rova RO-40A User manual

Digital Fingertip
Pulse Oximeter
RO-40A



Hi. Thank you for purchasing the Pulse Oximeter. Please read this user manual carefully and keep it
in a safe place for future reference. If you need any assistance, please contact Rova customer
Package Contents
1 x Pulse Oximeter
1 x Neck/Wrist Card
1 x Manual
Product Diagram
1. Light Emitting Diode
2. Finger
3. Photo Detector
4. Probe Body
Introduction

Front View
1. Pulse bar graph
2. SpO2 value
3. Pulse rate value
4. Operating button
Features
· The pulse oximeter is small in volumes, light in weight and easy to carry.
· Low in power consumption, 600 spot-checks on two AAA batteries.
· One button and easy to operate.
· There are three modes: power off, sleep and measure.
· Automatically turning into sleep ,mode within 8 seconds after there is no signal.
How to use
1. Remove the battery cover, and insert the two AAA batteries following polarity markings
indicated insides of the battery compartment, then reposition in the cover.
2. Hold the oximeter with the display facing toward you, slide our finger into the opening probe of
the device as showing below, until the fingertip touches the built-in stop guide. For best results,
/Min
4
3
2
1

make sure the finger is centered within the finger guide.
3. Press the button to activate the oximeter from the sleep mode, and then measurement interface
will appear in 3 seconds.
4. The measurement result will be read directly from the screen within 10 seconds.
5. The oximeter will turn into sleep mode within 8 seconds after the finger left the probe.
Press the button to open the probe Insert the finger Press the operating button to
activate the oximeter

Symbols
Measurement Specifications
Symbol
% Sp02
PR
IPX2 The product is protected against harmful effects of dripping water per IEC 60529
SpO2 Declared Accuracy
70%~100%:±2 digits
0%~69%: unspecified
1%
1 s
8 s
1 bpm
1 s
8 s
Range 25%~250%:±3 digits
Range
Resolution
Update Period
Averaging Time
Resolution
Update Period
Averaging Time
PR Declared Accuracy
Pulse Rate(BPM)
Power Indicator
Caution, Consult
Accompanying
Documents
Rx Only
Definition
Pulse Oxygen
Saturation(%)
Symbol Definition
Manufacturer
BF Type Applied Part
Not for Continuous Monitoring
(No Alarm for SpO2)
Federal law restricts this device
to sale by or on the order of a
licensed health care practitioner

Warnings
Necessary maintenance must be only performed by qualified service personal.
Users are not permitted to maintain the equipment by themselves.
There are not replaceable components in the equipment.
Warranty and Support
Warranty
Rova provides all our customers 24 months warranty from the date of its original purchase. If the
to launch a warranty claim. We will instruct you on how to return the product for repair/
replacement.
Temperature
Operating
Storage/
Transportation
Operating 70-106 kpa
50-107.4 kpa
Storage/
Transportation
Operating
Storage/
Transportation
+41°F to +104°F/
5℃to + 40℃10-95%
noncondensing
10-95%
noncondensing
-41°F to +140°F/
-40℃to + 60℃Humidity Atmosphere Pressure
Measurement Specifications

Statement
We can only provide after sale service for products purchased directly from Rova or Rova
authorized distributor. If you have purchased from a different seller, please contact them for any
exchange or request.
Contact Us
For all inquiries, please contact us with your order number for dealing with the issue quicker.
Appendix A
The equipment complies with the requirement of standard EN 60606-A-2:2007” Electromagnetic
Compatilibity-Medical Electrical Equipment”.

Guidance and manufacturer’s declaration - electromagnetic immunity
Immunity test IEC 60601 test level Compliance level Electromagnetic
environment - guidance
Electrostatic discharge
(ESD) IEC61000-4-2
Power frequency
(50/60Hz) magnetic field
NOTE UT is the a.c. mains voltage prior to application of the test level.
±6 KV contact
±8 KV, ±15KV air
30 A/m
±6 KV contact
±8 KV, ±15KV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location i na typical
commercial or hospital
environment.
30 A/m
The RO-40A Pulse Oximeter is intended for use in the electromagnetic environment specified below.The customer
or the user of the RO-40A Pulse Oximeter should ensure that it is used in such an environment.

Guidance and manufacturer’s declaration - electromagnetic immunity
Immunity test IEC 60601 test level Compliance level Electromagnetic
environment - guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHZ to 80MHz
3 V/m
80 MHz to 2.5 GHz
N/A
Portable and Mobile RF
communications equipment should
be used no closer to any part of
the RO-40A Pulse Oximeter, including
cables, than the recommended
separation distance calculated from
the equation appliance to the
frequency of the transmitter.
Recommended separation distance.
80MHz to 800MHz
800MHz to 2.5GHz
3 V/m
The RO-40A Pulse Oximeter is intended for use in the electromagnetic environment specified below.The customer
or the user of the RO-40A Pulse Oximeter should ensure that it is used in such an environment.

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
Where P is the maximum output
power rating of the transmitter in
watts(W) according to transmitter
manufacturer and d is the
recommended separation distance
in metres(m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey, a
should be less than the compliance
level in each frequency range. B
interference may occur in the
vicinity of equipment marked with
the following symbol:

Emissions test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11 Group 1
RF emissions
CISPR11 Class B
Harmonic emissions
IEC 61000-3-2 N/A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
N/A
Guidance and manufacturer’s declaration - electromagnetic immunity
The RO-40A Pulse Oximeter is intended for use in the electromagnetic environment specified below.The customer
or the user of the RO-40A Pulse Oximeter should ensure that it is used in such an environment.
The RO-40A Pulse Oximeter uses RF energy only for its
internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
The RO-40A Pulse Oximeter is suitable for use in all
establishments, including domestic establishments and
those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic
purposes.

0,01
0,1
1
10
100 ////0.06
0.18
0.57
1.8
0.11
0.35
1.1
3.5
150 kHz
to 80 MHz
/0.02
80 MHz
to 800 MHz
0.03
800 MHz
to 2.5 GHz
Recommended separation distance between portable and mobile RF
communications equipment and the Medical RO-40A Pulse Oximeter
Separation distance according to frequency of transmitter (m)
Rated maximum output
power of transmitter (W)
The RO-40A Pulse Oximeter is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the Medical RO-40A PULSE OXIMETER can help to prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters), and the RO-40A Pulse Oximeter as recommended below, according to the maximum output power of
the communications equipment.
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.

Einführung
Hallo! Vielen Dank für den Kauf Pulse Oximeter. Bitte lesen Sie diese Bedienungsanleitung
sorgfälting durch und bewahren im sicheren Platz für zufünfige Referenz. Sollten Sie jegliche Hilfe
brauchen, kontaktieren Sie bitte Rova Kunden-Support per E-mail unter support.all@rova-
health.com.
Packungsinhalt
1 x Pulse Oximeter
1 x Hals/Handwurzel Kordel
1 x Bedienungsanleitung
Produkt Diagramm
1. Licht emittierende Doide
2. Finger
3. Photodetektor
4. Sondekörper

Frontansicht
1. Pulse bar Balkendiagramm
2. SpO2 Wert
3. Pulsefrequenz Wert
4. Bedienknopf
Funktionen
· Die Pulse Oximeter ist klein in Volumen, leicht in Gewicht und einfach zu tragen.
· Wenig Stromverbrauch, 600 Stichproben mit 2 AAA-Batterien.
· Ein Knopf und einfach zu bedienen.
· Es gibt drei Modi: Ausschalten, Schlaf und Messen.
· Automatisch tritt es in Schlafmodus innerhalb 8 Sekunden, wenn kein Signal gibt.
Wie verwenden Sie
1. Entfernen Sie den Deckel der Batterie, und stecken zwei AAA Batterien nach Polarität Markierungen
die in der Innenseiten des Batterie Kompartment angegeben. Dann reponsitionieren Sie den Deckel.
/Min
4
3
2
1

2. Halten Sie das Oximeter mit dem Display zu Ihrer Richtung und schieben Sie Ihren Finger in die
Öffnung der Sonde des Geräts wie unten gezeigt, bis die Fingerspitz den eingebauten Anschlag
berührt. Für optimale Ergebnisse stellen Sie sicher, dass der Finger innerhalb der Finger Anschlag
zentriert ist.
3. Drücken Sie den Knopf zum Aktivieren aus Schlafmodus des Oximeter und dann erscheint
Messung Schnittstelle in 3 Sekunden.
4. Innerhalb von 10 Minuten wird das Messergebnis direkt vom Bildschirm gelesen.
5. Das Oximeter schaltet im Schlafmodus innerhalb von 8 Sekunden, nachdem der Finger von der
Sonde verlässt.
Drücken Sie den Knopf zum Öffnen der Sonde Stecken Sie den Finger Drücken Sie den
Bedienknopf zum Aktivieren des Oximeters.

Symbole
Messungsspezifikationen
Symbol
% Sp02
PR
IPX2 Das Produkt ist gegen die schädlichen Auswirkung vonTropfwasser pro IEC 60529 geschützt.
%Sp02 Deklarierte Genauigkeit
70%~100%:±2 digits
0%~69%: unspezifiziert
1%
1 s
8 s
1 bpm
1 s
8 s
Bereich 25-250:±3 digits
Bereich
Resolution
Update-Periode
Mittelungszeit
Lösung
Update-Periode
Mittelungszeit
PR Deklarierte Genauigkeit
Pulsfrequenz(BPM)
Power Indikator
Achtung,
konsultieren Sie
Begleitdokumente
Rx Only
Definition
Puls Sauerstoff
Sättigung(%)
Symbol Definition
Hersteller
BF Typ Teil angewendet
Nicht für kontinuierliche Überwachung
(Kein Alarm für %Sp02)
Bundesgesetz schränkt dieses Gerät zum
Verkauf von oder in der
Ordnung von einem lizentierten Arzt.

Warnungen
Notwendige Wartung muss nur von qualifizierten Service Personal ausgeführt werden.
Benutzer darf das Gerät nicht selbst warten.
Es gibt keine austauschbare Komponenten im Gerät.
Garantie und Unterstützung
Garantie
Rova bietet alle unsere Kunden 24 Monate Garantie ab dem Zeitpunkt des ursprüngliche Kaufs.
Wenn das Gerät aufgrund von Herstellerdefekte ist, kontaktiern Sie uns per Email an
einweisen, wie Sie das Produkt zur Reparatur/Austausch bringen können.
Temperatur
Bedienen
Lagerung/
Transport
Bedienen 70-106 kpa
50-107.4 kpa
Lagerung/
Transport
Bedienen
Lagerung/
Transport
+41°bis +104°F/
5°bis +40℃10-95%, Keine
Kondensation
10-95%, Keine
Kondensation
-40°+ 104°F/
-40°bis+60℃Fleuchtigkeit Atmosphärischer Druck
Ökologische Spezifikationen

Erklärung
Wir bieten nur nach Verkauf Service für Produkte, die gekauft von Rova oder Rova autoriserten
Händler. Wenn Sie von einem anderem Verkäufer gekauft haben, bitte kontaktieren Sie für
Austausch oder Anforderung.
Kontakt mit uns
Für alle Anfragen kontaktieren Sie uns bitte mit Ihrer Bestellnummer, mit der die Behandlung des
Problems schneller läuft.
Email: [email protected]
Table of contents
Popular Medical Equipment manuals by other brands

COMPASS HEALTH
COMPASS HEALTH PB6034 Assembly, installation and operating instructions

Ergotron
Ergotron StyleView SV32-S installation guide

Pressalit Care
Pressalit Care R8468 Operation and maintenance manual

Fukuda Denshi
Fukuda Denshi Dynascope 8000 Series Service manual

Nasco Healthcare
Nasco Healthcare 2271 instruction manual

ICU Medical
ICU Medical Diana DS1000 user manual