RADWAG MW-01 Series User manual

User manual
Weighing module MW-01
Manual number:
ITKU-29-05-08-14-A
BALANCES AND SCALES
RADWAG Wagi Elektroniczne, 26–600 Radom Bracka 28 Street - POLAND
Phone +48 48 38 48 800, phone/fax. +48 48 385 00 10
Selling department +48 48 366 80 06
www.radwag.com

2
AUGUST 2014

3
CONTENT
1. APPLICATION............................................................................................................................5
2. PRECAUTIONARY MEASURES................................................................................................5
3. WARRANTY CONDITIONS........................................................................................................5
4. TECHNICAL PARAMETERS .....................................................................................................6
4.1.
5. CONSTRUCTION .......................................................................................................................7
INPUT/OUTPUT parameters ................................................................................................7
5.1. External view.........................................................................................................................8
5.2. Dimensions ...........................................................................................................................8
5.3. Interfaces ..............................................................................................................................8
5.3.1.
5.4. Diagrams of connection cables...........................................................................................10
PROFIBUS ..................................................................................................................9
6. INSTALLATION of “MwMANAGER” software.......................................................................12
6.1. Minimal hardware requirements..........................................................................................12
6.2. Installation procedure..........................................................................................................13
7. PC PROGRAM DESCRIPTION................................................................................................16
7.1. Weighing window ................................................................................................................16
7.2. Application settings .............................................................................................................17
7.2.1. Connection settings...................................................................................................17
7.2.2. Language ..................................................................................................................19
7.2.3.
7.3. Parameters .........................................................................................................................20
Other .........................................................................................................................20
7.3.1. User Parameters .......................................................................................................21
7.3.2. Communication settings ............................................................................................22
7.3.3. Functions I/O .............................................................................................................24
7.4. Functions ............................................................................................................................26
7.4.1. Dosing .......................................................................................................................26
7.4.2. Checkweighing ..........................................................................................................30
7.4.3. Input/output state.......................................................................................................31
8. WEIGHING................................................................................................................................32
8.1. .......................................................................................................................32Terms of use
8.2. Zeroing................................................................................................................................33
8.3. Tarring.................................................................................................................................34
8.4. Weighing on dual range scales...........................................................................................34
8.5. Toggling between weight units............................................................................................35
9. SC LE PARAMETERS............................................................................................................36A
9.1. Autozero function ................................................................................................................36
9.2. Median filter ........................................................................................................................37
9.3. Filter ....................................................................................................................................37
10. CHECKWEIGHING.................................................................................................................38
10.1. limit..............................................................................................................................38LO
10.2. MIN/MAX limit ...................................................................................................................38
11. DOSING ..................................................................................................................................39
12. PARAMETERS IN FILE..........................................................................................................40
12.1. Saving to file......................................................................................................................41
1 .2. Uploading file data ............................................................................................................422
13. OFFLINE MODE .....................................................................................................................44

4
14. COMMUNICATION PROTOCOL............................................................................................45
14.1. General information ..........................................................................................................45
14.2. A set of commands ...........................................................................................................45
14.3. Respond message format.................................................................................................46
14.4. Command’s description ....................................................................................................46
14.4.1. Zeroing....................................................................................................................46
14.4.2. Tarring.....................................................................................................................47
14.4.3. Get tare value .........................................................................................................47
14.4.4. Set tare value..........................................................................................................47
14.4.5. esult in basic unit .........................................................................48Send the stable r
14.4.6. Send the result immediately in basic unit ...............................................................48
14.4.7. Send the stable result in current unit ......................................................................49
14.4.8. Send the result immediately in current unit.............................................................49
14.4.9. Switch on continuous transmission in basic unit.....................................................50
14.4.10. Switch off continuous transmission in basic unit...................................................50
14.4.11. Switch on continuous transmission in current unit................................................50
14.4.12. Switch off continuous transmission in current unit................................................51
14.4.13. Set lower threshold ...............................................................................................51
14.4.14. Set upper threshold...............................................................................................51
14.4.15. Read lower threshold ............................................................................................52
14.4.16. Read upper threshold ...........................................................................................52
14.4.17. Send all implemented commands.........................................................................52
15. COMUNICATION MODULE PROFIBUS................................................................................53
15.1. General information ..........................................................................................................53
15.2. work ..........................................................53Setting instrument’s address in a Profibus net
15.3. Memory map .....................................................................................................................54
15.3.1. Output address .......................................................................................................54
15.3.2. Input address ..........................................................................................................55
15.4. Description of variables.....................................................................................................55
15.4.1. riables......................................................................................................55Output va
15.4.2. Input variables ........................................................................................................57
16. MODBUS COMMUNICATION PROTOCOL...........................................................................60
16.1. Basic information ..............................................................................................................60
16.2. Addressing a device in ModBus network ..........................................................................61
16.3. Register tables in ModBus network ..................................................................................61
16.3.1. Table of output registers .........................................................................................62
16.3.2. Table of input registers ...........................................................................................63
16.4. Software for controller PLC SIEMENS S7-300 .................................................................63
17. ERROR MESSAGES..............................................................................................................64

5
1. APPLICATION
MW-01 weighing module is intended for conftruction of industrial load cell
scales. Depending on the needs, communication with the module may be
established by the following interfaces: RS232, RS485, Ethernet, Profibus
and ModBus protocol. MW-01 Module may cooperate with PUE 5, PUE 7
terminals or PC computer.
Operating of the weighing module MW-01 from a PC level is carried out with
a computer software “MwManager”. A detailed description of the application
is included in the further section of the MW-01 user manual.
2. PRECAUTIONARY MEASURES
A. Please, read carefully this user manual and use the device according
to its intended use;
B. Weighed loads should be placed possibly in the central setion of the
weighing platform;
C. The weighng platform should be loaded with objects which gross
weight does not exceed the maximal capacity of the scale;
D. Do not leave heavy loads on the weoghing patform for longer period
of time;
E. In case of failure, immediately disconnect the power supply;
F. Devices that are to be decommissioned should be utilized according
to valid legal regulations.
3. WARRANTY CONDITIONS
A. RADWAG is obliged to repair or change those elements that appear
to be faulty because of production and construction reason,
B. Defining defects of unclear origin and outlining methods of their
elimination can be carried out only in participation of the user and
the manufacturer representatives,
C. RADWAG does not take any responsibility connected with defecting
or loss deriving from unauthorized or inappropriate (not adequate
to manuals) production or service processes,

6
D. Warranty does not cover:
•Mechanical failures caused by inappropriate exploitation
of the device or failures of thermal or chemical origin or
•Defects caused by atmospheric discharge, overvoltage in mains
or other random event,
•Maintenance activities (cleaning of the weighing module).
E. Warranty loss appears if:
•A repair is carried out by an unauthorized service,
•Intrusion into mechanical or electronic construction
of unauthorized people,
•Removing or destroying protection stickers for the weighing module
F. The detailed warranty conditions are listed in the warranty certificate.
G. Contact with Authorized Service: +48 48 384 88 00 ext. 106 or 107.
4. TECHNICAL PARAMETERS
MODEL MW-01 ver.1 MW-01 ver.2
Interface RS232, RS485,
Ethernet RS232, PROFIBUS
Power Supply 8 ÷ 30VDC
Housing metal
IP rating IP 66
Working temperature -10°C ÷ 40°C
OIML class III
Number of verification intervals 10 000
Max. increase of input signal 19,5mV
Min. voltage per verification division 0,5µV
Min. load cell impedance 80Ω
Max. load cell impedance 1200Ω
Load cell excitation voltage 5VDC
Load cell connectivity 4 or 6 wire shielded cable
Number of weighing platforms 1
2INE/3OUT Standard
Transmission Speed RS232 RS485 Default 57600 bit/s

7
4.1. INPUT/OUTPUT parameters
The module features 2 optoisolated inputs, 2 reed relay outputs
and fast transistorized output type open collector.
OUT1 – OUT2 Parameters
Number of outputs 2
Type of outputs Reed operating contact
Conductor intersection 0.14 – 0.5mm2
Maximum switchable current 200mA
Maximum conducted voltage 50VDC
Output OUT3 Parameters
Number of outputs 1
Type of outputs Transistor
Umax 5VDC
Imax 100mA
Input Parameters
Number of inputs 2
Type of inputs Optocoupled
Conductor intersection 0.14 – 0.5mm2
Control voltage range 5 -24V DC
5. CONSTRUCTION
Weighing module comprises metal housing. Signal cables are fed through
cable glands. Communication with the peripheral devices is established by
one of the following interfaces: RS485, RS232, Ethernet, Profibus.
The weighing module enables cooperation with indicators or PC computers.
It features 2 optoisolated inputs, 2 reed relay outputs and one fast transistor
output. The weighing module is supplied form 8÷30VDC.

5.1. External view
Fig.1 Weighing module MW-01
5.2. Dimensions
Fig.2 Dimensions of the MW-01
5.3. Interfaces
MW-01 ver.1 has cables fed through cable glands.
MW-01 ver.2 has 2 sockets (PROFIBUS IN and PROFIBUS OUT),
RS232 is fed though cable gland.
8

Fig.3 Location and marking of interfaces for MW-01 main board
5.3.1. PROFIBUS
MW-01 module, version 2, is equipped with input and output socket for PROFIBUS.
The volatage, indispensable for correct operation of a terminator is 5VDC on an output
socket. Available socket standard is M12 5 pin B coded (for DP PROFIBUS).
Sockets’ topology:
PROFIBUS IN
(male)
Pin1 – NC
Pin2 – A
Pin3 – NC
Pin4 – B
Pin5 – NC
PROFIBUS OUT
(female)
Pin1 - +5V
Pin2 – A
Pin3 – GND
Pin4 – B
Pin5 – NC
9

5.4. Diagrams of connection cables
RS232 cable
Fig. 4 PT0145 Cable
RS485 cable
Fig. 5 PT0012 Cable
10

Ethernet cable
Fig. 6 PT0014 Cable
IN/OUT cable
Fig. 7 PT0146 Cable
11

12
6.INSTALLATION of “MwMANAGER” software
Caution:
•In order to install the program on the computer with older version
“MwManager” softwre, uninstall the previous version first,
•The installation instruction was made for Windows 7 and it is
compatibile with former versions of MS Windows
•Proper operation of the program reguires installing Microsoft .NET
Framework ver. 2.0 or higher. It is accessible on Microsoft website:
http://www.microsoft.com/downloads/details.aspx?displaylang=pl&
FamilyID=0856eacb-4362-4b0d-8edd-aab15c5e04f5
•Proper operation of the program reguires installing all accessible
ServicePacks for your operating system.
•Due to continuous software updates there may occur some
discrepancies between the instruction and the program,
•RADWAG takes no responsibility for the consequences of the
software operation, and any errors resulting from the incorrect
software operation,
•RADWAG takes no responsibility for loss or protection of any data
caused by the incorrect utilization of the software or computer.
6.1. Minimal hardware requirements
Proper operation of the software requires a computer with below specified
configuration:
•A PC computer with installed OS MS Windows 2000/XP/2003/Vista/7,
•processor 2.4 GHz or faster,
•RAM 512 MB or more (recommended 1 GB),
•At least 1 GB of free space on the hard drive,
•Display with minimum resolution 800 x 600,
•DVD drive.

6.2. Installation procedure
1. On receiving an installation file of MwManager x.x.x.x.exe software,
run installation file, select language version and press OK key.
2. Press in the welcome window:
13

3. Browse window:
If necessary, change the software insallation folder, and press
.
4. Window for selecting tasks:
On marking / unmarking available option, press key.
14

5. Installation window:
In order to install the software, press „Install” key.
6. Software installation completion window:
To finish installation process, press „Finish” key.
15

7. A shortcut will be created on computer’s desktop.
7. PC PROGRAM DESCRIPTION
Operation of the weighing module from PC computer level requires
application ”MwManager”. The application can operate in MS Windows
environment with installed add-on .NET framework 2.0. The program
enables calibrationg the module, displaying current mass indication, tarring,
zeroing, setting filters and functions for corresponding inputs and outputs.
Caution:
1. This manual complies with the software “MwManager” starting from
version 1.0.2.1 and weighing module program Prj215 from version 2.2.
2. The entered values are accepted by pressing Enter key. The changes
are saved in the weighing module on pressing Save key. All temporary
parameters settings that are not saved in the weighing module are
highlighte in red.
7.1. Weighing window
Fig.8 View of the software’s weighing window
On completing initializating procedure, the weighing window displays below
pictograms:
- precise zero indication
- stable measurement result
- weighing unit
- weighing platform number
16

If the indication is other than zero – press the zeroing key.
Function keys:
- Zeroing
- Taring
7.2. Application settings
Tab comprises the settings for conneting the weighing
module, choosing interface language version and other software options.
7.2.1. Connection settings
In tab , key opens a window for establishing
connection with the weighing module.
Fig.9 Connection settings’ window
17

18
Description:
Means of connection Interface for connecting to the weighing module
RS 232 Connection via port RS232
TCP/IP Ethernet connection
RS 485 RS 485 connection
Offline The offline mode is used for saving and editing all
indispensable parameters of the configuration file.
RS232:
Port Choosing a COM port number to which the weighing
module is plugged
Baud rate Baud rate for the RS232 interface. 57600 bps by default
Parity Parity bit parameter. “None” by default
Data bits Number of data bits. 8 by default
Stop bits Number of stop bits. 1 by default
TCP/IP:
Adres IP IP address of the device, default setting 192.168.0.2
Port Port set in the weight module, default setting 4001.
RS485:
Port Choosing a COM port number to which the weighing module is plugged
Baud rate Baud rate for the RS485 interface. 57600 bps by default
Address Weighing module address in the net
Caution:
1. The default baud rate of RS232 and RS485 interfaces in the MW-01
is set to 57600 bps;
2. If case of problems with establishing a connection to the weighing
module, set and check accessible baud rates or connect through
Ehternet interface;
3. If the display is set to port RS232, then connecting the computer software
through the same port is impossible. Display operation on this port is
restored on restarting the module (unplugging and repeated plugging
to mains).

Keys description:
Disconnect Communication disconnected
Reconnect Switching off and on (communication restart) communication with the device
Connect Establishing connection with the weighing module
7.2.2. Language
Use tab , and key to open a window enabling
changing the language of the software.
Fig.10 Language selection window
On selecting the language version, press “Apply” key to save the changes.
Accessible language versions:
•Czech
•English
•French
•German
•Polish
•Spanish
19

7.2.3. Other
Use tab , key to start other software options.
Fig.11 Other options window
Tick “autostart” option, so that on switching on the software automatically
connects to the weighing module, by the default of last activated means of
connection.
7.3. Parameters
Tab comprises all factory and user parameters, and also the
communication parameters of the weighig module and functions of 2 inputs
and 3 outputs.
20
Other manuals for MW-01 Series
2
This manual suits for next models
2
Table of contents
Other RADWAG Control Unit manuals
Popular Control Unit manuals by other brands

BroadMobi
BroadMobi BM818 Operation manual

GEM
GEM 740 Installation, operating and maintenance instructions

PLARAD
PLARAD X32 Series Translation of original operating manual

schmersal
schmersal SRB-E-232ST operating instructions

Apacer Technology
Apacer Technology EFC-D Series user manual

Barth
Barth Mini-PLC STG-850 manual