Operating Procedure
MK20402V1OHB20JAN
123
654
1 2 3
3
20~30 cm
For the controller cable, slide the joint and pull it
directly out.
For the camera cable, pull it out while turning the
ring of the joint.
For the camera head, turn the main unit to
remove/attach it (make sure to disconnect the
camera cable).
Advice on Cable Disconnection
Adjusting
the Viewing
Angle Position
If iris adjustment is required, turn
the dial to adjust the iris setting.
TIPS
If the remote control does not work due
to the deterioration of battery
performance, you can use the adapter’s
FOCUS dial to adjust the focus (For
emergency purposes only; be careful
not to touch the clean field.)
While viewing the chart displayed on the
monitor, adjust the focus by holding down
the NEAR and FAR buttons on the controller.
Now use the hexagonal wrench to
adjust the up, down, left, and right
positions so that your chart
appears in the center of the
monitor.
While viewing the monitor, adjust
the rotation by using a hexagonal
wrench to loosen the screw
securing the camera adapter and
camera head, then tighten the
screw with this wrench to secure
the screw again.
While looking through the
microscope’s ocular lens, centrally
place the chart printout (your chart
for adjusting the viewing angle
position). The recommended
working distance is 20~30 cm.
Focus wrong
Focus correct
Turn on power to the MCC-1000MD,
M525 and monitor to display your
chart on the monitor.
Microscope Connection Guide for the CCMA-2DAR Adapter
and the MCC-1000MD
Connecting to the Leica M525