
Operator Manual
Print No. XPB2-180G.620.20.03.02
5.3.5 Overhead patient hand grip 133
5.3.6 Tube unit position 133
5.3.7 Operating Height 133
5.4 IONTOMAT automatic exposure control 133
5.4.1 Performing IONTOMAT automatic exposure control 134
5.5 Additional Cu lter 134
5.5.1 Selecting additional lters 135
5.6 Light localizer and collimation 135
5.6.1 Switching on and o of the light localizer 136
5.7 Collimation 137
5.7.1 Collimating manually 137
5.7.2 Collimation / Automatic Collimation Size Sensing 137
5.7.3 Restoring the last collimation 138
5.8 Exposures onto the patient table 138
5.8.1 Vertical projection onto the patient table 139
5.8.2 Vertical projection onto the pulled-out detector 140
5.8.3 Oblique projection onto the patient table 141
5.9 Exposures onto the wall stand 142
5.9.1 Horizontal projection onto the wall stand 142
5.9.2 Auto Thorax Collimation 143
5.9.3 Vertical projection onto the wall stand 143
5.9.4 Oblique projection onto the wall stand 144
5.10 Free exposure onto the wireless detector 145
5.10.1 Preparations 146
5.10.2 Positioning the tube unit 146
5.10.3 SID 146
5.10.4 Positioning for vertical projection 147
5.10.5 Positioning for oblique projection 147
5.10.6 Horizontal projection onto the mobile detector 147
5.11 Ortho operation 148
5.11.1 Ortho on patient table 148
5.11.2 Ortho on wall stand 148
5.11.3 Organ program for Ortho 149
5.11.4 Preparing Ortho 149
5.11.5 Releasing the Ortho exposures 155
5.12 Pediatric Use: Summary 156
5.12.1 Introduction 156
5.12.2 Use of anti-scatter grid 157
5.12.3 Use of ltration 157
5.12.4 Automatic Exposure Control (AEC) 157
5.12.5 Specic organ programs for pediatric examinations 158
5.12.6 Pediatric positioning aids 158
6 Accessories and Auxiliary Devices 159
6.1 Preliminary remarks 159
6.1.1 Equipment with accessories 159
6.1.2 Proper use of the product 159
6.1.3 Safety 160
6.1.4 Orientation 161
6.2 Description 161
6.2.1 Trolley UM mobile patient table 161
6.2.2 Patient positioning mattress 165
6.2.3 Paper roll holder 166
6.2.4 Hand grip strip 166