Siemens SITRANS LG260 User manual

Operating Instructions • 01/2020
Guided Wave Radar
SITRANS LG260
Foundation Fieldbus
Rod and cable probe

2PBD-51041061
SITRANS LG260 - Operating Instructions
44575-EN-200212

3
PBD-51041061 SITRANS LG260 - Operating Instructions
44575-EN-20021244575-EN-200212
Contents
1 About this document ............................................................................................................... 5
1.1 Function ........................................................................................................................... 5
1.2 Target group ..................................................................................................................... 5
1.3 Symbols used................................................................................................................... 5
2 For your safety ......................................................................................................................... 6
2.1 Authorised personnel ....................................................................................................... 6
2.2 Appropriate use................................................................................................................ 6
2.3 Warning about incorrect use............................................................................................. 6
2.4 General safety instructions............................................................................................... 6
2.5 EU conformity................................................................................................................... 7
2.6 NAMUR recommendations .............................................................................................. 7
2.7 Installation and operation in the USA and Canada ........................................................... 7
3 Product description ................................................................................................................. 8
3.1 Conguration.................................................................................................................... 8
3.2 Principle of operation........................................................................................................ 9
3.3 Packaging, transport and storage..................................................................................... 9
3.4 Accessories.................................................................................................................... 10
4 Mounting................................................................................................................................. 11
4.1 General instructions ....................................................................................................... 11
4.2 Mounting instructions ..................................................................................................... 12
5 Connecting to power supply................................................................................................. 17
5.1 Preparing the connection ............................................................................................... 17
5.2 Connecting..................................................................................................................... 18
5.3 Wiring plan, single chamber housing.............................................................................. 19
5.4 Wiring plan, double chamber housing ............................................................................ 19
5.5 Switch-on phase............................................................................................................. 20
6 Set up with the display and adjustment module ................................................................ 21
6.1 Insert display and adjustment module............................................................................ 21
6.2 Adjustment system......................................................................................................... 22
6.3 Parameter adjustment - Quick setup .............................................................................. 23
6.4 Parameter adjustment - Extended adjustment................................................................ 23
6.5 Saving the parameterisation data................................................................................... 37
7 Setup with PACTware............................................................................................................. 38
7.1 Connect the PC.............................................................................................................. 38
7.2 Parameter adjustment with PACTware............................................................................ 38
7.3 Set up with the quick setup............................................................................................. 39
7.4 Saving the parameterisation data................................................................................... 41
8 Set up with other systems .................................................................................................... 42
8.1 DD adjustment programs ............................................................................................... 42
8.2 Field Communicator 375, 475 ........................................................................................ 42
9 Diagnostics and servicing .................................................................................................... 43
9.1 Maintenance .................................................................................................................. 43
9.2 Diagnosis memory ......................................................................................................... 43
9.3 Status messages............................................................................................................ 44
9.4 Rectify faults................................................................................................................... 47

4PBD-51041061
SITRANS LG260 - Operating Instructions
44575-EN-200212
9.5 Exchanging the electronics module................................................................................ 52
9.6 Exchange or shorten cable/rod ...................................................................................... 52
9.7 Software update ............................................................................................................. 54
9.8 How to proceed if a repair is necessary.......................................................................... 55
10 Dismount................................................................................................................................. 56
10.1 Dismounting steps.......................................................................................................... 56
10.2 Disposal ......................................................................................................................... 56
11 Supplement ............................................................................................................................ 57
11.1 Technical data ................................................................................................................ 57
11.2 Supplementary information Foundation Fieldbus........................................................... 68
11.3 Dimensions .................................................................................................................... 77
11.4 Trademark ...................................................................................................................... 82
Safety instructions for Ex areas
TakenoteoftheExspecicsafetyinstructionsforExapplications.
These instructions are attached as documents to each instrument
with Ex approval and are part of the operating instructions.
Editing status: 2020-01-24

5
PBD-51041061 SITRANS LG260 - Operating Instructions
44575-EN-200212
1 About this document
1.1 Function
This instruction provides all the information you need for mounting,
connection and setup as well as important instructions for mainte-
nance,faultrectication,theexchangeofpartsandthesafetyofthe
user. Please read this information before putting the instrument into
operation and keep this manual accessible in the immediate vicinity
of the device.
1.2 Target group
This operating instructions manual is directed to trained personnel.
Thecontentsofthismanualmustbemadeavailabletothequalied
personnel and implemented.
1.3 Symbols used
Information, note, tip: This symbol indicates helpful additional infor-
mation and tips for successful work.
Note: This symbol indicates notes to prevent failures, malfunctions,
damage to devices or plants.
Caution: Non-observance of the information marked with this symbol
may result in personal injury.
Warning: Non-observance of the information marked with this symbol
may result in serious or fatal personal injury.
Danger: Non-observance of the information marked with this symbol
results in serious or fatal personal injury.
Ex applications
This symbol indicates special instructions for Ex applications.
• List
The dot set in front indicates a list with no implied sequence.
1 Sequence of actions
Numbers set in front indicate successive steps in a procedure.
Battery disposal
This symbol indicates special information about the disposal of bat-
teries and accumulators.

6PBD-51041061
SITRANS LG260 - Operating Instructions
44575-EN-200212
2 For your safety
2.1 Authorised personnel
All operations described in this documentation must be carried out
onlybytrained,qualiedpersonnelauthorisedbytheplantoperator.
During work on and with the device, the required personal protective
equipment must always be worn.
2.2 Appropriate use
SITRANS LG260 is a sensor for continuous level measurement.
Youcannddetailedinformationabouttheareaofapplicationin
chapter "Product description".
Operational reliability is ensured only if the instrument is properly
usedaccordingtothespecicationsintheoperatinginstructions
manual as well as possible supplementary instructions.
2.3 Warning about incorrect use
Inappropriate or incorrect use of this product can give rise to applica-
tion-specichazards,e.g.vesseloverllthroughincorrectmounting
or adjustment. Damage to property and persons or environmental
contamination can result. Also, the protective characteristics of the
instrument can be impaired.
2.4 General safety instructions
This is a state-of-the-art instrument complying with all prevailing
regulations and directives. The instrument must only be operated in a
technicallyawlessandreliablecondition.Theoperatorisresponsi-
ble for the trouble-free operation of the instrument. When measuring
aggressive or corrosive media that can cause a dangerous situation
if the instrument malfunctions, the operator has to implement suitable
measures to make sure the instrument is functioning properly.
During the entire duration of use, the user is obliged to determine the
compliance of the necessary occupational safety measures with the
current valid rules and regulations and also take note of new regula-
tions.
The safety instructions in this operating instructions manual, the na-
tional installation standards as well as the valid safety regulations and
accident prevention rules must be observed by the user.
For safety and warranty reasons, any invasive work on the device
beyond that described in the operating instructions manual may be
carried out only by personnel authorised by the manufacturer. Arbi-
traryconversionsormodicationsareexplicitlyforbidden.Forsafety
reasons,onlytheaccessoryspeciedbythemanufacturermustbe
used.
To avoid any danger, the safety approval markings and safety tips on
the device must also be observed.

7
PBD-51041061 SITRANS LG260 - Operating Instructions
44575-EN-200212
2.5 EU conformity
ThedevicefullsthelegalrequirementsoftheapplicableEUdirec-
tives.ByaxingtheCEmarking,weconrmtheconformityofthe
instrument with these directives.
The EU conformity declaration can be found on our homepage.
Electromagnetic compatibility
Instruments in four-wire or Ex-d-ia version are designed for use in an
industrial environment. Nevertheless, electromagnetic interference
from electrical conductors and radiated emissions must be taken into
account, as is usual with class A instruments according to EN 61326-
1.Iftheinstrumentisusedinadierentenvironment,theelectromag-
netic compatibility to other instruments must be ensured by suitable
measures.
2.6 NAMUR recommendations
NAMUR is the automation technology user association in the process
industry in Germany. The published NAMUR recommendations are
acceptedasthestandardineldinstrumentation.
ThedevicefullstherequirementsofthefollowingNAMURrecom-
mendations:
• NE 21 – Electromagnetic compatibility of equipment
• NE53–Compatibilityofelddevicesanddisplay/adjustment
components
• NE107–Self-monitoringanddiagnosisofelddevices
For further information see www.namur.de.
2.7 Installation and operation in the USA and
Canada
This information is only valid for USA and Canada. Hence the follow-
ing text is only available in the English language.
Installations in the US shall comply with the relevant requirements of
the National Electrical Code (ANSI/NFPA 70).
Installations in Canada shall comply with the relevant requirements of
the Canadian Electrical Code.

8PBD-51041061
SITRANS LG260 - Operating Instructions
44575-EN-200212
3 Product description
3.1 Conguration
The scope of delivery encompasses:
• Sensor SITRANS LG260
• Optional accessory
The further scope of delivery encompasses:
• Documentation
– Operating instructions SITRANS LG260
– Instructions for optional instrument features
– Ex-specic"Safety instructions" (with Ex versions)
– Ifnecessary,furthercerticates
Information:
Optional instrument features are also described in this operating
instructions manual.The respective scope of delivery results from the
orderspecication.
This operating instructions manual applies to the following instrument
versions:
• Hardware from 1.0.0
• Software from 1.3.0
• OnlyforinstrumentversionswithoutSILqualication
Thetypelabelcontainsthemostimportantdataforidenticationand
use of the instrument:
Scope of delivery
Scope of this operating
instructions
Type label

9
PBD-51041061 SITRANS LG260 - Operating Instructions
44575-EN-200212
2
1
5
3
6
4
7
8
11
12
13
14
10
9
1
Fig. 1: Layout of the type label (example)
1 Instrument type
2 Product code
3 Process and ambient temperature
4 Process pressure
5 Signal output electronics
6 Voltage supply
7 Protection rating
8 Order number
9 Identicationcode
10 Serial number of the instrument
11 Symbol of the device protection class
12 Reminder to observe the instrument documentation
13 ID numbers, instrument documentation
14 Hardware and software version
3.2 Principle of operation
The SITRANS LG260 is a level sensor with cable or rod probe for
continuous level measurement, suitable for applications in bulk solids.
High frequency microwave pulses are guided along a steel cable or
a rod. Upon reaching the product surface, the microwave pulses are
reected.Therunningtimeisevaluatedbytheinstrumentandoutput
as level.
3.3 Packaging, transport and storage
Your instrument was protected by packaging during transport. Its
capacity to handle normal loads during transport is assured by a test
based on ISO 4180.
The packaging consists of environment-friendly, recyclable card-
board. For special versions, PE foam or PE foil is also used. Dispose
of the packaging material via specialised recycling companies.
Transport must be carried out in due consideration of the notes on the
transport packaging. Nonobservance of these instructions can cause
damage to the device.
Application area
Functional principle -
level measurement
Packaging
Transport

10 PBD-51041061
SITRANS LG260 - Operating Instructions
44575-EN-200212
The delivery must be checked for completeness and possible transit
damage immediately at receipt. Ascertained transit damage or con-
cealed defects must be appropriately dealt with.
Up to the time of installation, the packages must be left closed and
stored according to the orientation and storage markings on the
outside.
Unless otherwise indicated, the packages must be stored only under
the following conditions:
• Not in the open
• Dry and dust free
• Not exposed to corrosive media
• Protected against solar radiation
• Avoiding mechanical shock and vibration
• Storage and transport temperature see chapter "Supplement -
Technical data - Ambient conditions"
• Relative humidity 20 … 85 %
With instrument weights of more than 18 kg (39.68 lbs) suitable and
approved equipment must be used for lifting and carrying.
3.4 Accessories
The instructions for the listed accessories can be found in the down-
load area on our homepage.
The display and adjustment module LG Local Display Interface is
used for measured value indication, adjustment and diagnosis.
The LG Remote Interface is an external display and adjustment unit
for sensors with single chamber housing and Ex d double chamber
housing.
Screwedangesareavailableindierentversionsaccordingtothe
following standards: DIN 2501, EN 1092-1, BS 10, ASME B 16.5,
JIS B 2210-1984, GOST 12821-80.
If you are using an instrument with rod version, you can extend the
rod probe individually with curved segments and rod and cable exten-
sionsofdierentlengths.
All extensions used must not exceed a total length of 6 m (19.7 ft).
The extensions are available in the following lengths:
Rod: ø 16 mm (0.63 in)
• Basic segments: 20 … 5900 mm (0.79 … 232 in)
• Rod/cable segments: 20 … 5900 mm (0.79 … 232 in)
• Curved segments: 100 x 100 mm (3.94 … 3.94 in)
If you mount the SITRANS LG260 in a bypass tube or standpipe, you
have to avoid contact to the bypass tube by using a spacer at the
probe end.
Transport inspection
Storage
Storage and transport
temperature
Lifting and carrying
Display and adjustment
module
External display and
adjustment unit
Flanges
Rod components
Centering

11
PBD-51041061 SITRANS LG260 - Operating Instructions
44575-EN-200212
4 Mounting
4.1 General instructions
Ondeviceswithathreadedtting,thehexagonontheprocesstting
must be tightened with a suitable wrench.
See chapter "Dimensions"forwrenchsize.
Warning:
The housing or the electrical connection may not be used for screw-
ing in! Depending on the device version, tightening can cause dam-
age, e. g. to the rotation mechanism of the housing.
Protect your instrument against moisture ingress through the following
measures:
• Use a suitable connection cable (see chapter "Connecting to
power supply")
• Tighten the cable gland or plug connector
• Lead the connection cable downward in front of the cable entry or
plug connector
This applies mainly to outdoor installations, in areas where high
humidity is expected (e.g. through cleaning processes) and on cooled
or heated vessels.
Note:
Makesurethatthedegreeofcontaminationspeciedinchapter
"Technical data" meets the existing ambient conditions.
Note:
Make sure that during installation or maintenance no moisture or dirt
can get inside the instrument.
To maintain the housing protection, make sure that the housing lid is
closed during operation and locked, if necessary.
Metric threads
In the case of instrument housings with metric thread, the cable
glands are screwed in at the factory.They are sealed with plastic
plugs as transport protection.
You have to remove these plugs before electrical connection.
NPT thread
In the case of instrument housings with self-sealing NPT threads, it is
not possible to have the cable entries screwed in at the factory.The
free openings for the cable glands are therefore covered with red dust
protection caps as transport protection.The dust protection caps do
notprovidesucientprotectionagainstmoisture.
Prior to setup you have to replace these protective caps with ap-
proved cable glands or close the openings with suitable blind plugs.
Note:
For safety reasons, the instrument must only be operated within the
permissibleprocessconditions.Youcannddetailedinformationon
Screwing in
Protection against mois-
ture
Cable glands
Process conditions

12 PBD-51041061
SITRANS LG260 - Operating Instructions
44575-EN-200212
the process conditions in chapter "Technical data" of the operating
instructions or on the type label.
Hence make sure before mounting that all parts of the instrument ex-
posed to the process are suitable for the existing process conditions.
These are mainly:
• Active measuring component
• Processtting
• Process seal
Process conditions in particular are:
• Process pressure
• Process temperature
• Chemical properties of the medium
• Abrasionandmechanicalinuences
4.2 Mounting instructions
Mount SITRANS LG260 in such a way that the distance to vessel
installations or to the vessel wall is at least 300 mm (12 in). In non-
metallic vessels, the distance to the vessel wall should be at least
500 mm (19.7 in).
During operation, the probe must not touch any installations or the
vessel wall. If necessary, fasten the probe end.
In vessels with conical bottom it can be advantageous to mount the
sensor in the center of the vessel, as measurement is then possible
nearly down to the lowest point of the bottom. Keep in mind that
measurement all the way down to the tip of the probe may not be pos-
sible.Theexactvalueofthemin.distance(lowerdeadzone)isstated
in chapter "Technical data" of the operating instructions.
Fig. 2: Vessel with conical bottom
Plastic vessel/Glass vessel
The guided microwave principle requires a metallic surface on the
processtting.Therefore,inplasticvessels,etc.,useaninstru-
mentversionwithange(fromDN50)orplaceametalsheet
(ø>200mm/8in)beneaththeprocessttingwhenscrewingitin.
Makesurethattheplatehasdirectcontactwiththeprocesstting.
Installation position
Type of vessel

13
PBD-51041061 SITRANS LG260 - Operating Instructions
44575-EN-200212
When using the probes without metal vessel wall, e.g. in plastic
vessels,themeasuredvaluecanbeinuencedbystrongelectromag-
neticelds(emittedinterferenceaccordingtoEN61326:classA).
Use a probe in coax version for applications in liquids.
1 2
Fig. 3: Mounting in non-metallic vessel
1 Flange
2 Metal sheet
Concrete vessel
When mounting in thick concrete ceilings, SITRANS LG260 should
bemountedfrontushtotheloweredge.Inconcretesilos,thedis-
tance to the wall should be at least 500 mm (20 in).
ø >160mm
(ø >6.3")
Fig. 4: Mounting in concrete silo

14 PBD-51041061
SITRANS LG260 - Operating Instructions
44575-EN-200212
Ifpossible,avoidsockets.Mountthesensorushwiththevesseltop.
If this is not possible, use short sockets with small diameter.
Higher sockets or sockets with a bigger diameter can generally be
used.Theycan,however,increasetheupperdeadzone.Checkifthis
is relevant for your measurement.
In such cases, always carry out a false signal suppression after
mounting.Youcanndfurtherinformationunder"Setup procedure".
h
d
dh
DN25 ... DN150
> DN150 ... DN200
≤ 150 mm (5.91")
≤ 100 mm (3.94")
Fig. 5: Mounting socket
Whenweldingthesocket,makesurethatthesocketisushwiththe
vessel top.
1 2
Fig.6:Socketmustbeinstalledush
1 Unfavourable mounting
2 Socketush-optimummounting
Before beginning the welding work, remove the electronics module
from the sensor. By doing this, you avoid damage to the electronics
through inductive coupling.
Donotmounttheinstrumentsinorabovethellingstream.Makesure
thatyoudetecttheproductsurface,nottheinowingproduct.
Mounting socket
Welding work
Inowingmedium

15
PBD-51041061 SITRANS LG260 - Operating Instructions
44575-EN-200212
Fig.7:Mountingofthesensorwithinowingmedium
The reference plane for the measuring range of the sensors is the
sealingsurfaceofthethreadorange.
Keep in mind that a min. distance must be maintained below the refer-
ence plane and possibly also at the end of the probe - measurement
intheseareasisnotpossible(deadzone).Thelengthofthecable
can be used all the way to the end only when measuring conductive
products.Thesedeadzonesfordierentmediumsarelistedinchap-
ter "Technical data". Keep in mind for the adjustment that the default
setting for the measuring range refers to water.
Theprocessttingmustbesealedifthereisgaugeorlowpressurein
the vessel. Before use, check if the seal material is resistant against
the measured product and the process temperature.
Themax.permissiblepressureisspeciedinchapter"Technical
data" or on the type label of the sensor.
If there is a risk of the cable probe touching the vessel wall during
operation due to product movements or agitators, etc., the measuring
probeshouldbesecurelyxed.
There is an inside thread (M12) in the gravity weight, e.g. for an eye-
bolt (optional).
Make sure that the probe cable is not completely taut. Avoid tensile
loads on the cable.
Avoidundenedvesselconnections,i.e.theconnectionmustbe
eithergroundedreliablyorisolatedreliably.Anyundenedchangeof
this condition can lead to measurement errors.
If there is a danger of the rod probe touching the vessel wall, fasten
the probe at the bottom end.
Keep in mind that measurement is not possible below the fastening
point.
Measuring range
Pressure
Fasten

16 PBD-51041061
SITRANS LG260 - Operating Instructions
44575-EN-200212
1
2
1
2
Fig. 8: Fasten the probe
1 Measuring probe
2 Retaining sleeve
Incaseofdicultinstallationconditionsinliquidapplications,the
probe can be also mounted laterally. For this purpose, adapt the rod
with rod extensions or bow-shaped segments.
To compensate for the resulting changes in signal runtime, let the
instrument determine the probe length automatically.
The determined probe length can deviate from the actual probe
length when using curved or angled segments.
If internal installations such as struts, ladders, etc. are present on the
vessel wall, the measuring probe should be mounted at least 300 mm
(11.81 in) away from the vessel wall.
Youcanndfurtherinformationinthesupplementaryinstructionsof
the rod extension.
Incaseofdicultinstallationconditions,forexampleinasocket,the
probe can be suitably adapted with a rod extension.
To compensate for the resulting changes in signal runtime, let the
instrument determine the probe length automatically.
Youcanndfurtherinformationinthesupplementaryinstructionsof
the rod and cable components.
Lateral installation
Rod extension

17
PBD-51041061 SITRANS LG260 - Operating Instructions
44575-EN-200212
5 Connecting to power supply
5.1 Preparing the connection
Always keep in mind the following safety instructions:
• Carryoutelectricalconnectionbytrained,qualiedpersonnel
authorised by the plant operator
• If overvoltage surges are expected, overvoltage arresters should
be installed
Warning:
Onlyconnectordisconnectinde-energizedstate.
The instrument requires a operating voltage of 9 … 32 V DC. Operat-
ing voltage and the digital bus signal are carried on the same two-wire
connection cable. Power is supplied via the H1 power supply.
Connection is carried out with screened cable according to Fieldbus
specication.
Use cable with round cross section for instruments with housing and
cablegland.Toensurethesealeectofthecablegland(IPprotection
rating),ndoutwhichcableouterdiameterthecableglandissuitable
for.
Useacableglandttingthecablediameter.
Make sure that the entire installation is carried out according to the
Fieldbusspecication.Inparticular,makesurethatthebusistermi-
nated with suitable terminating resistors.
Metric threads
In the case of instrument housings with metric thread, the cable
glands are screwed in at the factory. They are sealed with plastic
plugs as transport protection.
Note:
You have to remove these plugs before electrical connection.
NPT thread
In the case of instrument housings with self-sealing NPT threads, it is
not possible to have the cable entries screwed in at the factory.The
free openings for the cable glands are therefore covered with red dust
protection caps as transport protection.
Note:
Prior to setup you have to replace these protective caps with ap-
proved cable glands or close the openings with suitable blind plugs.
On plastic housings, the NPT cable gland or the Conduit steel tube
must be screwed into the threaded insert without grease.
Max. torque for all housings, see chapter "Technical data".
Make sure that the cable screen and grounding are carried out ac-
cordingtoFieldbusspecication.Werecommendtoconnectthe
cable screening to ground potential on both ends.
Safety instructions
Voltage supply
Connection cable
Cable glands
Cable screening and
grounding

18 PBD-51041061
SITRANS LG260 - Operating Instructions
44575-EN-200212
In systems with potential equalisation, connect the cable screening
directly to ground potential at the power supply unit, in the connection
box and at the sensor. The screen in the sensor must be connected
directly to the internal ground terminal.The ground terminal outside
on the housing must be connected to the potential equalisation (low
impedance).
5.2 Connecting
The voltage supply and signal output are connected via the spring-
loaded terminals in the housing.
Connection to the display and adjustment module or to the interface
adapter is carried out via contact pins in the housing.
Information:
The terminal block is pluggable and can be removed from the
electronics.To do this, lift the terminal block with a small screwdriver
and pull it out. When reinserting the terminal block, you should hear it
snap in.
Proceed as follows:
1. Unscrew the housing lid
2. If a display and adjustment module is installed, remove it by turn-
ing it slightly to the left
3. Loosen compression nut of the cable gland and remove blind
plug
4. Remove approx. 10 cm (4 in) of the cable mantle, strip approx.
1 cm (0.4 in) of insulation from the ends of the individual wires
5. Insert the cable into the sensor through the cable entry
1 2
Fig. 9: Connection steps 5 and 6
1 Single chamber housing
2 Double chamber housing
6. Insert the wire ends into the terminals according to the wiring plan
Note:
Solidcoresaswellasexiblecoreswithwireendsleevesareinsert-
eddirectlyintotheterminalopenings.Incaseofexiblecoreswithout
end sleeves, press the terminal from above with a small screwdriver,
the terminal opening is then free.When the screwdriver is released,
the terminal closes again.
Connection technology
Connection procedure

19
PBD-51041061 SITRANS LG260 - Operating Instructions
44575-EN-200212
7. Check the hold of the wires in the terminals by lightly pulling on
them
8. Connect the screen to the internal ground terminal, connect the
external ground terminal to potential equalisation
9. Tighten the compression nut of the cable entry gland. The seal
ring must completely encircle the cable
10. Reinsert the display and adjustment module, if one was installed
11. Screw the housing lid back on
Theelectricalconnectionisnished.
5.3 Wiring plan, single chamber housing
The following illustration applies to the non-Ex, Ex-ia and Ex-d ver-
sion.
12
( ) (-)
1
5
0
1
0
1
+678
Bus
2
3
4
5
Fig. 10: Electronics and connection compartment - single chamber housing
1 Voltage supply, signal output
2 Contact pins for the display and adjustment module or interface adapter
3 Simulation switch ("1" = mode for simulation release)
4 For external display and adjustment unit
5 Ground terminal for connection of the cable screening
5.4 Wiring plan, double chamber housing
The following illustration applies to the non-Ex, Ex-ia and Ex-d ver-
sion.
Electronics and connec-
tion compartment

20 PBD-51041061
SITRANS LG260 - Operating Instructions
44575-EN-200212
5
0
1
0
1
+678
Bus
2
3
12
( ) (-)
11
Fig. 11: Electronics compartment - double chamber housing
1 Internal connection to the connection compartment
2 Contact pins for the display and adjustment module or interface adapter
3 Simulation switch ("1" = mode for simulation release)
Bus
5
12
+
( ) (-)
678
2
3
4
1
Fig. 12: Connection compartment - double chamber housing
1 Voltage supply, signal output
2 For display and adjustment module or interface adapter
3 For external display and adjustment unit
4 Ground terminal for connection of the cable screening
5.5 Switch-on phase
AfterconnectingSITRANSLG260tothebussystem,thedevicerst
performs a self-test:
• Internal check of the electronics
• Indication of the status message "F 105 Determine measured
value" on the display or PC
• Status byte goes to fault value
Then the actual measured value is output to the signal cable. The
value takes into account settings that have already been carried out,
e.g. default setting.
Electronics compartment
Connection compartment
Other manuals for SITRANS LG260
2
Table of contents
Other Siemens Radar manuals

Siemens
Siemens SITRANS LG270 User manual

Siemens
Siemens SITRANS LG250 User manual

Siemens
Siemens SITRANS LG270 User manual

Siemens
Siemens SITRANS LG270 User manual

Siemens
Siemens SITRANS LG250 User manual

Siemens
Siemens SITRANS LG250 User manual

Siemens
Siemens SITRANS LG240 User manual

Siemens
Siemens SITRANS LG250 User manual

Siemens
Siemens SITRANS LG260 User manual