Foreword
Warranty
Important information
1. Function Description
1.1 General description
2. Operation
2.1 When the patient sits down
2.2 Connection
2.3 Height adjustment
2.4 Adjusting the backrest
2.5 Adjusting the Trendelenburg position
2.6 Replacing the paper roll
2.7 Paper dispenser
2.8 The function of the castors
2.9 Side rails
2.10 IV stand
3. Maintenance & Cleaning
3.1 Routine examinations
3.2 Cleaning
3.3 Lubrication schedule
3.4 Spare parts
3.5 Accessories
3.6 Solutions for lubrication, cleaning and disinfecting
4. Technical specification
4.1 Dimensions
4.2 Motors and control units
4.3 Battery backup
4.4 Labels
5. EU-Declaration of Conformity
6. Contact details

300kg exam table Foreword, warranty, information
FOREWORD
Welcome as the owner of a product manufactured by Sjöbloms Sjukvårdsutrustning
AB. We hope that you will be pleased and satisfied with our quality product.
If a problem should arise with this or any of our products, you are always welcome
to contact us, and our service department will assist you with the expertise needed
to solve the problem.
The user manual contains all the information needed to use, maintain and service
the product. The document is continuously updated as the products are modified.
Therefore we reserve against possible changes in the user manual.
WARRANTY
Sjöbloms provides a two-year warranty for material, design and manufacturing
defects, provided the products are not subjected to tampering or excessive wear.
For the guarantee to apply, the product must have been used in accordance with
the user manual, and the maintenance performed as specified.
IMPORTANT INFORMATION
This product is intended for professional medical use. The user manual should be
read carefully before using the product, to avoid unnecessary risks for the patient
and operator.
The user manual contains warnings where there is a risk of personal injury, which
should be noted. Improper handling can be fatal.
A warning triangle is shown where particularly important information can be found.
Warning triangle

300kg exam table Functional description
General description
1. FUNCTIONAL DESCRIPTION
This product is meant for professional use in health care, for examination of the
patient.
It is designed to be a functional and user-friendly tool for the operator and patient.
Note that the product must only be used by competent health professionals who
read the user manual, for correct handling.
General description
2-sectioned examination table, electrically adjustable, chromium-plated or coated
all-welded steel tube chassis. Backrest and Trendelenburg adjusted by electric
motors. The cushions are stuffed with extra thick foam plastic and upholstered with
imitation leather.
125mm swivel central-locking double-castors, one with directional locking. Various
accessories such as neck pillow, IV stand etc. are also available.

300kg exam table Operation
When the patient sits down
2. OPERATION
2.1 When the patient sits down
We recommend to always have the castors locked when the patient sits down.
For maximum safety, have the patient sit down on the middle of the couch.
NOTE!
It is not possible to sit on the backrest.

300kg exam table Operation
Connection, height adjustment
2.2 Connection
The plug should be connected to a grounded 230V outlet. The product is now ready
for use.
2.3 Height adjustment
Electric, hand control
(standard)
Electric, foot control
(option)
NOTE!
Ensure that no unauthorized persons are in the couch's working range, as
a pinch hazard exists when raising and lowering the product.
Follow the symbols and
arrows to regulate the
couch.
Follow the symbols and
arrows to regulate the
couch.

300kg exam table Operation
Adjusting the backrest
2.4 Adjusting the backrest
Electric, hand control
Electric, foot control
For exam table with two motors:
The right pair regulates the
backrest.
The left pair regulates the height of
the table.
The right pair regulates the
backrest and/or the Trendelenburg
position. Press the middle switch to
change between the two functions.
NOTE!
Ensure that no unauthorized persons are in the couch's working range, as a
pinch hazard exists when raising and lowering the product.
Follow the symbols and arrows
to regulate the couch.

300kg exam table Operation
Adjusting Trendelenburg
2.5 Adjusting the Trendelenburg position
Electric, foot control
(optional)
The right pair regulates the
backrest and/or the
Trendelenburg position. Press
the middle switch to change
between the two functions.
The left pair regulates the
height of the table.
Electric, hand control
Follow the symbols and arrows
to regulate the couch.
NOTE!
Ensure that no unauthorized persons are in the couch's working range, as a
pinch hazard exists when raising and lowering the product.

300kg exam table Operation
Paper roll holder
2.6 Replacing the paper roll
2.7 Paper dispenser
Flip the tear-off edge open and pull out new paper. Cover the table with the paper.
Fix the paper into place by pressing down the tear-off edge. Tear off the old paper
on the tear-off edge, and the table is ready to be used again.
a) Unscrew the black ball.
b) Push the metal rod to the left.
c) Pull the rod gently towards you
until it comes loose on the right
side.
d) Pull the rod gently to the right.
The rod is now loose and you can
put on a new paper roll.
e) Put the rod back on the left side.
f) Put the rod back on the right
side, in the score.
g) Push the rod to the right.
h) Screw the ball tight again.

300kg exam table Operation
The Castors
2.8 The function of the castors
The table is equipped with central locking castors including a directional locking
function.
Press the red side to lock the castors.
Press the green side to activate the directional locking function.
Have the castors as shown above for a free mode.

1. Lift the black ball.
2. Fold the siderails using the other hand in the direction of the arrow.
3. Done!
4. When raising the siderails, make sure its mechanic clicks into position indicating
it is fully raised.
5.

3-sectioned exam table Maintenance & Cleaning
Routine check-ups
2.10 IV stand
NOTE!
Pinch hazard exists when raising and
lowering the side rails.
The IV stand may be adjusted in height.
Lock/open the knob to adjust the
height.

300kg exam table Operation
Side rails
3. MAINTENANCE & CLEANING
3.1 Routine checks
Routine inspections should be made to ensure proper functionality.
Important!
Failure to perform the following routine can cause the patient and/or operator to
be exposed to unnecessary risks and the couch may have a reduced life expectancy.
Please check all functions and parts on a regular basis.
Check the following:
Any unusual sounds.
Castors and castor locking mechanism.
Damages to electric cables.
The motor functions.
That joints and fittings are lubricated according to the lubrication chart.
3.2 Cleaning
Cleaning artificial leather
Water and a mild soap solution is recommended.
Disinfection
If the product needs to be disinfected, use a diluted alcohol solution. Too strong an
alcohol solution can have a drying effect and the imitation leather may deteriorate
and crack.
Chlorine
Chlorine-base solutions are not recommended as a general rule, since it effects
products and environment negatively, differently depending on concentration. If
chlorine-based solutions are used, always wipe afterwards with damp cloth.

300kg exam table Maintenance & Cleaning
Lubrication
3.3 Lubrication
Lubrication according to the schedule should be performed every 6 months or as needed.
1. First lock the castors.
2. Enlist the help of another person and tip the couch on the side, in order to
lubricate the moving joints.
The spray can in the pictures above indicates where to lubricate.

300kg exam table Maintenance & Cleaning
Spareparts
3.4 Spare parts
5-6
7
Nr Art nr Benämning
1 312100+3115004C Motor, ryggstöd
2 312100+1110004C Motor, chockläge
3 343100+0015004A Motor, lyft
4 CB9140AE3+0001J Styrenhet
5 12560150OT2LPV Hjul
6 12560150OT3LPV Hjul riktningsspärrat
7 HB834500003+11 Handkontroll Linak
8 010-8540 Pappersrullhållare inkl kula
1
2
3-4
8

300kg exam table Maintenance & Cleaning
Accessories
3.5 Accessories
Sidogrindar Infusionsstativ

300kg exam table Maintenance & Cleaning
Solutions for lubrication, cleaning and disinfecting
3.6 Solutions for lubrication, cleaning and disinfecting
R-1, Cleaning solution for artificial leather
5557001, Lubrication
C068, Disinfection solution, DAX Plus
Cleaning solution Lubrication Disinfection

300kg exam table Technical specifications
Dimensions
4. TECHNICAL SPECIFICATIONS
4.1 Dimensions
Length: 205 cm
Width: 80 cm
Height: 55-85 cm
Length of backrest: 75 cm
Weight: ca 90 kg
Max patient weight: 300 kg
SWL: 320 kg

300kg exam table Technical specifications
Motors, control units
4.2 Motors and control units
The motors comply with EMC directives in accordance with standard EN 60601-1-2.
At full load, the motors can be used a maximum of 6 minutes/hour. The actuator
manufacturer guarantees at least 10 000 cycles performance.
Control units
Max. usage time 10% or 6 min/h at continuous operation. Complies with directives
in accordance with standard EN 60601-1.

300kg exam table Technical specifications
Battery backup
4.3 Battery backup
(optional)
General
The batteries used in the control unit are sealed lead batteries.
The included power cord should be plugged into an electrical outlet with 230 volts.
The charger will recharge the battery in less than 12 hours. The charger can safely
be connected for charging for extended periods.
If a battery is not used for a long time, it will gradually lose its stored energy. It is
therefore recommended that the batteries are recharged at least 8 hours before
usage.
Battery life
Battery life depends on several factors including the extent of discharging. If the
battery is discharged too extensively between charges the lifespan will be reduced.
Charging as often as possible is recommended.
The built-in buzzer "Charge the battery" that is built into the controller will sound
when a discharge of 50% has occurred.
The life expectancy of the batteries is about 500-1200 charge cycles.
NOTE!
When the battery is depleted, the battery should be brought to a battery
collection, if applied in your country, to conserve the environment.