scigiene SMARTMETER User manual

SMARTMETER
PORTABLE TEMPERATURE METER - INSTRUCTIONS MANUAL –V1.0x
1INTRODUCTION
SmartMeter is a portable temperature meter with high-contrast LCD.
The Smart Meter housing enables industrial environment or bench-
top operation and provides a belt carrying fixture. It supports 2
temperature sensors with independent and concurrent display of
measured values. It provides the capability of reading 0-50 mV
electric signals with configurable indication scale.
Difference between channels, relative measure, minimum, maximum
and average values, hold and alarm functions are also available. It
also features an auto power off resource in case the device is idle, in
order to extend battery life.
The device is supplied with a K-type thermocouple with 1.5 m cable.
1.1 UNPACKING
The meter is supplied in proper packaging that offers suitable
protection during transit and storage. Unpack with care. Upon
delivery, make sure that there is no damage to the instrument, and
that there are no missing items.
Within packaging, you should find, besides the manual, the meter
and a K-type thermocouple sensor.
2SPECIFICATIONS
Input types and max. indication range
J thermocouple ..........................................................- 100 to 760 ºC
K thermocouple.......................................................-150 to 1.370 ºC
T thermocouple..........................................................-160 to 400 ºC
E thermocouple.............................................................-90 to 720 ºC
N thermocouple ......................................................-270 to 1,300 ºC
R thermocouple ........................................................-50 to 1,760 ºC
S thermocouple.........................................................-50 to 1,760 ºC
B thermocouple......................................................... 500 to 1,820 ºC
Pt100 .........................................................................-200 to 600 ºC
Voltage 0-50 mV...................................-1,999 to 9,999 (Adjustable)
All input types are factory calibrated. NBR 12771, Pt100 NBR
13773/97 (IEC-751) (385) calibrated thermocouples.
Accuracy: Thermocouples:..............................0.25 % F.S. 1 ºC
Pt100: ...............................................0.2 % F.S. 1 ºC
0-50 mV: ........................................ 0.2 % F.S. 1 digit
Resolution:...................................................................................... 0.1 ºC
Sampling rate: ............................................................ 2 timer per second
Input impedance: ...............................> 10 MΩ (Thermocouples, 50 mV)
Pt100 excitation current:............................................................0.165 mA
Power supply: ......................................................................9 Vdc battery
AC / DC adapter .............................................................................6 Vdc
Estimated autonomy:.................................... 400 hours (alkaline battery)
Dimensions:................................................................. 165 x 73 x 36 mm
Weight: ............................................................................................205 g
Protection: ..........................................................................................IP40
3CONNECTIONS
Female input connectors suitable for RTD / TC compensated
connectors, as shown below.
Figure 01 —Meter supported connectors
Only two input terminals are used to connect thermocouples. Input
terminals have different size holes, which prevent inappropriate
installations.
Figure 02 —Thermocouple connection
All three terminals are used for Pt100 (RTD)
Figure 03 –Pt100 connectors sensors
The Pt100 internal connections should be made as shown in figure
below.
Figure 04 —Internal connections of Pt100 sensors
For the 0-50 mV signal, installation should comply with the following
connections:
Figure 05 —0-50 signal connection

SmartMeter
2/4
4OPERATION
The meter features two input channels. Channel readouts are
updated two times per second. Channel 1 readouts are shown in the
largest digits and channel 2 readouts are shown in the smallest digits,
as Figure 07 shows. Channel 02 can be turned-off by operator.
Figure 06 —Meter front panel
Figure 07 —SmartMeter display
4.1 TURNING THE METER ON/OFF
Press
to turn on the meter. To turn off, press for 3 seconds.
The meter basic configuration is factory set, so it can be immediately
used with the sensor provided. Factory configuration is: two enabled
channels, K-type sensor, one decimal space, zero offset, sound
alarm not enabled, auto power off after five minutes, temperature unit
in degrees Celsius, 60 Hz filter and channels 1 and 2 for special
functions.
4.2 FIRST READOUTS
Insert the sensor provided with the meter in the terminals of the
required channel and take the sensor leads to the measurement
location. Check the indication displayed.
Use the keys to find special functions the meter offers. When using
other sensors, check if factory setup is still compliant. If not, access
the configuration cycle and change settings.
4.3 SPECIAL FUNCTIONS
HOLD FUNCTION
It “freezes” the measure reading indication. HOLD is then displayed. It
is activated by pressing key
HOLD
.To resume the meter operation,
press key
HOLD
again.
MIN FUNCTION
It allows the user to retrieve the minimum temperature value
measured within an interval. An indicator of such condition is
displayed in the meter. It is activated by pressing key
AVG
MAX MIN
.To
resume the meter operation, press key
AVG
MAX MIN
again.
MAX FUNCTION
It allows the user to retrieve the maximum temperature value
measured within an interval. An indicator of such condition is
displayed in the meter. It is activated by pressing key
AVG
MAX MIN
.To
resume the meter operation, press key
AVG
MAX MIN
again.
AVG FUNCTION (average)
It displays the average temperature value measured within an
interval. An indicator of such condition is displayed in the meter. It is
activated by pressing key
AVG
MAX MIN
.To resume the meter operation,
press key
AVG
MAX MIN
again.
Notes: Functions MIN, MAX and AVG are accessed via key
AVG
MAX MIN
.
Values displayed are the ones acquired from the start of an interval of
measurements. The interval starts after a RESET action.To reset the
meter, press and hold key
AVG
MAX MIN
for more than 10 seconds. When
reset button is pressed, MIN, MAX and AVG values are erased and a
new interval starts.
RELAT FUNCTION (relative)
When enabled, it will void the measured values and starts to indicate
variations from such zeroed values. An indicator of such condition is
displayed in the meter. It is activated by pressing key
RELAT
.To
resume the meter operation, press key
RELAT
again.
T1 –T2 FUNCTION
When enabled, it displays the highest value corresponding to the
difference between the value measured through channel 1 and the
value measured through channel 2. The indication corresponding to
channel 2 is not changed. An indicator of such condition is displayed
in the meter. It is activated by pressing key
T1 –T2
.To resume the
meter operation, press key
T1 –T2
again.
Note: This function is only enabled when both channels are enabled
and operating.
ALARMS
The meter has an audible alarm (BEEP) that can inform the operator
about an alarm condition. In each channel there can be two indication
values, when they are reached the audible alarm is triggered. A
visual indicator is also displayed in the meter when the indication is
under an alarm condition.
LOW BATTERY INDICATOR
A low meter indicator is displayed warning that the battery should be
replaced.
SET UP
To set up the meter, the user must access the Alarm and Meter Set
up cycles, where a series of parameters must be defined.
ALARM CYCLE
In the alarm cycle you find the setpoints parameters of the meter
alarms. Press
SET
to access this cycle. The meter displays spk.1,
the first parameter of the cycle. To access the remaining parameters
of this cycle, press
SET
again,until you reach the measures screen.
Keys
and
make possible to change the parameters defined.
The parameters of this cycle are:
spk.1 (Highest setpoint of channel 1)
Determines a temperature value in channel 1 that when exceeded
upwards triggers the meter alarm (BEEP). When the alarm is
triggered, the indication ALARM is displayed in the meter.
sp1.1 (Lowest setpoint of channel 1)
Determines a temperature value in channel 1 that when exceeded
downwards triggers the meter alarm (BEEP). When the alarm is
triggered, the indication ALARM is displayed in the meter.
spk.2 (Highest setpoint of channel 2)
Determines a temperature value in channel 2 that when exceeded
upwards triggers the meter alarm (BEEP). When the alarm is
triggered, the indication ALARM is displayed in the meter.
spk1.2 (Lowest setpoint of channel 2)
Determines a temperature value in channel 2 that when exceeded
downwards triggers the meter alarm (BEEP). When the alarm is
triggered, the indication ALARM is displayed in the meter.

SmartMeter
3/4
CONFIGURATION CYCLE
To enter the configuration cycle, press
BACK
and
SET
. The meter
displays typ.1, the first parameter of the cycle. To access the
remaining parameters ofthis cycle, press
SET
.Keys
and
make possible to change the parameters defined. To access the
measures screen, go to the end of this cycle or press keys
BACK
and
SET
again.
The parameters of the configuration cycle are:
typ.1
Selects the channel 1 sensor according to Table 1. Channel 1 sensor
is chosen regardless of the channel 2 sensor selection.
typ.2
Selects the channel 2 sensor according to Table 1. Channel 2 sensor
is chosen regardless of the channel 1 sensor selection. Channel 2
indication can be disabled.
Code
Input
oFF
Channel 2 - disabled
0-50
0-50 mV
Rtd
Pt100
Tc b
B thermocouple
Tc s
S thermocouple
Tc r
R thermocouple
Tc n
N thermocouple
Tc e
E thermocouple
Tc t
T thermocouple
tc h
K thermocouple
tc j
J thermocouple
Table 1
dP.P1 (Decimal Point Position)
Enables the display of decimal points for channel 1 measures.
Parameter not valid for thermocouples R, S and B. For the 0-50 mV
input, defines the decimal point position.
dP.P2(Decimal Point Position)
Enables the display of decimal points for channel 2 measures.
Parameter not valid for thermocouples R, S and B. For the 0-50 mV
input, defines the decimal point position.
In.l1 (Lowest indication limit of channel 1 for 0-50 mV)
Determines the inferior limit of the user-created meter indication
range, when the 0-50 mV input signal is selected. Parameter not
displayed for the other input types.
In.k1 (Highest indication limit of channel 1 for 0-50 mV)
Determines the upper limit of the user-created meter indication range,
when the 0-50 mV input signal is selected. Parameter not displayed
for the other input types.
In.l2 (Lowest indication limit of channel 2 for 0-50 mV)
Determines the inferior limit of the user-created meter indication
range, when the 0-50 mV input signal is selected. Parameter not
displayed for the other input types.
In.k2 (Highest indication limit of channel 2 for 0-50 mV)
Determines the upper limit of the user-created meter indication range,
when the 0-50 mV input signal is selected. Parameter not displayed
for the other input types.
0fs.1 (Indication Offset, channel 1)
Allows for an offset in the temperature value displayed in channel 1.
The offset value is simply added to the value measured by the
sensor. The offset can be of 10 % of the selected sensor maximum
range, upwards or downwards.
0fs.2 (Indication Offset, channel 2)
Allows for an offset in the temperature value displayed in channel 2.
The offset value is simply added to the value measured by the
sensor. The offset can be of 10 % of the selected sensor maximum
range, upwards or downwards.
beep (Beep –Sound alarm)
It enables or disables the sound alarm
Beep = no : disabled
Beep = yes : enabled
Note: The ALARM message is displayed regardless of the sound
alarm status (on or off). To stop beeping while an alarm condition is
present, press any key.
Sh.0f (shut-off –Auto power off)
It makes possible to define a time interval in which the meter powers
off automatically, provided no key is pressed within the interval.
Adjustable between 3 and 120 minutes.
vnit (Unit –Temperature indication unit)
It defines the temperature unit of the meter.
ºC for degrees Celsius and ºF Fahrenheit.
Filt (Noise filter)
It adapts the meter frequency to the mains voltage in order to
optimize the internal noise filter. Such adaptation reduces possible
temperature oscillations, resulting from the mains frequency. Options
are:
50 - for 50 Hz mains.
60 - for 60 Hz mains.
Fvnc (Special functions)
Defines whether special functions HOLD, AVG, MAX, MIN and
RELAT will be applied to channels 1 and 2 or only to channel 1.
Ch1 - special functions applied only to channel 1.
Ch1 - special functions applied to channels 1 and 2
4.4 CALIBRATION CYCLE
To enter the calibration cycle, press
BACK
and
SET
for 10 seconds. The
meter displays soft, the first parameter of the cycle. To access the
remaining parameters of this cycle, press
SET
.Keys
and
make possible to change the parameters defined. To access the
measures screen, go to the end of this cycle or press keys
BACK
and
SET
again. The parameters of the configuration cycle are:
Soft (Software version)
About information. Informs the software review installed.
8888 (Serial Number)
About information. Informs the meter serial number. The four larger
numbers are the most significant and the smaller the less significant.
L(.(1 (Low Calibration Channel 1)
INPUT OFFSET CALIBRATION: Adjusts the lowest limit of the input
signal indication range (offset). In order to change the indicated
value, it may be required to press
or
several times.
k(.(1 (High Calibration Channel 1)
INPUT SPAN CALIBRATION: Adjusts the highest limit of the input
signal indication range (span). In order to change the indicated value,
it may be required to press
or
several times.
(j.Lo
Adjusts the room temperature where the meter is held. Important
information for calculations and meter internal compensation.

SmartMeter
4/4
L(.(2 (Low Calibration Channel 2)
INPUT OFFSET CALIBRATION: Adjusts the lowest limit of the input
signal indication range (offset). In order to change the indicated
value, it may be required to press
or
several times.
k(.(2 (High Calibration Channel 2)
INPUT SPAN CALIBRATION: Adjusts the highest limit of the input
signal indication range (span). In order to change the indicated value,
it may be required to press
or
several times.
5CALIBRATION
The meter is factory calibrated, and recalibration by non-experienced
professionals is not recommended. If required, calibration must be
performed by expert professionals, who must proceed as it follows:
a) Configure the type of input to be calibrated: K thermocouple or
Pt100 or 0-50 mV.
Note: By calibrating K thermocouple, all the remaining
thermocouples will be automatically calibrated. Pt100 and 0-50
mV required their own calibration processes.
b) Access the Calibration cycle.
c) Apply to the input channel a signal corresponding to a known
indication, and slightly over the lowest limit of input type
indication.
d) To access parameter L(.(1 (or L(.(2). Use keys
and
until the meter indicates the value corresponding to the applied
signal.
e) Apply to the input channel a signal corresponding to a known
indication, and slightly below the highest limit of input type
indication.
f) Access parameter k(.(1 (or k(.(2. Use keys
and
until
the meter indicates the value corresponding to the applied signal.
g) Repeat cto funtil no adjustment is further required.
Note: While calibrating the meter, observe if the Pt100 excitation
current required by the calibrator used is compliant to the Pt100
excitation current used in this instrument. 0.165 mA.
6WARRANTY
The manufacturer products are covered by a 12-month warranty
provided the purchaser presents the sales receipt and the following
conditions are met:
1. Products are covered for one year from the original date of
purchase. Please retain the dated sales receipt as evidence
of the date of purchase. You will need it for any warranty
service.
2. Within this period, warranty against defects in material and
workmanship under normal use is free of charge.
3. For repair, send the product and the sales receipt to our
address. Expenses and transportation risks are under the
purchaser’s responsibility
4. This warranty does not cover any damage due to accident,
misuse, abuse, or negligence.
Other scigiene Measuring Instrument manuals

scigiene
scigiene CalCheck User manual

scigiene
scigiene TxIsoRail0-10 Vdc User manual

scigiene
scigiene MF-S-HTD User manual

scigiene
scigiene SP-850048 User manual

scigiene
scigiene SP-850001 User manual

scigiene
scigiene SP-840006 User manual

scigiene
scigiene RFP-0254-2 User manual

scigiene
scigiene ATP-4001 User manual

scigiene
scigiene SP-850055 User manual

scigiene
scigiene SP-800048 User manual
Popular Measuring Instrument manuals by other brands

Tractel
Tractel dynafor Pro 1 Installation, operating and maintenance manual

Triplett
Triplett PR600 user manual

Endress+Hauser
Endress+Hauser RID14 Brief operating instructions

Laserworld
Laserworld GREEN-6000 manual

ILX Lightwave
ILX Lightwave OMH-6722B user guide

BRUEL & KJAER
BRUEL & KJAER 1616 instruction manual