manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. RKI Instruments
  6. •
  7. Measuring Instrument
  8. •
  9. RKI Instruments OX-500 Configuration guide

RKI Instruments OX-500 Configuration guide

Manual Supplement
OX-500
P. 1 of 3
Released: 1/25/13
www.rkiinstruments.com
This supplement includes calibration instructions and a spare parts list for the OX-500.
Calibrating the OX-500 in Maintenance Mode
Preparing for Calibration
NOTE: The following procedure assumes the use of a calibration kit which includes a
100% nitrogen calibration gas cylinder, a zero air cylinder, a 2.0 LPM fixed flow
regulator with an on/off knob, a calibration cup, and a short piece of sample
tubing to connect the regulator to the calibration cup.
NOTE: The following procedure assumes that the OX-500 is not in a fresh air
environment. If a fresh air environment can be verified, applying zero air to the
detector is not necessary when setting the span adjustment.
1. Connect the calibration kit sample tubing to the calibration cup hose barb.
2. Connect the fixed flow regulator to the calibration cup using the sample tubing. Be
sure that you are using a 2.0 LPM fixed flow regulator.
WARNING: Failure to use the recommended gas flow rate will result in an inaccurate
reading.
Adjusting the Zero Reading
1. While in Measuring Mode, press and hold the MAINTE Switch and the MODE Switch
simultaneously for 3 seconds to enter Maintenance Mode. Release the switches when
the following screen appears.
NOTE: The MAINTE Switch is a small recessed push button switch located to the left of
the MODE Switch. It is labelled on page 5 of the attached OX-500 manual. A tool
such as a small tipped screwdriver or a pencil may be used to press the MAINTE
Switch.
2. Press and release the MODE Switch to proceed to item 2-2, Zero Adjustment Mode.
3. Press and release the RESET switch to select item 2-2. The display will indicate the
current gas reading.
4. Screw the 100% nitrogen cylinder into the fixed flow regulator.
5. Turn the regulator knob counterclockwise to open the regulator.
6. Hold the calibration cup against the case over the diffusion grill that covers the sensor
face in the lower left corner of the front.
7. Allow 100% nitrogen to flow for two minutes. The reading should decrease and
stabilize after two minutes.
2 - 1
Manual Supplement
OX-500
P. 2 of 3
Released: 1/25/13
www.rkiinstruments.com
8. Press and release the MODE switch to perform the zero adjustment. The reading will
momentarily flash and the OX-500 will perform a zero adjustment.
9. If the zero adjustment is successful, the buzzer will sound a double tone and the
display reading will be set to 0.0 %.
If the zero adjustment is not successful, the LCD will indicate the error message Err.
10. Remove the calibration cup from the front face of the OX-500.
11. Turn the regulator knob clockwise to close the regulator.
12. Unscrew the 100% nitrogen cylinder from the fixed flow regulator.
13. Press and release the RESET Switch to exit Zero Adjustment Mode and return to the
screen that indicates item 2-2.
Adjusting the Span Setting
1. Press and release the MODE Switch to proceed to item 2-3, Span Adjustment Mode.
2. Press and release the RESET Switch to select item 2-3. The display will indicate the
current gas reading.
NOTE: If you can confirm a fresh air environment, you do not need to apply zero air to
set the span.
3. Screw the zero air calibration gas cylinder into the fixed flow regulator.
4. Turn the regulator knob counterclockwise to open the regulator.
5. Hold the calibration cup against the case over the diffusion grill that covers the sensor
face in the lower left corner of the front.
6. Allow zero air calibration gas to flow for two minutes.
7. Adjust the gas reading so that it matches the calibration gas concentration using the
MAINTE Switch. The MAINTE Switch will only increase the gas reading. When the
reading reaches the maximum, it will “wrap around” and begin increasing from zero.
8. Press and release the MODE Switch to accept the span setting. The gas reading will
momentarily flash and the span setting will be saved.
9. If the span adjustment is successful, the buzzer will sound a double tone and the gas
reading will be displayed.
If the span adjustment is not successful, the display will indicate Err.
10. Remove the calibration cup from the front face of the OX-500.
11. Turn the regulator knob clockwise to close the regulator.
12. Unscrew the zero air calibration gas cylinder from the fixed flow regulator.
13. Press and release the RESET Switch to exit Span Adjustment Mode and return to the
screen that indicates item 2-3.
14. Press and hold the MAINTE Switch for 3 seconds to exit Maintenance Mode and
return to Measuring Mode.
15. Store the components of the calibration kit in a safe place.
Manual Supplement
OX-500
P. 3 of 3
Released: 1/25/13
www.rkiinstruments.com
Spare Parts List
Table 1 below lists spare parts for the OX-500.
Table 1: Spare Parts List
Part Number Description
06-1248RK-03 Calibration tubing, 3/16” ID x 5/16” OD, 3 foot length
81-OX500-LV Calibration kit, 2 LPM regulator, 100% nitrogen, 34 liter
81-0076RK Calibration cylinder, zero air, 17 liter
81-0076RK-01 Calibration cylinder, zero air, 34 liter
81-0078RK Calibration cylinder, 100% nitrogen, 17 liter
81-0078RK-01 Calibration cylinder, 100% nitrogen, 34 liter
81-1059RK-20 Regulator, with gauge and knob, 17L/34L, 2.0 LPM
81-1113RKP Calibration Cup
OS-B11 Oxygen replacement sensor

Other RKI Instruments Measuring Instrument manuals

RKI Instruments CO-04 User manual

RKI Instruments

RKI Instruments CO-04 User manual

RKI Instruments GX-6000 User manual

RKI Instruments

RKI Instruments GX-6000 User manual

RKI Instruments GX-8000 User manual

RKI Instruments

RKI Instruments GX-8000 User manual

RKI Instruments GX-2001 User manual

RKI Instruments

RKI Instruments GX-2001 User manual

RKI Instruments GD-K77D User manual

RKI Instruments

RKI Instruments GD-K77D User manual

RKI Instruments beacon 410 User manual

RKI Instruments

RKI Instruments beacon 410 User manual

RKI Instruments 03 Series User manual

RKI Instruments

RKI Instruments 03 Series User manual

RKI Instruments GX-3R User manual

RKI Instruments

RKI Instruments GX-3R User manual

RKI Instruments 01-Series User manual

RKI Instruments

RKI Instruments 01-Series User manual

RKI Instruments RP-3R User manual

RKI Instruments

RKI Instruments RP-3R User manual

RKI Instruments GasWatch 2 User manual

RKI Instruments

RKI Instruments GasWatch 2 User manual

RKI Instruments GX-2003 User manual

RKI Instruments

RKI Instruments GX-2003 User manual

RKI Instruments beacon 410 User manual

RKI Instruments

RKI Instruments beacon 410 User manual

RKI Instruments CX-04 User manual

RKI Instruments

RKI Instruments CX-04 User manual

RKI Instruments GX-6000 User manual

RKI Instruments

RKI Instruments GX-6000 User manual

RKI Instruments GP-03 User manual

RKI Instruments

RKI Instruments GP-03 User manual

RKI Instruments 65-2322RKSS User manual

RKI Instruments

RKI Instruments 65-2322RKSS User manual

RKI Instruments GX-3R User manual

RKI Instruments

RKI Instruments GX-3R User manual

RKI Instruments GP-03 User manual

RKI Instruments

RKI Instruments GP-03 User manual

RKI Instruments EAGLE 2 User manual

RKI Instruments

RKI Instruments EAGLE 2 User manual

RKI Instruments PS-2 User manual

RKI Instruments

RKI Instruments PS-2 User manual

RKI Instruments 72-RA-C User manual

RKI Instruments

RKI Instruments 72-RA-C User manual

RKI Instruments CO-04C User manual

RKI Instruments

RKI Instruments CO-04C User manual

RKI Instruments 65-2666RK-04 User manual

RKI Instruments

RKI Instruments 65-2666RK-04 User manual

Popular Measuring Instrument manuals by other brands

Futech MULTICROSS 5 SUPER Vision user manual

Futech

Futech MULTICROSS 5 SUPER Vision user manual

Bender ISOMETER isoPV425 with AGH420 manual

Bender

Bender ISOMETER isoPV425 with AGH420 manual

Silverline 727787 Original instructions

Silverline

Silverline 727787 Original instructions

Shimadzu MOC63 manual

Shimadzu

Shimadzu MOC63 manual

Novanta 32-1 Laser user manual

Novanta

Novanta 32-1 Laser user manual

LaMotte DC1500-CL manual

LaMotte

LaMotte DC1500-CL manual

ekwb EK-Quantum Scalar Flow Indicator Bottom-Top... installation manual

ekwb

ekwb EK-Quantum Scalar Flow Indicator Bottom-Top... installation manual

PCB Piezotronics 3711B11200G Installation and operating manual

PCB Piezotronics

PCB Piezotronics 3711B11200G Installation and operating manual

Madur GA-12plus operating manual

Madur

Madur GA-12plus operating manual

Soilmoisture Equipment 7201 operating instructions

Soilmoisture Equipment

Soilmoisture Equipment 7201 operating instructions

BNC RTSA7550 user guide

BNC

BNC RTSA7550 user guide

Kane 455 Quick reference guide

Kane

Kane 455 Quick reference guide

Diehl RAY FS WP installation guide

Diehl

Diehl RAY FS WP installation guide

wtw PhotoLab Spektral operating instructions

wtw

wtw PhotoLab Spektral operating instructions

LIMESA meters FG4000 Installations and specifications

LIMESA meters

LIMESA meters FG4000 Installations and specifications

PCE Health and Fitness VT 3800 user manual

PCE Health and Fitness

PCE Health and Fitness VT 3800 user manual

LCJ CAPTEURS CV7SF2-M user manual

LCJ CAPTEURS

LCJ CAPTEURS CV7SF2-M user manual

Hioki Memory HiCorder MR6000 instruction manual

Hioki

Hioki Memory HiCorder MR6000 instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.