Shel lab SCO5W User manual

WATER-JACKETED CO2INCUBATOR
110 – 120 Volts
Installation - Operation Manual
SCO5W

2| Page

3 | Page
SCO5W WATER-JACKETED CO2INCUBATORS
110 – 120 Voltage
Part Number (Manual): 4861732
Revision: January 2, 2018
SHEL LAB is a brand of Sheldon Manufacturing, INC.
Safety Certifications
These units are TÜV listed as water jacket incubators for professional, industrial, or educational use
where the preparation or testing of materials is done at an ambient air pressure range of 22.14 – 31.3
inHg (75 – 106 kPa), with no flammable, volatile, or combustible materials being heated.
These units have been tested to the following requirements:
CAN/CSA C22.2 No. 61010-1:2012
CAN/CSA C22.2 No. 61010-2-010:2004 Reaffirmed: 2014-07
UL 61010-1:2012-05
UL 61010A-2-010:2002-03
EN 61010-1:2010
EN 61010-2-010:2014
Supplemented by: UL 61010-2-010:2015

4| Page
TABLE OF CONTENTS
INTRODUCTION .........................................................................................................................................................7
Read this Manual.................................................................................................................................................................... 7
Safety Considerations and Requirements ...................................................................................................................... 7
Contacting Assistance ..........................................................................................................................................................8
Engineering Improvements..................................................................................................................................................8
CO2Gas Supply ......................................................................................................................................................................9
Reference Sensor Devices ................................................................................................................................................ 10
RECEIVING YOUR UNIT ...........................................................................................................................................11
Inspect the Shipment............................................................................................................................................................ 11
Orientation Image ................................................................................................................................................................ 12
Recording Data Plate Information................................................................................................................................... 13
INSTALLATION .........................................................................................................................................................15
Installation Checklist ........................................................................................................................................................... 15
Required Ambient Conditions........................................................................................................................................... 16
Required Clearances........................................................................................................................................................... 16
Environmental Disruption Sources .................................................................................................................................. 16
Power Source Requirements .............................................................................................................................................17
Lifting and Handling ............................................................................................................................................................ 18
Leveling................................................................................................................................................................................... 18
Install the Incubator ............................................................................................................................................................. 19
Deionized and Distilled Water.......................................................................................................................................... 19
Installation - Clean and Disinfect..................................................................................................................................... 19
Install Chamber HEPA Filter and Ceiling Air Duct ..................................................................................................... 20
Shelving Installation............................................................................................................................................................. 21
Access Port Stopper ........................................................................................................................................................... 23
Fill the Water Jacket........................................................................................................................................................... 23
GRAPHIC SYMBOLS ............................................................................................................................................... 25
CONTROL PANEL OVERVIEW .............................................................................................................................. 27
OPERATION.............................................................................................................................................................. 29
Theory of Operation ........................................................................................................................................................... 29
Put the Incubator into Operation ..................................................................................................................................... 31
Humidify the Incubator ...................................................................................................................................................... 33
Muting the Audible Temperature Alarm....................................................................................................................... 35
Automatic Door Cutoff ....................................................................................................................................................... 35
Start a Flow of CO2............................................................................................................................................................. 36
Muting the Audible CO2Alarm ........................................................................................................................................ 38
No Gas Supply Alarm......................................................................................................................................................... 38
Set the Over Temperature Limit...................................................................................................................................... 39
Load the Incubator.............................................................................................................................................................. 40
Accessory Compatibility.................................................................................................................................................... 40
Data Output Capabilities ................................................................................................................................................... 41
Condensation and the Dew Point .................................................................................................................................. 42
USER MAINTENANCE............................................................................................................................................. 43
Cleaning and Disinfecting................................................................................................................................................. 43
Minimizing Contamination Exposure ............................................................................................................................. 44
Gas Lines and HEPA Filters.............................................................................................................................................. 45
Storage of the Incubator ................................................................................................................................................... 45
Maintaining Atmospheric Integrity ................................................................................................................................. 45
Electrical Components....................................................................................................................................................... 45
Replace the Chamber HEPA Filter ................................................................................................................................. 46
Calibrate the Temperature display.................................................................................................................................47
Calibrate the CO2 Display .................................................................................................................................................. 51
Anode and Water Quality ................................................................................................................................................. 55

5 | Page
UNIT SPECIFICATIONS .......................................................................................................................................... 57
Weight......................................................................................................................................................................................57
Dimensions.............................................................................................................................................................................57
Capacity ................................................................................................................................................................................. 58
CO2.......................................................................................................................................................................................... 58
Temperature ......................................................................................................................................................................... 58
Power...................................................................................................................................................................................... 58
PARTS LIST............................................................................................................................................................... 59
Ordering Parts and Consumables.................................................................................................................................. 60
Accessories........................................................................................................................................................................... 60

6| Page
TABLE OF CONTENTS

7 | Page
INTRODUCTION
Thank you for purchasing a SHEL LAB product. We know you have many choices in today’s competitive
marketplace when it comes to constant temperature equipment. We appreciate you choosing ours. We
stand behind our products and will be here for you if you need us.
READ THIS MANUAL
Failure to follow the guidelines and instructions in this user manual may create a protection
impairment by disabling or interfering with the unit safety features. This can result in injury or death.
Before using the unit, read the manual in its entirety to understand how to install, operate, and
maintain the unit in a safe manner. Keep this manual available for use by all operators. Ensure all
operators are given appropriate training before the unit begins service.
SAFETY CONSIDERATIONS AND REQUIREMENTS
Follow basic safety precautions, including all national laws, regulations, and local ordinances in your
area regarding the use of this unit. If you have any questions about local requirements, please
contact the appropriate agencies.
SOPs
Because of the range of potential applications, this unit can be used for, the operator or their
supervisors must draw up a site-specific standard operating procedure (SOP) covering each
application and associated safety guidelines. This SOP must be written and available to all operators
in a language they understand.
Intended Applications and Locations
These incubators are intended for laboratory, industrial, and educational microbiological cultivation
applications. These incubators are not intended for use in hazardous or household locations.
Power
Your unit and its recommended accessories are designed and tested to meet strict safety
requirements.
•The unit is designed to connect to a power source using the specific power cord type
shipped with the unit.
•Always plug the unit power cord into a protective earth grounded electrical outlet
conforming to national and local electrical codes. If the unit is not grounded properly, parts
such as knobs and controls can conduct electricity and cause serious injury.
•Do not bend the power cord excessively, step on it, or place heavy objects on it.
•A damaged cord can be a shock or fire hazard. Never use a power cord if it is damaged or
altered in any way.
•Use only approved accessories. Do not modify system components. Any alterations or
modifications to your unit not explicitly authorized by the manufacturer can be dangerous
and will void your warranty.

8| Page
INTRODUCTION
CONTACTING ASSISTANCE
Phone hours for Sheldon Technical Support are 6 am – 4:30 pm Pacific Coast Time (west coast of
the United States, UTC -8), Monday through Friday. Please have the following information ready
when calling or emailing Technical Support: the model number and the serial number (see page 12).
EMAIL: [email protected]
PHONE: 1-800-322-4897 extension 4, or (503) 640-3000
FAX: (503) 640-1366
Sheldon Manufacturing, INC.
P.O. Box 627
Cornelius, OR 97113
ENGINEERING IMPROVEMENTS
Sheldon Manufacturing continually improves all of its products. As a result, engineering changes and
improvements are made from time to time. Therefore, some changes, modifications, and
improvements may not be covered in this manual. If your unit’s operating characteristics or
appearance differs from those described in this manual, please contact your SHEL LAB dealer or
customer service representative for assistance.

9 | Page
INTRODUCTION
Note: A CO2gas regulator must be purchased separately from the incubator.
CO2GAS SUPPLY
Supply Required
The incubator must be connected to a carbon dioxide gas supply system in order to
establish and maintain a CO2-enriched incubation chamber atmosphere. The supply
can be a building CO2gas system or one or two supply cylinders (tanks).
Supply Quality
Use medical or food grade CO2. Use of industrial CO2 risks introducing contaminants into the chamber,
may damage the incubator, and will void the manufacturing defect warranty.
Supply Source and Pressure
The incubator requires 15 - 20 psi of CO2gas pressure at the incubator intake port (labeled CO2 to
Chamber).
Two Stage Regulator
If connecting to a supply cylinder, always use a two-stage CO2pressure regulator. During normal
operations, the incubator uses small quantities of CO2to maintain the chamber gas concentration.
Under these conditions, precise regulation of the gas input flow is vital for the incubator
performance. Some single-stage regulators have two gauges. Make certain your regulator is a two-
stage regulator.
Regulator Not Included
See page 60 for ordering information.
Dual Cylinder Connection.
The incubator can be connected simultaneously to two CO2supply cylinders. The unit
automatically switches drawing from one cylinder to the other when the first cylinder
runs out of CO2. Two gas regulators are required for this configuration.

10 | Page
INTRODUCTION
REFERENCE SENSOR DEVICES
Must be purchased separately
Reference sensor devices or a combined device are required for calibrating the
incubator temperature and CO2displays.
Reference devices must meet the following standards:
•Accurate to at least 0.1°C
•Accurate to at least 0.1% CO2
The devices should be regularly calibrated, preferably by a third party.
Temperature Probes
For temperature, use a digital device with a wire thermocouple probe that can be
introduced into the incubation chamber through the unit access port. Select a
thermocouple suitable for the application temperature you will be calibrating at.
CO2 Sampling
For best CO2accuracy, use a digital gas analyzer with sample tubing connecting to
the incubator CO2sample port. The barbed adapter gas fitting on the port connects
to 3/16-inch (4.76mm) inner diameter (ID), 5/16-inch (7.94mm) outside diameter (OD)
tubing.
Why Probes and Tubing?
Reference readings taken outside the chamber using wire temperature probes and
CO2drawn through the sample port avoid chamber door openings. Openings disrupt
the chamber temperature and CO2 concentration level. Each disruption requires a
minimum 1-hour wait to allow the atmosphere to re-stabilize before continuing.
No Alcohol or Mercury Thermometers
Alcohol thermometers do not have sufficient accuracy to conduct accurate
temperature calibrations. Never place a mercury thermometer in the incubation
chamber. Always use thermocouple probes.
CO
2
Reference
Temperature
Reference

11 | Page
RECEIVING YOUR UNIT
INSPECT THE SHIPMENT
•When a unit leaves the factory, safe delivery becomes the responsibility of the carrier.
•Damage sustained during transit is not covered by the manufacturing defect warranty.
•Save the shipping carton until you are certain that the unit and its accessories function
properly.
When you receive your unit, inspect it for concealed loss or damage to its interior and exterior. If you
find any damage to the unit, follow the carrier’s procedure for claiming damage or loss.
1. Carefully inspect the shipping carton for damage.
2. Report any damage to the carrier service that delivered the unit.
3. If the carton is not damaged, open the carton and remove the contents.
4. Inspect the unit for signs of damage. See the orientation depiction on the next page as a
reference.
5. The unit should come with an Installation and Operation Manual.
6. Verify that the correct number of accessory items have been included.
7. The access and water fill stoppers ship installed in their respective ports on the unit. Verify
the presence of the stoppers.
8. Carefully check all packaging for accessory items before discarding.
Included Accessories
1 Copper Pan Token
1 Access Port Stopper
1 Ceiling Air Duct
1 Chamber HEPA Filter
1 Chamber Filter Cap
1 Fill Port Stopper
1 Humidification Pan
4 Leveling Feet
1 Power Cord
6 Shelf Slides
4 Shelf Standards
3 Shelves

12 | Page
RECEIVING YOUR UNIT
ORIENTATION IMAGE
SCO5W
Chamber
Gasket
Chamber
HEPA Filter
Door Latch
Viewing Door (Glass)
Chamber Door
Back of Unit
Fuse in Power Inlet
CO
2
In Ports
Chamber Door
Handle
Back of Unit
Fill
Water Jacket Ports
Siphon
Viewing Door Knob
Anode
Access Port
Door Gasket
CO
2
Sample
and Data Ports
Control Panel
Side of Unit
Shelf-Standard
Mounting Rail

13 | Page
RECEIVING YOUR UNIT
RECORDING DATA PLATE INFORMATION
The data plate contains the incubator model number and serial number. Tech Support will need this
information during any support call. Record it below for future reference.
•The data plate is located on the inside of the chamber door, top-right corner.
Model Number
Serial Number

14 | Page
RECEIVING YOUR UNIT

15 | Page
INSTALLATION
INSTALLATION CHECKLIST
Carry out the steps and procedures listed below to install the unit in a new workspace location and
prepare it for use. All procedures are found in the Installation section of this manual.
Pre-Installation
Check that the required ambient conditions and ventilation spacing for the incubator are met,
page 16
•Unit dimensions may be found on page 57
Check for performance-disrupting heat and cold sources in the environment, page 16
Check that a suitable electrical outlet and power supply is present, page 17
Procure a CO2gas supply for the incubator page 9
Install the Incubator in a suitable location
Review the lifting and handling instructions, page 18
Place the incubator in its workspace location, page 19
Make sure the incubator is level, page 18
Set up the incubator for use
Clean and disinfect the incubator and accessories (except the HEPA filter) that will be placed
in the incubation chamber, page 19
•Do not use deionized water to clean the unit or fill the water jacket, see page 19
Install the chamber HEPA filter and ceiling air duct in the incubation chamber, page 20
Install the shelving in the incubation chamber, page 21
Connect the incubator to the CO2gas supply source, page 22
Verify that the rubber stopper is installed in the access port inside the incubation chamber,
page 23
Fill the water jacket with 19 gallons (71 liters) of water, page 23

16 | Page
INSTALLATION
REQUIRED AMBIENT CONDITIONS
These units are intended for use indoors, at room temperatures between 15°C and 30°C (59°F and 86°F),
at no greater than 80% Relative Humidity (at 25°C / 77°F). Operating outside these conditions may
adversely affect its incubator temperature stability and effective operating range.
REQUIRED CLEARANCES
These clearances are required to provide air flows sufficient for ventilation and cooling.
4 inches (102mm) of clearance is required on the sides and back.
2 inches (51mm) of headspace clearance between the top of the unit and any overhead partitions.
ENVIRONMENTAL DISRUPTION SOURCES
When selecting a location to install the unit, look for sources of heat, cold, and moving air that can affect
the chamber temperature and atmospheric integrity:
•Ovens, autoclaves, and any device that produces significant radiant heat
•High-traffic areas
•Direct sunlight
•Heating and cooling ducts or other sources of fast-moving air currents
2” (51mm)
4” (102mm)
4” (102mm)
Access Port
2” (51mm)
4” (102mm)
Door Swing
23.5“ (597mm)
Power Cord
Gas Connections

17 | Page
INSTALLATION
POWER SOURCE REQUIREMENTS
When selecting a location for the unit, verify each of the following requirements is satisfied.
Power Source
The power source for the unit must match the voltage and match or exceed the amperage
requirements listed on the unit data plate. These units are intended for 110 - 120V 50/60 Hz
applications at 5 Amps
•Supplied voltage must not vary more than 10% from the data plate rating. Damage to the
unit may result if the supplied voltage varies more than 10%.
•The wall power source must be protective earth grounded.
•Use a separate circuit to prevent loss of the unit due to overloading or circuit failure.
•The recommended wall circuit breakers for these units are 15 amps.
•The wall power source must conform to all national and local electrical codes.
Power Cord
The unit must be positioned so that all users can quickly unplug the cord in the event of an
emergency.
•Each unit comes provided with a 115 volt 15 Amp, 9.2ft (2.86m) NEMA 5-15P power cord.
•Always use this cord or an identical replacement.
Fuses
These units ship with a fuse installed in a fuse holder next to the power cord inlet.
•The fuse must be installed and intact for the unit to operate.
•Always find and fix the cause of a blown fuse prior to putting the unit back into operation.
•Fuse type
o250V, T20, 5X20mm
NEMA 5-15R
wall socket

18 | Page
INSTALLATION
LIFTING AND HANDLING
The unit is heavy. Use appropriate lifting devices that are sufficiently rated for these loads. Follow these
guidelines when lifting the unit.
•Lift the unit only from its bottom surface.
•Doors, handles, and knobs are not adequate for lifting or stabilization.
•Restrain the unit completely while lifting or transporting so it cannot tip.
•Remove all moving parts, such as shelves and trays, and lock doors in the closed position
during transfers to prevent shifting and damage.
LEVELING
Install the leveling feet in the 4 corner holes on the bottom of the unit. The unit must be level and
stable for safe operation.
Note: To prevent damage when moving the unit, turn all 4 leveling feet so that the leg of each foot
sits inside the unit.

19 | Page
INSTALLATION
INSTALL THE INCUBATOR
Install the unit in a workspace location that meets the criteria discussed in the previous entries of the
Installation section.
DEIONIZED AND DISTILLED WATER
Do not use deionized water to clean or humidify the incubator or fill the water jacket. Use of
deionized water may corrode metal surfaces and voids the warranty. The manufacturer recommends
the use of distilled water in the resistance range of 50K Ohm/cm to 1M Ohm/cm, or a conductivity
range of 20.0 uS/cm to 1.0 uS/cm, for cleaning and humidifying and water jacketing applications.
INSTALLATION -CLEAN AND DISINFECT
Cleaning and disinfecting the unit incubation chamber, shelving components, and ceiling air duct now
reduces the risk of contamination. The chamber and shelving were cleaned and disinfected at the
factory, however, the unit may have been exposed to contaminants during shipping.
•Remove all protective wrappings from shelving components and the ceiling air duct prior to
cleaning.
•Do not clean the chamber HEPA filter
•See the Cleaning and Disinfecting entry on page 43 for information on how to clean and
disinfect without damaging the incubator or its components.

20 | Page
INSTALLATION
INSTALL CHAMBER HEPA FILTER AND CEILING AIR DUCT
The incubator must be unplugged while carrying out this procedure.
Exercise caution when installing the ceiling air duct. A plastic blower fan and the fragile head of the
temperature and CO2sensors are located just above the duct and just to the right of the duct.
Attach the HEPA filter to the mounting ring on the ceiling air duct. The filter should snap into
pace.
•It may be necessary to tilt the ring slightly to one side while doing so.
Attach the black plastic cap to the HEPA filter. Failure to do so will allow air to blow around
the filter instead of blowing through it.
Place the duct and filter into the chamber, slightly tilted to one side or the other for ease of
movement.
•The filter must be toward the back of the chamber.
Inset the tabbed feet on the duct into the slots on the tops of the shelf standard rails.
Step 4
Step 3
Step 2
Step 1
Back
Front
Ceiling Duct
HEPA Filter
Cap
Table of contents
Other Shel lab Accessories manuals

Shel lab
Shel lab SMI Series User manual

Shel lab
Shel lab 5215 User manual

Shel lab
Shel lab SCO5A User manual

Shel lab
Shel lab SCO6WE User manual

Shel lab
Shel lab SCO6AD-2 User manual

Shel lab
Shel lab SMI31-2 User manual

Shel lab
Shel lab SCO6AD User manual

Shel lab
Shel lab SMI2 User manual

Shel lab
Shel lab SCO6WE-2 User manual
Popular Accessories manuals by other brands

Emerson
Emerson Rosemount 328A Reference manual

APCO Aviation
APCO Aviation ChairBag 4 manual

IFM Electronic
IFM Electronic PA32xx Series installation instructions

Roll-A-Shade
Roll-A-Shade Eclipse installation instructions

Leviton
Leviton Acenti QuickPort AC642 installation instructions

EL
EL Moon M320 operating manual