1Purpose
This document describes the procedure for cleaning the RAM ION meter. In some cases, due to very high humidity in
the air, the background readings from the RAM ION will be unusually high. The silica gel, which is meant to protect
the internal electronics may not be sufficient to dry the air and sometimes even if the air is dried, there may remain
particles from the air inside the chamber which have now dried but are causing unusually high background readings.
The background of the RAM ION should not be more than 0.01 mR/h.
IMPORTANT
The cleaning of the RAM ION should be done in a clean, dustless environment
We recommend using compressed air as a cleaning agent.
Before opening the Chamber, please check that the flat cable of the Touch Panel is connected properly by following
the procedure below:
1. Open the 4 screws holding the Touch panel in place