Seko KRONOS 20 pH/ORP User manual

0000138014 Rev.1.0
EN
DE
ES
FR
IT
KRONOS pH/ORP
INSTALLATION MANUAL
HANDBUCH
MANUAL DE INSTALACION
MANUEL D’INSTALLATION
MANUALE D’INSTALLAZIONE

0000138014 Rev.1.0

0000138014 rev. 1.0 EN
1
Control panel – KRONOS 20 pH/ORP
Access to the programming menu. During programming it confirms entry to the various menu
levels and the modifications within the same.
Starts and stops the pump.
In the event of a level alarm (alarm function only), flow alarm and active memory alarm, it
deactivates the signal on the display.
Used to “exit” the various menu levels. Before definitively exiting the programming phase, you
will be asked if you wish to save the changes. During the normal operation, by pressing the
key and then the key you can adjust the contrast.
Scrolls the menu or changes the parameters during programming.
Used for priming and tube replacement.
CONNECTIONS DIAGRAM
5
6
7
P
T
1
0
0
PT100
5
6
7
P
T
1
0
0
PT100
Electrical connections
1
Power supply input
2
3
Pole -
pH or ORP probe
input (prewired on
BNC)
4
Pole +
5
PT100 temperature probe input (see
connection diagram)
6
7
8
Remote control input (start-stop)
9
10
Level control probe input
11
12
Flow sensor input
13
1 2
9 8
11
10
13
12
3 4 5 6 7
2-wire PT100 connection diagram
3-wire PT100 connection diagram

0000138014 rev. 1.0 EN
2
KRONOS 20 pH/ORP Programming Menu
Press the keys for more than three seconds to access the programming mode. Press the key to scroll
the menu items, then press the key to access the options. Whenever a menu item is editable, it flashes.
By default the pump is set for constant mode. The pump automatically returns to operating mode after 1 minute of
inactivity. In this case, the data entered will not be saved. Press the key to exit the programming levels. When
you exit the programming mode, the display shows:
to confirm your choice.
Exit
No Save
Exit
Save
PROG
Config.
Paus
N.Open
Alarms
Level
Password
****
Stats
T.Pump
Pump Cal
60.0ml
Setpoint
7.00pH
Set.Type
Acid
Pro.Band
0.50pH
AlarmBnd
0.50pH
OFA
0m
Calibr.
Probe
Temp.Man
25°C
Alarms
Flow
Config.
Funct.
Delay ON
Off
CALDelay
Off
Language
English
Funct.
pH
Funct.
Redox
Funct.
Manual
Main Menu
Settings Menu
Functions
Menu
Paragraph 11
Paragraph 10
Paragraph 9
Paragraph 8
Paragraph 7
Paragraph 6
Paragraph 5
Paragraph 4
Paragraph 2
Paragraph 3
Paragraph 1

0000138014 rev. 1.0 EN
3
Setting the language
Programming
Operation
Allows you to set the language. By default the pump is set to English.
Press to access the item, and then press to set the language.
Press to confirm and return to the main menu.
Paragraph 1 –Manual dosing
Programming
Operation
The pump operates in constant mode. The trigger for the consent is
connected to the Start-Stop input. You can also program:
- Time ON, the operating time of the pump.
- Time OFF, the functional stop of the pump.
Time ON and Time OFF alternate with each other.
Display during operation
–––
Function mode
•Man = Manual
Alarms and statuses
•Lev = Level Alarm
•Flw = Flow Alarm
Value of dosing on going
(depends on unit of measure
selected)
•Percentage, Frequency,
l/h, Gph, ml/m
MAN
P 100%
Pump status
•empty = pump in stating
•Stop = pump in stop
•Paus = pump in pause
PROG
Config.
Config.
Funct.
Delay ON
Off
CALDelay
Off
Language
English
PROG
Config.
Config.
Funct.
Funct.
Manual

0000138014 rev. 1.0 EN
4
Paragraph 2 – Proportional dosing for the pH measurement (factory default)
Programming
Operation
The pump measures and controls the pH value of a solution, by
programming the following parameters: set-point, set-point type,
proportional band and alarm band.
Set-point type: acid
014
pH
s/m
max
flow-rate
Ala rm Band Ala rm Band
Pr o p o r t i o n a l
Ba n d
set-point
Alarm On
Alarm On
Pump Off
P
ump On
Set-point type: alkaline
014
pH
s/m
max
flow-rate
Ala rm Ba nd Ala rm Ba nd
Pr o p o r t i o n a l
Ba n d
set-point
Alarm On
Alarm On
Pump Off
P
ump On
Can also be programmed:
- The O.F.A. (Over Feed Alarm) time in minutes, a set period of time after
which, if the measured pH does not reach the set-point, an alarm signal
is activated.
- Activation / deactivation of the calibration procedure
- Manual temperature value in °C (default) or °F
Display during operation
pH measure value
Value of dosing on going
(depends on unit of measure
selected)
•Percentage, Frequency,
l/h, Gph, ml/m
•
Alarms and statuses
•Cal = Calibration not
performed
•Lev = Level Alarm
•Flw = Flow Alarm
•Alm = Mesure out Band
Alarm
•OFA = O.F.A. Alarm
7.00pH
P 100%
Pump status
•empty = pump in stating
•Stop = pump in stop
•Paus = pump in pause
•
PROG
Config.
Config.
Funct.
Funct.
pH
Setpoint
7.00pH
Set.Type
Acid
Pro.Banda
0.50pH
AlarmBand
0.50pH
OFA
0m
Calibr.
Probe
Temp.Man
25°C
Temp.Man
25°C
Temp.Man
77°F

0000138014 rev. 1.0 EN
5
Paragraph 3 – Proportional dosing for the Redox (O.R.P.) measurement
Programming
Operation
The pump measures and controls the Redox value of a solution, by
programming the following parameters: set-point, set-point type,
proportional band and alarm band.
Set-point type: maximum
0
mV
s/m
max
flow-rate
Ala rm Ba nd Ala rm Ba nd
Pr o p o rt i o n a l
Ba n d
set-point
Alarm On
Alarm On
Pump Off
P
ump On
Set-point type: minimum
0
mV
s/m
max
flow-rate
Ala rm Ba nd Ala rm Ba nd
Pr o p o rt io n a l
Ba n d
set-point
Alarm On
Alarm On
Pump Off
P
ump On
Can also be programmed:
- The O.F.A. (Over Feed Alarm) time in minutes, a set period of time after
which, if the measured Redox value does not reach the set-point, an
alarm signal is activated.
- Activation / deactivation of the calibration procedure
Display during operation
Allarmi e stati
•Cal = calibrazione non
eseguita
•Liv = Allarme livello
•Fls = Allarme flusso
•Alm = Misura fuori
Banda allarme
•OFA = Allarme O.F.A.
Valore misura del
potenziale Redox.
Stato della pompa
•Vuoto = pompa in start
•Stop = pompa ferma
•Paus = pompa in pausa
Valore del dosaggio in corso
(dipende dall’unità di misura
selezionata)
•Percentuale, Frequenza,
l/h, Gph, ml/m
465mV
P 100%
PROG
Config.
Config.
Funct.
Funct.
Redox
Setpoint
0mV
Set. Type
High
ProBand
100mV
AlarmBnd
20mV
OFA
0m
Calibr.
Probe

0000138014 rev. 1.0 EN
6
Paragraph 4 – Setting the Power On Delay
Programming
Operation
Allows you to set a pump operation delay at the start up of
the pump. This delay takes effect only if the pump is
switched off and then switched on by disconnecting the
power supply.
The delay can be disabled, Off (default) or can be set from 1
to 60 minutes.
With the delay enabled, during the set time the LED flashes
(1 sec On – 1 sec Off) and the display shows the countdown
in seconds. If the pump is in Stop mode the display shows
only the flashing LED. During the delay time the function can
be disabled by accessing the menu and setting the time to
Off.
Press to access the item, and then press to set
the value. Press to confirm and return to the main
menu.
Paragraph 5 – Setting the Calibration Delay
Programming
Operation
Allows you to set a pump operation delay after the probe
(Redox or pH) calibration
The delay can be disabled, Off (default) or can be set from 1
to 60 minutes.
With the delay enabled, during the set time the LED flashes
(1 sec On – 1 sec Off) and the display shows the countdown
in seconds. If the pump is in Stop mode the display shows
only the flashing LED. During the delay time the function can
be disabled by accessing the menu and setting the time to
Off.
Press to access the item, and then press to set
the value. Press to confirm and return to the main
menu.
Paragraph 6 – Flow Rate Calibration
Programming
Operation
On the main menu appears, the value in mL in memory
since the last calibration or the default value if it is the first..
By pressing , the pump will dose for 30 seconds and
the display will show the countdown; during that time, by
means of a graduated container you can measure the
dispensed quantity.
After 30 seconds, the display will show the value in memory,
press the key and then to change the value with
the actual amount measured by the graduated container and
confirm with .
The data entered will be used in flow calculations.
PROG
Config.
Config.
Funct.
Delay ON
Off
PROG
Config.
Config.
Funct.
Delay ON
Off
CAL Delay
Off
PROG
Config.
Pump Cal
60.0ml
Quantity
60.0 ml
Quantity
65.0 ml
Pump Cal
ENTER
WAIT
30s

0000138014 rev. 1.0 EN
7
Paragraph 7 – Statistics
Programming
Operation
From the main menu, by pressing the key you can
access the statistics:
- T.Pump = operating time of the pump.
- T.Tube = operating time since the last tube replacement.
- Q.Pump = quantity dosed by the pump in liters; this data is
calculated based on the memorized calibration value.
- Activations = number of pump activations.
- Alm.Meas = number of alarms relating to measure.
- Reset = use the key to reset the counters (YES) or
otherwise (NO), then confirm by pressing .
Press the key to return to the main menu.
Paragraph 8 – Password
Programming
Operation
By setting the password, the programming section can be
accessed to view all the setup parameters, but every time
you try to change the settings you will be prompted for the
password.
The flashing line indicates the editable number; press
to select the number (from 1 to 9), then press to select
the number to modify, and then press to confirm.
By setting “0000” (default), the password will be eliminated.
PROG
Config.
Stats
T.Pompa
T.Pump
0h12m
T.Tube
0h5m
Q.Pump
0,79l
Power
12
Reset Reset?
Yes
Stats
T.Pompa
Alm.Meas
1
PROG
Config.
Password
****
Password
0000

0000138014 rev. 1.0 EN
8
Paragraph 9 – Flow Alarm
Programming
Operation
This allows you to activate (deactivate) the flow sensor.
When activated (On), press the key to set the sensor
as N. Open or N. Closed.
Press the key to return to the main menu.
Confirm and return to the main menu by pressing the
key.
Paragraph 10 – Level Alarm
Programming
Operation
This allows you to activate the level sensor alarm, which
means that you can decide whether to stop the dosage
(Stop) or simply activate the alarm signal without stopping
the dosage.
Changes can be made by pressing the key, then
using the key to set the alarm type. Confirm by
pressing the key.
When activated, press the key to set the sensor as N.
Open or N. Closed.
Press the key to return to the main menu
Confirm and return to the main menu by pressing the
key.
Paragraph 11 – Setting the Pause
Programming
Operation
Remote input to pause the pump.
By default, the system is set to Normally Open.
Press to access the item, and then press to set
the value (N. OPEN or N. CLOSED).
Press to confirm and return to the main menu.
PROG
Config.
Alarms
Flow
Flow
Off
Flow
On
FlowConf
N.Open
Alarms
Flow
PROG
Config.
Alarms
Flow
Flow
Off
Flow
On
FlowConf
N.Open
Alarms
Flow
PROG
Config.
Paus
N.Open
Paus
N.Closed

0000138014 rev. 1.0 EN
9
pH Calibration Menu
You can choose the automatic or manual calibration mode, in both cases the calibration at pH 7 is automatic.
- Automatic calibration:
When the display shows the value of the buffer solution, insert the probe into the bottle, press and the
display will show the countdown of the 60 seconds required to complete the calibration. If the quality of the
alignment is less than 50%, an error appears on the display; press to exit the calibration (after 4 seconds the
pump will automatically exit the calibration); if the quality is higher than 50%, the value appears on the display;
pressing you will be prompted for the pH 4 or 9 buffer solution; at this point the procedure is similar to the
previous calibration.
- Manual calibration:
When the display shows the value of the buffer solution, insert the probe into the bottle, press and the
display will show the countdown of the 60 seconds required to complete the calibration. If the quality of the
alignment is less than 50%, an error appears on the display; press to exit the calibration (after 4 seconds the
pump will automatically exit the calibration); if the quality is higher than 50%, the value appears on the display;
pressing on the display will flash the pH 7.00 value; press to insert the value of my solution, then press
to confirm and start the calibration procedure as described previously.
PROG
Config.
Calibr.
Probe
Calibr.
Auto.
Auto Cal
7pH
WAIT
60 s.
Error
7 pH
7 pH
100% ok
Auto Cal
4/9.25pH
WAIT
60 s.
Error
4 pH
4 pH
100% ok
Calibr.
Manual
Man Cal
7pH
WAIT
60 s.
Error
7.00 pH
7 pH
100% ok
Man Cal
7.00 pH
WAIT
60 s.
Error
4.00 pH
4 pH
100% ok

0000138014 rev. 1.0 EN
10
Redox (O.R.P.) Calibration Menu
You can choose the automatic or manual mode.
- Automatic calibration:
When the display shows the value of the buffer solution, insert the probe into the bottle, press and the
display will show the countdown of the 60 seconds required to complete the calibration. If the quality of the
alignment is less than 50%, an error appears on the display; press to exit the calibration (after 4 seconds the
pump will automatically exit the calibration); if the quality is higher than 50%, the value appears on the display;
press and the procedure is completed.
- Manual calibration:
When the display shows the value of the buffer solution, insert the probe into the bottle, press and the
display will show the value 465mV flashing, insert the probe into my solution, press to display the value of my
solution, then press to confirm and start the calibration procedure as described previously.
PROG
Config.
Calibr.
Probe
Calibr.
Auto.
Auto Cal
475mV
WAIT
60 s.
Error
475mV
475mV
100% ok
Calibr.
Manual Man Cal
475mV
WAIT
60 s.
Error
475mV
475mV
100% ok

0000138014 rev. 1.0 EN
11
Priming
With the pump in Stop (press the key), by pressing the key, the pump will start to operate at maximum
speed for 30 seconds. To stop the pump before the 30 seconds passed, press again the key.
Tube replacement
With the pump in Stop (press the key), by holding down for at least 3 seconds the key, the pump will start
to operate at minimum speed for 60 seconds, with operation and pause intervals, to allow the tube replacement. To
stop the pump before the 60 seconds passed, press again the key.
Alarms
Dislay
Cause
Remedy
Alarm LED on
“Lev” icon flashing
End level alarm, without pump
operation interruption.
Restore the liquid level.
Alarm LED on
“Lev” and “Stop” icons flashing
End level alarm, with pump
operation interruption.
Restore the liquid level.
Alarm LED on
“Flw” icon flashing
Flow alarm activated, the pump has
not received from the flow sensor
the programmed number of signals.
Press the key.
“OFA” icon flashing
“Stop” icon flashing
O.F.A. alarm
Press the key to stop the
flashing of the “Stop” icon, press
the key again to restart the pump.
“Alm” icon flashing
The value read by the probe is out
of the alarm band set
Check the "Alarm Band" parameter
for the correct settings in
programming mode.
“Cal” icon flashing
Probe not calibrated alarm
Perform the probe calibration
procedure.
Flashing words Mtr and Stop
Motor block alarm.
Power cycle the system, if the alarm
persists check the installation or the tube
position.
Flashing words Brk and Stop
Tube break alarm.
Reset the tube.

0000138014 rev. 1.0 DE
1
Kontrolpanell - KRONOS 20 pH/ORP
Zugang zum Programmiermenü. Während des programmierens bestätigen Sie mit dieser
Taste den Zugang zu den verschiedenen Menüebenenen und Modifikationen.
Startet und stoppt die Pumpe.
Sollte ein Levelalarm,Flowalarm oder Memoryalarm anliegen, können Sie diesen mit ESC
deaktivieren.
Wird benutzt um die Menüebenen zu verlassen. Bevor Sie das Menü endgültig verlassen,
werden Sie gefragt, ob Sie vorgenommene Änderungen speichern möchten. Während des
normalen Betriebes könn Sie mit halten dieser Taste und aschließend dieser
Taste, den Kontrast einstellen.
Zum Scrollen im Menü oder zur Änderung der Parameter während der Porgrammierung. Die
Befüllung oder das wechseln des Schlauches wird mit dieser Taste aktiviert.
ANSCHLUSSPLAN
5
6
7
P
T
1
0
0
PT100
5
6
7
P
T
1
0
0
PT100
Electrical connections
1
Stromversorgungseingang
2
3
Pole -
Eingang der pH oder
Redox Sondeneingang
(vorverdrahtet mit BNC)
4
Pole +
5
PT100 Eingang der Temperatursonde
(siehe Anschlussplan )
6
7
8
Fernsteuereingang (start-stopp)
9
10
Levelregler Sondeneingang
11
12
Durchflusssensoreingang
13
1 2
9 8
11
10
13
12
3 4 5 6 7
2-Draht PT100 Anschlussplan
3-Draht PT100 Anschlussplan

0000138014 rev. 1.0 DE
2
KRONOS 20 pH/ORP Menü Programmierung
Drücken Sie die Tasten für mindestens drei Sekunden um in das Menü Programmierung zu gelangen.
Drücken Sie die Taste um durch das Menü zu navigieren und dann die Taste um die Optionen zu
öffnen. Immer wenn ein Menüeintrag blinkt kann er bearbeitet werden.
Die Pumpe ist standardmäßig auf Dauerbetrieb eingestellt. Die Pumpe kehrt nach 1 Minute Inaktivität automatisch
in den Betriebsmodus zurück. In diesem Fall werden die eingegebenen Daten nicht gespeichert. Drücken Sie die
Taste um das Menü Programmierung zu verlassen. Wenn Sie diesen verlasen werden Sie folgendes auf dem
Bildschirm sehen:
um Ihre Auswahl zu bestätigen.
Verlass.
Verwrf.
Verlass.
Sichern
PROG
Konfig.
Paus
N.Offen
Allarmi
Fuellst
Passwort
****
Statist.
Z.Pumpe
Pmp Kalib
60.0ml
Sollwert
7.00pH
Sollw Typ
Acid
Prop Band
0.50pH
Alrm Band
0.50pH
OFA
0m
Kalibrt
Sonda
Temp.Man
25°C
Allarmi
Fluss
Konfig.
PmpFunkt
Verz.Pwr
Off
Verz.Kal
Off
Sprache
Deutsch
PmpFunkt
pH
PmpFunkt
Redox
PmpFunkt
Manuale
Main
menu
Settings
menu
Function
menu
Paragraph 11
Paragraph 10
Paragraph 9
Paragraph 8
Paragraph 7
Paragrafo 6
Paragrafo 5
Paragraph 4
Paragraph 2
Paragraph 3
Paragraph 1

0000138014 rev. 1.0 DE
3
Sprache einstellen
Programmierung
Bedienung
Ermöglicht die Sprache zu ändern. Die Pumpe ist standardmäßig auf
Englisch eingestellt.
Drücken Sie die Taste um den Menüpunkt zu öffnen, drücken Sie
danach um die Sprache einzustellen. Drücken Sie die um zu
bestätigen und zurück zum Hauptmenü zu gelangen.
Paragraph 1 – Manuelle Dosierung
Programmierung
Bedienung
Die Pumpe arbeitet im konstanten Modus. Das Signal zur Freigabe wird
am Start-Stopeingang angeschlossen. Zusätzlich können Sie wie folgt
programmieren:
Time ON, ist die Zeit in der die Pumpe dosiert,
Time OFF, ist die Zeit in der die Pumpe zwar an ist, aber nicht dosiert.
Display während des Betriebs
–––
Funktionsmodus
•Man = Manuell
Alarm and Status
•Lev = Level Alarm
•Flw = Flow Alarm
Wert der laufenden Dosierung
(abhängig von gewählter
Maßeinheit)
•Prozent, Frequenz, l/h,
Gph, ml/m
Pumpenstatus
•Leer = Pumpe am Saugen
•Stopp = Pumpe im Stopp
•Pause = Pumpe in Pause
MAN
P 100%
PROG
Konfig.
Konfig.
PmpFunkt
Verz.Pwr
Off
Verz.Kal
Off
Sprache
Deutsch
PROG
Konfig.
Config.
PmpFunkt
PmpFunkt
Bedien
Zeit ON
0’00”
Zeit Off
0’00”

0000138014 rev. 1.0 DE
4
Paragraph 2 – Proportionale Dosierung für die pH Messung (Werkseinstellungen)
Programmierung
Bedienung
Die Pumpe misst und kontrolliert den pH Wert der Lösung,
nachdem folgende Parameter eingestellt werden: Sollwert,
Sollwert-Typ. Proportionales Band und Alarm Band.
Sollwert-Typ: Säure
014
pH
s/m
max
flow-rate
Ala rm Ba nd Ala rm Band
Pr o p o r t i o n a l
Ba n d
set-point
Alarm On
Alarm On
Pump Off
P
ump On
Sollwert-Typ: Alkalisch
014
pH
s/m
max
flow-rate
Ala rm Ba nd Ala rm Ba nd
Pr o p o r t i o n a l
Ba n d
set-point
Alarm On
Alarm On
Pump Off
P
ump On
Auch programmierbar:
- Der O.F.A (Over Feed Alarm), ist ein Zeitraum in Minuten. Dieser
schlägt nach einem bestimmten Zeitraum Alarm, wenn der
gemessene pH Wert nicht dem eingestellten Wert entspricht.
- Aktivierung/Deaktivierung der Kalibrierung
- Der manuelle Temperaturwert in °C (Standard) oder °F
Display während des Betriebs
pH Messwert
Wert der laufenden Dosierung
(abhängig von gewählter
Maßeinheit)
•Prozent, Frequenz, l/h,
Gph, ml/m
Alarm and Statuse
•Cal = Kalibrierung nicht
vorgenommen
•Lev = Level Alarm
•Flw = Flow Alarm
•Alm = Messung außerhalb
des Bands Alarm
•OFA = O.F.A. Alarm
Pumpenstatus
•Leer = Pumpe am Saugen
•Stopp = Pumpe im Stopp
•
Pause = Pumpe in Pause
7.00pH
P 100%
PROG
Konfig.
Konfig.
PmpFunkt
PmpFunkt
pH
Sollwert
7.00pH
Sollw Typ
Acido
Prop. Band
0.50pH
Alrm Band
0.50pH
OFA
0m
Kalibrt
Probe
Temp.Man
25°C
Temp.Man
25°C
Temp.Man
77°F

0000138014 rev. 1.0 DE
5
Paragraph 3 – Proportionale Dosierung für die Redox (O.R.P) Messung
Programmierung
Bedienung
Die Pumpe misst und kontrolliert den Redox Wert der Lösung, nachdem
folgende Parameter eingestellt werden: Sollwert, Sollwert-Typ.
Proportionales Band und Alarm Band.
Sollwert-Typ: Maximum
0
mV
s/m
max
flow-rate
Ala rm Ba nd Ala rm Ba nd
Pr o p o rt i o n a l
Ba n d
set-point
Alarm On
Alarm On
Pump Off
P
ump On
Sollwert-Typ: Minimum
0
mV
s/m
max
flow-rate
Ala rm Ba nd Ala rm Ba nd
Pr o p o rt i o n a l
Ba n d
set-point
Alarm On
Alarm On
Pump Off
P
ump On
Auch programmierbar:
- Der O.F.A (Over Feed Alarm), ist ein Zeitraum in Minuten. Dieser schlägt
nach einem bestimmten Zeitraum Alarm, wenn der gemessene Redox
Wert nicht dem eingestellten Wert entspricht.
- Aktivierung/Deaktivierung der Kalibrierung
Display während des Betriebs
465mV
P 100%
Redox Messwert
Wert der laufenden Dosierung
(abhängig von gewählter
Maßeinheit)
•Prozent, Frequenz, l/h,
Gph, ml/m
Alarm and Statuse
•Cal = Kalibrierung nicht
vorgenommen
•Lev = Level Alarm
•Flw = Flow Alarm
•Alm = Messung außerhalb
des Bands Alarm
•OFA = O.F.A. Alarm
Pumpenstatus
•Leer = Pumpe am Saugen
•Stopp = Pumpe im Stopp
•
Pause = Pumpe in Pause
465mV
P 100%
PROG
Konfig.
Konfig.
PmpFunkt
PmpFunkt
Redox
Sollwert
0mV
Solle Typ
Hoch
PropBand
100mV
Alrm Band
20mV
OFA
0m
Kalibrt
Probe

0000138014 rev. 1.0 DE
6
Paragraph 4 – Einstellen der Einschaltverzögerung
Programmierung
Bedienung
Ermöglicht Ihnen eine verzögerte Inbetriebnahme der
Pumpe beim Start. Diese Verzögerung kann nur aktiv
werden, wenn Sie die Pumpe ausschalten und dann wieder
an, indem Sie die Stromzufuhr trennen.
Die Verzögerung kann ausgeschaltet werden, Off
(Standard), oder von 1 zu 60 Minuten eingestellt werden.
Wenn die Verzögerung eingeschaltet ist blinkt die LED
während dieser Zeit (1 Sek. An - 1 Sek. Aus) und der
Bildschirm zeigt der Countdown an. Falls die Pumpe im
Stopp Modus ist blinkt lediglich die LED. Während der
Verzögerungszeit kann man diese auch im Menü wieder
ausstellen.
Drücken Sie die Taste um den Menüpunkt zu öffnen,
drücken Sie danach um den Wert einzustellen.
Drücken Sie die um zu bestätigen und zurück zum
Hauptmenü zu gelangen.
Paragraph 5 – Einstellen der Kalibrierungsverzögerung
Programmierung
Bedienung
Ermöglicht Ihnen eine verzögerte Inbetriebnahme der
Pumpe nach der Sonden- (Redox oder Ph) Kalibrierung.
Die Verzögerung kann ausgeschaltet werden, Off
(Standard), oder von 1 zu 60 Minuten eingestellt werden.
Wenn die Verzögerung eingeschaltet ist blinkt die LED
während dieser Zeit (1 Sek. An - 1 Sek. Aus) und der
Bildschirm zeigt der Countdown an. Falls die Pumpe im
Stopp Modus ist blinkt lediglich die LED. Während der
Verzögerungszeit kann man diese auch im Menü wieder
ausstellen.
Drücken Sie die Taste um den Menüpunkt zu öffnen,
drücken Sie danach um den Wert einzustellen.
Drücken Sie die um zu bestätigen und zurück zum
Hauptmenü zu gelangen.
Paragraph 6 – Kalibrierung der Fließrate
Programmierung
Bedienung
Im Hauptmenü erscheint ein Wert in ml, welcher seit der
letzten Kalibrierung oder noch als Default-Wert gespeichert
ist.
Durch Drücken dieser Taste, dosiert die Pumpe für 30
Sekunden und zeigt den Countdown im Display
an.Währendessen wird mittels eines abgemessenen
Behälters ein Wert bestimmt, nachdem Sie die dosierte
Menge ermitteln können.
Nach 30 Sekunden,zeigt das Display den gespeicherten
Wert an , drücken Siec diese Taste, danach drücken
Sie , um den Wert auf den aktuell gemessenen Wert zu
verändern und drücken diese Taste zur Bestätigung.
Die eingegebenen Werte werden zur Flow-Berechnung
verwendet.
PROG
Konfig.
Konfig.
PmpFunkt
Verz.Pwr
Off
PROG
Konfig.
Konfig.
PmpFunkt
Verz.Pwr
Aus
Verz.Kal
Aus
PROG
Konfig.
Pmp Kalib
60.0ml
Pmp Kalib
ENTER
Wait
30s
Anzahl
60.0ml
Anzahl
65.0ml

0000138014 rev. 1.0 DE
7
Paragraph 7 – Statistiken
Programmierung
Bedienung
Vom Hauptmenü ausgehend, gelangen Sie durch Drücken
dieser Taste zu den Statistiken:
- Z.Pumpe = Laufzeit der Pumpe
- Z.Tube = Laufzeit der Pumpe, seit Schlauchtausch.
- A.Pumpe = Dosierte Menge der Pumpe, welcher vom
gespeicherten Kalibrierungswert abhängt.
- Netzstr.= Zahl der Aktivierungen der Pumpe
- Alarm = Zahl der gemessenen Alarme
- Ruckset = benutzen Sie diese Taste, um die Zähler zu
reseten. Wählen Sie ,,Ja“ oder ,,Nein“ und bestätigen Sie mit
.
Drücken Sie diese Taste, um zum Hauptmenü zurück
zu kehren.
Paragraph 8 – Passwort
Programmierung
Bedienung
Wenn Sie ein Passwort anlegen können Sie den Bereich
Programmierung und alle Setup Parameter einsehen, aber
jedes Mal wenn Sie etwas ändern möchten werden Sie
aufgefordert das Passwort einzugeben.
Die blinkende Reihe zeigt die editierbare Zahl an; drücken
Sie die Taste um eine Zahl auszuwählen (1 bis 9),
danach drücken Sie die Taste um die zu bearbeitende
Zahl auszuwählen und danach die Taste zum
bestätigen.
Wenn Sie "0000" (Standard) wird das Passwort gelöscht.
PROG
Konfig.
Statist.
Z.Pumpe
Z.Pumpe
0h12m
Z.Tube
0h5m
A.Pumpe
0,79l
Netzstr.
12
Ruckset RstStat?
Ja
Statist.
Z.Pumpe
Alarm
2
PROG
Konfig.
Passwort
****
Passwort
0000
Table of contents
Languages:
Other Seko Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Frontier Labs
Frontier Labs BAR-LT quick start guide

Sonel
Sonel CMP-1010 user manual

DEVA Broadcast
DEVA Broadcast Radio Explorer II Maintenance and operation instruction manual

Datacolor
Datacolor 45 Quick start instructions

TASKING
TASKING iSYSTEM iC5700 user manual

Komshine
Komshine KLS-35m instruction manual