Pure Processing FlexiPump User manual

www.pure-processing.com Support: 877-718-6868 Manual FPIFS 0001/April 2015
FlexiPump™ Independent Flushing
System User Manual
Pure Processing, LLC
470 Mission Street, Unit 2
Carol Stream, Illinois 60188

© Pure Processing, Inc. 2015. All rights reserved. FlexiPump™Independent Flushing System User Manual; April 2015
2
TABLE OF CONTENTS
Page(s)
3
3
4
4
5
5
6
6
7
8
9
9
10
10
11
11
11
12
13
14
15
16
Section
OVERVIEW: FlexiPump™ Independent Flushing System
SYSTEM DESCRIPTION
OPERATION: Set-up with Nest Accessory
OPERATION: Set-up with Sink Rail Accessory
OPERATION: Set-up on a wall
OPERATION: Set-up on PureStation™ Sink Insert
OPERATION: Set-up tube sets
OPERATION: General guidelines for speed control
OPERATION: Connecting devices to the FlexiPump system
OPERATION: Connecting flexible endoscopes to the FlexiPump system
OPERATION: PSI Settings
OPERATION: General rules and guidelines for flexible endoscope speed
control OPERATION: FlexiPump Cycles
OPERATION: Keypad Operation
MAINTENANCE: Decontaminating the tube set
MAINTENANCE: Cleaning FlexiPump Surfaces
MAINTENANCE: Replacing the Tube Set
MAINTENANCE: Instructions for replacing the Tube Set
TROUBLESHOOTING
REPLACEMENT PARTS
REPLACEMENT ADAPTERS
ADDITIONAL ACCESSORIES
DAILY ASSURANCE LOG 17

© Pure Processing, Inc. 2015. All rights reserved. FlexiPump™Independent Flushing System User Manual; April 2015
3
OVERVIEW: FlexiPump™Independent Flushing System
The FlexiPump system is designed for use in a medical reprocessing area. It is engineered to provide optimal flushing
capacity, installation flexibility, and easy use and maintenance. It can be used independently or with a Pure Station™
Sink Insert, and has been fully tested for use with a wide variety of reusable lumened devices, including rigid and flexible
endoscopes, minimally invasive surgical devices, ocular and robotic devices, and suction tips.
The FlexiPump System is adaptable for counter, sink or wall installation near a water source. Accessories are available
from Pure Processing to adapt the system to multiple additional uses. The FlexiPump tube sets connect to most devices,
with or without luer locks and/or ports. A pressure relief feature is included to help protect scopes during flushing. The
FlexiPump system can be used alone or can augment existing pumps. Up to three devices can be flushed at once through
each FlexiPump unit.
DESCRIPTION
Components
Removable cap with side tabs
Pump housing
Electric cord
Key Pad
Impact Cycle button
Stop Cycle button
Forward Flush button
Speed Control button
Internal Components
Squeeze tube
Rotor
Supply tube (connects to medical devices)

© Pure Processing, Inc. 2015. All rights reserved. FlexiPump™Independent Flushing System User Manual; April 2015
4
OPERATION
Set-up: new FlexiPump Independent Flushing System
Your FlexiPump system does not come with a tube set installed. To install your first tube set, please see the
Maintenance section in this manual (page 12).
Set-up: with Nest Accessory
The Nest Accessory can be set up on any counter near a sink. Simply insert the FlexiPump system into the Nest with
tubes facing the labeled front of the accessory. Plug the power supply cord into an electrical outlet.
NOTE: Do not place pumps directly under a water supply.
Set-up: with Sink Rail Accessory
The Sink Rail accessory (sold separately) adheres to the top inside surface of a clean sink at the top edge. Once the rail is
attached, lock the back of the FlexiPump system onto the Sink Rail track, then plug the FlexiPump power supply cord
into a wall outlet.
Use extra caution with sink wall installations to avoid impact with hard surfaces.

© Pure Processing, Inc. 2015. All rights reserved. FlexiPump™Independent Flushing System User Manual; April 2015
5
OPERATION (continued)
Set-up: on a wall
The Nest Accessory can be mounted on a wall using the two through-holes in the back of the accessory. Simply hang the
nest on a wall near a sink and insert the FlexiPump unit. Then plug the cord into an outlet. (Hardware not included.)
Mounting holes
Set-up: on Pure Station™Medical Devices Pre-Cleaning System
The FlexiPump system can be inserted into the corner well of the PureStation Sink Insert, or along the side ridges
molded into the sink.
The FlexiPump unit is not meant to be submerged; use caution when installing near water.

© Pure Processing, Inc. 2015. All rights reserved. FlexiPump™Independent Flushing System User Manual; April 2015
6
OPERATION (continued)
Set-up: tube sets
The drop-down tube on each new tube set is 30 inches long. If the tube is too long for your sink, it can be cut
to shorten it. If you cannot position it so it is submerged under the cleaning solution, use a tissue clip or other
medical device to add weight and rest it at the bottom of the sink where solution is drawn from. A ceramic
weight can also be purchased from Pure Processing.
Each FlexiPump tube set comes with three (3) leads, each of which can be attached to an instrument. You
can use all three at once. If you have fewer than three devices to flush, you can cap off the unused lead(s) to
ensure appropriate pressure into the lead(s) with attached devices. Use the luer end cap provided with your
tube set. You may also re-use end caps from other FlexiPump tube sets.
General guidelines for speed control
Because fluid will take the path of least resistance, it is important that instrument channels attached to a
FlexiPump system are approximately the same diameter, to assure consistent flushing into all the devices.
DO NOT ATTACH A LARGE-DIAMETER LUMEN TO ONE LEAD AND A SMALL-DIAMETER LUMEN TO ANOTHER
ON THE SAME PUMP SYSTEM.
Low speed: this setting is designed for small channels, 1 to 3 millimeters (mm) in size. This setting
will flush 500 milliliters (ml) of fluid in one minute into all three channels (166 ml through each
lead). DO NOT TURN UP THE SPEED when you hear the motor slowing down. The smaller channels
cannot accommodate any higher flow volume. Trying to force it to go faster will cause back up in
the motor and will make the pressure release valve drain fluid.
Medium speed: This setting is designed for channels that are 2 to 3 mm in diameter. This setting
will flush 750 ml of fluid in one minute into all three channels (250 ml through each lead). DO NOT
TURN UP THE SPEED if you hear the motor slowing down. However, if there is fluid draining from
the pressure relief valve, you can consider slowing the pump speed.
High Speed: This setting will flush 1,000 ml of fluid in one minute into all three channels (333 ml
through each lead). This is used for channels larger than 4 mm in diameter.
Too much pressure and speed for small devices can result in tube set damage.

© Pure Processing, Inc. 2015. All rights reserved. FlexiPump™Independent Flushing System User Manual; April 2015
7
OPERATION (continued)
Connecting devices to the FlexiPump system
Many lumened/channeled devices can be properly connected and flushed with the FlexiPump system.
For devices with luer locks:
Twist the FlexiPump tube collar onto the device
with a luer, until locked
Flush the device following manufacturer’s
recommendations
For devices with ports:
Pull the tube collar back
Insert the slip tip into the device
Flush according to manufacturer’s
recommendations
For devices without luer locks or ports:
Use a cup adaptor to create a tight seal
Flush according to manufacturer’s
recommendations
For devices with female luer locks:
Use a female-to-female luer lock adapter
Flush according to manufacturer’s
recommendations

© Pure Processing, Inc. 2015. All rights reserved. FlexiPump™Independent Flushing System User Manual; April 2015
8
OPERATION (continued)
Connecting flexible endoscopes to the FlexiPump Independent Flushing System
It is important to follow the manufacturer’s instructions for use (IFU) for each flexible endoscope. The FlexiPump system
replaces the syringe flushing method in the pre-cleaning sections of these IFU.*
The injector tube (or suction channel tube in this example) is used to
inject reprocessing fluids into the instrument channel, suction
channel, and air/water channels of the endoscope. It is also used to
flush air through these channels to expel fluids.
Follow the device manufacturer’s IFU to attach the injector tube to
the scope.
Attach the FlexiPump tube set luer locks to the suction channel port
and air/water port of the injector tube. Directions may vary
depending on the scope and the manufacturer.
Cap off all unused tubes on the FlexiPump Tube Set. If there are two
channels on the endoscope, place a FlexiPump seal cap on the end of
the third FlexiPump tube lead. If you are flushing only one channel or
lumen, cap off the other two FlexiPump tube leads with luer end caps.
Select the appropriate flush function and speed (page 9) and flush the
device.
IMPORTANT:
Most endoscope manufacturers specify how much cleaning solution to flush through the device. Follow IFU's to
prepare the cleaning solution. Carefully monitor all device exit points for flowing cleaning solution to ensure copious
flushing.
Be sure to rinse devices per manufacturer’s IFU.
To eject solutions from a scope, remove the drop-down tube from the reprocessing fluid container. Allow pump to
run. This will inject air into the scope and purge fluid from the device. Then continue to next reprocessing step.

© Pure Processing, Inc. 2015. All rights reserved. FlexiPump™Independent Flushing System User Manual; April 2015
9
OPERATION (continued)
Flushing flexible endoscopes with the FlexiPump Independent Flushing System
PSI Settings
Please refer to the manufacturer’s IFU for each endoscope’s pressure requirements. The FlexiPump system
can be used with a low-pressure tube set or a higher-pressure tube set. In general, smaller flexible devices
(urology scopes and bronchoscopes, for example) require lower pressure.
General rules and guidelines for flexible endoscope speed control
There are three (3) tubes on each FlexiPump tube set. IF ANY TUBES ARE NOT IN USE, BE SURE THEY ARE
SECURED WITH END CAPS PROVIDED IN THE TUBE SET.
When using all three tubes of the tube set for flushing, the MEDIUM or HIGH FlexiPump speed should
be used.
When using two tubes, make sure the third tube is capped and the FlexiPump is set for LOW, MEDIUM
or (in some cases) HIGH speed.
When using only one of the three tubes of the FlexiPump tube set, cap off the other two tubes and set
the FlexiPump speed to LOW, MEDIUM or (in rare cases) HIGH speed.
If the pump sounds or looks stressed, turn the speed setting down to avoid unnecessary
wear on the motor.
PLEASE NOTE: the FlexiPump Independent Flushing System is intended for pre-cleaning and flushing of lumened devices.
It is not for final treatment and does not replace manual brushing and rinsing. Refer to each medical device
manufacturer’s instructions for use for complete pre-cleaning instructions.

© Pure Processing, Inc. 2015. All rights reserved. FlexiPump™Independent Flushing System User Manual; April 2015
10
Too much pressure and speed can cause tube set damage.
Key Pad Operation
Impact: press for clogged devices. Press the button ONCE for two (2)
minutes, TWICE for four (4) minutes, and THREE TIMES for six (6)
minutes.
STOP: press to stop all pump action and cancel any cycle
Forward Flush: press the button ONCE for 30 seconds, TWICE for 60
seconds and THREE TIMES for 90 seconds
Speed Control: Push the button ONCE for low (500 ml per minute),
TWICE for medium (750 ml per minute), and THREE TIMES for high
speed (1,000 ml per minute)
The Forward flush cycle is used for flushing channels. There are three options for this cycle: 30 seconds,
60 seconds and 90 seconds.
The Impact cycle is used ONLY for devices that are clogged. The FlexiPump tube set must be
decontaminated after each Impact Cycle run.
The Speed Indicator shows one, two or three lights, to indicate the pump speed selected.

© Pure Processing, Inc. 2015. All rights reserved. FlexiPump™Independent Flushing System User Manual; April 2015
11
MAINTENANCE
Decontaminating the tube set
Disinfect the tube sets at least daily, preferably at every shift change.
Step 1. Pour ready-to-use disinfectant into an open, clean container.
Step 2. Submerge the drop-down tube in the container with disinfectant.
Step 3. Set speed control on Low (1 LED light).
Step 4. Press the START button to begin fluid flow.
Step 5. Press STOP when the tube set is completely full of disinfectant.
Step 6. You may need to place end cap(s) on the tube(s) to hold the disinfectant inside the tube set for the
full
contact time.
Step 7. Disinfectant should remain inside each tube for five (5) minutes.
Step 8. Fill a cleaning solution caddy with clean water.
Step 9. Place drop-down tube into the water and press START to rinse out the tubing.
Cleaning FlexiPump surfaces
The surfaces of the system should be cleaned daily and whenever there is visible soil.
Step 1. Wipe all external surfaces with a ready to use disinfectant, including the tubing, and
allow them to air-dry.
Step 2. Wipe the surfaces of the Nest accessory and/or Rail accessory (if used) and let them air-dry.
Replacing the tube set
There are several important reasons why it is necessary to replace the tube set every month:
1. Infection prevention: hard water deposits could potentially build up on inside surfaces of the tube set
2. Reduced flow rate: the squeeze tube can lose its elasticity with time and use
3. Reduced pump life: the factory-applied tube set lubricant eventually wears away, resulting in more resistance
and more work for the pump. This can reduce the pump’s overall useful life
Bleach or alcohol disinfectants are not compatible with the FlexiPump; it must be water based.

© Pure Processing, Inc. 2015. All rights reserved. FlexiPump™Independent Flushing System User Manual; April 2015
12
MAINTENANCE (continued)
Instructions for replacing the tube set:
1. Remove the Pump Cap by pressing on both side tabs
and pulling gently.
2. Unthread the tubing from around the squeeze tube
rotor.
3. Thread the new tube set around the rotor, making
sure it is properly aligned and the tube fittings are
properly situated in the cap fittings.
4. Place the cap back onto the base, making sure it snaps
down fully.
Flow validation and decontamination records:
These daily activities should be recorded and maintained as part of the department’s quality system. A daily
log sheet template is included in this manual (page 17) for convenient recording of daily flow and
decontamination testing.

© Pure Processing, Inc. 2015. All rights reserved. FlexiPump™Independent Flushing System User Manual; April 2015
13
TROUBLESHOOTING
Problem
Recommendation
Notes
1. No power to unit
Disconnect the black and gray cords at the barrel connector. Reconnect
making sure they are tight. If there is no power, slightly turn the adapter
to engage the two and tighten. Repeat as necessary.
2. Power to unit but
buttons don’t work
Disconnect the black and gray cords at the barrel connector. Reconnect
making sure they are tight. If there is no power, slightly turn the adapter
to engage the two and tighten. Repeat as necessary.
3. Pump is not working
Disconnect the power cord from outlet and from the FlexiPump unit.
Reconnect the cord and plug in. Press a control button to see if it works
If it still doesn’t work, go to #4
4. Pump does not work
Reset the unit by hitting stop, and then press the Forward Flush 30-
second button.
If it still does not operate,
contact Pure Processing
Customer Service.
5. Rotor is spinning but no
solution is coming out
the end of the device
Make sure the drop-down tube is fully submerged. Also check for kinks in
the tubing. Make sure you have the correct end of the tubing in the
solution.
6. Luer lock is broken or
separated from the
tubing
Change the tube set.
7. Devices are not clean
Check the device manufacturer’s cleaning recommendations. Check the
disinfectant instructions for use. Confirm that all directions have been
properly followed.
8. Impact Cycle is not
working
Impact Cycle works in conjunction with detergent. Ensure that proper
detergent is used for the type of soil in the device (protein-based enzyme
for blood, e.g.).
If still not working, see #7
9. Impact Cycle is not
working
Conduct a flow validation test:
Place drop-down tube into container of clean water
Press the Forward Flush button once for a 30-second flush
When a steady stream of water exits the connector, press the Stop
button
10. Water is leaking from
under the cap
Check the tube set for cracks
Change the tube set if cracked
11. One or more
components on the tube
set are broken
Replace the entire tube set
12. Unit is not running and I
have a red blinking light
Allow the unit to cool down for 30 minutes, and then resume use.
Contact us if this does not remedy the issue.
This can occur after prolonged
use.
13. Speed control is stuck on
HIGH
Allow the unit to reset for 15 seconds. Press the speed
control to test.
14. Unit makes lots of noise
Use a surface liner under the nest to reduce vibration
15. Unit stopped while
flushing
Unplug the unit to reset it
Repeat interrupted flushing cycle
16. Pump cap will not stay
locked to the base
Replace the cap
Probably cracked or broken
where locks engage
If it still does not operate,
contact Pure Processing
Customer Service.

© Pure Processing, Inc. 2015. All rights reserved. FlexiPump™Independent Flushing System User Manual; April 2015
14
FLEXIPUMP REPLACEMENT PARTS
Item
Product Code
FlexiPump™ Irrigation Tube
Set Two-Lead, 15 psi or 30 psi
FXITS-0215psi6
FlexiPump™ Irrigation Tube
Set Three-Lead, 15 psi or 30
psi FXITS-0330psi6
FlexiPump™ Power Supply
FPPS-01
FlexiPump™ Pump Cap
FLX-CAP-01
FlexiPump™ Pump Rotor
FLX-ROTOR-01
FlexiPump™ Nest Accessory
FLX-NEST
FlexiPump™ Sink Well Adapter
FPSWA-01
FXITS-0230psi6
FXITS-0315psi6

© Pure Processing, Inc. 2015. All rights reserved. FlexiPump™Independent Flushing System User Manual; April 2015
15
REPLACEMENT ADAPTERS
Item
Product Code
End Cap (to seal FlexiPump tube)
PSFP-LLSC-10
Female-to-Female Luer Lock
FFLL-10
FFLL-30
FFLL-50
Large Rubber sealing bridge
Rblg-01
Rubber Cup Adapter
RCA-01
Small Sealing Bridge
Rbsm-01
XL Slip Tip Adapter
XLSLIP-10
XLSLIP-20
XLSLIP-30

© Pure Processing, Inc. 2015. All rights reserved. FlexiPump™Independent Flushing System User Manual; April 2015
16
ADDITIONAL ACCESSORIES
These products are available from Pure Processing, Inc. and are compatible for use with the FlexiPump system:
Item
Product Code
PureChannel™ Flush Instrument Channel Cleaner
•1 gallon size
•32 oz. size PSCF-32oz
Pro-Spray™ Ready-to-Use Disinfectant/Cleaner, 24 oz.
PSC240-1
Pro-Spray™ Ready-to-Use Disinfectant/Cleaner, 1 GAL.
PSC128-1
PureStation™ Sink Insert
PSFP1616-EYE
PSFP1626-EYE
PureStation™ Sink Insert with FlexiPump(TM) Independent Flushing
System (large)
PS1626FP
PureStation™ Sink Insert (large)
PS1626
PureStation™ Sink Insert with FlexiPump(TM) Independent Flushing
System (small)
PS1616FP
PureStation™ Sink Insert (Small)
PS1616
PSCF-08

© Pure Processing, Inc. 2015. All rights reserved. FlexiPump™Independent Flushing System User Manual; April 2015
17
Day
Decontamination Procedure
Initials
1
Yes
No
2
Yes
No
3
Yes
No
4
Yes
No
5
Yes
No
6
Yes
No
7
Yes
No
8
Yes
No
9
Yes
No
10
Yes
No
11
Yes
No
12
Yes
No
13
Yes
No
14
Yes
No
15
Yes
No
16
Yes
No
17
Yes
No
18
Yes
No
19
Yes
No
20
Yes
No
21
Yes
No
22
Yes
No
23
Yes
No
24
Yes
No
25
Yes
No
26
Yes
No
27
Yes
No
28
Yes
No
29
Yes
No
30
Yes
No
31
Yes
No
Month and Year
Unit Number
FlexiPump™ Independent Flushing System
Daily Assurance Log
Tube Replacement Name: Date:
Day
Flow Validation Test
Initials
1
Pass
Fail
2
Pass
Fail
3
Pass
Fail
4
Pass
Fail
5
Pass
Fail
6
Pass
Fail
7
Pass
Fail
8
Pass
Fail
9
Pass
Fail
10
Pass
Fail
11
Pass
Fail
12
Pass
Fail
13
Pass
Fail
14
Pass
Fail
15
Pass
Fail
16
Pass
Fail
17
Pass
Fail
18
Pass
Fail
19
Pass
Fail
20
Pass
Fail
21
Pass
Fail
22
Pass
Fail
23
Pass
Fail
24
Pass
Fail
25
Pass
Fail
26
Pass
Fail
27
Pass
Fail
28
Pass
Fail
29
Pass
Fail
30
Pass
Fail
31
Pass
Fail
Pure Processing Customer Support 1-877-718-6868
Other manuals for FlexiPump
1
Table of contents
Other Pure Processing Medical Equipment manuals
Popular Medical Equipment manuals by other brands

BMB MEDICAL
BMB MEDICAL TIM VELVET SWANN user manual

Invacare
Invacare Birdie EVO Series user manual

Mercado Medic
Mercado Medic REAL 8200 PLUS EL 24V SCREENING Instructions for use

Medisana
Medisana HS 680 instruction manual

Dräger Medical
Dräger Medical Inhalog 1 plus Instructions for use

Henry Schein
Henry Schein COLPOSCOPE I user manual