1
1.0 INTRODUCTION
The Savant®DNA SpeedVac®Model DNA120, is a dedicated centrifugal
vacuum concentration system for drying low volume ethanol or isopropanol-
water precipitates of DNA and RNA. The Savant DNA120 OP allows for
additional drying down of synthetic oligos in Ammonium Hydroxide
(NH4OH). This design incorporates a glass cover, ammonia post-trap and
Ammonia Neutralizing Solution for fully automated, unattended, odor-free
drying. The post-trap is connected on the back, left side of the unit. It is
easily maintained and replaced. The Savant DNA SpeedVac combines
centrifugal force with vacuum for bump-free sample drying. Vacuum is
supplied by an integral, oil-free vacuum pump with an automatic bleeder
valve.
The Savant DNA120 SpeedVac has a chamber heater that counteracts
evaporative sample cooling and accelerates solvent evaporation rates to
shorten drying times. The operator can select LOW (ambient),
MEDIUM (43 °C), or HIGH (65°C) drying rates with the 3-position switch
on the front panel.
Several different microcentrifuge tube rotors are available for drying ethanol
precipitates of DNA/RNA, isopropanol-water mixtures, amplified reaction
products in buffer, aliquots of radiochemicals, or other low-volume samples
(see section 5). A unique swing-out rotor (RD2MP) for microplates is avail-
able as well.
The Savant DNA SpeedVac is equipped with polypropylene vacuum fittings
to avoid leaks and resist corrosion. A chemical trap kit (DTK120R) and dis-
posable cartridge (DC120R) should be used for trapping volatile radioactivi-
ty when the system is dedicated for drying down radiolabeled materials.
NOTE: To assure safe operation and best results, read this manual in its
entirety before operating this instrument.
2.0 INSTALLATION
RECEIVING Inspect the shipping carton upon receipt. If the carton is
damaged in anyway, do not accept delivery. Contact
Thermo Electron.
UNPACKING Carefully remove the instrument from its shipping carton.
Compare the packing list to the box contents. If there is
a discrepancy, contact Thermo Electron.
INSPECTION Inspect the unit for any damage that may have occurred
during shipment. Should there be any damage, report it
to the carrier and contact Thermo Electron immediately.
Make sure the carrier inspects the damage and leaves an
inspection report. Register any claims for shipping
damage against the carrier or his agent. Save the
shipping carton in the event a return is necessary.
104-3003-00 Rev. C