WATER VIAL
SOLUTION VIAL
WHITE
CLEAR
Scenario B
Keep the WATER vial on the bottom.
Remove the caps from the device
and the syringe.
Connect the syringe to the device.
Pull the plunger back SLOWLY to
withdraw any air from the vials.
Turn device (with vials attached)
over so the WATER vial is on top.
SLOWLY press plunger to inject
all air back into the vials. This will
force water into the bottom vial.
Swirl gently until powder is
completely dissolved.
Turn the device (with vials attached) over
so that the PRODUCT vial is on top.
SLOWLY withdraw all air into the syringe by
pulling the plunger back.
This draws mixed solution into the bottom vial.
Disconnect the syringe.
Press the plunger to expel all the air
from the syringe.
Reconnect the syringe to the device.
Turn the device (with vials
attached) over so that the vial
containing mixed solution is on top.
SLOWLY pull back the plunger to
withdraw the prescribed amount of
mixed solution into the syringe.
Disconnect the syringe.
The WATER vial is INCORRECTLY
connected to the WHITE side of
the BAXJECT II device