sewerin ATS 503 User manual

14.01.2021 a – 108537 – en
ATS 503/501
Test set
Operating Instructions

ATS 503/501
Canister holder
Fig. 1: ATS 503 – front view
Display
Fig. 2: ATS 503/501 – control panel
Locking mechanism
Device holder
Control panel
Menu key
Up key
Enter key
Down key
ON/OFF key

ATS 503/501
Fig. 3: ATS 503 – view from above
Test gas connections
Connection for power
supply
Fig. 5: ATS 503/501 – rear (section)
Cable guide
Air connection
Fig. 4: ATS 503 – tilted position
Supporting bracket
USB port

Illustration of warnings in this document
NOTICE!
Risk of damage to property.

Contents│ I
1 Introduction .............................................................................1
1.1 Information about this document...............................................1
1.2 Purpose.....................................................................................2
1.3 Intended use .............................................................................2
1.4 General requirement for use .....................................................2
1.5 Safety information .....................................................................3
1.6 Qualicationofusers.................................................................4
2 Product description ................................................................6
2.1 Product variants ........................................................................6
2.2 Features ....................................................................................6
2.3 Visual and audible signals.........................................................7
2.4 Power supply.............................................................................7
2.5 Menu .........................................................................................8
3 Preparing for start-up ...........................................................10
3.1 Suitable environment ..............................................................10
3.2 Position of use.........................................................................10
3.2.1 Overview ..............................................................................10
3.2.2 Tilted position ....................................................................... 11
3.2.3 Wall mounting.......................................................................12
4 Operation ...............................................................................15
4.1 Switching on the test set .........................................................15
4.2 Switchingothetestset .........................................................17
4.3 Placing the device in the test set.............................................18
4.4 Navigating ...............................................................................19
4.4.1 Orientation aids on the display.............................................19
4.4.2 Key functions........................................................................21
4.4.3 Switching between levels .....................................................22
4.4.4 Cancel the action or process................................................22
4.4.5 Scrolling ...............................................................................22
4.4.6 Select values........................................................................23
4.4.7 Answer questions.................................................................23
4.4.8 Lists – view detailed information ..........................................24
5 Test gases..............................................................................25
5.1 Test gas connections for the test set.......................................25
5.1.1 ATS 503 ...............................................................................25
5.1.2 ATS 501 ...............................................................................26

II │Contents
5.2 Utilisable test gases ................................................................26
5.3 Available test gases ................................................................27
5.4 Planning for gas assignment...................................................27
5.5 Connecting test gases.............................................................28
5.5.1 Screwing on the test gas can ...............................................28
5.5.2 Connecting a test gas bottle.................................................29
5.6 Assigning test gases to test gas connections (gas
assignment).............................................................................29
5.7 Non-use of test set ..................................................................30
6 Settings ..................................................................................31
6.1 General information about the settings ...................................31
6.2 Settings using the GasCom software......................................31
6.3 Settings using the test set .......................................................32
6.3.1 Possible settings ..................................................................32
6.3.1.1 Gas assignment ................................................................32
6.3.1.2 ATS mode .........................................................................32
6.3.1.3 Device inspection mode....................................................33
6.3.1.4 Timer .................................................................................34
6.3.2 Changing settings ................................................................34
7 Device inspections ...............................................................36
7.1 Ways of carrying out the procedure ........................................36
7.2 Requirements..........................................................................36
7.3 Features ..................................................................................37
7.3.1 All operating modes .............................................................37
7.3.2 Manual operating mode: Special features and variants of
the device inspection............................................................37
7.3.3 Automatic and Timer operating modes: Special features.....38
7.4 Subtests ..................................................................................39
7.5 Due date..................................................................................40
7.6 Presentation of results ............................................................40
7.7 Performing the device inspection ............................................41
7.7.1 Manual operating mode .......................................................41
7.7.2 Automatic and Timer operating modes ................................43
8 Adjusting devices ................................................................45
8.1 Ways of carrying out the procedure ........................................45
8.2 Requirements..........................................................................45
8.3 Frequency ...............................................................................45

Contents│ III
8.4 Features ..................................................................................46
8.5 Presentation of results ............................................................46
8.6 Performing an adjustment .......................................................47
9 Protocols and information ...................................................48
9.1 Protocols .................................................................................48
9.2 Information ..............................................................................50
10 Maintenance of the test set ..................................................52
10.1 Servicing .................................................................................52
10.2 Care ........................................................................................52
10.3 Replacing the rubber gasket ...................................................53
11 Faults and problems .............................................................54
11.1 Messages in the event of faults...............................................54
11.2 Monitoring test gas pressure...................................................58
11.3 Resolution of problems ...........................................................58
12 Appendix................................................................................61
12.1 Technical data .........................................................................61
12.2 Preset test gases ....................................................................63
12.3 Limit values for device inspection ...........................................63
12.4 Symbols ..................................................................................64
12.4.1 Symbols on the housing.......................................................64
12.4.2 Symbols on the display ........................................................64
12.5 Accessories and consumables................................................67
12.6 Advice on disposal ..................................................................68
12.7 Declaration of conformity ........................................................69
12.8 Abbreviations ..........................................................................69
12.9 Technical terms .......................................................................69
12.10 Conversion of the concentration data .....................................69
13 Index.......................................................................................70

1 Introduction│ 1
1 Introduction
1.1 Information about this document
This document is a component part of the product.
●Read the document before putting the product into operation.
●Keep the document within easy reach.
●Pass this document on to any subsequent owners.
● Unlessotherwisespecied,theinformationinthisdocument
refers to the product as delivered (factory settings) and applies
to all product variants.
●
Notethefollowingterminologicalspecicationsthatapplyto
this document:
SEWERIN product name Name in document
ATS 503/501 – Test set
– Product
EX-TEC
PM 580/550/500/400
– Device
– Gas concentration measuring
device
●
The product is described with all device features. Not all of
the functions described may be available on the product you
are using. Please contact the SEWERIN sales department for
further information.
●Contradictory national legal regulations take precedence over
the information in this document.
Translations
Translations are produced to the best of our knowledge. The
original German version is authoritative.
Right of reproduction
Nopartofthisdocumentmaybeedited,duplicatedorcirculated
in any form without the express consent of Hermann Sewerin
GmbH.

2 │1Introduction
Registered trademarks
Registered trademarks are generally not indicated in this docu-
ment.
1.2 Purpose
The ATS 503 or ATS 501 test set can be used to perform the
following maintenance activities:
●Device inspection
●Adjustment
The test set is suitable for the following gas concentration meas-
uring devices:
●EX-TEC PM 580/550/500/400
●EX-TEC PM 580/550/500/400 CSA
The test set can also be used:
●for charging the rechargeable batteries of devices
●for reading protocols from devices to a computer
1.3 Intended use
The product is suitable for the following uses:
●professional
●industrial
●commercial
Theproductmustonlybeusedfortheapplicationsspeciedin
section 1.2.
1.4 General requirement for use
Only device inspections and adjustments that are technically pos-
sible can be performed using the test set.
Whether a device inspection/adjustment is technically possible
depends on:
−Device: variant and features
−Test set: current gas assignment

1 Introduction│ 3
1.5 Safety information
This product was manufactured in keeping with all binding legal
and safety regulations.
The product is safe to operate when used in accordance with the
instructionsprovided.However,whenhandlingtheproduct,there
mayberiskstopersonsandproperty.Forthisreason,observe
the following safety information without fail.
●Observe all the applicable safety standards and accident pre-
vention regulations.
●Use the product only as intended.
●
Donotmakeany changes or modications totheproduct
unless these have been expressly approved by Hermann
Sewerin GmbH.
●
Only use accessories and consumables approved by Hermann
Sewerin GmbH.
●Always observe the permitted operating and storage temper-
atures.
● Handletheproductcarefullyandsafely,bothduringtransport
and when working.
●Do not use the product if it is damaged or faulty.
● Protecttheportsandsocketsagainstdirt,andelectricalports
in particular against moisture.
●Never carry the test set by the locking mechanism or the sup-
porting bracket.
●Only operate the test set in well ventilated rooms.
●
Do not operate the test set in potentially explosive or oxy-
gen-enriched atmospheres.
●Handle test gases in a professional manner.
●
The test set complies with the requirements of Directive
2014/30/EU (EMC). When using the product close to devices
with radio transmitters:
−Comply with the information in the operating instructions for
the devices regarding emission of interference.

4 │1Introduction
1.6 Qualication of users
The appropriate specialist knowledge is required for using this
product.Inaddition,certainactivitiesmaybeperformedonlyby
qualiedpersons.
A distinction is made between the following user groups:
Trained person
−
Trainedpersonsknowtheexternalstructure,functionsand
operating instructions of the devices and the test set.
−Trained persons can detect defects or changes to the device
and the test set.
−Trained persons may operate the test set and connect or re-
place test gases.
Specialist (device inspector)
−Specialists have received professional training and thus have
fundamental knowledge of the measuring principles used in
gas concentration measuring devices.
−
Specialistsareexperiencedusersandthereforehavesucient
knowledge of the use of gas concentration measuring devices.
−
Specialists are allowed to carry out the activities of trained per-
sonsand,inaddition,commissionthetestsetandcongure
the GasCom software.
Competent person
SEWERIN service personnel and people trained by SEWERIN
are competent persons.
−Competent persons have knowledge of the applicable regula-
tions and guidelines as well as the tasks and functions of gas
concentration measuring devices.
−Competent persons can evaluate whether the condition of de-
vices and the test set renders them safe to operate and can
perform the necessary operations without supervision.
−Competent persons are appointed by their company in writing.
Theirresponsibilitiesaredenedwhentheyareappointed.
They are authorised by the manufacturer.

1 Introduction│ 5
−Competent persons must receive regular training.
−Competent persons are allowed to carry out the activities of
specialists and also to service the test set.

6 │2Productdescription
2 Product description
2.1 Product variants
The following product variants of the test set are available:
●ATS 503
●ATS 501
Theproductvariantsdierinthenumberoftestgasconnections.
ATS 503 ATS 501
Test gas connections (total) 3 1
– for test gas cans 2 1
– for test gas bottles 1 –
Air connection 1 1
2.2 Features
Youwillndanoverviewwiththenamesofthepartsinsidethe
frontcover(g.1tog.5).
Device holder and locking mechanism
The device holder is a positive-locking recess in the housing of
the test set for holding a device.
The device is secured in the device holder by the locking mecha-
nism.Duringadeviceinspectionoradjustment,testgasorclean
air is supplied to the device via the locking mechanism.
Test gas connections
The test gases used to perform device inspections or adjust-
ments are connected to the test gas connections.
Detailed information on test gas connections can be found in
(section 5.1).

2 Product description│ 7
Air connection
Clean air is drawn in via the air connection during a device in-
spection or adjustment.
The air can be drawn in:
●Directly
No accessories are connected to the air connections.
●Using an ambient air hose
Anambientairhoseisconnectedtotheairconnection,withthe
free end positioned in a place where there is clean air.
Forexample,theambientairhosecanberoutedthroughan
open window so that ambient air is drawn in instead of room air.
The ambient air hose can be purchased as an accessory.
The air connection is equipped with an internal carbon dioxide
lter.ThelteristestedbySEWERINServiceaspartofthean-
nual maintenance.
Electrical connections
The test set has the following electrical connections:
●USB port
−for data exchange with a computer
●Connection for power supply
−for powering the test set
−for charging the rechargeable batteries of a device
2.3 Visual and audible signals
The test set does not emit any signals. If visual signals are visible
andaudiblesignalsareaudible,theycomefromthedevicethat
is being used.
2.4 Power supply
The test set is powered by an external power source. The follow-
ing is required for connection:
●AC/DC adapter M4

8 │2Productdescription
2.5 Menu
Themenu(g.6)isaccessedusingthemenukey.Whichmenu
level is visible depends on the situation.
In the menu the user can:
●Perform actions
●Perform settings
●View information
Device inspection
………….. All
Warning
Warning ECO
Measuring
Structure
Adjustment
………….. All
Warning
Measuring
Structure
Settings
………….. Gas assignment ………….. Gas 1
Gas 2
Gas 3
Operating mode (ATS)
………….. Manual
Automatic
Timer
Device inspection mode
………….. Without adjustment
With adjustment
in case of error
Timer
………….. Days of the week
Start time
Protocols
………….. Display
Delete
Information
………….. ATS
Gas status
Connected device (PM)
PIN code
Service
Explanation:
top menu level
Menu item is only displayed in certain situations
Fig. 6: Menu structure

2 Product description│ 9
Information on some menu items
●Warning,Warning ECO,Measuring,Structure
The menu items are only displayed if a device inspection/ad-
justment is technically possible (section 1.4).
Ifadeviceinspection/adjustmentisnottechnicallypossible,
the Warning menu item will be displayed. If this menu item is
thenselected,anerrormessagewillappear.
●All
This menu item is only displayed if more than one device in-
spection/adjustment is due.
●Service
This menu item is protected by a PIN code and only accessible
to authorised users.

10 │3Preparingforstart-up
3 Preparing for start-up
Note:
Start-up may only be carried out by specialist technicians.
3.1 Suitable environment
The test set may only be operated in spaces that meet the fol-
lowing requirements:
●clean ambient air
●well ventilated
●dry
●dust-free
●vibration-free
●no direct sunlight on test gas cans
If the set-up location does not meet the ambient air and ventila-
tionrequirements,thetestsetcanstillbeoperatedatthisloca-
tion,providedthatanambientairhoseisused.Thehosemust
be positioned so that clean air can be drawn in.
3.2 Position of use
3.2.1 Overview
The test set may be operated in the following operating positions:
● placedonahorizontal,atsurface
−upright
−tilted and supported by the bracket
● mountedonavertical,load-bearingsurface(wallmounting)

3 Preparing for start-up│ 11
3.2.2 Tilted position
The test set can be tilted. It is supported by the supporting brack-
et.
NOTICE!
Whentilted,thesupportingbracketisnotdesignedforforces
applied from above.
●Never exert pressure on a test set that is tilted.
Fig. 7: Setting up the test set in a tilted position
(test set shown without canister holder and locking mechanism)

12 │3Preparingforstart-up
Setting up the test set in a tilted position (g. 7)
1. Pull the supporting bracket down vertically. Remove the sup-
porting bracket. 1
2.
Plug the device plug of the AC/DC adapter into the power
supply connection of the test set.
3. Slide the supporting bracket into the test set. 2
−Insert the pins into the slots on both sides.
4. Push the supporting bracket upwards as far as it will go. 3
5. Pivot the bracket as far as it will go. Tilt the test set until it is
securely supported by the bracket. 4
3.2.3 Wall mounting
Forwallmounting,thesupportingbracketisscrewedtoaver-
tical,load-bearingsurface.Thetestsetisthenclickedintothe
supporting bracket.
The following mounting materials are included in the scope of
delivery:
−3 wood screws 4.5 x 60 mm and 3 universal plugs 5 x 30 mm
for attaching the supporting bracket to the surface
−2 self-tapping screws
for securing the test set
NOTICE!
The mounting materials supplied may be unsuitable for certain
surfaces. If the mounting material is nevertheless used in such
cases,thesupportingbracketmayripawayfromthesurfaceafter
the test set has been inserted.
●Only use the supplied mounting material if it is suitable for the
surface intended for mounting.

3 Preparing for start-up│ 13
Fig. 8: Mounting the test set on a vertical surface
(test set shown without canister holder and locking mechanism)
This manual suits for next models
1
Table of contents