Preparing for use
R&S®FS-SNS 18/26/40/55/67/90/110
3Manual 1338.8050.02 ─ 06
2 Preparing for use
The R&S®FS-SNS is designated for use in industrial, administrative, and labora-
tory environments. Use the R&S®FS-SNS only for its designated purpose.
Observe the usage instructions documented in this manual, as well as operating
conditions and performance limits stated in the data sheet. The product documen-
tation helps you use the R&S®FS-SNS safely and efficiently. Keep the product
documentation in a safe place and pass it on to subsequent users.
●Before use, inspect the RF and interface connectors of the R&S®FS-SNS for
damage. If there is any damage, contact Rohde & Schwarz for repair.
●Before use, inspect the R&S®SNSCABLE and R&S®SNSCABLE-Y cables for
damage. If there is any damage, contact Rohde & Schwarz for repair.
●Always use an appropriate torque wrench when connecting the R&S®FS-SNS
cables to an instrument to avoid excessive force.
●Ensure the instrument that you want to connect the R&S®FS-SNS to has an
appropriate RF input connector.
–R&S®FS-SNS18: SMA male
–R&S®FS-SNS26: APC 3.5 mm male (compatible with SMA)
–R&S®FS-SNS40: 2.92 mm male (compatible with SMA)
–R&S®FS-SNS55/67: 1.85 mm male (compatible with 2.4 mm)
–R&S®FS-SNS90: WR12
–R&S®FS-SNS110: WR10
3 Measurement setup
Depending on the available connector types on the analyzer, connect the
R&S®FS-SNS to the instrument as described below.
For sensitive DC-coupled amplifiers, we recommend using a DC blocker.
The identification and setup procedure after connecting the R&S®FS-SNS
can take up to 10 seconds.