DissoPrep X8
1. regular cleaning
A periodical cleaning of the DissoPrep X8 has be be arranged. The intervals of the
cleaning depend on the following individual factors:
• the kind of used media (especially if using foaming media the cleaning interval should
be arranged shorter)
• the frequency of using the DissoPrep X8
• the whole volume of media, which is prepared with the DissoPrep X8
By a regular cleaning the growth of bacteria is avoided.
For cleaning can be used DI-water in the AUTOWASH-function. To prevent possible
bacterial growth, it´s possible to use very effective 0,2% sodium-hypo-chlorid or more
compatible 5% hydrogen-peroxid as washing liquid. This cleaning is harmless fot the
RIGGTEK – Dissolution Test Systems maintenance of DissoPrep X8 06/2011 3
RIGGTEK recommends in general the weekly cleaning with DI-water and the
AUTOWASH-function. Only if a bacterial growth is very probably, the monthly cleaning
with 5 liters of 5% hydrogen-peroxid as inlet media
• for one complete run (PREFILL, FILL, DISPENSE, EMPTYING) at 37°C or
• for theAUTOWASH-function with 2 cleaning-cycles with each 2500ml cleaning volume.
For rinsing RIGGTEK recommends to use 5l DI-water for an additional complete run at
37°C or for an additional use of the AUTOWASH-function with 2 cl eaning-cycles with
each 2500ml cleaning volume.
filter at the inlet channel and for the transparent PVC-tubings, which only may become a
little bit milky. For rinsing DI-water is recommended.
The cleaning can be done with
1. two complete runs (PREFILL, FILL, DISPENSE, EMPTYING). In the first run with the
cleaning liquid in the inlet media. In the second run with DI-water in the inlet media for
rinsing.
2. or with the AUTOWASH-function (if available).
(see also at DPX8-Operation-manual Version > 7.111, chapter 10)