QUICK REFERENCE GUIDE
To place an order, call +1.941.556.1850,
email info@RPSdetectors.com, or visit www.RPSdetectors.com
Rapid Pathogen Screening, Inc.
7227 Delainey Court • Sarasota, FL 34240 USA • +1.941.556.1850 • RPSdetectors.com
©2013 Rapid Pathogen Screening, Inc. All Rights Reserved. FORM-MKT-014.2 Aug 2013
1. For in vitro diagnosc use only.
2. Keep the test cassee and sample collector in their foil pouches unl just before use.
3. The Dacron material used in the sampling eece may cause allergic reacons for some people.
4. Do not use AdenoPlus past the expiraon date.
5. All specimens should be considered potenally hazardous and handled in the same manner
as an infecous agent. Proper handling and disposal methods should be established according to
local, state, and federal regulaons.
6. Wear disposable gloves while handling samples and wash hands aer the test is complete.
7. Both AdenoPlus and the buer vial are single use items. Do not reuse with mulple specimens.
8. AdenoPlus requires a visual readout. Do not interpret the test result if you have color-impaired vision.
9. Result interpretaon requires a brightly lit environment.
10. Do not use the same AdenoPlus test kit on more than one paent.
WARNINGS AND
PRECAUTIONS
Once the background within the result window is white and 10 minutes have elapsed,
the test may be accurately read.
Note: Do not interpret the test results before compleng 10 minutes of development me.
A purple uid wave may be observed moving across the result window while the test is
running.
The results of the test are indicated through two lines which appear in the result window:
the control line and the result line. The control line appears as a BLUE line in the control
zone. It indicates the correct applicaon and performance of the test, and must appear for
the test to be valid.
POSITIVE RESULT
The presence of both a BLUE line in the control zone and a RED line in the result zone
indicates a posive result. Even if the RED line is faint in color, incomplete over the width
of the test strip, or uneven in color, it must be interpreted as posive. A posive result
indicates the presence of Adenovirus angens.
NEGATIVE RESULT
The presence of only a BLUE line in the control zone indicates a negave result. A negave
result is indicave of an absence of Adenovirus angens.
INVALID RESULT
If a BLUE line does not appear, the test may be invalid. Reimmerse the absorbent
p into the buer vial for an addional 10 seconds. If a BLUE line sll does not
appear, the test must be discarded and the subject retested by resampling the eye using a
new AdenoPlus test kit.
THE TEST RESULTS
POSITIVE
NEGATIVE
INVALID