Salter Labs PRO2 Check User manual

L
CAUTION

30 Spur Dr
El Paso, TX 79906
Phone: 1.800.421.0024



PRO
2
check ®
Ultrasonic Oxygen Indicator
WARRANTY
Salter Labs® will extend the following warranty to the original purchaser of
the PRO2 check® Ultrasonic Oxygen Indicator:
If the unit becomes inoperable prior to two years from date of purchase for
any reason other than:
1.
Unusual, abusive use or handling
2.
Dead, defective battery
3.
Water or water vapor has been introduced into unit.
Salter Labs® will, at its option, have the unit repaired or replaced. If repair
is authorized, Salter Labs® will not charge for parts or labor. Purchaser
will be responsible for all shipping charges to return the unit to Salter
Labs. Salter Labs will be responsible for normal shipping charges for the
return of the unit to purchaser.
All warranties are voided if the PRO2 check® unit has been opened or
tampered with in any way.
If claim is made after two years from the date of purchase and unit is
inoperable, all costs involved in such repair are to be paid by the purchaser.
In the event that the unit becomes inoperable, the purchaser should
promptly notify Salter Labs® for instructions in handling repair/return of
the unit. Any return without prior authorization from Salter Labs® is at
purchaser's expense and risk. Salter Labs® does not assume any other
liability except as stated above.
15

PRO2 check ®
Ultrasonic Oxygen Indicator
PRO2 check ®
Ultrasonic Oxygen Indicator
Operation Manual
CONTENTS
I. Warning Notes 1
II. Introduction 2
III. Initial Setup 3
IV. Instructions For Checking O2 Concentrators 4
To Turn Unit On and Off 4
Measuring Oxygen Concentrations 4
V. Instructions For Spot Checking Liquid Oxygen
Vessels or Oxygen Cylinders 5
VI. Verification & Calibration Instructions 6
Pure Oxygen Verification Procedure 6
Quick "One Step" Pure Oxygen Calibration 6
Full "Two Step" Calibration Procedure 7
VII. Maintenance 8
Battery Replacement 8
Cleaning 8
VIII. Environmental Effects
Temperature 9
Barometric Pressure 9
Vibration 9
Relative Humidity 9
IX. Unit Specifications 10
X. Frequently Asked Questions 11 - 12
XI. Trouble Shooting Check List 13
XII. Warranty 15
Manufactured by:
Salter Labs® - El Paso, Texas

PRO
2
check ®
Ultrasonic Oxygen Indicator
WARNING NOTES
USAGE: The PRO2 check® Ultrasonic Oxygen Indicator is designed for
intermittent usage to determine the oxygen concentration produced by an
oxygen concentrator. It may also be utilized to do spot checks of gaseous or
liquid oxygen. This instrument has no clinical diagnostic or therapeutic
functions.
WARNING: The unit is not designed nor intended for use in anesthesia
applications, or for monitoring oxygen concentration from any source other
than a conventional oxygen concentrator using molecular sieve beds. This unit
will not withstand fluid spillage. The unit will not withstand mechanical shock
or vibration.
NOTE: This equipment has been tested and found to comply with the limits
for a Class B digital device, pursuant to Part 15 of the FCC Rules. These
limits are designed to provide reasonable protection against harmful
interference in a residential installation. This equipment generates, uses, and
can radiate radio frequency energy and if not installed in accordance with the
instructions, may cause harmful interference to radio communications.
However, there is no guarantee that interference will not occur in a particular
installation. If this equipment does cause interference to radio or television
reception, which can be determined by turning the equipment off and on, the
user is encouraged to try and correct the interference by one or more of the
following measures:
Reorient or relocate the receiving antenna
Increase the distance between the equipment and receiver.
Connect the equipment into an electrical outlet on a different
circuit from which the receiver is connected.
Consult the dealer or an experienced radio/TV technician for help.
1

PRO2 check ®
Ultrasonic Oxygen Indicator
Trouble Shooting Checklist (Continued)
Complaint ProbableCause
Suggested
Corrective Action
Display shows "err"
duringcalibration
Calibration was
performedincorrectly
or wrong calibration
gas was used.
Ensure U.S.P. oxygen
(> 99.0%) is being
used and re-perform
procedure according to
directions in the
OperationManual.
Display shows "err"
during use
Unit has made a reading
outside normal ranges.
Checkcalibration.
Make sure gas being
measured is from
oxygenconcentrator.
"LO BATT"
appears in display
screen when unit is
turned on
Battery is low.
Change battery.
Display shows "err"
or a reading other
than 100% ± 2% after
full calibration
Unit may have been
used to check a
concentrator with a
humidifier bottle in line
or water has gotten
inside of unit.
Call Salter Labs'
Customer Service
Department at
(800) 235-4203 for
authorization to return
unit for repair.
14

PRO2 check ®
Ultrasonic Oxygen Indicator
Trouble Shooting Checklist
Complaint ProbableCause
Suggested
Corrective Action
Display is blank when
unit is turned on.
Bad batteries.
Corroded
battery
terminals.
Replacebatteries.
Clean battery terminals.
When unit is turned on
displayflashes
"PRO O2 @ CAL"
and then turns off.
Unit
requires
calibration.
Perform"Full
CalibrationProcedure"
outlined in Operation
Manual.
Display reads 91.8% on
a source of U.S.P.
oxygen (> 99.0%).
Unit
is operating in
mode
to check
concentrators
and is
adjusting
for
concentrator's
output
of
argon.
With unit still connected to
source of U.S.P. oxygen (>
99.0%) push
and continue
to hold
down “pure
O2/calibrate” button to
temporarilyremove argon
correction from unit's
With unit still connected to
source of U.S.P. oxygen (>
Display reads between
98% to 102% (but not
99.9%) on a source of
U.S.P.oxygen
(> 99.0%) with button
held down.
Unit
requires quick
calibration.
99.0%) push and hold
down “pure O2/
calibrate” button until
display reads 99.9% - 101
(approximately20
seconds). The unit is then
recalibrated.
Unit is out of 98% to
102% range after quick
calibration.
Unit
requires full
calibration.
Perform"Full
CalibrationProcedure"
outlined in User Manual.
"---" appears on display
while unit is being
calibrated.
Incomplete
calibration.
Release the "pure
O /calibrate" button
2
and restart calibration
process.
Display shows "-U-".
Indicates
an out of
range
reading.
Perform"Full Calibration
Procedure" outlined in
User Manual.
13

PRO
2
check ®
Ultrasonic Oxygen Indicator
INTRODUCTION
Thank you for purchasing a PRO2 check ® Ultrasonic Oxygen Indicator. The
designers of this product have years of experience in the field of oxygen
concentrator repairs. The unreliability and high cost of operation of fuel
cell analyzers led to the invention of the PRO2 check ® Ultrasonic Oxygen
Indicator.
Byrelying on a method already used in many oxygen concentrators to
trigger "Low O2" alarms – ultrasound – Salter Labs has produced a more
cost effective product to check the output of oxygen concentrators. The
PRO2 check ® Ultrasonic Oxygen Indicator responds faster, has no fuel
cell
to replace and is simpler to use than a fuel cell analyzer.
Directions for Use:
There are two (2) controls for the PRO2 check® unit:
1.
Pressing and releasing the “on/off” button will turn the unit on or off.
2.
The “pure O2 calibrate” button when pressed and held down
while connected to a compressed gas or liquid U.S.P. oxygen
(> 99.0%) source verifies that the unit is properly calibrated, thus
eliminating the need to have an outside source calibrate the unit.
Since the PRO2check® device uses ultrasound technology to check the con-
centrator operation, there has to be two modes of operation:
1.
When the unit is turned on, the software enables the unit to
measure the output of an oxygen concentrator adjusting for the
argon that passes through the sieve beds.
2.
When using the unit with the “pure O2 calibrate” button
depressed the software measures oxygen not adjusting for the
argon. When the “O2 calibrate” button is released, the unit
automatically returns to the concentrator mode and will then again
adjust for the argon and read incorrectly for pure oxygen.
We are sure that the PRO2 check® Ultrasonic Oxygen Indicator will
provide years of fast, reliable operation.
2

PRO2 check ®
Ultrasonic Oxygen Indicator
INITIAL SETUP
1.
Remove the PRO2 check® Ultrasonic Oxygen Indicator from the
package. Make sure you have the following items:
•
PRO2 check® indicator unit
•
9-volt battery (factory installed)
•
2 foot connecting tubing
•
Carrying case and strap
•
OperationManual
2.
The PRO2 check®Ultrasonic Oxygen Indicator was calibrated at
the
factory. However, the proper calibration of the unit's oxygen
concentration measurement mode can be quickly and easily
verified by following the "Pure Oxygen Verification" procedure
on page 6 of this manual.
3.
If accreditation organizations or internal policies and procedures
require routine calibration, instructions on performing both a
"Quick One Step Calibration Procedure" to U.S.P. oxygen (>
99.0%) can be found on page 6 of this manual. A "Full Two Step
Calibration Procedure" to both U.S.P. oxygen (> 99.0%) and
room air can be found on page 7 of this manual.
Low Battery Indicator:
When the unit displays “LO” and “BAT” in the lower left corner,
battery is low and should be replaced. See page 8.
3

PRO
2
check ®
Ultrasonic Oxygen Indicator
FREQUENTLY ASKED QUESTIONS (continued)
When I am in the Concentrator Mode and hook the unit up to a U.S.P.
oxygen(> 99.0%) source the LCD does not read 99% (± 2), why?
When the unit is in its default (concentrator) mode the device's software
enables the unit to measure the output of an oxygen concentrator by
compensating for the argon that passes through the sieve beds. In this
mode, it will display a false reading (usually ± 90%) if connected to an
oxygen source other than a concentrator. Pressing and holding the “pure
O2/calibrate” button removes the argon compensation and allows the
device to measure the oxygen purity with ±2% of compressed gaseous or
liquid oxygen.
When I am in the Concentrator Mode and hook the unit up to a U.S.P.
oxygen (> 99.0%) source the LCD the display reads - "Concentrator O2
91.8" (approx). What's wrong?
Nothing. When in the Concentrator Mode, the electronic components of
the indicator automatically compensates for the argon found in the gas
produced by an oxygen concentrator. As a result the display will read
approximately90% unless this function is over-ridden bypushing the
“pure O2/calibrate” button.
Whenever I check concentrators with the PRO2 check®, I consistently get
an O2 concentration reading of 96.0 on every concentrator. Why?
Concentrators work at their maximum efficiency (96.0%) at lower flow
rates. When the flow rate is increased the output of the concentrator
typically will decrease slightly. If you question whether the device is
functioning properly, stabilize it to room air and then retest with the
concentrator set on the maximum flow rate or beyond. Be sure to allow
the concentrator to stabilize at the higher flow rate before testing. This
may take as long as 10 minutes.
Will the unit lose calibration when I change batteries?
No. The unit holds calibration even when the battery is removed.
12

PRO
2
check ®
Ultrasonic Oxygen Indicator
FREQUENTLY ASKED QUESTIONS
Thank you for purchasing the PRO2 check® Ultrasonic Oxygen Indicator.
It
will give you years of trouble-free operation. The following are some of
the frequently asked questions about the operation of this device.
Do I have to calibrate my PRO2 check® unit daily?
No. The PRO2 check® indicator was calibrated at the factory.
How do I know that my PRO2 check® device is accurately
measuringconcentrations?
To verify, simply connect the unit to a source of either compressed gas or
liquid U.S.P. oxygen (> 99.0%) at 2 LPM. Then turn on the device and
hold down the “pure O2/calibrate” button. As long as the display
reads between 99% and 101%, the unit is properly calibrated.
How often should I calibrate my PRO2 check® unit?
As long as you can verify (via the Pure Oxygen Verification Procedure)
that the unit is holding calibration against a U.S.P. oxygen (> 99.0%)
source, the unit does not require calibration.
My accreditation organization as well as our internal policies require
periodic calibration of this type of equipment. Can I calibrate the PRO2
check® device to comply with these requirements?
Yes, if accreditation organizations or internal policies and procedures
require routine calibration, instructions on calibrating the device are
found on pages 6 - 7 of this manual.
Does the unit have to be turned off before verifying or calibrating?
Yes. The unit must be turned off and evacuated to room air before you
begin the oxygen verification or calibration procedures.
Why doesn't the unit drop immediately to room air?
Oxygen is heavier than room air, so it will linger inside the unit. When
there is a flow (either from room air or another concentrator), the reading
will change immediately.
Can I exhale (blow) into the unit to reach room air?
No. You may apply a small suction or flow of room air, but do not exhale
into the unit.
11

PRO
2
check ®
Ultrasonic Oxygen Indicator
INSTRUCTIONS FOR CHECKING O2CONCENTRATORS
To turn unit on and off: To turn the unit on or off press and release the
“on/off” button on the lower left-hand side of the unit. (EXAMPLE:
Concentrator O2 20.9) When the unit is turned on the device defaults to the
mode in which the software enables the unit to measure the output of an
oxygen concentrator compensating for the argon that passes through the
sieve beds. The display will read "PRO", then "O2" and finally
"Concentrator O2" and a number. If the PRO2 check® unit is hooked up
to
an operating concentrator, the device will automatically measure the
concentration of the oxygen being produced and show the value on the
LCD display. The unit's default operating mode automatically compensates
for the argon that passes through the sieve beds, it will display a false
reading (approximately 90%) if connected to an oxygen source other than
a concentrator unless the specific actions discussed on pages 6 - 7 of this
manual are taken. If the device is not hooked up to an oxygen source it will
measure and display the concentration of oxygen in the ambient air
(usually 20.9%). Reading may vary if sample gas remains in unit.
To turn the PRO2 check® device off, press and release the “on/off”
button.
The unit will display “OFF” and shut down. To conserve battery
life, it
automatically shuts down if any one of the unit's functions is not
utilized
for approximately 5 minutes.
Measuring OxygenConcentrations:
Follow the steps below to measure the concentration of oxygen being
supplied byan oxygen concentrator:
1.
Make sure the concentrator has been running long enough (follow
concentrator manufacturer's guidelines) to reach full operational output.
2.
Make sure flow rate is above 2 LPM - preferably at the concentrator's
highest output.
3.
Remove any humidifier fromconcentrator.
4.
Connect supply tubing to the concentrator oxygen outlet and the inlet
port on the PRO2 check®.
5. Turn on the PRO2 check® device (See above).
6. The indicator will stabilize in about 10 seconds and provide an
accurate reading on the LCD display.
7. If there is any question to the accuracy of the reading, perform the
"Pure Oxygen Verification Procedure" and if necessary the “Quick
One Step Pure Oxygen Calibration Procedure on page 6.
CAUTION: Do not check concentrator with humidifier in place.
Humidity can affect unit's readings and damage the unit.
NOTES: Never check a concentrator while holding down the "pure O2 /calibrate"button.
You will get a false reading. You must hold down the "pure O2 /calibrate" button to
check unit against a source of U.S.P. oxygen (> 99.0%) cylinder or liquid oxygen.
4

PRO
2
check ®
Ultrasonic Oxygen Indicator
2
INSTRUCTIONS FOR SPOT CHECKING
LIQUID OXYGEN VESSELS or OXYGEN CYLINDERS
Although the PRO2 check® Ultrasonic Oxygen Indicator is designed, as its
name implies, to check the various functions of an oxygen concentrator, it
can also be used to perform spot checks (± 2%) of gaseous or liquid
oxygen.
To utilize the PRO2 check® device for this purpose follow this procedure:
1.
Connect the inlet port on the PRO2 check® unit to a gaseous or
liquid
oxygen source (cylinder or liquid oxygen) via a supply tube.
2.
Turn the oxygen flow on to 2 LPM.
3.
Press the “on/off” button to turn the PRO2 check® unit on.
4.
When unit stabilizes the unit display will read "Concentrator O2
91.8" (approx concentration.)
5.
Press and continue to hold down the “pure O /calibrate” button.
6.
Unit will display "Pure O2" and a numeric reading.
7.
If the source gas is oxygen, the display should read between 99%
and 101%.
8.
If the unit display reads "err", "---", "-U-" or any other reading
other than 99.9 (± 2%), the source gas may not be pure oxygen and
should be quarantined for retesting using a more sophisticated method
such as a Servomex* oxygen analyzer.
NOTE: PRO2 check® Ultrasonic Oxygen Indicator can be used to spot
check
the identity and purity of gaseous or liquid oxygen only, it cannot be
used to
performor verify U.S.P. oxygen (> 99.0%) testing since it cannot
provide
the required U.S.P. oxygen (> 99.0%) accuracy of ± 0.1%.
* Servomex Inc. - Sugarland, TX
5

PRO2 check ®
Ultrasonic Oxygen Indicator
UNIT SPECIFICATIONS
General Specifications
PhysicalData
Dimensions: 3.60" W x 5.75" H x 1.29" D
(9.14 cm W x 14.6 cm H x 3.28 cm D)
Weight: 9 ounces (255.15 Grams)
Connector: ¼" anodized aluminum hose barb
Case: Solid color ABS plastic with threaded metal bolt channels
Battery: The battery capacity is a standard 9-volt "transistor" alkaline
type battery with polarized terminals.
A rechargeable type battery is not recommended.
Operating Temperature: 32º F to 105º F (0º C to 41º C)
Storage Temperature: -29º F to 160º F (-34º C to 71º C)
Specifications - Concentration Mode
Measured Oxygen Concentration Range: 20.9% to 100%
Response Time: 10 seconds
Accuracy: ± 2%, assuming proper calibration with a sample gas at a
temperature of 32º F to 105º F (0º C to 41º C), and a flow rate of 2 to 5 LPM
to the unit during calibration.
Input Gas Flow Rate: Use a flow rate of .5 to 10 LPM for concentrator
sampling.
Sampling Frequency: Continuous
Vents: The unit vents through the side vent
port.
Linearity: ± 2% of Full Scale
Sensor: Ultrasonic
10

PRO
2
check ®
Ultrasonic Oxygen Indicator
ENVIRONMENTAL EFFECTS
Temperature
Under normal operating conditions 32º F to 105º F (0º C to 41º C), the
PRO2 check ®unit's microprocessor and temperature circuit will
compensate
for variations in temperature. The temperature during storage
and shipping
should not reach below -29º F (-34º C) or above 160º F (71º
C). If the unit
has been stored at other than normal room temperatures,
allow unit to
stabilize at room temperature for 15 to 30 minutes. Unit will stabilize quicker
if turned on.
BarometricPressure
The PRO 2 check® multifunctional unit's microprocessor makes it
unnecessary
to recalibrate during minor altitude changes. Please verify
calibration if used
over 5,000 feet.
Note: If readings do not seem accurate, recalibrate the unit.
Vibration
It is recommended that the unit be used in a stable position to prevent the
displayreading from fluctuating.
Note: Unit will not withstand excessive shock or vibration.
Relative Humidity/Water Vapor
The relative humidity (RH) of the gas being sampled will affect the reading
being displayed. As the RH increases, the sample gas becomes diluted with
water vapor. This decreases the percentage of all gases measured including
oxygen. Gases from high-pressure cylinders and oxygen concentrators are
basically dry (< 0.5% RH). Gas measured after a humidifier or that has a
high RH can cause a reading up to 10% lower than the actual reading.
Note: For accurate results, never measure oxygen percentage downstream
from a humidifier. This can damage the unit.
9

PRO2 check ®
Ultrasonic Oxygen Indicator
VERIFICATION & CALIBRATION INSTRUCTIONS
PURE OXYGEN VERIFICATION PROCEDURE
To quickly verify that the indicator unit is accurately measuring the
concentration of oxygen being produced bya concentrator, performthe
following steps:
1.
Connect the inlet port on the PRO2 check ® unit to a USP oxygen
source
(cylinder or liquid oxygen) via a supply tube.
2.
Turn the O2 delivery flow to 2 LPM.
3.
Press the “on/off” button to turn unit on.
4.
When unit stabilizes the display will read - "Concentrator O2 91.8"
(approx.concentration.).
NOTE: When in the Concentrator Mode, the electronic components of the
PRO2 check ® indicator automatically compensate for the argon found in
the gas produced by an oxygen concentrator. As a result the display will
read 90% (± 2%) unless this function is over-ridden by pushing the
"pure O2/calibrate" button).
5.
Press and continue to hold down the “pure O2/calibrate” button.
6.
Unit will display “Pure O2” and a numeric reading.
7.
If the unit is properlycalibrated, the display should read between
99% and 101%. The PRO2 check is ready to use.
8.
If the unit display reads "err, "---", "-U-" or any other reading
other than 99.9 (± 2%), the unit should be recalibrated (See below).
QUICK “ONE STEP” PURE OXYGEN CALIBRATION
The following procedure can be used to quickly and easily calibrate the
Ultrasonic Oxygen Concentrator Indicator to pure oxygen:
1.
Follow steps 1 through 6 of "Pure Oxygen Verification Procedure"
above.
2.
When the unit displays "Pure O2" and a numeric reading, continue
holding down the “pure O2/calibrate” button.
3.
After a few seconds, the LCD display (while continuing to read
"Pure O2") will start to alternately display "CAL" and number.
4.
Continue to hold the “pure O2/calibrate” button down until
the display stops alternating and simply displays “Pure O2 100”.
5.
At that point, the unit is calibrated to pure oxygen and ready for use.
NOTE: If unit displays an O2 concentration reading outside of the ± 2%
tolerance range (eg. < 98% or >102%) unit will not “Quick Cal” and the
Full Two Step procedure on the following page must be followed.
6

PRO
2
check ®
Ultrasonic Oxygen Indicator
FULL "TWO STEP" CALIBRATION PROCEDURE
The following procedure may be used to calibrate the PRO2 check® indicator
to both room air and U.S.P. oxygen (> 99.0%):
1. Clear analyzer gas chamber by inhaling through inlet. Use clean tubing and
discard after use.
2. With analyzer off, Push and Hold Calibration/Pure O2 button while turning on power.
3. When three dashed lines appear (---) release Calibration/Pure O2 button. Display
should read Cal Air (Cal Air).
NOTE: Do not exhale breath into unit. Moisture will affect calibration; a slight suction
can be used to evacuate the sampling chamber.
4. Press and release Calibration/Pure O2 button. You will see three dashed lines (---)
scroll in circles. Then the scrolling will stop and you will see CAL O2
5. Connect to O2 Cylinder at 2 LPM, then press and release Calibration/Pure O2 button.
6. You will see three dashed lines (---) scroll in circles. Then the scrolling will stop and
then you will see Cal End and read 91.7 Note: Reading may range from 91 to 92.
This is not a concern.
7. Push the Calibration/Pure O2 button. You should get a reading between 99% and 101%.
8. Unit is calibrated and ready for use.
9. If step 7 above is not achieved, try one more time. If 99% to 101% is still not achieved,
call Salter Labs® Customer Service to send unit back for service by a Technician at
Salter Labs® Service Dept.
NOTE: If unit displays "CAL ERR" during verification or calibration
procedures, check to ensure that you are using U.S.P. oxygen (> 99.0%) and
that the flow is turned on to at least 2 LPM, then repeat procedure.
NOTE: If unit displays "---" during verification or calibration procedures, release
the "pure O2/calibrate” button or the unit will shut off automatically.
If any problems are encountered during the calibration sequences,
contact Salter Labs at 1-800-421-0024 or 1-661-854-3166.
7

PRO2 check ®
Ultrasonic Oxygen Indicator
MAINTENANCE
BatteryReplacement
When the LCD displays "LO BATT" in the lower left corner, the battery
is low and should be replaced.
NOTE: The PRO2 check® must be removed from the carrying case to
replace battery.
Install a new 9-volt battery by sliding the battery door (located on the rear
of the unit) away from the case. Connect the 9-volt battery to the standard
9-volt battery connection and lay the battery sideways in the battery
compartment. Insert battery door by sliding door back onto unit in the
reverse direction of opening.
Cleaning
The PRO2 check® unit may be cleaned by wiping the case with a mild
detergent or standard topical disinfectant and a soft cloth.
CAUTION: Do not let cleaning solution spill on or inside
unit. Unit is not intended for any type of sterilization.
8
Table of contents
Other Salter Labs Measuring Instrument manuals