex tank while the
moving mechanism is working.
Pay attention that hairs or cloths of the user and
other persons involved do not touch the rotating
spindle.
The moving mechanism is made of stainless steel
and is permanently installed on the tank. It consists
of three arms with the following designations:
A = horizontal arm, fixed,
B = moving vertical arm and
C = moving horizontal arm with slider.
The moving mechanism is suitable for both horizon-
tal and vertical radiation. Holding devices for detec-
tors to move in the field are screwed to the slider of
the C-arm. The slider and the arms are moved by
stepper motors that permit a minimum step width of
0.1 mm. The reproducibility of the geometric posi-
tioning is also ± 0.1 mm.
The moving range limits can be adjusted for each
axis and stored in the control unit to ensure that the
detector does not hit the Perspex walls.
The components of the moving mechanism, particu-
larly the guides and moving parts, are not adversely
affected by water. Bending or contortion due to
water absorption or corrosion are thus excluded.
8
1
2
3
7
5
6
4

MP3-PL Therapy Beam Analyzer T41050
Operating Manual –The Components of the MP3-PL System
D850.131.05/01 en
19
1.3 Positioning Device
The positioning device allows the Perspex tank to
be aligned, using the laser system of the accelera-
tor and the lines on the walls of the tank.
The positioning device has three adjustable legs
that allow the height to be adjusted by ± 10 mm.
Additionally, the three adjusting screws allow the
tank to be adjusted horizontally by ± 10 mm, or by
approx. 1° on a vertical axis (micro-adjustment). A
mechanism allows the tank to be rotated by
45°or 90° (refer to Figure 6 for center of rotation).
Horizontal alignment of the tank is easy when it is
filled to the horizontal lines and the water surface is
used as a level.
Figure 6: Positioning device for tank alignment
1 Transportation lock
2 Center of rotation
3 Plate
4 Locking device
5 Adjusting screw (3x)
6 Washer 6.4 (4x)
7 Screw M6x16 (4x)
8 Adjustable leg (3x)
8
7
6
5
4
2
3
1

MP3-PL Therapy Beam Analyzer T41050
Operating Manual –The Components of the MP3-PL System
20
D850.131.05/01 en
1.4 SCANLIFT Carriage and
Reservoir
CAUTION
Improper handling.
Equipment damage!
If you are pumping water, you must not remove
the plastic hose between SCANLIFT and Per-
spex tank.
The positioning device is screwed together with the
chamber cable holding device and the Perspex tank
to the triangle holder of SCANLIFT (refer also to
section 2). For mounting and handling of SCANLIFT
please refer to the user manual of SCANLIFT.
1.5 Control Unit and Control
Pendant
The MP3-PL system is operated via the control unit
and the control pendant.
The control unit (TBA CONTROL UNIT ) is an
interface that connects the MP3-PL water phantom
to the computer. It supplies the power for the step-
per motors and controls the detector movement.
The control pendant (TBA CONTROL PENDANT)
can be connected either to the control unit or to the
junction box (at the Perspex tank). The control pen-
dant indicates the detector coordinates A, B and C
and allows the detector to be moved at two different
speeds (10 mm/s and 50 mm/s). Additionally, the
limits for the detector movement, the null point in
the central beam and the reference points can be
determined via the control pendant. Please note
that some of the control pendant functions are
available only when it is directly connected to the
junction box at the Perspex tank (refer to the user
manual of the control unit and the control pendant).
Therefore it is recommended to connect the control
pendant to the junction box.
NOTE
During adjustment work with the TBA CONTROL
PENDANT the TBA CONTROL UNIT can be
disturbed by a simultaneous communication
between the control unit and the COM-Server
via the RS232 interface or
with the control unit via the COM-Server via
the TCP/IP interface.
Disturbances are for instance that determining
the limits or switching the speed from FAST to
SLOW is not functioning properly.
In order to avoid disturbances during adjustment
work you have to disconnect the RS232 connec-
tion or the TCP/IP connection to the COM-
Server.
This manual suits for next models
1
Table of contents
Other PTW Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Pitney Bowes
Pitney Bowes DM475 series Operator's guide

T&D
T&D RTR-574 Introductory guide

Endress+Hauser
Endress+Hauser Proline Promass 84 Safety instructions

Vanguard Instruments
Vanguard Instruments CT-8000 S3 user manual

Endress+Hauser
Endress+Hauser Proline Prosonic Flow93C operating instructions

X-Rite
X-Rite Ci61 user guide