Electrical connection:
- Your IMPLANT CENTER 2 must be connected to the
electric power supply by a certified dental
installation technician.
- Warning: To avoid any risk of electric shock, this
device must be connected to an electrical
power supply with a protective ground.
- The electric supply to which the device is
connected must comply with the standards in
force in your country.
- If loss of electrical power during use is likely to
generate an unacceptable risk, the device must
be connected to a suitable power source (e.g.
UPS).
Using the device:
- Do not use the device if it appears to be
damaged or faulty.
- Turn the device off before unplugging the power
cord.
- To unplug the power cord, grip the cord plug
and hold the wall socket.
- Never use any other irrigation solution
containers than those intended for suspension
from the supplied brackets.
- The device must only be used with bottles or
bags of physiological saline or sterile water.
- The capacity of the irrigation solution
containers used must not exceed one litre.
- When the device is not to be used for a long
period of time, unplug the device from the
electric supply.
- Do not exert excessive force on the screen.
- Do not move IMPLANT CENTER 2 during use.
Environment:
- Do not cover the device or obstruct the
ventilation vents.
- Do not immerse the device in liquid, and do not
use it outdoors.
- Do not tilt the device at an angle greater than
5°.
- Do not place the device near a heat source.
- Make sure that the cords are not in a traffic
path.
- The device must be stored in its original
packaging in an appropriate and safe place.
- The device is not designed for use in the
presence of anesthetic gases or any other
flammable gas.
- Do not expose the device to water vapor, or
splashes.
- Any condensation inside an electrical device is
potentially dangerous.
- If the device is to be moved from a cold
environment to a hot one, it must not be used
until it has reached room temperature.
- The device is not designed to work near ionizing
radiation.
- Do not insert metal objects into the device (risk
of electric shock, short-circuit or projection of
hazardous substances).
Maintenance:
- Before and after each use, your IMPLANT CENTER 2
must be disinfected with SATELEC-approved
agents.
- Before each procedure, it is essential to make
sure that the accessories to be used have been
cleaned, disinfected and sterilized.
Accessories:
- The device has been designed and developed
for use only with SATELEC accessories to ensure
maximum safety and performance.
6