
20 • 65-2322-04SS Oxygen Transmitter
7. Place the new amplifier in the same position as the old amplifier. A foam gasket that
orients the amplifier and keeps it from rotating is installed on the bottom of the
amplifier. Make sure the amplifier is seated flat in the junction box.
8. Install the new amplifier into the junction box with the screw, lock washer, and flat
washer you removed in Step 5.
9. Install the detector and controller terminals strips into their sockets on the new
amplifier as shown in Figure 6 and Figure 7. If controller leads or detector leads were
removed during this procedure, refer to Figure 6 and Figure 7 for the detector and
amplifier connections.
NOTE: When a transmitter is first powered up with a new amplifier, the initial output
may be either high or low depending on the setting of the span pot. Be sure to
make arrangements so that this does not cause unwanted alarms.
10. Reinstall the amplifier junction box cover.
11. Turn on or plug in power to the controller.
12. Turn on the controller and place it into normal operation.
13. Allow the transmitter to warm up for 5 minutes.
14. Calibrate the oxygen transmitter as described in the “Calibration” on page 21.
Calibration Frequency
Although there is no particular calibration frequency that is correct for all applications, a
calibration frequency of every 3 months is adequate for most oxygen transmitter
applications. Unless experience in a particular application dictates otherwise, RKI
Instruments, Inc. recommends a calibration frequency of every 3 months for the oxygen
transmitter.
If an application is not very demanding, for example detection in a clean, temperature
controlled environment, and calibration adjustments are minimal at calibration, then a
calibration frequency of every 6 months is adequate.
If the application is very demanding, for example if the environment is not well
controlled, then more frequent calibration than every 3 months may be necessary.
Determining Response Time
Since the detector junction box is installed remotely from the amplifier, the response time
of the detector will depend on the length of flexible tubing that is attached to the
calibration adapter/splash guard. To determine the response time based on the length of
tubing connected:
1. Place the controller into its calibration mode or disable external alarms.
2. Note the current gas reading displayed at the controller.
3. Screw the regulator into the calibration cylinder. Do not use a zero air cylinder for this
operation.
4. Use the flexible tubing coming from the calibration adapter/splash guard to connect
the regulator to the calibration cup.