Analog Output Adjustment Metis MS/MI
Maintenance Manual Metis MS / MI
4
*)Defining the beginning of temperature range: The potentiometer must be set a few degrees above the
beginning of the temperature range in order to be able to set a current value of 0.2 mA. The necessary
temperature must be calculated, it depends on the pyrometer temperature range as well as a possibly
adjusted sub range.
Attention: Matching with the potentiometers affects each other, so these steps must be repeated until the
output is set for both currents.
4 Assembly
The assembly takes place in reverse order of disassembly:
Remove pyrometer connector and flat cable connector at rear cover
Push the optics unit with the printed circuit boards into the housing (the laser sticker points to the front) and
tighten the hexagon screws corresponding to the tightening torques for aluminum castings.
Connect the plug of the flat cable to the printed circuit board on the back cover (if necessary, solder the
cable from the video module (same color) and insulate with shrink tubing.
Carefully push the rear cover into the housing and screw it with the 4 hexagon socket screws.
Slider to the left to set the current for the beginning of temperature range
Analog output 0-20 mA: adjust until the amperemeter indicates 0.2 mA
Slider to the right to adjust the current for the end of the measuring range:
adjust until the amperemeter indicates 20 mA (+16 µA)
Move slider to the left (a few
degrees above the beginning
of the measuring range *)) or
right (or enter simulation
temperature)