Spectrum SRO-SYS 1.3 Manual

Installation, Service &
Operation Manual
SRO-SYS 1.3 - 21
Spectrum
TM
Reverse Osmosis Systems

2
Spectrum
TM
Contents
Introduction p.3
Safety p.4
Feed Water & Operation Specications p.5
Rejection, Recovery & Flow Rates p.6
System Requirements & Operation Guidelines p.7
Electrical p.8
Component Identication p.9
Membrane Installation, Removal & Replacement p.10
System Purging p.11
Initial Start-up p.12
System Specications p.13
Operation & Maintenance p.14
Flushing the System p.15
Preparation for Storage & Shipment p.16
Troubleshooting p.17
Membrane Temperature Correction Factors p.18
Service Assistance p.18
System Drawings & Dimensions p.19
Electrical Schematics p.20
Warranty p.22

3
Spectrum
TM
Introduction
SPECTRUM SRO Systems are durable and high performance, designed to offer years of reliable
operation providing the proper care is taken. This manual outlines installation, operation, maintenance
and troubleshooting details vital to sustained performance.
If a system is altered at the site of operation, or if the feed water conditions change contact your local
dealer or distributor to determine the proper recovery for the application.
NOTE: In order to maintain the manufacturer’s warranty an operating log must be maintained
and copies will need to be sent to your local dealer or distributor for review.
NOTE: Prior to operating or servicing the Reverse Osmosis System this manual must be
read and fully understood. Keep this and other associated information for future reference
and for new operators or qualied personnel near the system.
CAUTION
CAUTION

4
Spectrum
TM
• DO NOT USE WHERE THE WATER IS MICROBIOLOGICALLY UNSAFE OR OF UNKNOWN
QUALITY WITHOUT ADEQUATE DISINFECTION BEFORE OR AFTER THE SYSTEM.
• PRETREATMENT MUST BE SUFFICIENT TO ELIMINATE CHEMICALS, ORGANICS OR
INORGANICS THAT COULD ATTACK THE MEMBRANE MATERIAL.
• ALWAYS TURN OFF THE UNIT, SHUT OFF THE FEED WATER AND DISCONNECT THE
ELECTRICAL POWER BEFORE WORKING ON THE UNIT.
• NEVER ALLOW THE PUMP TO RUN DRY.
• NEVER START THE PUMP WITH THE CONCENTRATE VALVE CLOSED.
• NEVER ALLOW THE UNIT TO FREEZE OR OPERATE WITH A FEED WATER TEMPERATURE
ABOVE 38°C.
Safety
The safety section of this manual outlines the various safety headings used throughout this manual’s
text and are enhanced and dened below:
!
CAUTION: Indicates statements that are used to identify conditions or practices that could
result in equipment or other property damage.
WARNING: Indicates statements that are used to identify conditions or practices that could
result in injury or loss of life. FAILURE TO FOLLOW WARNINGS COULD RESULT IN SERIOUS
INJURY OR EVEN DEATH.
CAUTION
CAUTION
CAUTION
DO NOT UNDER ANY CIRCUMSTANCE REMOVE ANY CAUTION, WARNING OR OTHER
DESCRIPTIVE LABEL FROM THE SYSTEM.
PLEASE READ THE ENTIRE MANUAL BEFORE PROCEEDING WITH THE INSTALLATION
AND STARTUP. FAILURE TO FOLLOW THE ATTACHED INSTRUCTIONS OR OPERATING
PARAMETERS MAY LEAD TO THE PRODUCT’S FAILURE, WHICH CAN CAUSE PROPERTY
DAMAGE AND/OR PERSONAL INJURY.

5
Spectrum
TM
Feed Water &
Operation Specifications
Feed water quality has the greatest effect on the operation of Reverse Osmosis Systems, therefore it is
important to ensure that the feed is within the operating parameters.
Feed Temperature (ºC) 4 - 29
System Inlet Pressure (bar) 3.1 - 5.8
Maximum Operating Pressure (bar) at Temperature (ºC)13.8 at 25
Maximum SDI Rating < 3
Maximum Turbidity 1 NTU
Maximum Free Chlorine and/or Chloramines 0 PPM
pH (continuous) 5 - 11
pH (cleaning - 30 min) 2 - 12
Operating Limits
Operating Parameters:*
Feed Water Requirements:*
*If any of the feed water parameters are not within the limits given, consult the local dealer or distributor for assistance.
NOTE: Higher TDS and/or lower temperatures will reduce the system’s production.
NOTE: It is very important to meet the minimum feed water requirements. Failure to do so
will cause the membranes to foul and void the manufacturer’s warranty.
!
!

6
Spectrum
TM
Rejection, Recovery &
Flow Rates
The amount of total dissolved solids (TDS) rejected by the membrane is expressed as a percentage. For
example, a 98.5% rejection rate means that 98.5% of the total dissolved solids do not pass through the
membrane. To calculate the % rejection use the following formula:
% Rejection = [(Feed TDS - Product TDS) / Feed TDS] x 100
Example: 98.5% = [(550-8.25) / 550] x 100
NOTE: All TDS gures must be expressed in the same units, typically parts per million (ppm)
or milligrams per litre (mg/l).
SPECTRUM Reverse Osmosis Systems are designed to reject up to 98.5% NaCl unless computer projections
have been provided or stated otherwise.
The amount of permeate water recovered for use is expressed as a percentage. To calculate % recovery
use the following formula:
% Recovery = (Product Water Flow Rate / Feed Water Flow Rate) x 100
Example: 50% = (1.02/2.04) x 100
NOTE: All ow rates must be expressed in the same units, typically litres per minute (lpm).
Contents
!
!

7
Spectrum
TM
System Requirements &
Operation Guidelines
Plumbing
The membranes and high pressure pumps used on the SPECTRUM Systems require continuous ow of
water with a minimum feed pressure of 3.1 bar and must not exceed 38°C.
Feed Water Connection
1. Locate the 3/4" FNPT solenoid valve feed water inlet.
2. Attach the inlet piping to the 3/4" FNPT solenoid valve feed water inlet.
3. Be certain that all of the components of the feed water are soluble at the concentrations attained in
the system.
NOTE: Feed line must be a minimum of 1/2".
Permeate (Product Water) Connection
Locate the threaded 1/2" MNPT tting on the ow meter labelled “permeate” and attach to storage tank.
The permeate line can be run to the holding tank with PVC tting or other FDA approved materials. This is
so the material being used does not dissolved into the permeate water.
NOTE: The pH of the reverse osmosis permeate water will typically be 1 or 2 pH units lower
than the feed water pH. A low pH can be very aggressive to some plumbing materials such
as copper piping.
Concentration (Waste Water) Connection
Locate the 1/2" MNPT connector on the ow meter labelled “concentrate” and attach the line to a drain.
Run the concentrate line to an open drain in a free and unrestricted manner (no back pressure).
CAUTION
NOTE: Any restrictions or blockage in the drain line can cause back pressure, which
will increase the system’s operating pressure. This can result in damage to the system’s
membranes and components.
Contents
!
!

8
Spectrum
TM
Electrical
The motor used on the SPECTRUM Systems are pump and motor combinations. The motor is available in
220V 50Hz 1 Phase. Each SPECTRUM System is equipped with an 2.4 metre electrical cord w/o plug.
Ensure that the electrical circuit supplying the system is compatible with the requirements of the specic
SPECTRUM model being installed.
NOTE: It is recommended that a qualified electrician wire the system in accordance with
thelocal and national electrical codes.
CAUTION NOTE: To reduce the risk of electrical shock the incoming power supply must include a
protective earth ground.
SPECTRUM Systems are controlled with a high pressure switch in the permeate line. The liquid pressure
switch turns the system on when the permeate water pressure drops below 2.8 bar and off when the
permeate water pressure exceeds 4.1 bar.
Pre-ltration
SPECTRUM Systems are supplied with a carbon block lter. Change the cartridge once a month or when a
1 bar differential exists between the two pre-lter gauges.
NOTE: The system must be operated on ltered water only.
Pump
The pump type used on the SPECTRUM Systems is multi-stage centrifugal stainless steel. If any damage
occurs to the system’s pump a re-build kit may be available. Contact the local dealer or distributor and
inform them of the system pump model.
Mounting
The free standing system should be bolted down or securely fastened in compliance with local regulation
standards.
!
!

9
Spectrum
TM
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
18
17
19
Component Identification
No. Description
1Carbon Cartridge in Housing Bowl
2Filter Housing
3Post Filter Gauge
4Pre Filter Gauge
5Concentrate Valve
6Concentrate Recycle Valve
7Concentrate Recycle Flow Meter
8Permeate Flow Meter
9Concentrate Flow Meter
10 High Pressure Switch
11 Motor
12 C23 Computer Controller
13 Digital TDS-DM2 Meter
14 Pump Filter Gauge
15 Solenoid Valve
16 Membrane
17 Membrane Housing
18 Throttle Valve
19 Low Pressure Switch
Figure 1.

10
Spectrum
TM
Membrane Installation,
Removal & Replacement
Installing and replacing membranes in the membrane housings is an easy process if the correct information
and tools are at hand. Refer to the following instructions when removing or replacing membrane elements:
NOTE: All pressure gauges must read zero before proceeding. Before attempting, disconnect
the power from the system and bleed all water pressure from the system.
1. Remove the black end cap from the top of the membrane housing. This is done by removing the yellow
locking clip and then the white snap ring on the membrane housing end.
2. Remove the replacement membrane element(s) from the shipping box; the membrane(s) should be
contained within a plastic oxygen barrier bag.
3. Cut the bag open as close as possible to the seal at one end of the bag, so the bag may be re-used if
necessary.
4. Make sure that all parts are clean and free from dirt. Examine the brine seal and permeate tube for
nicks or cuts.
5. Flow directions should be observed for installation of each element into their respective membrane
housings.
NOTE: Wear gloves for the following steps in order not to contaminate the membrane.
Replacing the Membrane Element
NOTE: The brine seal must be in the same position for each membrane element housing.
The brine seal is a rubber seal that protrudes on one side of the membrane and is always on
the feed side of the membrane element. On SPECTRUM Systems the brine seal should be
on the top side of the membrane housing.
1. Remove one membrane element at a time from the top of the membrane housings. Long nose pliers
may be necessary to pull the old membrane element out of the membrane housing.
2. Lubricate the brine seal with a non-petroleum based lubricant. Do not use a petroleum-based lubricant.
3. Install membranes with brine seal location depicted (Figure 2).
4. With a smooth and constant motion, push the membrane element into the housing so the brine seal
enters the housing without coming out of the brine seal groove.
5. Once the membrane has been fully inserted into housing re-install black end plug, white snap ring
(ensure the white snap ring is fully seated all the way around end plug) and yellow locking clip.
6. Reconnect any ttings that may have been disconnected when the membrane housing were
disassembled.
7. To start-up the system refer to the initial start-up section of this manual.
!
!

11
Spectrum
TM
SPECTRUM 5.2 SPECTRUM 1.3
SPECTRUM 10
SPECTRUM 2.6
SPECTRUM 16
SPECTRUM 21
System Purging
Carefully inspect the system before initial start-up. Check that all plumbing and electrical connections are
not loose or have come undone during shipment. A lter housing wrench is included with the system.
NOTE: Leave the power to the system off for this procedure.
1. Redirect permeate water to the drain for this procedure.
2. Fully open the concentrate valve (Figure 1, Item 5, Counter Clockwise).
3. Fully close the concentrate recycle valve (Figure 1, Item 6, Clockwise).
4. Fully open the throttle valve (Figure 1, Item 18, Counter Clockwise).
5. To activate the solenoid bypass feature, place the three way push button located on the front panel of
the C-23 controller to the FLUSH position (Figure 3).
6. Turn the feed water on the let the system purge until no visible bubbles appear from concentrate ow
meter.
7. Place the three way push button in OFF position (Figure 3).
Membrane Installation,
Removal & Replacement
Figure 2.
!

12
Spectrum
TM
1. Keep the permeate water line to drain for this procedure.
2. Fully open the concentrate valve (Figure 1, Item 5, Counter Clockwise).
3. Fully close the concentrate recycle valve (Figure 1, Item 6, Clockwise).
4. Adjust the throttle valve at 50% open (Figure 1, Item 18, Counter Clockwise).
5. Turn the RO system on and adjust the concentrate (waste) valve, recycle valve and the throttle valve to
the designed ow and pressure (refer to System Specications on page 13).
6. Inspect the system for leaks.
7. Allow the system 30 minutes to ush the preservative solution from the system.
8. After 30 minutes shut down the system.
9. Re-direct the permeate water back to the tank and then turn the system back on.
10. Record the readings daily for a week; after one week record the readings once a week.
Initial Start-up
Figure 3.

13
Spectrum
TM
Design SRO-SYS-1.3 SRO-SYS-2.6 SRO-SYS-5.2 SRO-SYS-10 SRO-SYS-16 SRO-SYS-21
Conguration Single Pass Single Pass Single Pass Single Pass Single Pass Single Pass
Feed Water Source (TDS) Up to 2,000 Up to 2,000 Up to 2,000 Up to 2,000 Up to 2,000 Up to 2,000
System Recovery (%)* 50 - 75 50 - 75 50 - 75 50 - 75 50 - 75 50 - 75
Rejection and Flow Rates
Nominal Salt Rejection (%) 99 99 99 99 99 99
Permeate Flow Rate (lpm)* 1.29 2.61 5.26 10.52 15.79 21.05
Minimum Concentrate Flow Rate (lpm) 3.78 3.78 11.36 11.36 11.36 11.36
Concentrate Recycle Flow Rate (lpm) Up to 3.78 Up to 3.78 Up to 7.57 Up to 18.93 Up to 18.93 Up to 18.93
Connections
Feed Connection (FNPT ") ¾¾¾¾¾¾
Permeate Connection (MNPT ") ½½½½½ 1
Concentrate Connection (MNPT ") ½½½½½½
Membranes
Membranes Per Vessel 111111
Membrane Quality 231234
Membrane Size 2521 2521 4040 4040 4040 4040
Vessels
Vessel Array 1:1 1:1:1 1 1:1 1:1:1 1:1:1:1
Vessel Quantity 231234
Pumps
Pump Type Multi-Stage Multi-Stage Multi-Stage Multi-Stage Multi-Stage Multi-Stage
Motor HP ¾¾¾¾ 1 1
RPM at 60Hz 2850 2850 2850 2850 2850 2850
Electrical
Controller C23 C23 C23 C23 C23 C23
Standard Voltage 230V 50Hz 230V 50Hz 230V 50Hz 230V 50Hz 230V 50Hz 230V 50Hz
System Dimensions
Dimensions - H x W x D (mm) 1213 x 460 x 460 1213 x 460 x 460 1339 x 460 x 460 1339 x 460 x 460 1339 x 460 x 460 1339 x 460 x 460
Approx. Weight (kg) 45 48 50 55 63.5 73
System Specifications
* Product ow and recovery rates based on feed water conditions of 1000ppm TDS and 25°C. Treatment ability of the RO System is dependent on
feed water quality. Higher TDS and/or lower temperatures will reduce product ow. Performance projections should be run for each installation.

14
Spectrum
TM
Operation & Maintenance
DO:
•Change the cartridge lters regularly
•Monitor the system and keep a daily log
•Run the system as much as possible on a continuous basis
•Adjust the system recovery to the recommended value
•Always feed the pump with ltered water
DON’T:
•Permit chlorine to enter or be present in the feed water
•Shut down the system for extended periods
•Close the throttle valve completely
•Operate the system with insufcient feed ow
•Operate the pump dry
The reverse osmosis process causes the concentration of impurities. The impurities may precipitate (come
out of solution) when their concentration reaches saturation levels.
Check the feed water chemistry and pre-treat the water and/or reduce the system’s recovery as required.
If necessary consult the local dealer or distributor.
Pre-lter Pressure Gauges
These gauges measure the feed water pressure when it enters and exits the pre-lter. A pressure differential
of 0.68 - 1 bar or more on the pressure gauge indicates the pre-lter requires servicing. For example, if the
inlet pressure is 2.75 bar the lter should be changed when the outlet pressure is 2 bar or below.
Permeate (product) Flow Meter and Concentrate (waste) ow meter
These ow meters indicate the ow rates of the permeate and concentrate water. The measurements, when
added together, also indicate the feed water ow rate or (total ow rate); if the system is not equipped with
a concentrate recycle valve.
NOTE: Precipitation can scale or foul membranes and must be prevented.
NOTE: Excessive recycling may cause premature fouling /scaling of the membrane elements.
Low Pressure Switch
The low pressure switch shuts off the system when the feed water pressure drops below 1 bar, preventing
damage to the pump. The system restarts automatically when there is a constant pressure of 2.4 bar or more.
NOTE: If you notice the pressure uctuating and the system cycling off and on, turn the
system off and ensure that proper feed ow and pressure are available to the system.
!
!
!

15
Spectrum
TM
High Pressure Switch
The high pressure switch shuts off the system when the product water pressure exceeds 4.1 bar, preventing
damage to the system. The system restarts automatically when the pressure drops below 2.75 bar.
Pump Throttle Valve
The pump throttle valve is installed as a standard feature on SPECTRUM systems. It provides an adjustment
for pump pressure. As the feed water temperature decreases and/or the feed TDS increases, the system
will require a higher operating pressure to produce the specied permeate ow.
Adjusting the Throttle Valve
To decrease the pressure, turn the handle clockwise. To increase the pressure, turn the handle counter
clockwise, shown in Figure 4.
Figure 4.
Flushing the System
The system should be ushed weekly to remove sediment from the surface of the membranes. To manually
ush the system follow the steps shown below:
1. The system must be operating during the ush procedure.
2. Fully open the concentrate valve (Figure 1, Item 5).
3. Allow the system to run for 10 - 20 minutes.
4. After 10 - 20 minutes, close the concentrate valve to its previous setting. Ensure the proper concentrate
ow rate is going to drain.
5. The system is now ready to operate.
Operation & Maintenance

16
Spectrum
TM
NOTE: Prior to shipping or storing a system for more than three weeks, the system should
be cleaned with an appropriate cleaner, ushed with water and protected from biological
attack with an appropriate solution for membrane elements. The membrane housing(s)
and plumbing lines of the system must be completely drained. Any water remaining in the
plumbing of a system may freeze, causing serious damage.
Preparation for Storage
or Shipment
Preparing the System for Storage
1. Totally immerse the elements in the membrane housing in a solution of 2% M-100 solution, venting the
air outside of the housings. Use the overow technique: circulate the M-100 solution in such a way that
the remaining air in the system is minimised after the recirculation is completed. After the membrane
housing is lled, the M-100 solution should be allowed to overow through an opening located higher
than the upper end of the highest housing being lled.
2. Separate the preservation solution from the air outside by closing all valves. Any contact with oxygen
will oxidise the M-100 solution.
3. Check the pH once a week. When the pH becomes 3 or lower, change the preservation solution.
4. Repeat this process at least once a month.
During the shutdown period, the plant must be kept frost-free and the temperature must not exceed 45°C.
Preparing unit for shipment:
5. Disconnect the inlet, concentrate, pre-lter and permeate plumbing.
6. Drain all water from the pre-lter cartridge housings by unscrewing the housings, removing the pre-
lter cartridge and draining the water from the housings.
7. Disconnect the tubing from the connectors on the permeate and concentrate inlets and outlets.
8. Fully open the concentrate valve.
9. Drain the ow meters.
10. Allow the system to drain for a minimum of eight hours or until the opened ports stop dripping.
After draining is complete, reconnect all plumbing.
!

17
Spectrum
TM
Symptoms Possible Causes Corrective Action
Low Inlet Pressure
Low supply pressure Increase inlet pressure
Cartridge lters plugged Change lters
Solenoid valve malfunction Replace solenoid valve and/or coil
Motor may not be drawing correct current Use clamp-on amp meter to check motor amp draw
Concentrate valve might be damaged Replace needle valve
Leaks Fix any visible leaks
Low Permeate Flow
Low inlet ow Adjust concentrate valve
Cold feed water See temperature correction sheet
Low operating pressure See low inlet pressure
Defective membrane brine seal Inspect and/or replace brine seal
Fouled or scaled membrane Clean membranes
High Permeate Flow
Damaged product tube O-rings Inspect and/or replace
Damaged or oxidised membrane Replace membrane
Exceeding maximum feed water temperature See temperature correction sheet
Poor Permeate Quality
Low operating pressure See low inlet pressure
Damaged product tube O-rings Inspect and/or replace
Damaged or oxidised membrane Replace membrane
Membrane Fouling
Metal oxide fouling Improve pretreatment to remove metals.
Clean with acid cleaners
Colloidal Fouling Optimise pretreatment for colloid removal.
Clean with high pH anionic cleaners
Scaling (CaSO4, CaSO3, BaSO4, SiO2)
Increase acid addition and antiscalant dosage for CaVO3and
CaCO4.
Reduce recovery. Clean with acid cleaners.
Biological fouling
Shock dosage of Sodium Bi-Sulfate.
Continuous feed of Sodium Bi-Sulfate at reduced pH.
Chlorination and de-chlorination. Replace cartridge lters
Organic fouling Activated carbon or other pretreatment.
Clean with high pH cleaner.
Chlorine oxidation Check chlorine feed equipment and de-chlorination system.
Abrasion of membrane by crystalline material Improve pretreatment.
Check all lter for media leakage.
Troubleshooting
Abnormal Permeate Flow
Permeate ow should be within 20% of the rated production, after correcting the feed water temperatures
above or below 25°C. Check the permeate ow meter to determine the permeate ow rate.
NOTE: To determine the temperature correction factor, locate the temperature correction
table in this manual and follow the directions.
!

18
Spectrum
TM
Membrane Temperature
Correction Factors
Find the temperature correction factor (TCF) in the table below. Divide the rated permeate ow at 25°C by
the temperature correction factor. The result is the permeate ow at the desired temperature (see example).
If the system is rated to produce 18.9 lpm of permeate
water at 25°C, the same system will produce more
water at a higher temperature. It will also produce less
water at a lower temperature. Use the temperature
correction table to obtain the correct ow.
Example:
18.9 lpm @ 15°C (18.90 ÷ 1.422 = 13.29)
18.9 lpm @ 25°C (18.90 ÷ 1.000 = 18.90)
18.9 lpm @ 28.9°C (18.90 ÷ 0.892 = 21.19)
If service assistance is required, please complete the following process:
Contact the local dealer or distributor. Prior to making the call, have the following information available:
system installation date, serial number, daily log sheets, current operating parameters (e.g. ow, operating
pressures, pH, etc.) and a detailed description of the problem.
Service Assistance
Temperature (°C) Temperature Correction Factor
10 1.711
11 1.648
12 1.588
13 1.530
14 1.475
15 1.422
16 1.371
17 1.323
18 1.276
19 1.232
20 1.189
21 1.148
22 1.109
23 1.071
24 1.035
25 1.000
26 0.971
27 0.943
28 0.915
29 0.889

19
Spectrum
TM
Example:
18.9 lpm @ 15°C (18.90 ÷ 1.422 = 13.29)
18.9 lpm @ 25°C (18.90 ÷ 1.000 = 18.90)
18.9 lpm @ 28.9°C (18.90 ÷ 0.892 = 21.19)
Dimension
(mm) SRO-SYS-1.3 SRO-SYS-2.6 SRO-SYS-5.2 SRO-SYS-10 SRO-SYS-16 SRO-SYS-21
A1213 1213 1339 1339 1339 1339
B1149 1149 1149 1149 1149 1149
C460 460 460 460 460 460
D460 460 460 460 460 460
E497 497 497 497 649 649
System Drawings &
Dimensions
A
B
C
D
E

20
Spectrum
TM
C23 Controller
Input Wiring Diagram 1
This manual suits for next models
5
Table of contents
Other Spectrum Water Dispenser manuals
Popular Water Dispenser manuals by other brands

Everpure
Everpure CGT-20S Specification sheet

Elkay
Elkay VRC Series Installation, care & use manual

Excalibur
Excalibur EWS S130 Installation and user guide

WaterLogic
WaterLogic WL500 - Sparkling Operating, installation and service manual

AQUA-LITY
AQUA-LITY A/C Installation, use and maintenance handbook

Blue Star
Blue Star GA Series user manual